← Product Code [FBK](/productcode/FBK) · K092090

# ERBELIFT HAND PUMP AND FLEXIBLE PROBE (K092090)

_Byrne Medical, Inc. · FBK · Nov 30, 2009 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K092090

## Device Facts

- **Applicant:** Byrne Medical, Inc.
- **Product Code:** [FBK](/productcode/FBK.md)
- **Decision Date:** Nov 30, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The ERBELIFT™ Hand Pump in conjunction with an ERBELIFT Flexible Probe is indicated for the induction of sterile normal saline into the submucosa to lift polyps or other mucosal lesions using direct visualization through a flexible endoscope.

## Device Story

The ERBELIFT system consists of a reusable nylon/stainless steel hand pump and a sterile, single-use PEEK/nylon flexible probe. The device is used during endoscopic procedures to lift polyps or mucosal lesions. A control syringe filled with sterile normal saline is attached to the probe via a luer lock and loaded into the hand pump. The probe is inserted through the working channel of an endoscope. Under direct visualization, the probe tip is placed against the target tissue; the hand pump trigger is squeezed to deliver a high-pressure jet stream of saline (145-650 psi) into the submucosa. This needle-free, mechanical fluid injection creates a tissue lift, facilitating lesion removal. The system reduces tissue trauma compared to traditional needle-based injection methods due to the smaller diameter of the fluid stream and lack of a mechanical needle puncture.

## Clinical Evidence

No clinical data. Evidence consists of functional bench testing and animal lab testing to demonstrate safety and effectiveness of the submucosal lift. Biocompatibility testing was performed for the specific materials used in the flexible probe.

## Technological Characteristics

Hand pump: nylon and stainless steel; 8-3/4" x 6" x 2-1/2". Flexible probe: PEEK and nylon; 260 cm length, 1.3 mm OD, 120 µm jet stream diameter. Energy source: manual mechanical compression. Pressure range: 145-650 psi. Flow rate: 17-38 ml/minute. Sterilization: Ethylene Oxide (probe).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Endoscopic Injection Needle ([K924102](/device/K924102.md))
- Endoscopic Injection Needle/Snare ([K040961](/device/K040961.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K092090

#### Byrne Medical Incorporated

Traditional 510(k): ERBELIFT™ Hand Pump and Flexible Probe

### 510(k) SUMMARY

Submitted By:

Byrne Medical Incorporated 3150 Pollock Drive Conroe, TX 77303 Tel: 936-539-0391 Fax: 936-539-2381

NOV 3 0 2009

Chris Hierholzer Contact Person: Quality Control Manager

April 22, 2009 Date Prepared:

Common Name: Iniector

ERBELIFT™ Hand Pump and Flexible Probe Trade/Proprietary Name:

Classification Name:

Product Code:

FBK

Endoscopic Injection Needle (21 CFR Part 876.1500)

Legally Marketed Predicate Device:

Endoscopic Injection Needle 510(k) Number: K924102 and Endoscopic Injection Needle/Snare 510(k) Number: K040961

### Device Description:

The ERBELIFT™ Hand Pump and Flexible Probe are used for the delivery of pressurized sterile normal saline to the submucosa to lift polyps or other mucosal lesions. A control svringe is filled with sterile normal saline. The filled svringe is attached to the Flexible Probe via a hub with a standard luer lock and loaded into the Hand Pump. The Flexible Probe is primed with the saline and then the distal end of the Probe is placed down the working channel of the endoscope. Under direct visualization, the Probe's tip is placed against the target tissue without puncturing Then, the trigger of the Hand Pump is squeezed to produce the mucosa. pressurized saline from the control syringe and through the Probe into the bed of the polyp or lesion to achieve a desired tissue lift.

{1}------------------------------------------------

Kogzogo

# Byrne Medical Incorporated Traditional 510(k): ERBELIFT™ Hand Pump and Flexible Probe

The Hand Pump is made of nylon and stainless steel. It is 8-3/4" (22 cm) long, 6" (15 cm) wide, and having a thickness of 2-1/2" (6 cm). The Pump per squeeze of the handle delivers approximately 0.75 ml of normal saline. The Flexible Probe is made of Polyetheretherketone (PEEK) and nylon (Note: Medical grade inks and reducer are used to imprint logos on the device.). It has a 1.3 mm Outer Diameter and length of 260 cm. The distal tip is tapered to approximately 0.005" and delivers a jet stream of 120 um. The Hand Pump with the Flexible Probe can generate pressures in the approximate range of 145 to 650 psi which correlates to flow rates of 17 to 38 ml/minute.

The Hand Pump is reusable with the Flexible Probe being provided as sterile and single use.

# Intended Use:

The ERBELIFT™ Hand Pump in conjunction with an ERBELIFT Flexible Probe is indicated for the induction of sterile normal saline into the submucosa to lift polyps or other mucosal lesions using direct visualization through a flexible endoscope.

Similarities and Differences of the Proposed Device to the Current Devices (Predicate Comparison/Substantial Equivalence):

#### Similarities

The ERBELIFT Hand Pump and Flexible Probe have an intended use that is within the intended uses of the predicates. Also, like the predicates, the purpose is the same in that the devices are for the injection of fluids into tissue for endoscopic purposes. The lifting medium (i.e. sterile normal saline) is also a part of the types of solutions identified for use with the predicates. The materials used for the proposed device are also similar to the predicates. Additionally for accompanying various types and sizes of endoscopes, the working length of the Flexible Probe is comparable (260 cm) to the predicates (230 cm). Both predicates also require the need for a syringe connected via standard luer lock like the proposed device (Note: A designated control syringe is used to ensure sufficiency in regards to strength and The control syringe is identified in the Instructions For Use for the durabilitv. proposed device.). Furthermore, the Flexible Probe is sterilized via Ethylene Oxide, single use, and disposable as are the predicates.

