SMARTTRACK SYSTEM

K091803 · Nexus6, Ltd. · CAF · Oct 9, 2009 · Anesthesiology

Device Facts

Record IDK091803
Device NameSMARTTRACK SYSTEM
ApplicantNexus6, Ltd.
Product CodeCAF · Anesthesiology
Decision DateOct 9, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5630
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The SmartTrack System is intended for single-patient use as an electronic data capture accessory for recording and monitoring actuations of prescribed MDI usage. This may be used in the following applications: - In clinical trials, where researchers need to know when a patient has actuated their trial MDI medication; - In clinical practice, where specialists, general practitioners, and nurse educators need to know if a patient has actuated their prescribed MDI medication. The SmartTrack is not intended to indicate remaining quantity of medication in an MDI and does not include a dose counting function.

Device Story

SmartTrack System is a compliance monitoring accessory for Metered Dose Inhalers (MDI). It consists of a clip-on module, an optical docking station, and PC-based analysis software. The clip-on module attaches to the MDI exterior and uses an optical sensor to detect and log the date and time of MDI actuations. Data is transferred via the docking station to a PC. The Respiratory Analyzer software allows clinicians, researchers, or nurse educators to review, export, and analyze usage history in table or graph formats. The device provides objective data on medication adherence to inform clinical decision-making and support clinical trial monitoring. It does not count doses or indicate remaining medication quantity.

Clinical Evidence

Bench testing only. No clinical data was required or provided. Functional verification and performance testing of the MDI actuation sensor system confirmed accurate logging of MDI usage. Compliance with IEC 60601 (safety/EMC) and IEC 60068 (environmental) standards was performed.

Technological Characteristics

Clip-on module for MDI; optical sensor for actuation detection; microprocessor-controlled; internal battery power; USB-connected optical docking station; PC-based offline software for data reporting. Complies with IEC 60601 and IEC 60068 series standards.

