K091796 · Zap Lasers, LLC · GEX · Jul 31, 2009 · General, Plastic Surgery
Device Facts
Record ID
K091796
Device Name
SOFTLASE PRO, ORTHOLASE AND HYGIENELASE
Applicant
Zap Lasers, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 31, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SoftLase Pro™/OrthoLase™/HygieneLase™ is to provide the ability to perform intraoral soft tissue maxilla-facial dental. general, oral and i cosmetic surgery. The SoftLase ProTM/OrthoLase™/HygieneLase™ is intended for ablating, incising, excising, vaporizing and coagulation of soft tissues using a contact fiber optic delivery system. The device will be used in the following areas: general and cosmetic dentistry otolaryngology, arthroscopy, gastroenterology, general surgery, dermatology & plastic surgery, neurosurgery, genecology, urology, opthamology and pulmonary surgery. The following are the oralpharneeal indications for use for which the device will be marketed: -Excision and Incision Biopsies -Hemostatic assistance -Treatment of Apthous Ulcers -Frenectomy -Frenotomy -Gingival Incision and Excision -Gingivectomy -Gingivoplasty -Incising and Draining of Abscesses -Operculectomy -Oral Papillectomy -Removal of Fibromas -Soft Tissue Crown Lengthening -Sulcular Debridement (removal of diseased or inflamed soft tissue in the periodontal pocket) -Tissue retraction for Impression -Vestibuloplasty -Light activation of bleaching materials for teeth whitening -Laser-assisted bleaching/whitening of teeth
Device Story
Surgical diode laser system; uses contact fiber optic delivery system to ablate, incise, excise, vaporize, and coagulate soft tissue; also used for light-activated teeth whitening. Operated by physicians and dentists in clinical/healthcare settings. Output is laser light energy; healthcare provider controls energy delivery to target tissue to achieve surgical outcomes (e.g., hemostasis, tissue removal). Benefits include precise soft tissue management and minimally invasive surgical capabilities.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and performance verified through compliance with IEC60601-1, IEC60601-2-22, and IEC60825-1 standards, alongside internal quality systems and process certifications.
Technological Characteristics
Surgical diode laser system; contact fiber optic delivery; Class II device. Complies with IEC60601-1:1995+A1+A2, IEC60601-2-22:1995, and IEC60825-1:1993+A1+A2. Utilizes ISO materials standards and cGMP/ISO 9000 quality regulations.
Indications for Use
Indicated for intraoral soft tissue maxillofacial, dental, oral, and cosmetic surgery, including biopsies, hemostasis, aphthous ulcer treatment, frenectomy/frenotomy, gingival procedures, abscess drainage, operculectomy, papillectomy, fibroma removal, crown lengthening, sulcular debridement, tissue retraction, and laser-assisted teeth whitening. Used in dentistry, otolaryngology, arthroscopy, gastroenterology, general/plastic surgery, neurosurgery, gynecology, urology, ophthalmology, and pulmonary surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
STYLA MICROLASER/STYLAORTHO Diode Laser System (K081214)
Submission Summary (Full Text)
{0}------------------------------------------------
## SoftLase Pro OrthoLase HygieneLase 510K
CONFIDENTIAL
KC091796
# JUL 81 2009
## Appendix D PREMARKET NOTIFICATION |510(k)] Summary
510(K) Summary of Safety and Effectiveness Premarket Notification 510(k)
Zap Lasers, LLC 2621-B Pleasant Hill Road Pleasant Hill, Ca. 94523
## Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92
#### 1. Device Name:
SoftLase Pro/OrthoLase/HygieneLase family of products Trade Name: Common Name(s): Surgical Laser System Classification Name(s): Laser, Surgical
2. Establishment Name & Registration Number: Name: Zap Lasers, LLC Number: 3005212363
#### 3. Classification(s):
\$878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology. (a) Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide. (2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b) Classification. Class II.
| Device Class: | Class II for all requested indications |
|-----------------------|----------------------------------------|
| Classification Panel: | General and Plastic Surgery & Others |
| Product Code(s): | GEX |
.
#### 4. Section 514 Compliance
Zap Lasers, LLC intends to comply fully with the general controls authorized under Sections 501, 502, 510, 516, 518, and 520 of the Food, Drug, and Cosmetic Act.
{1}------------------------------------------------
#### 5. Performance Standards
United States Food and Drug Administration mandated performance standards for this device exist and are provided under Sections 21 CFR 1010 & 1020, with permissible deviations relative to Laser Notice 50, dated July 26, 2001. The device also complies with IEC60601-1:1995+A1+A2, IEC60601-2-22:1995, and IEC60825-1:1993+A1+A2. In addition, various voluntary performance standards are utilized. Voluntary standards utilized include Standard Operating Procedures, vendor & process certification and qualification procedures, Quality Systems Regulations, ISO materials standards and cGMP & ISO 9000 series quality regulations.
