K092107 · Nidacon International AB · MQL · Jun 15, 2010 · Obstetrics/Gynecology
Device Facts
Record ID
K092107
Device Name
VITRIBLAST, THERMOBLAST, MODEL VB-010, TB-010
Applicant
Nidacon International AB
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Jun 15, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K092107 · Jun 15, 2010
VITRIBLAST, THERMOBLAST, MODEL VB-010, TB-010
Nidacon International AB
Clinical records from The Fertility Centre at Carlanderska Hospital
Retrospective analysis of clinical outcomes (blastocyst survival, implantation, and pregnancy rates) for patients treated with the device in a routine clinical setting.
Retrospective clinical performance at The Fertility Centre; Retrospective cohort analysis of routine clinical treatments; Follow-up/Duration: Data collected from 2007 to 2010; Study Period: 2007-2010
Patients undergoing IVF treatment with excess blastocysts; Sample Size: 424 thawed blastocysts; Number of Sites: 1
VitriBlast is intended for ultra-rapid freezing (vitrification) of human blastocysts. This kit is designed for use with Nidacon's ThermoBlast kit for optimal recovery of specimens. This product is used for assisted reproduction-technology procedures. ThermoBlast is intended for the recovery of human blastocysts that have undergone ultra-rapid freezing (vitrification) using Nidacon's VitriBlast kit. It is designed for optimal thawing and recovery of the specimens. This product is used for assisted reproduction-technology procedures.
Device Story
VitriBlast and ThermoBlast are cryopreservation media kits for human blastocysts. VitriBlast (5 vials) facilitates vitrification; ThermoBlast (4 vials) facilitates thawing. Solutions contain cryoprotectants (ethylene glycol, DMSO), osmotic agents (sucrose), and polymers (Ficoll) in a Hepes-buffered base with human serum albumin (hSA). Used in clinical embryology laboratories; operated by trained embryologists. Embryos are exposed to sequential solutions to manage osmotic stress and prevent intracellular ice formation during rapid cooling/warming. Artificial shrinkage of blastocysts via laser may be performed prior to vitrification. Output is a cryopreserved or thawed blastocyst suitable for transfer. Benefits include high survival rates and successful implantation in ART patients.
Clinical Evidence
Clinical data from 424 thawed blastocysts at the Fertility Centre (Carlanderska Hospital) showed 92% survival rate (391/424). Implantation rate was 45.5% (178/391) and pregnancy rate was 45.7% (176/385). Abortion rate was 9%. No malformations or adverse effects reported in 125 live births. Bench testing included pH, osmolality, endotoxin levels, and mouse embryo assays (MEA).
Technological Characteristics
Hepes-buffered media containing ethylene glycol, DMSO, Ficoll, sucrose, and human serum albumin (hSA). Formulated as multi-vial kits for sequential use. Non-automated, manual preparation. Sterility and endotoxin testing performed per batch. No electronic components or software.
Indications for Use
Indicated for ultra-rapid freezing (vitrification) and subsequent thawing/recovery of human blastocysts in patients undergoing assisted reproductive technology (ART) procedures.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
K092107
Image /page/0/Picture/1 description: The image shows the word "Page" at the top, followed by a symbol that looks like a power button. Below the symbol, the word "Nidacon" is written in a serif font. The text and symbol appear to be part of a logo or heading.
JUN 1 6 2010
Flöjelbergsgatan 16 B SE-431 37 Mölndal Sweden Tel: +46-31-703 06 30 Fax: +46-31-40 54 15 E-mail: contact(@nidacon.se
#### 510(k) Summary
Tradename: VitriBlast™ and ThermoBlast™
Common Name: Solutions for the vicrification of human blastocysts and for the thawing of vitrified human blastocysts.
Classification name: Reproductive Media (21 CFR, 886.6180, Product code: MQL)
#### Intended use:
- VitriBlast is intended for ultra-rapid freezing (vitrification) of human blastocysts. This kit is designed for use . with Nidacon's ThermoBlast kit for optimal recovery of specimens. This product is used for assisted reproduction-technology procedures.
- ThermoBlast is intended for the recovery of human blastocysts that have undergone ultra-rapid freezing . (vitrification) using Nidacon's VitriBlast kit. It is designed for optimal thawing and recovery of the specimens. This product is used for assisted reproduction-technology procedures.
