VICTORY XL 130 FOUR WHEEL SCOOTER, MODEL SC713

K092088 · Pride Mobility Products Corp. · INI · Jul 16, 2009 · Physical Medicine

Device Facts

Record IDK092088
Device NameVICTORY XL 130 FOUR WHEEL SCOOTER, MODEL SC713
ApplicantPride Mobility Products Corp.
Product CodeINI · Physical Medicine
Decision DateJul 16, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 890.3800
Device ClassClass 2

Indications for Use

The intended use of the Pride Mobility Products Corp. Victory XL 130, Four Wheel Scooter / Model # SC713 is to provide mobility to disabled persons having limited walking capabilities.

Device Story

The Victory XL 130 (Model SC713) is a heavy-duty, battery-operated, four-wheel mobility scooter. It features a digital controller, electrical system, charger, transaxle/motor, removable 12V 55AH batteries, and a 2-piece frame. The device includes electronic regenerative and electromechanical disc brakes, anti-tip rear wheels, and a foldable, rotating contour seat with fore-aft sliders. Operated by the user, the scooter provides mobility assistance for individuals with limited walking capabilities. It is designed for use by consumers and provided through retailers. The device transforms battery power into mechanical propulsion via the transaxle/motor, controlled by the user through the digital controller. The output is the physical transport of the user. It is intended for both prescription and over-the-counter use.

Clinical Evidence

No clinical data. Bench testing only, including compliance with RESNA WC Vol. 1 & 2 (2008), ANSI/RESNA WC Vol. 2-1998 Section 21 (EMC), IEC 601-1-1 (Safety), and CAL 117 (Flammability).

Technological Characteristics

Heavy-duty, battery-operated, four-wheel scooter. Features: digital controller, transaxle/motor, 12V 55AH batteries, electromechanical disc/regenerative brakes, 2-piece frame, rotating contour seat. Standards: RESNA WC Vol. 1/2, ANSI/RESNA WC Vol. 2-1998 Section 21, IEC 601-1-1, CAL 117.

Indications for Use

Indicated for disabled persons with limited walking capabilities requiring a motorized mobility scooter.

