DK50 DS

K091871 · Ekom S.R.O. · BTI · Oct 23, 2009 · Anesthesiology

Device Facts

Record IDK091871
Device NameDK50 DS
ApplicantEkom S.R.O.
Product CodeBTI · Anesthesiology
Decision DateOct 23, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.6250
Device ClassClass 2

Indications for Use

Indicated for supplying compressed air for medical ventilators.

Device Story

DK50 DS is a portable, oil-free piston air compressor for medical ventilators. It operates as a primary or standby air source. In standby, it monitors central facility air pressure; if pressure drops, it automatically activates. It features a single-phase electric motor, double-stage air filtration, and a pressure regulator to maintain constant output. The device includes an integrated air tank (2L) and provides indicators/alarms for output pressure, operation hours, power status, drying efficiency, and battery condition. Alarms trigger for high temperature (>80°C), low output pressure (<2.1 bar), and power loss. Used in hospitals or sub-acute institutions, it ensures continuous air supply to respiratory equipment, enhancing patient safety during potential central air distribution failures.

Clinical Evidence

No clinical data. Substantial equivalence is based on bench testing and technical comparison of specifications, including output flow, pressure regulation, alarm functionality, and environmental performance.

Technological Characteristics

Oil-free piston compressor; single-phase electric motor; double-stage 5-micron air filtration; automatic condensed water separation; pressure regulator; 2L air tank. Output: 40 Lpm (max), 200 Lpm (peak). Pressure range: 5-6.5 bar. Connectivity: DISS outlet connection. Alarms: Acoustic/optical for high temperature (>80°C), low pressure (<2.1 bar), and power loss. Dimensions: 510 x 480 x 470 mm. Weight: 30 kg. Continuous (S1) operation mode.

Indications for Use

Indicated for supplying compressed air for medical ventilators in hospital or sub-acute institutional settings.

