The EverOn 1.0 system is intended for continuous measurement of respiration rate, heart rate and movement, in an automatic contact-less manner, at home, hospital or clinic setting. The system is indicated for use in children, adolescents and adults. The operation of the EverOn has been studied in children (weight ≥10 Kg) and adults (weight ≤111 Kg) during sleep and resting condition.
Device Story
EverOn 1.0 System provides continuous, contact-less monitoring of respiration rate, heart rate, and movement. System components include an under-mattress piezoelectric sensing unit and a bedside control unit. Sensing unit converts mechanical body movements into electrical signals; control unit processes signals to calculate parameters, log data, and trigger audible/visible alerts for threshold breaches or bed exits. Used in home, hospital, or clinic settings; operated by clinicians or patients. Output allows healthcare providers to monitor patient status during sleep or rest, facilitating timely clinical intervention and patient safety.
Clinical Evidence
Clinical testing validated performance in children (≥10 kg) and adults (≤111 kg). Accuracy compared against gold standard and predicate (Compass). Results: EverOn vs. Gold Standard (Adults 95.6% accuracy, Children 94.8% accuracy); EverOn vs. Predicate (Adults 92.3% accuracy, Children 90.7% accuracy). No adverse events reported.
Technological Characteristics
Under-mattress piezoelectric sensor; bedside control unit. Measures mechanical movement to derive respiration and heart rate. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, AAMI/ISO 14971-1, IEC 68-2-27 (shock), and IEC 68-2-6/34 (vibration).
Indications for Use
Indicated for continuous monitoring of respiration rate, heart rate, and movement in children (≥10 kg), adolescents, and adults (≤111 kg) in home, hospital, or clinic settings during sleep and rest.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
MAY 2 4 2010
# 510(K) SUMMARY EarlySense Ltd. EverOn 1.0 System
# 7.1.1 Applicant's Name:
EarlySense Ltd. 12 Tzvi st. Ramat Gan 52504, Israel Tel: +972-3-7522330 Fax: +972-3-7522340
# 7.1.2 Contact Person:
Dalia Argaman
7.1.3 Date Prepared:
April, 2010
7.1.4 Trade Name:
EverOn 1.0 System
- 7.1.5 Classification Name:
Breathing frequency monitor (868.2375)
7.1.6 Classification:
Class II; Product Code- BZQ
# 7.1.7 Predicate Devices
EarlySense Ltd. is relying on the combination of the following predicate devices for the EverOn's movement capability substantial equivalence discussion.
- . The EverOn System, cleared under K082465, breathing frequency - 868.2375, Class II; Product Code- BZQ
{1}------------------------------------------------
- Compass F10 system { MedCare) cleared under K041904 (Ventilatory . Effort Recorder, Class II - Regulation number 866.2375- MNR)
- PAM 3000 (Wireless 2000 RF & UWB Technologies Ltd) cleared under . K082626 - Heart Rate and Respiration rate monitor (monitor, cardiac (incl. cardiotachometer & rate alarm monitor, breathing frequency -870 2300 & 868 2375, Class II - DRT)
# 7.1.8 Device Description:
The EverOn 1.0 System is designed for continuous and contact-less monitoring of respiration rate, heart rate and movement. The system automatically starts measuring whenever the patient is in bed. The EverOn can provide alert notification (audible and visible) if either parameters exceed predefined thresholds set by the user. The EverOn can also provide patient out-of-bed (Bed Exit) alert notification to the user.
The EverOn System consists of the following main components:
- A Sensing Unit placed under the mattress or mattress pad. 다
- A Control Unit (Bedside Unit).
- . Proprietary recording and data analysis software
The under mattress Sensing Unit includes a piezoelectric sensor, which converts mechanical movements into an electric signal. The Control Unit receives the electric signals, processes them and finally calculates, logs, displays the patient's parameters, and generates alerts as per set thresholds. when needed.
## 7.1.9 Intended Use:
The EverOn 1.0 system is intended for continuous measurement of respiration rate, heart rate and movement, in an automatic contact-less manner, at home, hospital or clinic setting. The system is indicated for use in children, adolescents and adults. The operation of the EverOn has been studied in children (weight ≥10 Kg) and adults (weight ≤111 Kg) during sleep and resting condition.
