FRONTIER DEVICES CRANIOMAXILLOFACIAL, CRANIO CLOSURE, ORTHOGNATHIC, MANDIBLE RECONSTRUCTION AND MESH SYSTEMS
K091812 · Frontier Devices, Inc. · JEY · May 7, 2010 · Dental
Device Facts
Record ID
K091812
Device Name
FRONTIER DEVICES CRANIOMAXILLOFACIAL, CRANIO CLOSURE, ORTHOGNATHIC, MANDIBLE RECONSTRUCTION AND MESH SYSTEMS
Applicant
Frontier Devices, Inc.
Product Code
JEY · Dental
Decision Date
May 7, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Frontier Devices Maxillofacial System is intended for use in selective trauma of the mid-face and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin. Frontier Devices Orthognathic System is intended for use in selective trauma of the mid-face and maxillofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin. The Frontier Devices Mandible Reconstructive System is intended for use in selective trauma of the mid-face and maxillofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin. Frontier Devices Mesh System is intended for use in selective trauma of the mid-face and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
Device Story
System comprises titanium bone plates and screws for maxillofacial/orthognathic surgery. Plates (unalloyed titanium, ASTM F67) and screws (titanium alloy, ASTM F136) provide rigid fixation for bone fractures, reconstruction, and orthognathic procedures. Screws are self-tapping or self-drilling. Surgeon selects plate/mesh size and shape based on anatomical site (mandible, maxilla, zygomatic, orbital, sinus) and defect severity. Devices implanted during surgery to stabilize bone segments; facilitate healing; restore facial structure. No electronic components or software.
Clinical Evidence
Bench testing only; no clinical data provided. Performance tests on representative screws demonstrated results similar to predicate devices.
Technological Characteristics
Materials: Unalloyed titanium (ASTM F67) for plates/meshes; Titanium 6Al 4V alloy (ASTM F136) for screws. Plate thickness: 0.4–1.7 mm. Screw diameter: 1.5–3.0 mm. Mechanical fixation via plates and screws. No energy source or software.
Indications for Use
Indicated for patients requiring fixation or reconstruction of the mid-face, maxillofacial skeleton, mandible, or maxilla due to trauma, tumor resection, infection, or orthognathic surgery. Includes use for comminuted fractures, bone loss, and contour irregularities.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
BioPlate Rigid Fixation Bone Plating System for Craniomaxillofacial Surgery (K030806)
Synthes Craniofacial Plate and Screw System (K050608)
Submission Summary (Full Text)
{0}------------------------------------------------
### Revised 510K Summary . K091812
# MAY - 7 2010
[JEY] (DZL) .
The following information is provided as required by 21 CFR 807.87 for the 510(k) premarket notification for five The following mormation is provided as required of all of the one of the mathic System and Frontier Devices Mandible Reconstruction System and Mesh System.
Date Prepared:
## May 6, 2010
Frontier Devices
Sponsor :
| | 153A Cahaba Valley Parkway |
|------------------------------------|---------------------------------------------------|
| | Pelham, AL 35124 |
| FDA Registration #: | 3006803588 |
| Contact Person: | Don Petersen, Ph.D. |
| | Director Quality Assurance and Regulatory Affairs |
| | 205-733-0901 (phone) |
| | 205-733-8445 (fax) |
| | dpetersen@frontierdevices.com |
| Proprietary Names: | Frontier Devices Maxillofacial System |
| | Frontier Devices Orthognathic System |
| | Frontier Devices Mandible Reconstruction System |
| | Frontier Devices Mesh System |
| Common Names: | Bone plates and screws |
| Classification Names: | Bone plate (872.4760) |
| (Regulation Number) [Product Code] | Intraosseous fixation screw (872.4880) |
| Device Classification: | Class II |
|------------------------|----------|
| Panel: | Dental |
#### Predicate Devices
Frontier Devices Maxillofacial System BioPlate Rigid Fixation Bone Plating System for Craniomaxillofacial Surgery (K030806)
Frontier Devices K091812 - 510(k) Response Letter
## Appendix D Page D-1
{1}------------------------------------------------
## Frontier Devices Orthognathic System Ortrautek Orthognathic System (K031989)
Frontier Devices Mandible Reconstruction System BioPlate Mandible Fixation System (K012910)
Frontier Devices Mesh System Synthes Craniofacial Plate and Screw System (K050608)
#### Device Descriptions
#### Frontier Devices Maxillofacial System
The devices in the Maxillofacial System are implantable bone screws for oral and maxillofacial procedures including fractures, orthognathic, reconstructive procedures and revisions procedures when other treatments of devices have failed.
