← Product Code [EWO](/productcode/EWO) · K092111

# ENTOMED SCREENING AUDIOMETER, MODEL  SA201,SA202, ENTOMED DIAGNOSTIC AUDIOMETER MODEL SA203,SA204 (K092111)

_Entomed AB · EWO · Oct 2, 2009 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K092111

## Device Facts

- **Applicant:** Entomed AB
- **Product Code:** [EWO](/productcode/EWO.md)
- **Decision Date:** Oct 2, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.1050
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Pediatric

## Indications for Use

The Entomed SA 201/202/203/204 series of audiometers are screening and diagnostic audiometers used to determine hearing thresholds and assist to diagnose hearing loss. The audiometer can be used on the majority of adults and children (from 4 years of age and upwards). The test person has to be able to understand the basic principles of the test method when explained by the operator.

## Device Story

Electroacoustic audiometers (SA201/202 screening; SA203/204 diagnostic) generate controlled test tones and signals for hearing threshold determination. Inputs include manual operator controls and patient response switches; SA203/204 include synchronized masking. Digital signal processor generates frequencies; outputs delivered via headphones (TDH39, Sennheiser HDA200, EARtone 5A) or bone vibrators (B71). Used in clinical settings by audiologists, ENT specialists, or trained nurses. Device assists clinicians in diagnosing hearing loss by measuring patient response to stimuli. Modernized digital generation replaces traditional analog tone generation; calibration and operational software stored on removable Compact Flash (CF) cards to enhance security and prevent unauthorized calibration changes.

## Clinical Evidence

Bench testing only. Performance verified against IEC 60645-1, IEC 60645-2, ISO 389 series, and ISO 8253 series standards for audiometric equipment and test methods.

## Technological Characteristics

Digital signal processor-based tone generation; RoHS compatible. Dimensions: 37.6x26.4x7.3 cm; Weight: 1.15-1.25 kg. Power: External 100-240V. Connectivity: CF card slots for calibration/operation. Transducers: TDH39, Sennheiser HDA200, EARtone 5A, B71 bone vibrator. Standards: IEC 60645-1/2, ISO 389-1/2/3/4/7, ISO 8253-1/2/3, IEC 60601-1, IEC 60601-1-2.

## Regulatory Identification

An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.

## Predicate Devices

- Interacoustic Diagnostic Audiometer AD229b ([K982249](/device/K982249.md))
- GSI 61 Clinical Audiometer

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# OCT - 2 2009

/

1092111

## Summary of 510(k) owners administrative data

COMPANY NAME AND ADDRESS:

CONTACT PERSON:

EMAIL:

1

TELEPHONE NUMBER:

FAX:

TRADE NAME OF DEVICE:

COMMON NAME:

CLASSIFICATION:

PREDICATE DEVICE:

DATE SUMMARY PREPARED:

Entomed AB Bariumgatan 29 213 64 Malmö Sweden

Charlotte Mattisson

charlotte.mattisson@entomed.se

+ 46 (40) - 210090

+ 46 (40) - 210092

Entomed SA201/202 Screening Audiometer and Entomed SA203/204 Diagnostic Audiometer

Audiometer

Class II Medical Device, Reg no 874.1050, Product Code EWO

Interacoustic Diagnostic Audiometer AD229b and GSI 61 Clinical Audiometer

July 10 2009

{1}------------------------------------------------

### 2 Entomed AB

Entomed is a Swedish specialist company within the ENT (Ear-Nose-Throat) sector, founded 1986, with home production of audiometers, and an extensive agency business for Welch Allyn and diagnostical instruments from Grason-Stadler (GSI), Maico, and Oriola.

Up to year 2007 Entomed products were sold and distributed by GSI under the brand name GSI 66, GSI 67 and GSI 68. Since 2008 the products are sold directly from Entomed through a worldwide distributor network, under the brand names SA201TM SA202TM, SA 203TM and SA204TM

The SA20X series audiometer presented in this 510(k) submission is a new modernized generation of the well reputed product line. The new generation is digital and RoHS compatible.

## 3 Description of the device

The SA 201/202/203/204 audiometers are electroacoustic devices that produces controlled levels of test tones and signals.