#### Description:
VitriBlast consists of 5 separate vials as described hereunder:
- VitriBlast 1: 10mL based on PureSperm Wash (K002630), with additional hSA.
- VitriBlast 2: 10mL same as solution 1, plus Ethylene glycol (7.5%) and DMSO (7.5%) both provided for separate addition.
- VitriBlast 3: 10mL same as solution 1, including Sucrose (0.67 M) and Ficoll (0.14 mM), plus Ethylene glycol (15%) and DMSO (15%), both provided for separate addition just prior to use.
ThermoBlast consists of 4 separate, 10 mL vials as follows:
- ThermoBlast 4: based on VitriBlast I with additional sucrose (0.33 M).
- ThermoBlast 5: based on VitriBlast I with additional sucrose (0.21 M).
- ThermoBlast 6: the same as VitriBlast | (2 vials).
#### Device function/Scientific Concept
During the cryopreservation of embryos, they are at risk of injury from chilling and crystallisation, the toxicity of the cryo protectant, extracellular ice, intracellular ice, fracture damage, osmotic swelling and osmotic shrinkage. To obtain high rates of survival, all these problems must be circumvented.
Among the injuries, the damage caused by the formation of intracellular ice crystals during cooling and or warming is one of the greatest obstacles to overcome. It is desirable to freeze without forming crystals within the cell and this can be achieved by vitrification (glass transition without crystals). Vitrification is ultimately the fact that a liquid cannot have more order than its corresponding crystal. As the temperature of a liquid substance is lowered, its entropy is reduced more rapidly than the entropy of the corresponding crystalline form of the substance. If the liguid does not freeze, a thermodynamic conundrum is approached as the entropy of the liquid approaches the entropy of the crystal.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the text "page 2 of 3" at the top. Below the text is a logo that appears to be a stylized symbol. The word "Nidacon" is written below the logo. The image appears to be a page from a document.
Flöjelbergsgatan 16 B SE-431 37 Mölndal Sweden Tel: +46-31-703 06 30 Fax: +46-31-40 54 15 E-mail: contact(@nidacon.se
For preventing intracellular ice from forming, rapidly permeating cryo-protectants, such as ethylene glycol and DMSO, are suitable since the rapid permeation by these cryo-protectants is essential to vitrify the embryo. To prevent osmotic swelling during removal of the cryoprotectants during rapidly permeating agents are also suitable, because the faster the diffusion of the intracellular cryo-proteccant out of the lower the risk of osmotic over-swelling. Ficoll is included as a non-permeating polymer (macromolecules promote vicrification) because it has high solubility, low viscosity and low toxicity. It has been shown that even large amounts of Ficoll are virtually non-toxic. A large amount of polymer occupies a significant proportion of the solution volume. Therefore, its inclusion must increase the proportion of permeating cryoprotectant per water volume. This may be one mechanism that promotes vitrification of the solution contains not only ethylene glycol and Ficoll, but also sucrose as a non-permeating small sugar which has considerable osmotic effect. The inclusion of sucrose is quite effective at reducing the apparent toxicity of ethylene glycol, probably because sucrose promotes the shrinkage of embryos, thereby restricting excess permeation of ethylene glycol and, thus, reducing its toxic effect. In addition, sucrose helps prevent over-swelling during thawing and the removal of the ethylene glycol.
#### Products claimed for substantial equivalence:
- VitKit-Freeze and VitKit-Thaw from Irvine Scientific (K060168) (DMSO, Ethylene glycol, Sucrose) .
- Vitri Freeze Kit from FertiPro (K070135) (same as above + Ficoll) .
#### Comparison to predicate device
The two predicate devices were compared in the following areas and were found to have similar technological characteristics:
Nidacon's products use the same cryoprotectants (DMSO, Ethylene glycol and Sucrose) but the devices have different base media. Irvine uses medium 199 and FertiPro uses PBS, while Nidacon uses PureSperm®Wash. We have also added Ficoll, a polymer macromolecule (as also used in the FertiPro product). Nidacon has the same concentrations of the cryoprotectants Ethylene glycol (7.5%) in vitrification solution number 2 and 15% of Ethylene glycol (EG) and DMSO in vitrification solution number 3, as with the Irvine products.