Regulatory Classification

Identification

A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K092088 **Pride** Mobility Products Corp. JUL 1 6 2009 Research & Development 182 Susquehanna Ave Exeter, PA 18643 570-655-5574 FAX 655-2990 www.pridemobility.com ### Exhibit 1 # 510(k) Summary Pride Mobility Products Corporation Victory Fride Mobility Victory Vicel Scooter / Model #SC713 # Submitter's Name & Address: Pride Mobility Products Corporation 182 Susquehanna Avenue Exeter, Pa. 18643 Phone: (570) 655-5574 Facsimile: (570) 655-2990 ### Contact Person: Kimberly Blake Official Correspondent Image /page/0/Picture/23 description: The image shows a solid black circle. The circle is centered on a white background. The circle is a simple geometric shape with no other features. ### Date Prepared: 6/16/09 # Name of Device and Proprietary Name: Name of Device and Prophetary Name: Victory XL 130, Four Wheel Scooter / Model # SC713/ Pride Mobility Products Corporation Common or Usual Name: Four Wheel Power Scooter Classification Name: Classincation Names Physical Medicine / Motorized Three - Wheeled Vehicle Product Code: INI # Comparison to Predicate Devices: Comparison to Predicate Devices: The Victory XL 130, Four Wheel Scooter / Model # SC713 is substantially equivalent The Victory XL 130, Four Wheel Scooter / Model #SC445, (K9 The Victory XL 130, Four Wheel SC / Model #8C445 (K944939) when comparing, to the Pride Mobility Celebrity - Life and structure - The performance characteristics to the Pride Mobility Celebrity AL 7 Model #00=10 (1. s reformance characteristics performance, maneuverability, stability, and structure, similar to achieve the performance, maneuverability, stablily, and drive mechanisms are similar to achieve the and the position of the electioniss and drive moonline on maintain optimum stablity without same intended use function that enables the user to maintain optimum (130) Four Wh same intended use function that enables me assets to maintain of AL 130, Four Wheel hindering performance. The major differences between Victory AL 130, Four Wheel hindering performance. The major unferences between between head. Scooter / Model # SC713 to the Celebrity XL SC445 (K944939) are in the Control Mechanisms. Image /page/0/Picture/21 description: The image shows a black circle on a white background. The circle is located in the upper portion of the image. The bottom of the circle is cut off by the edge of the image. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Pride Mobility Products Corp. The word "Pride" is written in a bold, stylized font, with a horizontal line running through the middle of the word. Below the word "Pride" is the text "Mobility Products Corp." in a smaller, less bold font. The logo is black and white. Research & Development 182 Susquehanna Ave Exeter, PA 18643 570-655-5574 FAX 655-2990 www.pridemobility.com ### Device Description: The Victory XL 130, Four Wheel Scooter / Model # SC713 is a Heavy Duty batteryoperated scooter having a Digital Controller, electrical system, Charger, Transaxle / Motor, Batteries, Seating and 2-Piece Frame. The Victory XL 130 Scooter is equipped with electronic regenerative and electromechanical disc brakes, off-board battery charger, removable 12 Volt 55AH batteries, and anti-tip rear wheels. The scooter has various size foldable Contour seats that are removable. The seat interface is equipped with a clover leaf design allowing for seat rotation and locking at 90° intervals mounted on stationary seat post. The seat is also equipped with sliders for fore-aft adjustment. The Victory XL 130, Four Wheel Scooter / Model # SC713 is designed with ultimate safety, stability, and performance in mind. The scooter is designed for, but not limited to Pride Mobility Products Corporation, providers/retailers and their consumers. ### Intended Use: The intended use of the Pride Mobility Products Corp. Victory XL 130, Four Wheel Scooter / Model # SC713 is to provide mobility to disabled persons having limited walking capabilities. ### Non-Clinical Testing: Compliance to applicable Testing Standards is as follows: RESNA WC Vol.1 2008 DRAFT - Requirements and Test Methods for Wheelchairs (Including Scooters) RESNA WC Vol. 2 2008 DRAFT - Additional Requirements for Wheelchairs (Including Scooters) with Electrical Systems ANSI/RESNA WC Vol. 2-1998 Section 21 - Requirements and Test Methods for Electromagnetic Compatibility. IEC 601-1-1 Medical Electrical Equipment, General Requirements for Safety CAL 117 - Flammability Testing ### Discussion of Clinical Testing Performed: N/A ### Conclusions: The Victory XL 130, Four Wheel Scooter / Model # SC713 has the same intended use and similar technological characteristics as the Celebrity XL SC445 (K944939) moreover, the non-clinical testing and the predicate comparisons demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the Victory XL, Four Wheel Scooter / Model # SC713 is substantially equivalent to the predicate device, has passed all the necessary testing, and is considered to be safe for user operation. Image /page/1/Picture/18 description: The image shows the text "2 of 2". This likely indicates that it is the second page of a two-page document. The text is written in a simple, sans-serif font. The background is plain white. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the words "Public Health Service" in a simple, bold, sans-serif font. The text is arranged on a single line. The words are clearly legible and the overall impression is clean and straightforward. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Pride Mobility Products Corporation % Ms. Kimberly Blake Official Correspondent 182 Susquehanna Avenue Exeter, Pennsylvania 18643 JUL 1 6 2009 Re: K092088 Trade/Device Name: Victory XL 130, Four Wheel Scooter/Model # SC713 Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized three-wheeled vehicle Regulatory Class: II Product Code: INI Dated: June 19, 2009 Received: July 9, 2009 Dear Ms. Blake: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ### Page 2 - Ms. Kimberly Blake If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Exhibit 3 # Indications for Use ## 510(k) Number (if known): K Device Name: Victory XL 130, Four Wheel Scooter / Model # SC713 #### Indications for Use: The intended use of the Pride Mobility Products Corporation Victory XL 130, Four The intended use of the hobility to provide mobility to disabled persons having limited walking capabilities. AND / OR Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use_ X (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division of Surgical, Orthopedic, and Restorative Devices Page 1 of 1 \$10(k) Number K092088 19(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...