The SA201/202 are IEC 60645 Type 4 Pure Tone Audiometers and the SA203/204 Diagnostic Audiometers are IEC 60645 Type 3 Audiometers.

The SA201/202 are manual pure tone air conduction, screening audiometers. SA202 has the additional capability to perform automatic pure tone audiometry.

The SA203/204 are both automatic air and bone conduction audiometers with synchronized masking possibility in order to avoid overhearing to the not tested ear. The SA 204 has the additional capability of using speech-testing. Both can be operated manually if required.

All models can be used for determination of hearing thresholds and assist to diagnose hearing loss.

All models, 201/202/203/204 have a built-in microphone for easier communication with the test person. Furthermore they all have outputs for TDH39, Sennheiser HDA200 and EARtone 5A insert ear phones. SA203/204 have outputs for B71 Bone Conductor and Free-field system (optional).

All models should be operated by people with education/training within the field of audiometry such as Audiologists. Hearing Aid dispensers, Health Care and School nurses with appropriate training in audiology and Ear Nose Throat specialists.

{2}------------------------------------------------

The performance characteristics for the SA 201/202/203/204 are such that they meet or exceed the following standards:

| Standard                       | Name                                                                                                                                                                              | SA201 | SA202 | SA203 | SA204 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|-------|-------|-------|
| IEC 60645-1 (1992)             | Audiometers Part1; Pure Tone Audiometers                                                                                                                                          | x     | x     | x     | x     |
| IEC 60645-2(1993)              | Audiometers Part 2; Equipment for Speech Audiometry                                                                                                                               | NA    | NA    | NA    | x     |
| ISO 389:1991                   | Calibration of Air Conduction Audiometers                                                                                                                                         | x     | x     | x     | x     |
| ISO 389-2                      | Calibration of Insert Phones                                                                                                                                                      | x     | x     | x     | x     |
| ISO 389-3                      | Calibration of Bone Vibrators                                                                                                                                                     | NA    | NA    | x     | x     |
| ISO 389-4                      | Calibration of Narrow Band Masking Noise                                                                                                                                          | NA    | NA    | x     | x     |
| ISO 389-7                      | Calibration of Free Field Audiometers                                                                                                                                             | NA    | NA    | x     | x     |
| IEC 60601-1 2nd edition 1988   | Medical Electrical Equipment Part1: General Requirements for Safety                                                                                                               | x     | x     | x     | x     |
| IEC 60601-1-2 3rd edition 2007 | Medical Electrical Equipment Part1-2: General Requirements for Basic Safety and Essential Performance-Collateral Standard: Electromagnetic Compatibility - Requirements and Tests | x     | x     | x     | x     |
| ISO 8253-1                     | Audiometric Test Methods Part1; Basic Pure Tone Air and Bone Conduction Threshold Audiometry                                                                                      | x     | x     | x     | x     |
| ISO 8253-2                     | Audiometric Test Methods Part 2; Sound Field Audiometry with Pure Tone and Narrow Band Test Signals                                                                               | NA    | NA    | x     | x     |
| ISO 8253-3                     | Audiometric Test Methods Part 3 Speech Audiometry                                                                                                                                 | NA    | NA    | NA    | x     |

.

NA = Not Applicable

、

{3}------------------------------------------------

### Indication for use. 4

The Entomed SA 201/202/203/204 series of audiometers are screening and diagnostic audiometers used to determine hearing thresholds and assist to diagnose hearing loss.

The audiometer can be used on the majority of adults and children (from 4 years of age and upwards). The test person has to understand the basic principles of the test method when explained by the operator.

## 5 Summary of Performance Comparison between SA 204 and Interacoustic AD 229b (predicate) and GSI 61 (predicate)

The predicates selected for the SE discussion in this 510(k) are the Interacoustic Diagnostic Audiometer AD229b and GSI 61 Clinical Audiometer. The technical features of the SA204 are very similar to the Interacoustic AD 229b. However, the technical design using two separate, identical channels differs between the SA204 and the Interacoustic AD 229. The SA204 has 2 identical channels. By adding another predicate to the SE discussion that also has 2 identical channels, the GSI 61, it is shown that the technical solution to use two identical channels is well known and accepted.