The EG and DMSO in the Nidacon kit are to be added by the clinic just prior to use. In Irvine's product, these compounds are added during products contain Human Serum Albumin (hSA). Nidacon has not included any vitrification device but, instead, provides recommendations for both closed and open systems (open systems only outside US). Irvine's Vitrification media contain antibiotics, which are not present in either Nidacon's or Fertipro's products.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the text "Page 3 of", followed by a logo that looks like a stylized letter "N" or an abstract symbol. Below the logo, the word "Nidacon" is printed in a clear, sans-serif font. The text and logo appear to be part of a document or letterhead.
Flöjelbergsgatan 16 B SE-431 37 Molndal Sweden Tel: +46-31-703 06 30 Fax: +46-31-40 54 15 E-mail: contact@nidacon.sc
| | Nidacon | Irvine | FertiPro |
|--------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | VitriBlast is intended for<br>ultra-rapid freezing<br>(vitrification) of human<br>blastocysts. This kit is<br>designed for use with<br>Nidacon's ThermoBlast kit<br>for optimal recovery of<br>specimens. This product is<br>used for ART procedures.<br><br>ThermoBlast is intended for<br>the recovery of human<br>blastocysts that have<br>undergone ultra-rapid<br>freezing (vitrification) using<br>Nidacon's VitriBlast kit. It is<br>designed for optimal<br>recovery of specimens. This<br>product is used forART<br>procedures. | VitriKit Freeze is<br>intended for ultra-rapid<br>freezing and containment<br>of human blastocysts for<br>ART procedures<br><br>VitriKit -Thaw is<br>intended for the recovery<br>of human blastocysts that<br>have undergone ultra-<br>rapid freezing and<br>containment using Vitri<br>Kit - Freeze | VitriFreeze is intended<br>for ultra-rapid freezing<br>using of human<br>blastocysts for ART<br>procedures<br><br>VitriThaw is intended for<br>the recovery of human<br>blastocysts that have<br>undergone ultra-rapid ·<br>freezing using VitriFreeze,<br>for ART procedures |
| Formulation | Hepes buffered media<br>containing ethylene glycol,<br>DMSO, Ficoll, Sucrose and<br>HSA | Media 199 containing<br>ethylene glycol, DMSO,<br>Sucrose, HSA and<br>antibiotics | PBS containing ethylene<br>glycol, DMSO, Sucrose,<br>Ficoll and HSA |
| Shelf-life | 6 months | l year | 6 months |
The products from the three companies (Nidacon, Irvine and FertiPro) are based on the same technologies and scientific concepts, and can therefore be considered equivalent.
#### Performance testing
#### Bench
Satisfactory safety of the products has been determined during all batch productions and during establishment of the shelf-life. The following tests were utilised:
- pH testing
- Osmolality control
- One cell MEA
- MEA according to intended use
- Endotoxin level measurements
#### Animal Non-Clinical Trials
Nidacon applies two different types of MEA testing, one according to the FDA recommended method (2 cell MEA %blast 96 h) and one where the performance of the solutions is evaluated (testing according to intended use). Both results will be stated in the QA-Certificate and the product insert.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the text "Page 4 of" at the top. Below this text is a logo that appears to be a stylized symbol. The word "Nidacon" is written below the logo in a clear, legible font.
Flöjelbergsgatan 16 B SE-431 37 Mölndal Sweden Tcl: +46-31-703 06 30 Fax: +46-31-40 54 15 E-mail: contact@nidacon.se
#### Human Clinical Trials
Fertility Clinic: The Fertility Centre at Carlanderska Hospital in Gothenburg, Sweden, was started in 1987 by Dr. Matts Wikland. Since the start, they have been one of the world leading clinics in the ART field. They were the first European laboratory to be accredited according to ISO/IEC 17025 in 1998. The clinic has been using the vicrification methodology since 2005. The products from Nidacon called VitriBlast™ and ThermoBlast™ were developed during scientific collaboration between Nidacon International AB and The Fertility Centre.
Patient group: When going through an IVF treatment, one might have excess potential blastocysts which need to be stored for the patient/couple for possible future treatment in case of failure of pregnancy or the desire for more children. Today these excess blastocysts are routinely stored by cryo-preservation. Nidacon became the manufacturer of the vitrification media used at The Fertility centre, using their same recipe, two years after the clinical testing was commenced, and had become routine at the clinic. Therefore, no specific test group was selected for the Nidacon vitrification media.