Surgical locations for device placement includes fracture and reconstruction sites in the mandible, maxilla, zycomatic bone and orbital socket.
Plates range in shape and size to aid the surgeon in repairing bone. Plate shapes include Oblique I, T, Curve, Straight, Y, Double Y, Box, X, H and Z. The sizes for the plates range from 0.4 to 1.0 mm thick and up to 100 mm in length. Plates are made from medical grade unalloyed titanium (commercially pure Titanium grade 4, ASTM F-67).
The screws are made from alloyed titanium (Titanium 6Al 4V, ASTM F-136). The screws have two designs, a selftapping screw that is to be used after pre-drilling pilot holes and a self-drilling screw that does not require a predrilled hole. Screws have diameters ranging from 1.5 to 2.3 mm and lengths from 3 to 24 mm.
#### Frontier Devices Orthognathic System
The Frontier Devices Orthognathic System consists of titanium bone plates and screws of various shapes and sizes. The plates are made from unalloyed titanium that conforms to the ASTM F67 standard. Plates range in shape and size to aid the surgeon in repairing the maxilla and mandible. Plate shapes include L-Tooth, L-Zigomatic, and Straight. The sizes for the plates range from 0.8 to 1.0 mm thick and up to 42 mm in length.
The screws are self-tapping screws that require a pre-drilled pilot hole and are made from a titanium alloy, Ti-6Al-4V, that conforms to the ASTM F136 standard. Screws are 2.0 to 2.7 mm in diameter and range in length from 6 to 22 mm.
This system provides a wide variety of specific screws and plates intended for use for maxillofacial trauma, maxillofacial, orbital reconstruction and specialized orthognathic surgery. The indications include mandible and maxilla trauma, reconstruction of the mandible or maxillary fixation and orthognathic surgery.
Surgical locations for device placement includes fracture and reconstruction sites in the mandle.
{2}------------------------------------------------
### Frontier Devices Mandible Reconstruction System
The Frontier Devices Mandible Reconstruction System consists of titanium bone plates and screws of various shapes and sizes intended for use for maxillofacial, orbital reconstruction and specialized orthognathic surgery. The indications include mandible trauma, such as, severe comminution and bone loss and unstable mandible fractures; mandible reconstruction such as, bridging defects after tumor resection or severe infection; intermaxillary fixation, and orthognathic surgery.
Surgical locations for device placement includes fracture and reconstruction sites in the mandible.
The plates are made from unalloyed titanium that conforms to the ASTM F67 standard. Plates are 1.25 to 1.70 mm thick and up to 56 mm in length. The plates come in a variety of shapes and sizes to aid the surgeon in repairing the maxilla and mandible. Plate shapes include straight and angled.
The screws are self-tapping screws that require a pre-drilled pilot hole and are made from a titanium alloy, Ti-6Al-4V, that conforms to the ASTM F136 standard. Screws are 2.3 to 3.0 mm in diameter and range in length from 5 to 18 mm.
### Frontier Devices Mesh System
The Frontier Devices Mesh System consists of titanium meshes and screws of various shapes and sizes. Reconstruction of maxillofacial defects can be carried out with titanium mesh to reconstruct the missing framework of the maxillofacial area. The defects can be caused by acute trauma, tumor removal, and previous operations. The Mesh System is indicated for immediate reconstruction in the primary treatment of comminuted fractures with bone loss in non load-bearing areas, treatment of contour irregularities (possibly in combination with bone or cartilage grafts), and applicability for orbital and sinus defects. The desired strength, contour ability and desired location are all factors that determine which size and thickness of mesh the surgeon selects.
Surgical locations for device placement includes fracture and reconstruction sites in the mandible, maxilla, orbital and sinus regions.