#### 7.1.10 Performance Standards:
The EverOn 1.0 System complies with voluntary standards such as IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, AAMI / ISO 14971-1. IEC 68-2-27: Shock; IEC 68-2-6: Sinusoidal Vibration ; IEC 68-2-34 Random Vibration (Wide band)
{2}------------------------------------------------
# 7.1.12 Performance Data & Substantial Equivalence
The EverOn 1.0 System is substantially equivalent in all aspects, e.g., technological characteristics, mode of operation (contact-less monitoring), performance characteristics (accuracy), intended use, etc., to the EverOn system cleared under K082465.
In addition EverOn 1.0 is substantially equivalent to its other predicates (Compass F10 system and PAM 3000 system) due to the following reasons: Both the EverOn and its predicate devices are intended for the purpose of detecting and recording physiological signals. The specific movement indication claimed by the EverOn 1.0 system is among other indications claimed by the predicates. Similar environment of use is claimed for both the EverOn and the PAM 3000 system. Both are indicated for use in the general healthcare market. The general principles of operation are also similar for both the EverOn system and its predicates in the way that all systems detect and convert physiological parameters and the body motion signals into electrical signals using sensing elements. EverOn is also similar to its predicate PAM 3000, since both systems accomplish the continuous measurement of the motion in a contact free manner. In addition the equivalence of the EverOn to its predicates regarding its capability to safely and effectively detect movement was demonstrated through a set of validation testing whereas one of the predicate was the reference device, as well as by a set of Software validation tests.
# 7.1.12.1 Non-Clinical Testing
In-order to demonstrate the equivalence of the EverOn 1.0 to its predicates and to demonstrate that its performance meets the requirements of its predefined acceptance criteria and its intended use the EverOn 1.0 System was subjected to the following tests:
- Electrical safety and Electromagnetic Compatibility tests
- 트 Software verification and validation tests, to ensure the system software meet all software requirement specifications.
- . Mechanical Vibration (including random and sinusoidal vibration) and Shock resistance
- 트 Environmental tests (including operating and storage temperature and humidity conditions)
{3}------------------------------------------------
- Risk analysis activities were performed in compliance with the . requirements of ISO 14971-1 "Application of risk management to medical devices" (2007).
The results of the testing indicated that the EverOn 1.0 performs according to its specifications.
# 7.1.12.1 Clinical Testing
The Clinical testing of the EverOn to validate its performance was performed among both children and adult populations. The system's accuracy and its capability to detect motion were compared to two reference methods: Predicate and Gold Standard.
Accuracy of the EverOn in detecting and displaying movement vs. the gold standard and the predicate are as following:
| EverOn vs. Gold Standard | |
|--------------------------------|-------|
| Overall Accuracy - Adults | 95.6% |
| Overall Accuracy - Children | 94.8% |
| EverOn vs. Predicate (Compass) | |
| Overall Accuracy - Adults | 92.3% |
| Overall Accuracy - Children | 90.7% |
No Adverse events or complications of any sort were observed during the clinical testing.
## 7.1.12 Conclusions
. . .
Based on the design verification and validation processes, non-clinical and the clinical testing performed it was concluded that the EverOn 1.0 System, is substantially equivalent to its predicates, without raising new safety and/or effectiveness issues.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with a bird-like figure in the center. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Dalia Argaman VP, Clinical and Regulatory Affairs EarlySense Limited TZVI 12 Ramat - Gan Israel 52504
MAY 2 4 2010
Re: K092062
Trade/Device Name: EverOn 1.0 System Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: BZQ, DRT Dated: April 22, 2010 Received: April 22, 2010
Dear Ms. Argaman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Ms. Dalia Argaman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS),regulation (21 CER Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Min for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known):_
Device Name: EverOn_1.0 System
Comments of the country of the country of the country of
#### : : : : : : : : : : : : Indications for Use:
The EverOn 1.0 system is intended for continuous measurement of respiration rate, heart rate and movement, in an automatic contact-less manner, at home, hospital or clinic setting. The system is indicated for use in children, adolescents and adults. The operation of the EverOn has been studied in children (weight ≥10 Kg) and adults (weight ≤111 Kg) during sleep and resting condition.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 C.F.R. 801 Subpart D)
510(k) Number:
AND/OR
Over-The-Counter Use (Part 21 C.F.R. 807 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) 510(k) Number: Intection Control, Dental Devices Distinctions in Supersional to moisiviO (DO-upis noisivio) U at (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 7-7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.