Test period: Today VitriBlast™ and ThermoBlast™ are the only vitrification solutions used in patient treatments at The Fertility Centre and the figures below illustrate all results from the clinic since 2007, when the solutions made by Nidacon were first used.
Method: The protocol is according to the package insert for VitriBlast™, and using the cryoloop as device. A more comprehensive methodological description using the cryo-loop is attached.
#### Results from Fertility Centre;
| Age of woman | Average : 35,9 range 24-45 |
|-------------------------------|------------------------------------------------------------------------|
| Infertility Reason | 1/3 female factors, 1/3 male factors and 1/3 unexplained infertility |
| No of thawed blastocysts | 424 |
| No of survived | 391 (391/424= 92%) |
| No of transferred | 391 |
| No of transfers | 385 |
| No of single embryo transfers | 379 (98.4%) |
| No of gestational sacs | 178 |
| Implantation rate | $178/391 = 45.5%$ |
| No of clinical pregn, | 176 |
| Pregnancy rate | $176/385= 45.7%$ |
| Abortion rate | 9 % (calculated by born babies - amount positive heart beats week 6-7) |
1) It is standard procedure at the Göteborg clinic to use blastocysts for vitrification (usually at Day 5 or 6).
2) All the blastocysts used for vitrifying were artificially shrunk by laser.
To date, 125 babies have been born using this freezing technique and these Nidacon formulations, VitriBlast™ and ThermoBlast™. No malformations in the offspring have been observed and no pregnancy terminations have occurred. Moreover, there were no multiple pregnancies.
Based on the clinical data presented and our experience with VitriBlast™ and ThermoBlast™, we feel that the safety and effectiveness of the products for their intended use are shown in the present submission and the products are substantially equivalent to the predicate devices from Irvine (K060168) and FertiPro ((K070135). A follow up study was performed on the children born after using Nidacon's Vitrification media, and from this study it
{4}------------------------------------------------
# K092107
Image /page/4/Picture/1 description: The image shows the text "Page 5 of" at the top. Below the text is a logo that appears to be a stylized symbol. The word "Nidacon" is written below the logo. The image is in black and white.
Flöjelbergsgatan 16 B SE-431 37 Mölndal Sweden Tel: +46-31-703 06 30 Fax: +46-31-40 54 15 E-mail: contact(@nidacon.se
could be concluded that there were no adverse effects or complications, and that Nidacon's products were better than clinically equivalent.
#### Additional Information
For the release of each production batch of this product, an endotoxin test, two mouse embryo assays and sterility testing will be performed. The reported on a Quality Assurance Certificate available for all batches.
### Collapsing of blastocysts prior to Vitrification
Artificial shrinkage of expanded and hatching blastocysts will be recommended. However, this technique is optional when vitrifying early blastocysts with smaller blastocoelic cavities.
#### Conclusion
The testing performed, using the animal model and the testing proves that these products meet the design requirements, and that they have substantial equivalence to the Irvine and FertiPro products and is therefore suitable for the intended use.
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle with its wings spread. The eagle is facing to the left.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
NidaCon International AB c/o Daniel Kamm, P.E. Submission Correspondent Kamm & Associates 8870 Ravello Court NAPLES FL 34114
# JUN 1 6 2010
Re: K092107
> Trade/Device Name: VitriBlast and ThermoBlast Reproductive Media Regulation Number: 21 CFR \$884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: MOL Dated: May 25, 2010 Received: June 2, 2010
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{6}------------------------------------------------
### Page 2 -
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use 510(k) Number (if known): K09 2 | 0 7
Device Name: VitriBlast
Indications For Use:
VitriBlast is intended for ultra-rapid freezing of human blastocysts. This kit is designed for use with Nidacon's ThermoBlast Thaw Kit for optimal recovery of specimens.
This product is used for Assisted Reproductive Technology (A.R.T.) procedures.
Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number 92107
Page 1 of 2
{8}------------------------------------------------
Indications for Use 510(k) Number (if known): K09 2107
Device Name: ThermoBlast
Indications For Use:
ThermoBlast is intended for the recovery of human blastocysts that have undergone ultra-rapid freezing using Nidacon's VitriBlast Kit. It is designed for optimal recovery of specimens. This product is used for Assisted Reproductive Technology (A.R.T.) procedures.
Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.