## 6. Special Controls:
All Class II devices are subject to Special Controls.
#### 7. Labeling:
.
The laser system discussed in this premarket notification will be manufactured by Zap Lasers, LLC and labeled as such. Zap Lasers, LLC will market the system exclusively to healthcare facilities, physicians and dentists. In addition to the usual package and identification labeling, the following additional Warnings, Cautions & Precautions statements are displayed as appropriate on or within the device packaging. They are repeated here for ease of use.
Warning: Federal (United States) Law restricts this device to sale by or on the order of a physician or dentist only.
### 9. Predicate Device (legally marketed comparison device)
Zap Lasers, Inc. believes that the following surgical laser systems are substantially equivalent to the SoftLase Pro/OrthoLase/HygieneLase Family of products surgical diode system.
- 1. SoftLase G2 (K021227, Zap Lasers, LLC)
- 2. ODYSSEY NAVIGATOR (K062258, Ivoclar Vivadent, Inc.)
- 3. STYLA MICROLASER/STYLAORTHO Diode Laser System (K081214, Zap Lasers LLC.)
## 10. Summary of Equivalence:
There are no unique applications, material or specifications presented herein. Evidence of equivalence has been demonstrated through:
- The SoftLase Pro ™/OrthoLase™/HygieneLase™ family of products intended use and indications . for use were previously cleared by FDA for the predicate devices.
- The technical characteristics of the SoftLase Pro ™/OrthoLase™/HygieneLase™ family of products . are similar to those of the cleared SoftLase G2, Odyssey Navigator and the Styla MicroLase.
- Laser output values of the SoftLase Pro ™/OrthoLase™/HygieneLase™ family of products are well . within previously cleared values of the predicate dental laser system as described.
- The predicate devices and other previously cleared laser systems with similar power outputs have a . proven safety and effectiveness in the treatment of the claimed indications.
- . Safety and performance testing.
Therefore, the SoftLase Pro/OrthoLase Family of products is substantially equivalent to its predicate devices cited above and raises no new safety and/or effectiveness issues.
Page 46 of 72
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ZAP LASERS, LLC % Mr. Jay Goble, DDS President 2621-B Pleasant Hill Road Pleasant Hill, California 94523
JUL 3 1 2009
Re: K091796
Trade/Device Name: Softlase Pro™/OrthoLase™/HygieneLase™ Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery
and in dermatology
Regulatory Class: II Product Code: GEX Dated: July 15, 2009 Received: July 22, 2009
Dear Mr. Goble:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
#### Page 2 - Mr. Jay Goble, DDS
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
> Sincerely yours, Mark N.Melkerso Director Division of Surgical, Orthopedic and Restorative Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### CONFIDENTIAL
## Tab 2 Indications for Use Statement
#### Device Name: SoftLase Pro™/OrthoLase™/HygieneLase™
Indication for Use:
The SoftLase Pro™/OrthoLase™/HygieneLase™ is to provide the ability to perform intraoral soft tissue maxilla-facial dental. general, oral and i cosmetic surgery. The SoftLase ProTM/OrthoLase™/HygieneLase™ is intended for ablating, incising, excising, vaporizing and coagulation of soft tissues using a contact fiber optic delivery system.
The device will be used in the following areas: general and cosmetic dentistry otolaryngology, arthroscopy, gastroenterology, general surgery, dermatology & plastic surgery, neurosurgery, genecology, urology, opthamology and pulmonary surgery. The following are the oralpharneeal indications for use for which the device will be marketed:
-Excision and Incision Biopsies
-Hemostatic assistance
-Treatment of Apthous Ulcers
-Frenectomy
-Frenotomy
-Gingival Incision and Excision
-Gingivectomy
-Gingivoplasty
-Incising and Draining of Abscesses
-Operculectomy
-Oral Papillectomy
-Removal of Fibromas
-Soft Tissue Crown Lengthening
-Sulcular Debridement (removal of diseased or inflamed soft tissue in the periodontal pocket)
-Tissue retraction for Impression
-Vestibuloplasty
-Light activation of bleaching materials for teeth whitening
-Laser-assisted bleaching/whitening of teeth
| Prescription Use | X |
|-----------------------------|---|
| (21 CFR Part 801 Subpart D) | |
| And/Or | |
| Over the Counter Use | . |
| (21 CFR Part 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
| (Division Sign-Off) | |
|---------------------|--|
Division of Surgical, Orthopedic, and Restorative Devices
Page 6 of 72
6/15/09
510(k) Number K091796
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.