Similarities between the predicates are presented in Tables 1a-1d and differences between the predicates are presented in Tables 2a - 2d.

{4}------------------------------------------------

| Comparison<br>Parameter                                              | SA204                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Interacoustic AD229b                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | GSI 61                                                                                                                                                                                                                                                                                                                                  |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use                                                         | "The Entomed SA<br>201/202/203/204 series of<br>audiometers are screening<br>(201/202) and diagnostic<br>(203/204) audiometers used<br>to determine hearing<br>thresholds and assist to<br>diagnose hearing loss<br>The audiometer can be used<br>on the majority of adults<br>and children (from 4 years<br>of age and upwards). The<br>test person has to be able to<br>understand the basic<br>principles of the test method<br>when explained by the<br>operator" | "The Interacoustic Model AD229<br>Diagnostic Audiometer is indicated<br>for use in conducting diagnostic<br>hearing evaluations and assisting<br>in the diagnosis of possible<br>otologic disorders.<br>Because of its master hearing aid<br>capability this device is also<br>indicated for simulating a hearing<br>aid during audiometric testing thus<br>it may help in the selection and<br>adjustment of a patient's hearing<br>aid". (statement obtained from<br>510(k) summary from 510(k)<br>application k982249) | Determine hearing thresholds and<br>assist to diagnose hearing loss. It<br>can also be used in the selection<br>and adjustment of a patient's<br>hearing aid.                                                                                                                                                                           |
| Air<br>Conduction<br>Pure Tone -<br>Frequency<br>Range and<br>Format | Range:<br>125, 250, 500, 750, 1000,<br>1500, 2000, 3000, 4000,<br>6000, 8000 Hz<br>Format:<br>Steady,<br>Pulsed (2.5 pulses/sec),<br>Frequency Modulation<br>(Warble: ± 5 % triangular<br>10 Hz at 8 kHz, 8 Hz at 6<br>kHz and 5 Hz at 125 Hz - 4<br>kHz)                                                                                                                                                                                                             | Range:<br>125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz.<br>Format:<br>Steady,<br>Multiple Pulses (250-5000 msec.<br>on/off),<br>Frequency Modulation (Warble: ±<br>5 % , 5Hz true sine wave)                                                                                                                                                                                                                                                                                                        | Range:<br>125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz).<br>Format:<br>Steady,<br>Pulsed (Tone pulsed 200 msec ON,<br>200 msec OFF),<br>Frequency Modulation (Tone<br>modulated ± 5 % of centre<br>frequency at a rate of 5 Hz) |
| Bone<br>Conduction -<br>Frequency<br>Range                           | 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000,<br>8000 Hz                                                                                                                                                                                                                                                                                                                                                                                                      | 250, 500, 750, 1000, 1500, 2000,<br>3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 250, 500, 750, 1000, 1500, 2000,<br>3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                           |
| Narrow Band<br>Noise Range                                           | 125, 250, 500, 750, 1000,<br>1500, 2000, 3000, 4000,<br>6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                                                 | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz)                                                                                                                                                                                   |
| Speech Noise<br>Range                                                | 125, 250, 500, 750, 1000,<br>1500, 2000, 3000, 4000,<br>6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                                                 | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz)                                                                                                                                                                                   |
| Synchronous<br>Masking                                               | Locks the noise attenuator<br>to the tone attenuator                                                                                                                                                                                                                                                                                                                                                                                                                  | Locks channel 2 attenuator to<br>channel 1 attenuator                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Locks the noise attenuator to the<br>tone attenuator                                                                                                                                                                                                                                                                                    |
| Speech<br>Channel L and<br>R                                         | Microphone: Input for live<br>speech testing and<br>communications<br>IN1 and IN2: Inputs for<br>recorded speech material<br>from external stereo<br>CD/tape player                                                                                                                                                                                                                                                                                                   | Microphone: Input for live speech<br>testing and communications<br>Channel 1 and Channel 2: Inputs<br>for 2 channels for prerecorded<br>material<br>Speech Noise: Narrow Band noise<br>or White noise                                                                                                                                                                                                                                                                                                                     | Microphone: Input for live speech<br>testing and communications<br>External A and External B:<br>Accepts recorded speech material<br>from external stereo tape cassette<br>or CD player<br>Speech Noise: White noise                                                                                                                    |

| Table 1a Similarities between SA204 and Interacoustic AD229b and GSI 61 |  |  |  |  |  |  |  |
|-------------------------------------------------------------------------|--|--|--|--|--|--|--|
|-------------------------------------------------------------------------|--|--|--|--|--|--|--|

.

