← Product Code [FXX](/productcode/FXX) · K092012

# PRIMAGARD DELTA FIT 80,120, 160  SURGICAL FACE MASK (SMALL, MEDIUM, LARGE) MODEL PG4-7061,-7071,-7081, PG4-7062, 7072-70 (K092012)

_Primed Medical Products, Inc. · FXX · Sep 25, 2009 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K092012

## Device Facts

- **Applicant:** Primed Medical Products, Inc.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Sep 25, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

PRIMAGARD™ Delta Surgical Masks are surgical apparel as identified in 21 CFR 878.4040 and are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

## Device Story

PRIMAGARD Delta Surgical Mask is a deltoid-shaped, waterfall-pleated surgical mask. Device is worn by operating room personnel during surgical procedures. Function is to provide a barrier against microorganisms, body fluids, and particulate material transfer between the patient and the wearer. Device utilizes non-woven materials (polypropylene and wet-laid cellulose) to balance breathability and filtration. Elastomeric non-woven material is used for piping and headbands to secure the mask. Device is a passive barrier; no electronic or mechanical components.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Deltoid-shaped, waterfall-pleated surgical mask. Materials: non-woven polypropylene and wet-laid cellulose. Elastomeric non-woven material used for piping and headbands. Passive barrier device; no energy source, software, or connectivity.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Mask ([K081633](/device/K081633.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KO9 2012

# 510 (k) Summar

| Submitted by:        | priMED<br>#900, 10707-100 Ave.<br>Edmonton, Alberta<br>780-497-7600 |
|----------------------|---------------------------------------------------------------------|
| Contact Person:      | Dr. Raymond G. Marusyk                                              |
| Date Prepared:       | June 10, 2009                                                       |
| Proprietary Name:    | PRIMAGARD DELTA Surgical Mask                                       |
| Common Name:         | Surgical Mask                                                       |
| Classification Name: | Surgical Mask, 878.4040; Product code FXX                           |

Predicate Device:

· Surgical Mask 510(k) #K081633

Description of the Device: These surgical masks are deltoid shape, waterfall-pleated devices manufactured from selected non-woven materials (polypropylene and wet-laid cellulose) designed to provide optimal breathability, particulate filtration and a fluid-penetration barrier relative to the degree of protection required during intended use.

Intended Use of the Device: PRIMAGARD™ Delta Surgical Masks are surgical apparel as identified in 21 CFR 878.4040 and are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

Technological Characteristics: The PRIMAGARD™ Delta Surgical Masks are technologically similar to the predicate device (K081633) in that both devices consist of non-woven barrier materials selected and arranged in such a manner as to provide, at the time of design, and under the conditions of use, optimal breathability and particulate filtration. The PRIMAGARD™ Delta Masks incorporate a deltoid design and include an elastomeric non-woven material as piping and headband

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the symbol. The logo is black and white.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Mr. Raymond G. Marusyk Director, Design and Development PriMED Medical Products, Incorporated #900, 10707-100 Avenue Edmonton, Alberta CANADA T5J-3M1

SEP 2 5 2009

Re: K092012

Trade/Device Name: PRIMAGARD Delta Surgical Masks Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: September 15, 2009 Received: September 17, 2009

Dear Mr. Marusyk: .

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

#### Page 2- Mr. Marusyk

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to .

http://www.fda.gov/MedicalDevices/Safety/Reporta Problem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm,

Sincerely yours,

Century 29. Nov. 62
Susan Reynolds R.R.S. M.A.

Susan Runner, D.D.S., M.A Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{3}------------------------------------------------

## Indications for Use Statement

K092012

S10(k) Number (if known)

Device Name PRIMAGARD Delta Surgical Masks

Indications PRIMAGARD Delta Surgical Masks are surgical apparel as identified in 21 for Use CFR 878.4040 and are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particloste material.

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801. 109) OR

Over-The-Counter Use

Lahip Rangubi

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K092012

---

**Source:** [https://fda-staging.innolitics.com/device/K092012](https://fda-staging.innolitics.com/device/K092012)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