Regulatory Classification

Identification

A portable air compressor is a device intended to provide compressed air for medical purposes, e.g., to drive ventilators and other respiratory devices.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K091871 ### 510(k) Summary Page 1 of 3 12-Aug-09 OCT 28 2009 EKOM spol s.r.o. Priemyselna 5031/18 921 01 Piestany SLOVAKIA Tel - +011 421 33 7967246 Fax - +011 421 33 7967223 | Official Contact: | Ernest Koppon - Quality and Regulatory Manager | |----------------------------|-------------------------------------------------------------| | Proprietary or Trade Name: | DK50 DS | | Common/Usual Name: | Portable air compressor | | Classification Name/Code: | BTI - portable air compressor<br>21 CFR 868.6250 | | Device: | DK50 DS | | Predicate Devices: | K060781 - Ekom Model DK50 D<br>K041406 - Newport Model C250 | #### Device Description: The Ekom model DK50 DS is a portable air compressor designed to supply air to medical equipment, i.e., ventilators. The Model DK50 DS contain an oil-free piston compressor driven by a low maintenance single phase electric motor. Compressed air is cooled where condensed water is separated into a vessel. Incoming air passes through 2 filters and undergoes double filtration as it passes through the system. Constant output pressure is maintained by a pressure regulator. There is a built-in air tank which allows peak air consumption of 200 Lpm. The device may be used as a standby source of air. In this configuration, the respiratory equipment is supplied with compressed air from the facilities central air distribution. Air pressure in this central distribution is sensed by the Ekom pressure sensor. If the pressure is sufficient, the compressor stays in the STANDBY mode. If the pressure falls the compressor automatically starts and becomes the main air supply. When the compressor is used as the main source, the control unit determines its operation according to the current need for air. If air consumption is zero, the device switches to STANDBY. The compressor is equipped with indicators for output pressure, operation hours, power status, drying efficiency and battery condition as well as acoustic and optic alarms to warn of high operating temperature, low output pressure and loss power. {1}------------------------------------------------ #### 510(k) Summary Page 2 of 3 12-Aug-09 #### Indications for Use: Indicated for supplying compressed air for medical ventilators. # Environment of Use: Hospital or sub-acute institution # Summary of substantial equivalence: | Model Name: | Modified device<br>DK50 DS | Unmodified device<br>Predicate - K060781<br>DK50 D | |-------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------| | Specification | | | | Output flow | Max. 40 Lpm | Max. 40 Lpm | | Peak Flow | 200 Lpm for 2 sec | 200 Lpm for 2 sec | | Power | 120V /60 Hz | 120V /60 Hz | | Nominal current (amps) | 5.6 A | 5.6 A | | Air filtration | 5 micron | 5 micron | | Pressure dew point @ 40 Lpm 20°C | 5°C below ambient temperature | 5°C below ambient temperature | | Outlet connection | DISS | DISS | | Sound level | < 51 dB(A) | < 51 dB(A) | | Mode of operation | Continuous - S1 | Continuous - S1 | | Separation of condensed water | Automatic | Automatic | | Operating pressure of safety valve | 8 bar (116 psig) | 8 bar (116 psig) | | Adjustment of pressure output | Pressure regulator | Pressure regulator | | Alarm for cooling failure / high<br>temperature | Acoustic and optical if increase in internal<br>temperature > 80°C (176°F) | Acoustic and optical if increase in<br>internal temperature > 80°C (176°F) | | Automatic turn-on pressure | When central distribution pressure < 2.8<br>bar (40.6 psig) | When central distribution pressure < 2.8 bar (40.6 psig) | | Output pressure | 3 bar (43.5 psig) | 3 bar (43.5 psig) | | Output pressure indicator | Display | Pressure gauge | | Alarm - loss of power | Yes | No, but instructions require<br>connection to equipment with this<br>alarm | | Indication of drying | LEDs | Pressure gauge | | Alarm for low pressure | Internal when output pressure < 2.1 bar (30<br>psig) | None instructions require<br>connection to equipment with this<br>alarm | | Air tank capacity | 2L | 5L | | Pressure range | 5 to 6.5 bar (72.5 - 94 psig) | 5 to 7 bar (72.5 – 101.5 psig) | | Dimensions (mm) | 510 x 480 x 470 mm (17.5 x 14 x 17 in.) | 500 x 530 x 870mm<br>16 x 17 x 30.5 in | | Weight | 30 kg (66 lbs) | 46 kg (101 lbs) | # Table 1 – Table of the Similarities and Modifications from Predicate vs. new Model {2}------------------------------------------------ ### 510(k) Summary Page 3 of 3 12-Aug-09 | Model Name: | New device<br>DK50 DS | Predicate - K041406<br>Newport Model 250C | |------------------------|-------------------------------------------------------|-------------------------------------------------| | Specification | | | | Alarm - loss of power | Yes | Red light and audible alarm<br>when no power | | Alarm for low pressure | Triggers when output pressure<br>< 2.1 bar ((30 psig) | Triggers at 30 + 2 psig<br>(2.0 $ ± $ 0.13 bar) | ### Table 2. Comparison of Alarms between Predicate and Proposed Model It is our view that there are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate device. #### Substantial Equivalence The Ekom DK50 DS is viewed as substantially equivalent to the predicate devices because: #### Indications - - Identical to predicate K060781 Ekom DK50 D . #### Technology - - Similar technology used -. - K060781 Ekom DK50 D o - o K041406 -- Newport Model C250 for alarms #### Materials - - The materials in patient contact are identical to predicate device, K060781 Ekom . DK50 D. #### Environment of Use - - Identical to predicate K060781 Ekom DK50 D � {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, rendered in black. The overall design is simple and conveys a sense of official government authority. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 Ekmo S.R.O. C/O Mr. Paul E. Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134-2958 OCT 2 3 2009 Re: K091871 Trade/Device Name: Model DK50 DS Regulation Number: 21 CFR 868.6250 Regulation Name: Portable Air Compressor Regulatory Class: II Product Code: BTI Dated: September 22, 2009 Received: September 23, 2009 Dear Mr. Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2 - Mr. Dryden Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Wh for Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {5}------------------------------------------------ #### Indications for Use Statement Page 1 of 1 510(k) Number: K091871 (To be assigned) Device Name: Model DK50 DS Indications for Use: Indicated for supplying compressed air for medical ventilators. Prescription Use XX (Part 21 CFR 801 Subpart D) or Over-the-counter use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Schulte (Division Sign-Off) (Division Sign-chi) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K091871
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...