1

:

·

{5}------------------------------------------------

|                                                                      | Speech Noise: White noise<br>filtered to a low and middle<br>frequency band, simulating<br>the average spectrum of<br>conversational speech                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                         |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Communicatio<br>n<br>Operator/Test<br>subject                        | Talk Forward, Talk back<br>and Monitor / Live Speech<br>monitor headset                                                                                                                                                                                                                                                                                                                                                                                               | Talk Forward, Talk back and<br>Monitor / Live Speech monitor<br>headset                                                                                                                                                                                                                                                                                                                                                                    | Talk Forward, Talk back and<br>Monitor / Live Speech monitor<br>headset                                                                                                                                                                                                                                                                 |
| Transducers                                                          | TDH39,<br>EARTone 5A,<br>Sennheiser HDA 200,<br>Radioear B71 bone vibrator                                                                                                                                                                                                                                                                                                                                                                                            | TDH39,<br>EARTone 5A,<br>Sennheiser HDA 200,<br>Radioear B71 bone vibrator                                                                                                                                                                                                                                                                                                                                                                 | TDH50P,<br>EARTone 3A,<br>Single Insert Phone $(470 \Omega)$<br>Sennheiser HDA 200,<br>Radioear B71 bone vibrator                                                                                                                                                                                                                       |
| Tone<br>Presentation                                                 | Manual<br>Single Pulse<br>Multiple Pulses<br>Frequency Modulation                                                                                                                                                                                                                                                                                                                                                                                                     | Manual or Reverse<br>Single Pulse<br>Multiple Pulses<br>Frequency Modulation                                                                                                                                                                                                                                                                                                                                                               | Manual or Reverse<br>Single Pulse<br>Multiple Pulses<br>Frequency Modulation                                                                                                                                                                                                                                                            |
| Patient signal                                                       | Test subject response switch<br>push button                                                                                                                                                                                                                                                                                                                                                                                                                           | Reed switch push button                                                                                                                                                                                                                                                                                                                                                                                                                    | Subject response hand switch                                                                                                                                                                                                                                                                                                            |
| Store<br>Function:                                                   | Internal Memory for AC<br>L/R and BC L/R                                                                                                                                                                                                                                                                                                                                                                                                                              | Internal Memory for AC L/R and<br>BC L/R and full speech curve                                                                                                                                                                                                                                                                                                                                                                             | No internal memory for either AC<br>L/R, BC L/R or speech curve                                                                                                                                                                                                                                                                         |
| Test: Auto<br>Threshold                                              | Patient controlled Hughson<br>Westlake procedure<br>according to ISO 8253-1                                                                                                                                                                                                                                                                                                                                                                                           | Patient controlled Hughson<br>Westlake procedure according to<br>ISO 8253-1 or OSHA procedure<br>with automatic re-check (US<br>edition only)                                                                                                                                                                                                                                                                                              | No auto threshold function<br>available                                                                                                                                                                                                                                                                                                 |