The meshes are made from unalloyed titanium that conforms to the ASTM F67 standard. Meshes are 0.45, 0.60 and 0.867 mm thick and come in a variety of shapes and sizes that are designed for mesh applications. The 0.4 mm thick mesh is less rigid and more contour-able and intended for small voids/defects. The 0.6 mesh is more rigid and is intended for larger voids/defects. The 0.87 mm mesh offers the greatest strength and is intended for larger defects.
The screws are self-drilling screws that do not require a pre-drilled pilot hole and are made from a titanium alloy, Ti-6Al-4V, that conforms to the ASTM F136 standard. Screws are 1.5 to 2.3 mm in diameter and range in length from 3 to 6 mm.
## Intended Use:
## Frontier Devices Maxillofacial System
Frontier Devices Maxillofacial System is intended for use in selective trauma of the mid-facial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
{3}------------------------------------------------
#### Frontier Devices Orthognathic System
Frontier Devices Orthognathic System is intended for use in selective trauma of the mid-facial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
#### Frontier Devices Mandible Reconstructive System
The Frontier Devices Mandible Reconstructive System is intended for use in selective trauma of the mid-face and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
#### Frontier Devices Mesh System
Frontier Devices Mesh System is intended for use in selective trauma of the mid-face and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
#### Technology Characteristics
The plates and screws of the five different Frontier Devices systems have similar dimensions, designs and materials as the predicate devices The plates are between 0.4 and 1.7 mm thick, made from unalloyed titanium (ASTM F67) and have similar design shapes and sizes as the predicate devices. The screws are all made from titanium alloy (ASTM F136) with diameters between 1.5 and 3.0 mm and lengths of 3-22mm which are similar to the predicate devices.
#### Performance Tests
Performance tests on representative screws showed them to have similar results as predicate screws.
#### Substantial Equivalence
The Frontier Devices Maxillofacial, Orthognathic, Mandible Reconstruction and Mesh Systems consist of plates and screws that are identical in material composition and have the same indications for use as the relevant predicate devices. All of the technical characteristics are substantially equivalent to the corresponding characteristics of the predicate devices. There may be slight differences in dimensions and shapes between the Frontier Devices Systems and the predicate devices, however, these minor differences raise no new issues of safety and efficacy of the device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three bars forming its body and wings. The eagle is oriented diagonally, facing upwards and to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Don Peterson, Ph.D. Director of Quality Assurance and Regulatory Affairs Frontier Devices, Incorporated 153A Cahaba Valley Parkway Pelham, Alabama 35124
MAY - 7 2010
Re: K091812
Trade/Device Name: Frontier Devices Maxillofacial System Frontier Devices Orthognathic System Frontier Devices Mandible Reconstruction System Frontier Devices Mesh System
Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY, DZL Dated: April 30, 2010 Received: May 4, 2010
Dear Dr. Peterson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2- Dr. Peterson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.ida.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K091812
Device Name: Frontier Devices Maxillofacial System
Indications for Use:
Frontier Devices Maxillofacial System is intended for use in selective trauma of the mid-face and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kein Muly for MSR
Page 1 of 1
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K091812
Frontier Devices K091812 - 510(k) Response Letter
Appendix C Page C-1
{7}------------------------------------------------
Indications for Use
510(k) Number (if known): K091812
Device Name: Frontier Devices Orthognathic System
Indications for Use:
Frontier Devices Orthognathic System is intended for use in selective trauma of the mid-face and maxillofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kei Muhy for MSR
Division Sign-Off) Tivision of Anesthesiology, General Hospital ifection Control, Dental Devices
Page 1 of 1
) ਪ 510(k) Number:
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K091812
# Device Name: Frontier Devices Mandible Reconstruction System
Indications for Use:
The Frontier Devices Mandible Reconstructive System is intended for use in selective trauma of the mid-face and maxillofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Kein Muler for NSR
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K091812
{9}------------------------------------------------
# Indications for Use
510(k) Number (if known): K091812
Device Name: Frontier Devices Mesh System
Indications for Use:
Frontier Devices Mesh System is intended for use in selective trauma of the mid-face and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kein Muhn for NSR
Page *1* of *1*
Ovision Sion-Off vivision of Anesthesiology, General Hoapital nfection Control, Dental Deviews
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.