Indications for Use

Indicated for single-patient use to record and monitor MDI actuations in clinical trials or clinical practice settings for patients prescribed MDI medication.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K091803 Image /page/0/Picture/1 description: The image shows the logo for Nexus 6. The logo consists of a plus sign made up of four squares, followed by the word "nexus6" in lowercase letters. The "6" is slightly raised and has a small "tm" symbol next to it, indicating a trademark. OCT - 9 2009 Nexus 6 Limited Level 2, 109 Cook Street, Auckland 1010, New Zealand PO Box 106-612, Auckland 1143, New Zealand Phone +64 9 307 2771 Fax +64 9 307 2773 6 July 2009 # 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92: (a)(1) Refer to information above and concluding this summary. ## (a)(2) Name of the Device | Model Number / Name: | SmartTrack System | |----------------------|-------------------------------------------------------------------------------------------------| | Classification Name: | Nebulizer (Direct Patient Interface)<br>Anesthesiology Devices, 21 CFR §868.5630, Class II, CAF | ## (a)(3} Identification of Legally Marketed Devices | K990185 | MDILog Model MDC-512 | Medtrac Technologies Inc | |---------|----------------------|--------------------------| | K970344 | MDILog Model MDC-511 | Medtrac Technologies Inc | | K935955 | Doser MDI Counter | Newmed Corporation | #### (a)(4) Description of the Device The SmartTrack System consists of three modules used to provide a compliance monitoring function for use of a Metered Dose Inhaler (MDI). - · SmartTrack is a clip-on module that attaches externally around the enclosure of an MDI, using an optical sensor to detect MDI actuation, to log the usage history of the MDI. The device includes an LCD display and control buttons to review basic data and adjust device settings. Various device status conditions are indicated to the user, and logged to provide potentially relevant context for later review of MDI usage history data. - SmartTrack Docking Station is an optical interface cradle used to download data stored on the SmartTrack module, via USB cable connection to a PC, and allow for interaction by the offline software with device settings and status monitoring. - Respiratory Analyzer is a software application for use on a PC that provides for offline review of the MDI compliance data logged by the SmartTrack module. Data is stored in database files with patient, medication, and prescribed dose information, and is available for review and export in various table and graph formats to highlight MDI usage characteristics. {1}------------------------------------------------ #### (a)(5) Statement of the Intended Use The SmartTrack System is intended for single-patient use as an electronic data capture accessory for recording and monitoring actuations of prescribed MDI usage. This may be used in the following applications: - In clinical trials, where researchers need to know when a patient has actuated their trial MDI medication; - . In clinical practice, where specialists, general practitioners, and nurse educators need to know if a patient has actuated their prescribed MDI medication. The SmartTrack is not intended to indicate remaining quantity of medication in an MDI and does not include a dose counting function. ## (a)(6) Technological Characteristics Summary Technological characteristics of the SmartTrack system are largely equivalent to the predicate devices listed above. Equivalent features between the devices include: configuration in attaching to the outside of an MDI enclosure; microprocessor control and use of an internal clock, to log date and time of MDI actuations; power supply from an internal battery; interface to a personal computer via an optical docking station to download MDI usage data; and offline review software providing for data handling, device interaction, and patient data reporting functions. The sensor technology used to detect MDI actuation is different from the predicate devices, and this aspect of the device has been verified by non-clinical testing to establish equivalent performance to the predicate devices. #### (b)(1) Discussion of the Non-Clinical Tests Non-clinical testing of the SmartTrack system has been carried out to cover functional verification and device performance. This included completion of software verification procedures, with performance testing of the MDI actuation sensor system to ensure data is logged accurately for MDI usage. This established correct functionality of the SmartTrack System according to requirements. Third party testing of the SmartTrack System for compliance to IEC 60601 series standards for general safety and electromagnetic compatibility, and IEC 60068 series standards for environmental testing, will be completed by an accredited laboratory before marketing of the device. ## (b)(2) Discussion of the Clinical Tests Clinical testing was not required to demonstrate the safety and effectiveness of the SmartTrack System. The product functionality has been adequately assessed by bench testing as above. #### (b)(3) Conclusions Demonstrating Safety, Effectiveness and Performance Hardware testing carried out for the SmartTrack System indicates it meets design and performance functional requirements. Software verification demonstrates that device features are effective, and the system configuration functions equivalently to the predicate devices. The device will meet standard requirements for electrical safety and electromagnetic compatibility before marketing. This information indicates that the SmartTrack System is equivalent to the predicate devices in terms of device safety and effectiveness. Garth Sutherland Chief Operating Officer Nexus6 Ltd {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing left. The eagle is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the circumference of the circle. The text is in all capital letters and is arranged to follow the curve of the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-0609 Silver Spring, MD 20993-0002 Nexus6 Limited C/O Mr. Morten Simon Christensen Responsible Third Party Official Underwriters Laboratories, Incorporated 455 East Trimble Road San Jose, California 95131-1230 OCT - 9 2009 Re: K091803 Trade/Device Name: SmartTrack System Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: September 22, 2009 Received: September 24, 2009 Dear Mr. Christensen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Mr. Christensen Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. h for Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Nexus6 LTD. The logo consists of a plus sign made up of four squares, with a white square in the center. To the right of the plus sign is the word "nexus6" in lowercase letters, with "LTD" in superscript. Nexus6 Limited Level 2, 109 Cook Street, Auckland 1010, New Zealand PO Box 106-612, Auckland 1143, New Zealand Phone +64 9 307 2771 Fax +64 9 307 2773 9 October 2009 # INDICATIONS FOR USE 510(k) Number: K091803 SmartTrack System Device Name: Indications for Use: The SmartTrack System is intended for single-patient use as an electronic data capture accessory for recording and monitoring actuations of prescribed MDI usage. This may be used in the following applications: - In clinical trials, where researchers need to know when a patient has actuated their . trial MDI medication; - In clinical practice, where specialists, general practitioners, and nurse educators need . to know if a patient has actuated their prescribed MDI medication. The SmartTrack is not intended to indicate remaining quantity of medication in an MDI and does not include a dose counting function. Prescription Use ✓ (21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Schulten (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K091803
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...