| Comparison<br>Parameter                                              | SA203                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Interacoustic AD229b                                                                                                                                                                                                                                                                                                                                                                                                                       | GSI 61                                                                                                                                                                                                                                                                                                                                  |
| Intended use                                                         | "The Entomed SA<br>201/202/203/204 series of<br>audiometers are screening<br>(201/202) and diagnostic<br>(203/204) audiometers used<br>to determine hearing<br>thresholds and assist to<br>diagnose hearing loss<br>The audiometer can be used<br>on the majority of adults<br>and children (from 4 years<br>of age and upwards). The<br>test person has to be able to<br>understand the basic<br>principles of the test method<br>when explained by the<br>operator" | "The Interacoustic Model AD229<br>Diagnostic Audiometer is indicated<br>for use in conducting diagnostic<br>hearing evaluations and assisting<br>in the diagnosis of possible<br>otologic disorders.<br>Because of its master hearing aid<br>capability this device is also<br>indicated for simulating a hearing<br>aid during audiometric testing thus<br>it may help in the selection and<br>adjustment of a patient's hearing<br>aid". | Determine hearing thresholds and<br>assist to diagnose hearing loss. It<br>can also be used in the selection<br>and adjustment of a patient's<br>hearing aid.                                                                                                                                                                           |
| Air<br>Conduction<br>Pure Tone -<br>Frequency<br>Range and<br>Format | Range:<br>125, 250, 500, 750, 1000,<br>1500, 2000, 3000, 4000,<br>6000, 8000 Hz<br>Format:<br>Steady,<br>Pulsed (2.5 pulses/sec),<br>Frequency Modulation<br>(Warble: ± 5 % triangular<br>10 Hz at 8 kHz, 8 Hz at 6<br>kHz and 5 Hz at 125 Hz - 4<br>kHz)                                                                                                                                                                                                             | Range:<br>125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz<br>Format:<br>Steady,<br>Multiple Pulses (250-5000 msec.<br>on/off ),<br>Frequency Modulation (Warble: ±<br>5 % , 5Hz true sine wave)                                                                                                                                                                                                                         | Range:<br>125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz).<br>Format:<br>Steady,<br>Pulsed (Tone pulsed 200 msec ON,<br>200 msec OFF),<br>Frequency Modulation (Tone<br>modulated ± 5 % of centre<br>frequency at a rate of 5 Hz) |
| Bone<br>Conduction -<br>Frequency<br>Range                           | 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000,<br>8000 Hz                                                                                                                                                                                                                                                                                                                                                                                                      | 250, 500, 750, 1000, 1500, 2000,<br>3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                              | 250, 500, 750, 1000, 1500, 2000,<br>3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                           |
| Narrow Band<br>Noise Range                                           | 125, 250, 500, 750, 1000,<br>1500, 2000, 3000, 4000,<br>6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                                                 | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                         | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz)                                                                                                                                                                                   |
| Speech Noise<br>Range                                                | Speech Noise not available                                                                                                                                                                                                                                                                                                                                                                                                                                            | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                         | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz)                                                                                                                                                                                   |
| Synchronous<br>Masking                                               | Locks the noise attenuator<br>to the tone attenuator                                                                                                                                                                                                                                                                                                                                                                                                                  | Locks channel 2 attenuator to<br>channel 1 attenuator                                                                                                                                                                                                                                                                                                                                                                                      | Locks the noise attenuator to the<br>tone attenuator                                                                                                                                                                                                                                                                                    |
| Speech<br>Channel L and<br>R                                         | Speech Channels L and R<br>are not available                                                                                                                                                                                                                                                                                                                                                                                                                          | Microphone: Input for live speech<br>testing and communications<br>Channel 1 and Channel 2: Inputs                                                                                                                                                                                                                                                                                                                                         | Microphone: Input for live speech<br>testing and communications<br>External A and External B:                                                                                                                                                                                                                                           |

