# STERILE, POWDER FREE NITRILE EXAMINATION GLOVE, BLUE, TESTED FOR USE WITH CHEMOTHERAPY DRUGS (K091851)

_Gx Corporation Sdn Bhd · LZA · Dec 28, 2009 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K091851

## Device Facts

- **Applicant:** Gx Corporation Sdn Bhd
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Dec 28, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

This glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This glove has been tested for use with specific chemotherapy drugs listed below.

## Device Story

Sterile, powder-free nitrile examination gloves; blue color. Designed for use by healthcare professionals during patient examinations to provide a barrier against contamination. Tested for permeation resistance against specific chemotherapy agents (e.g., Dacarbazine, Mitomycin C, Cyclophosphamide, Doxorubicin, 5-Fluorouracil, Cisplatin, Etoposide, Paclitaxel, Vincristine, Methotrexate, Mitoxantrone, Ifosfamide). Device provides physical barrier protection; breakthrough detection times measured for chemotherapy agents. Warning provided against use with Thiotepa and Carmustine due to insufficient breakthrough times. Used in clinical settings to protect both patient and examiner.

## Clinical Evidence

Bench testing only. Breakthrough detection times for various chemotherapy drugs were measured and reported in minutes.

## Technological Characteristics

Nitrile material; powder-free; sterile; blue color. Form factor is a disposable hand-worn examination glove. No electronic or software components.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name arranged in a circular pattern around the perimeter. In the center is a stylized caduceus symbol, which is often associated with healthcare and medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Mr. Cho Sow Fong Quality Assurance Manager GX Corporation SDN BHD Lot 6491-A, Batu 5%, Sementa, Off Jalan Kapar, 42100 Klang Selangor Darul Ehsan MALAYSIA

DEC 2 8 2009

Re: K091851

Trade/Device Name: Sterile, Powder Free Nitrile Examination Gloves, Blue Tested for with Chemotherapy Drugs Regulation Number: 880.6250 Regulation Name: Patient Examination Glove ... Regulatory Class: 1 Product Code: LZA, LZC Dated: December 15, 2009 Received: December 17, 2009

Dear Mr. Fong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2- Mr. Fong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony O. Rutan

Anthony D. Watson, BS, MS, MBA Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known): K091851

Device Name: Sterile, Powder Free Nitrile Examination Gloves, Blue Tested For Use With Chemotherapy Drugs.

Indication For Use:

This glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This glove has been tested for use with specific chemotherapy drugs listed below.

Chemotherapy Drug Permeation (Breakthrough Detection Time in Minutes)

| Dacarbazine (DTIC) (10.0 mg/mL)         |
|-----------------------------------------|
| Mitomycin C (5.0 mg/mL)                 |
| Carmustine (BCNU) (3.3 mg/mL)           |
| Cyclophosphamide (Cytoxan) (20.0 mg/mL) |
| Doxorubicin Hydrochloride (2.0 mg/mL)   |
| 5-Fluorouracil (50.0 mg/mL)             |
| Cisplatin (1.0 mg/mL)                   |
| Etoposide (20.0 mg/mL)                  |
| Paclitaxel (taxol) (6.0 mg/mL)          |
| Thio-Tepa (10.0 mg/mL)                  |
| Vincristine Sulfate (50.0 mg/mL)        |
| Methotrexate (25.0 mg/mL)               |
| Mitoxantrone (100.0 mg/mL)              |
| Ifosfamide (Ifex) (50.0 mg/mL)          |

Warning: Do Not Use with Thiotepa and Carmustine

- Prescription Use -(Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ (21 CFR 801 Subpart C)

> 240 > 240 3.57 > 240 > 240 > 240 > 240 > 240 > 240 24.39 > 240 > 240 > 240 > 240

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE (F NEEDED)

AND/OR

Concurrence of CDRH, Office Of Device Evaluation (ODE)

Elizabet P. Clauss-Will

(Division Sign-Off) (Division of Anesthesiology, General Heapital Infection Control, Dental Devices

Page 1 of

510(k) Number: K09183

---

**Source:** [https://fda-staging.innolitics.com/device/K091851](https://fda-staging.innolitics.com/device/K091851)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).