← Product Code [OCW](/productcode/OCW) · K091832

# ENDOGASTRIC SOLUTIONS STOMAPHYX SYSTEM WITH SEROSAFUSE FASTENER (K091832)

_Endogastric Solutions, Inc. · OCW · Jul 22, 2009 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K091832

## Device Facts

- **Applicant:** Endogastric Solutions, Inc.
- **Product Code:** [OCW](/productcode/OCW.md)
- **Decision Date:** Jul 22, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The EndoGastric Solutions, StomaphyX System with SerosaFuse Fastener is intended for transoral tissue approximation, ligation and full-thickness plication in the G.I. tract.

## Device Story

The StomaphyX system is a flexible, single-use endoscopic delivery device used for tissue approximation in the GI tract. It utilizes vacuum suction to invaginate tissue into a distal chamber. The device features a fastener delivery subsystem consisting of a stylet (sharpened wire), a pusher tube, and internal lumens. The operator loads a sterile polypropylene fastener into a proximal cartridge, which snaps onto the stylet. Upon manual actuation of a handle/knob mechanism, the stylet pierces the tissue, and the pusher deploys the H-shaped fastener to secure the plication. The device is operated by a physician under endoscopic visualization. It enables transoral tissue ligation and full-thickness plication, providing a minimally invasive alternative to traditional surgical tissue approximation.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on design, performance, and biocompatibility bench testing.

## Technological Characteristics

Flexible, ergonomic fastener delivery device; sterile polypropylene fasteners; stainless steel stylet; vacuum-assisted tissue invagination; manual handle/knob actuation; single-use; sterile packaging.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- EndoGastric Solutions StomaphyX Device and Accessories ([K062875](/device/K062875.md))
- EndoGastric Solutions StomaphyX Device and Accessories ([K073644](/device/K073644.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### SECTION 12. 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number: _K091832

JUL 2 2 2009

### Applicant Information:

Date Prepared: June 5, 2009

Name: EndoGastric Solutions, Inc. Address: 8210 154" Avenue N.E. Redmond, WA 98052 Phone: 425 307 9200 Fax: 425 307 9201

| Contact Person:   | Steve Hoffman        |
|-------------------|----------------------|
| Phone Number:     | Office: 425-307-9226 |
| Facsimile Number: | (425) 307-9201       |

#### Device Information:

| Classification:      | Class II                                                                        |
|----------------------|---------------------------------------------------------------------------------|
| Trade Name:          | EndoGastric Solutions StomaphyX® Delivery Device, Fasteners and<br>Accessories  |
| Common Name:         | Endoscopic Tissue Approximation Device, Implantable Fastener and<br>Accessories |
| Classification Name: | Endoscope and Accessories (21 CFR 876.1500, Product Code OCW)                   |

#### Predicate Devices:

The EndoGastric Solutions StomaphyX Device and Implantable Fasteners is substantially equivalent in intended use and method of operation to a combination of the following predicate device:

K062875 - EndoGastric Solutions StomaphyX Device and Accessories K073644 - EndoGastric Solutions StomaphyX Device and Accessories

## Device Description:

The EndoGastric Solutions StomaphyX Delivery System and Implantable Fasteners consist of an ergonomic, flexible fastener delivery device and sterile polypropylene fastener implants delivered through a fastener cartridge. The unit is provided sterile and is a single use device. The polypropylene fasteners are proprietary and function only with the StomaphyX device,

EndoGastric Solutions, Inc. Confidential

{1}------------------------------------------------

The device uses vacuum to invaginate tissue through a port into a chamber and fastens it using Hshaped polypropylene fasteners.

The fastener delivery subsystem is comprised of 3 elements: stylet, pusher, and internal lumens. The three elements run the length of the device, inside the outer shaft. The pusher is a hollow tube that rides over the length of the stylet provides a "pilot" hole for the fastener to pass through the tissue and both ride inside the lumen. The stylet is a wire three-faceted, sharpened tip at the distal end to pierce tissue.

A fastener is loaded for deployment by depressing the lever on the removable, fastener cartridge located at the proximal end of the device. This action automatically snaps the fastener onto the stylet. When the pusher tube is pushed by the operator, the stylet carnes the fastener down the lumen which runs from the proximal handle assembly to the distal tissue port where it will eventually be deployed into the tissue.

#### Indications for Use:

The EndoGastric Solutions StomaphyX® system with SerosaFuse® Fastener is intended for transoral tissue approximation, ligation and full-thickness plications within the G.I. tract.

#### Comparison to Predicate Device:

The design of the EndoGastric Solutions StomaphyX System with SerosaFuse® Fastener is identical to the predicate listed in that they are devices designed to reach the desired suture location under endoscopic visualization, grasp tissue in some fashion and place sutures/fasteners in a desired location. The products are re-loadable for repeat fastener/suture placement. The products share common features such as a sterile, stainless steel needle called a stylet, housed in a fastener loading unit. They deliver a fastener/suture through soft tissue by manually actuating the needle with a handle/knob mechanism. The devices are packaged sterile and are for single patient use. Further, the EndoGastric Solutions StomaphyX System with SerosaFuse Fastener and the predicate device have the same intended use, which is to place fasteners/sutures/clips to approximate soft tissue under endoscopic visualization.

#### Summary:

Based upon the intended use, product technical information, performance and biocompatibility information provided in this pre-market notification, the EndoGastric Solutions StomaphyX system with SerosaFuse Fastener is substantially equivalent to the currently marketed predicate device in terms of design, performance and intended use.

46

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare. The caduceus in the logo is a modern, abstract design.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Steven J. Hoffman Director, Ouality & Regulatory Affairs EndoGastric Solutions, Inc. 8210 154th Avenue, N.E. REDMOND WA 98052

JUL 222009

Re: K091832

Trade/Device Name: EndoGastric Solutions StomaphyX Device with SerosaFuse Fastener Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCW Dated: June 18, 2009 Received: June 22, 2009

Dear Mr. Hoffman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical

{3}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours

adde

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Ericlosure

{4}------------------------------------------------

# 7.1 Indications for Use

### 510(k) Number (if known): K091832

Device Name: EndoGastric Solutions StomaphyX Device with SerosaFuse Fastener

#### Indications For Use:

The EndoGastric Solutions, StomaphyX System with SerosaFuse Fastener is intended for transoral tissue approximation, ligation and full-thickness plication in the G.I. tract.

Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Reproductive, Abdominal,
and Radiological Devices

510(k) Number _

EndoGastric Solutions, Inc. Confidential

14

---

**Source:** [https://fda-staging.innolitics.com/device/K091832](https://fda-staging.innolitics.com/device/K091832)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).