.

・

{6}------------------------------------------------

:

.

·

Table 1b. Similarities between SA203 and Interacoustic AD 229b and GSI 61

. .

{7}------------------------------------------------

|                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | material<br>Speech Noise: Narrow Band noise<br>or White noise                                                                                                                                                                                                                                                                                                                                                                              | from external stereo tape cassette<br>or CD player<br>Speech Noise: White noise                                                                                                                                                                                                                                                         |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Communicatio<br>n<br>Operator/Test<br>subject                        | Talk Forward                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Talk Forward, Talk back and<br>Monitor / Live Speech monitor<br>headset                                                                                                                                                                                                                                                                                                                                                                    | Talk Forward, Talk back and<br>Monitor / Live Speech monitor<br>headset                                                                                                                                                                                                                                                                 |
| Transducers                                                          | TDH39,<br>EARTone 5A,<br>Sennheiser HDA 200,<br>Radioear B71 bone vibrator                                                                                                                                                                                                                                                                                                                                                                                            | TDH39,<br>EARTone 5A,<br>Sennheiser HDA 200,<br>Radioear B71 bone vibrator                                                                                                                                                                                                                                                                                                                                                                 | TDH50P,<br>EARTone 3A,<br>Single Insert Phone (470 Ω )<br>Sennheiser HDA 200,<br>Radioear B71 bone vibrator                                                                                                                                                                                                                             |
| Tone<br>Presentation                                                 | Manual<br>Single Pulse<br>Multiple Pulses<br>Frequency Modulation                                                                                                                                                                                                                                                                                                                                                                                                     | Manual or Reverse<br>Single Pulse<br>Multiple Pulses<br>Frequency Modulation                                                                                                                                                                                                                                                                                                                                                               | Manual or Reverse<br>Single Pulse<br>Multiple Pulses<br>Frequency Modulation                                                                                                                                                                                                                                                            |
| Patient signal                                                       | Test subject response switch<br>push button                                                                                                                                                                                                                                                                                                                                                                                                                           | Reed switch push button                                                                                                                                                                                                                                                                                                                                                                                                                    | Subject response hand switch                                                                                                                                                                                                                                                                                                            |
| Store<br>Function:                                                   | Internal Memory for AC<br>L/R and BC L/R                                                                                                                                                                                                                                                                                                                                                                                                                              | Internal Memory for AC L/R and<br>BC L/R and full speech curve                                                                                                                                                                                                                                                                                                                                                                             | No internal memory for either AC<br>L/R, BC L/R or speech curve                                                                                                                                                                                                                                                                         |
| Test: Auto<br>Threshold                                              | Patient controlled Hughson<br>Westlake procedure<br>according to ISO 8253-1                                                                                                                                                                                                                                                                                                                                                                                           | Patient controlled Hughson<br>Westlake procedure according to<br>ISO 8253-1 or OSHA procedure<br>with automatic re-check (US<br>edition only)                                                                                                                                                                                                                                                                                              | No auto threshold function<br>available                                                                                                                                                                                                                                                                                                 |
| Comparison<br>Parameter                                              | SA202                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Interacoustic AD229b                                                                                                                                                                                                                                                                                                                                                                                                                       | GSI 61                                                                                                                                                                                                                                                                                                                                  |
| Intended use                                                         | "The Entomed SA<br>201/202/203/204 series of<br>audiometers are screening<br>(201/202) and diagnostic<br>(203/204) audiometers used<br>to determine hearing<br>thresholds and assist to<br>diagnose hearing loss<br>The audiometer can be used<br>on the majority of adults<br>and children (from 4 years<br>of age and upwards). The<br>test person has to be able to<br>understand the basic<br>principles of the test method<br>when explained by the<br>operator" | "The Interacoustic Model AD229<br>Diagnostic Audiometer is indicated<br>for use in conducting diagnostic<br>hearing evaluations and assisting<br>in the diagnosis of possible<br>otologic disorders.<br>Because of its master hearing aid<br>capability this device is also<br>indicated for simulating a hearing<br>aid during audiometric testing thus<br>it may help in the selection and<br>adjustment of a patient's hearing<br>aid". | Determine hearing thresholds and<br>assist to diagnose hearing loss. It<br>can also be used in the selection<br>and adjustment of a patient's<br>hearing aid.                                                                                                                                                                           |