{2}------------------------------------------------

R 092090

### Byrne Medical Incorporated

Traditional 510(k): ERBELIFT™ Hand Pump and Flexible Probe

### Differences

The Hand Pump is an additional component for the ERBELIFT™ system as compared to the predicates. The Pump provides the force to the control syringe to induce the pressurized fluid into the submucosa as compared to the force of the needle and fluid hand pushed from a syringe to lift polyps or other mucosal lesions.

The technology employed in the ERBELIFT product has been widely used in medical devices since the 1900s. Needle-free injection devices, first called "iet injectors," were developed in the 1930s and used extensively for over 50 years in mass human vaccination programs for smallpox, polio, and measles.123 Usina mechanical compression to force fluid through a small orifice, these devices produced a high-pressure stream 76 to 360 µm in diameter (compared to 810 µm for a 21-gauge needle) that could penetrate skin and subcutaneous tissue4. These devices are used today to inject insulin, cardiovascular dves, etc. The manual fluid injectors are classified by FDA as product code. KZE.

Functional bench and animal lab testing has demonstrated that the ERBELIFT Hand Pump and Flexible Probe lifts the submucosa safely and effectively. See Section III, Product Data - Performance Testing.

The types of materials used for the Flexible Probe are similar to the predicates but specific materials are slightly different. Therefore, biocompatibility of the specific materials for the Probe was demonstrated. See Section III, Product Data -Biocompatibility Study,

Besides the use of a Hand Pump with the ERBELIFT system to provide the force needed to lift the mucosa in comparison to force of a needle and fluid pressure from the predicates. the working diameter and tip/fluid stream of the ERBELIFT Flexible Probe is much smaller (i.e. a working diameter of 1.3 mm with a tipped Inner Diameter of approximately 0.120 mm as compared to a working diameter of 2.5/3.0 mm respectively with a 25 G needle/tip of 0.508 mm Outer Diameter and 0.241 mm Inner Diameter for the predicates). For the ERBELIFT Flexible Probe the smaller diameter and fluid stream with no mechanical needle injection causes less penetration tissue damage. Also, because fluid seeks the path of least resistance it is easier to find the tissue plane and easier to create a lift, See Section III. Product Data - Performance Testing.

{3}------------------------------------------------

### Byrne Medical Incorporated

Traditional 510(k): ERBELIFT™ Hand Pump and Flexible Probe

# Conclusion:

The ERBELIFT Hand Pump and Flexible Probe's intended use is a part of the predicate's indications in the previously cleared 510(k)s. The ERBELIFT system has slightly different principles of operation and technological characteristics as the predicate devices. The use of an amplified fluid force via the Hand Pump as compared to the force of a needle and pressure from pushing a syringe by hand with the predicates (i.e. mechanical trauma of a needle with less fluid pressure) was found to be less traumatic to tissue because of the penetration size being much smaller as well as it being easier to do because of fluid dynamics. The modifications of having a Hand Pump instead of a needle as well as a Flexible Probe with a smaller working diameter and tip configuration did not raise any new safety or efficacy concerns. In conclusion, all the changes were verified or validated. As a result, the ERBELIFT Hand Pump and Flexible Probe did not adverselv affect safety or effectiveness.

| Reis EC, Jacobson RM, Tarbell S, Weniger BG, Taking the sting out of shots; control of vaccination-associated pain and adverse reactions. Pediatr Ann. 1998:27:375-386.

2 Hingson RA, Davis HS, Rosen M. The historical development of jet injection and envisioned uses in mass immunization and mass therapy based upon two decades experience. Mil Med. 1963;128:516-524.

3 Chase CCL, Daniels CS, Garcia R, et al. Needle-free injection technology in swine: Progress toward vaccine efficacy and pork quality. J Swine Health Prod. 2008:16(5):254-261. (http://www.aasv.org/shap/issues/v16n5/v16n5p254.htm

4 Mitragotri S. Immunization without needles. Nat Rev Immunol, 2005;5;905-916

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Byrne Medical, Inc. % Mr. Daniel W. Lehtonen Sr. Staff Engineer - Medical Devices Intertek Testing Services 2307 East Aurora Road, Unit B7 TWINSBURG OH 44087

# NOV 3 0 2009

Re: K092090

Trade/Device Name: Erbelift Hand Pump and Flexible Probe Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FBK Dated: September 28, 2009 Received: September 29, 2009

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical.

:

{5}------------------------------------------------

Page 2 -

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR·Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: Byrne Medical, Inc.'s ERBELIFT™ Hand Pump and Flexible Probe

Indications For Use:

The ERBELIFT Hand Pump in conjunction with an ERBELIFT Flexible Probe is indicated for the induction of sterile normal saline into the submucosa to lift polyps or other mucosal lesions using direct visualization through a flexible endoscope.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Huthenie

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device 092091 510(k) Number_

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K092090](https://fda-staging.innolitics.com/device/K092090)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