| Air<br>Conduction<br>Pure Tone -<br>Frequency<br>Range and<br>Format | Range:<br>125, 250, 500, 750, 1000,<br>1500, 2000, 3000, 4000,<br>6000, 8000 Hz<br>Format:<br>Steady,<br>Pulsed (2.5 pulses/sec),<br>Frequency Modulation<br>(Warble: ± 5 % triangular<br>10 Hz at 8 kHz, 8 Hz at 6<br>kHz and 5 Hz at 125 Hz - 4<br>kHz)                                                                                                                                                                                                             | Range:<br>125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz<br>Format:<br>Steady,<br>Multiple Pulses (250-5000 msec.<br>on/off ),<br>Frequency Modulation (Warble: ±<br>5 % , 5Hz true sine wave)                                                                                                                                                                                                                         | Range:<br>125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz).<br>Format:<br>Steady,<br>Pulsed (Tone pulsed 200 msec ON,<br>200 msec OFF),<br>Frequency Modulation (Tone<br>modulated ± 5 % of centre<br>frequency at a rate of 5 Hz) |
| Bone<br>Conduction -<br>Frequency<br>Range                           | Bone conduction not<br>available                                                                                                                                                                                                                                                                                                                                                                                                                                      | 250, 500, 750, 1000, 1500, 2000,<br>3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                              | 250, 500, 750, 1000, 1500, 2000,<br>3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                           |
| Narrow Band<br>Noise Range                                           | Narrow Band Noise not<br>available                                                                                                                                                                                                                                                                                                                                                                                                                                    | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                         | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz)                                                                                                                                                                                   |
| Speech Noise<br>Range                                                | Speech Noise not available                                                                                                                                                                                                                                                                                                                                                                                                                                            | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                         | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz)                                                                                                                                                                                   |
| Synchronous<br>Masking                                               | Synchronous Masking not<br>available                                                                                                                                                                                                                                                                                                                                                                                                                                  | Locks channel 2 attenuator to<br>channel 1 attenuator                                                                                                                                                                                                                                                                                                                                                                                      | Locks the noise attenuator to the<br>tone attenuator                                                                                                                                                                                                                                                                                    |
| Speech<br>Channel L and<br>R                                         | Speech Channel L and R not<br>available                                                                                                                                                                                                                                                                                                                                                                                                                               | Microphone: Input for live speech<br>testing and communications<br>Channel 1 and Channel 2: Inputs<br>for 2 channels for prerecorded                                                                                                                                                                                                                                                                                                       | Microphone: Input for live speech<br>testing and communications<br>External A and External B:<br>Accepts recorded speech material                                                                                                                                                                                                       |
|                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | material<br>Speech Noise: Narrow Band noise<br>or White noise                                                                                                                                                                                                                                                                                                                                                                              | from external stereo tape cassette<br>or CD player<br>Speech Noise: White noise                                                                                                                                                                                                                                                         |
| Communicatio<br>n<br>Operator/Test<br>subject                        | Talk Forward                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Talk Forward, Talk back and<br>Monitor / Live Speech monitor<br>headset                                                                                                                                                                                                                                                                                                                                                                    | Talk Forward, Talk back and<br>Monitor / Live Speech monitor<br>headset                                                                                                                                                                                                                                                                 |
| Transducers                                                          | TDH39,<br>EARTone 5A,<br>Sennheiser HDA 200                                                                                                                                                                                                                                                                                                                                                                                                                           | TDH39,<br>EARTone 5A,<br>Sennheiser HDA 200,<br>Radioear B71 bone vibrator                                                                                                                                                                                                                                                                                                                                                                 | TDH50P,<br>EARTone 3A,<br>Single Insert Phone (470 Ω )<br>Sennheiser HDA 200,<br>Radioear B71 bone vibrator                                                                                                                                                                                                                             |
| Tone<br>Presentation                                                 | Manual<br>Single Pulse<br>Multiple Pulses<br>Frequency Modulation                                                                                                                                                                                                                                                                                                                                                                                                     | Manual or Reverse<br>Single Pulse<br>Multiple Pulses<br>Frequency Modulation                                                                                                                                                                                                                                                                                                                                                               | Manual or Reverse<br>Single Pulse<br>Multiple Pulses<br>Frequency Modulation                                                                                                                                                                                                                                                            |
| Patient signal                                                       | Test subject response switch<br>push button                                                                                                                                                                                                                                                                                                                                                                                                                           | Reed switch push button                                                                                                                                                                                                                                                                                                                                                                                                                    | Subject response hand switch                                                                                                                                                                                                                                                                                                            |
| Store<br>Function:                                                   | Internal Memory for AC<br>L/R                                                                                                                                                                                                                                                                                                                                                                                                                                         | Internal Memory for AC L/R and<br>BC L/R and full speech curve                                                                                                                                                                                                                                                                                                                                                                             | No internal memory for either AC<br>L/R, BC L/R or speech curve                                                                                                                                                                                                                                                                         |
| Test: Auto<br>Threshold                                              | Patient controlled Hughson<br>Westlake procedure<br>according to ISO 8253-1                                                                                                                                                                                                                                                                                                                                                                                           | Patient controlled Hughson<br>Westlake procedure according to<br>ISO 8253-1 or OSHA procedure<br>with automatic re-check (US edition only)                                                                                                                                                                                                                                                                                                 | No auto threshold function<br>available                                                                                                                                                                                                                                                                                                 |
| Comparison<br>Parameter                                              | SA201                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Interacoustic AD229b                                                                                                                                                                                                                                                                                                                                                                                                                       | GSI 61                                                                                                                                                                                                                                                                                                                                  |
| Intended use                                                         | "The Entomed SA<br>201/202/203/204 series of<br>audiometers are screening<br>(201/202) and diagnostic<br>(203/204) audiometers used<br>to determine hearing<br>thresholds and assist to<br>diagnose hearing loss<br>The audiometer can be used<br>on the majority of adults<br>and children (from 4 years<br>of age and upwards). The<br>test person has to be able to<br>understand the basic<br>principles of the test method<br>when explained by the<br>operator" | "The Interacoustic Model AD229<br>Diagnostic Audiometer is indicated<br>for use in conducting diagnostic<br>hearing evaluations and assisting<br>in the diagnosis of possible<br>otologic disorders.<br>Because of its master hearing aid<br>capability this device is also<br>indicated for simulating a hearing<br>aid during audiometric testing thus<br>it may help in the selection and<br>adjustment of a patient's hearing<br>aid". | Determine hearing thresholds and<br>assist to diagnose hearing loss. It<br>can also be used in the selection<br>and adjustment of a patient's<br>hearing aid.                                                                                                                                                                           |
| Air<br>Conduction<br>Pure Tone -<br>Frequency<br>Range and<br>Format | Range:<br>125, 250, 500, 750, 1000,<br>1500, 2000, 3000, 4000,<br>6000, 8000 Hz<br>Format:<br>Steady,<br>Pulsed (2.5 pulses/sec),<br>Frequency Modulation<br>(Warble: ± 5 % triangular<br>10 Hz at 8 kHz, 8 Hz at 6<br>kHz and 5 Hz at 125 Hz - 4<br>kHz)                                                                                                                                                                                                             | Range:<br>125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz<br>Format:<br>Steady,<br>Multiple Pulses (250-5000 msec.<br>on/off ),<br>Frequency Modulation (Warble: ±<br>5 % , 5Hz true sine wave)                                                                                                                                                                                                                         | Range:<br>125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz).<br>Format:<br>Steady,<br>Pulsed (Tone pulsed 200 msec ON,<br>200 msec OFF),<br>Frequency Modulation (Tone<br>modulated ± 5 % of centre<br>frequency at a rate of 5 Hz) |
| Bone<br>Conduction -<br>Frequency<br>Range                           | Bone conduction not<br>available                                                                                                                                                                                                                                                                                                                                                                                                                                      | 250, 500, 750, 1000, 1500, 2000,<br>3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                              | 250, 500, 750, 1000, 1500, 2000,<br>3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                           |
| Narrow Band<br>Noise Range                                           | Narrow Band Noise not<br>available                                                                                                                                                                                                                                                                                                                                                                                                                                    | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                         | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz)                                                                                                                                                                                   |
| Speech Noise<br>Range                                                | Speech Noise not available                                                                                                                                                                                                                                                                                                                                                                                                                                            | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000 Hz                                                                                                                                                                                                                                                                                                                                                                         | 125, 250, 500, 750, 1000, 1500,<br>2000, 3000, 4000, 6000, 8000,<br>12000 Hz<br>(High frequency as "Optional<br>accessories" 8000 Hz to 20 000<br>Hz)                                                                                                                                                                                   |
| Synchronous<br>Masking                                               | Synchronous Masking not<br>available                                                                                                                                                                                                                                                                                                                                                                                                                                  | Locks channel 2 attenuator to<br>channel 1 attenuator                                                                                                                                                                                                                                                                                                                                                                                      | Locks the noise attenuator to the<br>tone attenuator                                                                                                                                                                                                                                                                                    |
| Speech<br>Channel L and<br>R                                         |…

---

**Source:** [https://fda-staging.innolitics.com/device/K092111](https://fda-staging.innolitics.com/device/K092111)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com