K091736 · Daniels Corporation Pty, Inc. · MMK · Oct 30, 2009 · General Hospital
Device Facts
Record ID
K091736
Device Name
SHARPSMART DISPOSABLE SHARPS CONTAINER, MODEL S2
Applicant
Daniels Corporation Pty, Inc.
Product Code
MMK · General Hospital
Decision Date
Oct 30, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
Sharpsmart™ S2 disposable sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps. This device is intended for "Over-the-Counter" use.
Device Story
Sharpsmart S2 is a 1-quart, two-piece, injection-molded polypropylene copolymer container; used for safe disposal of hazardous medical sharps. Device features closure mechanism allowing multiple lid re-openings while ensuring security between uses; includes optional bracket for vertical/horizontal surface mounting. Used in patient rooms, medication carts, ORs, and emergency vehicles. Healthcare professionals or small quantity waste generators insert sharps vertically through top flaps; flaps lock for final disposal. Translucent design with fill line and overfill warning allows visual monitoring of capacity. Device provides puncture resistance, leak-proof sides/bottom, and stability to prevent toppling. Benefits include safe containment of hazardous sharps, reducing risk of accidental needle sticks during disposal.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified via standardized testing: UN Recommendations on the Transport of Dangerous Goods (15th Ed. 6.1.5.3) and AS/NZS 4031:1992. Passed tests for puncture resistance, leak resistance, impact, handle integrity, drop, stacking, and vibration.
Technological Characteristics
Polypropylene copolymer construction; injection-molded two-piece design. 1-quart capacity (1.6 qt total volume, 1.16 qt fill capacity). Dimensions: 7.5"h x 4.5"w x 5.5"d. Features top-loading flap mechanism for sharps insertion and locking closure. Non-sterile. Standalone or bracket-mounted. Complies with AS/NZS 4031:1992 and UN Transport of Dangerous Goods standards.
Indications for Use
Indicated for safe disposal of hazardous sharps in healthcare facilities, nursing stations, medication carts, labs, dental offices, ERs, surgical rooms, treatment rooms, emergency vehicles, and veterinary offices. Intended for over-the-counter use by healthcare professionals and small quantity waste generators.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
KC091736
510(k)
### Daniels Sharpsmart
Page 3
## 510(k) Summary
Manufacturer Name:
The Daniels Corporation PTY, LTD
Company Address:
Representative
135 S. LaSalle St. Ste 2850 Chicago, IL 60603
Andrea Arredondo Daniels Sharpsmart, Inc. 2133-126 Upton Dr., #436 Virginia Beach, VA 23454 (540) 818-0783 (757) 299-8363 (fax) aarredondo@danielsinternational.com
OCT 3 0 2009
Summary Preparation Date: Sept 29, 2009
DEVICE NAME
| Trade/Proprietary Name: | Sharpsmart™ S2 disposable sharps container |
|-------------------------|------------------------------------------------------------------|
| Common Name: | Disposable sharps disposal container |
| Classification Name: | Accessory to hypodermic single lumen needle;<br>Sharps Container |
# PREDICATE DEVICE IDENTIFICATION:
21CFR880.5570
Sharps Container
Product Code MMK
Device Class: II
## LEGALLY MARKETED EQUIVALENT DEVICE:
| Company | Product | 510(k)# |
|----------------|----------------------------------|---------|
| Solutions, Inc | Sharps Away Disposable Container | K072667 |
{1}------------------------------------------------
#### 510(k)
### DESCRIPTION OF DEVICE:
The Sharpsmart™ S2 disposable sharps container is intended to be used in patient rooms, medication rooms, operating rooms, physicians' offices or any other patient care area requiring the use of a sharps container. It is a two-piece injection-molded container made from a polypropylene copolymer. The container is designed with a closure mechanism that allows multiple lid re-openings, while ensuring secure between uses. Both handle and lid can be secured out of the way for safe, unobstructed disposal of sharps. The S2 container is capable of standalone use. A Unique bracket design is offered as an additional measure to secure mounting on vertical or horizontal surfaces to prevent toppling. The bracket is not required for the use of the container. It only offers a secondary option for the user for ease of use.
The containers are closable, puncture resistant, leak proof on the sides and bottom and stable. The label, containing an overfill warning and an easily visible fill line, is adhered to the container at the factory. The containers are translucent and are red or yellow. The label is white and red with the biohazard warning.
Sharpsmart™ S2 disposable sharps containers are available in a 1 quart size. Each product has two parts, which assemble together to form a unit. Assembly is easily visualized. They can be nested together to reduce storage space.
The container is designed and certified to comply with penetration requirements, as represented though product testing for Test methods: AS/NZS 4031:1992 and Test Method: roprosence and the Transport of Dangerous Goods. 15th revised Ed. 6.1.5.3The Sharpsmart™ S2 disposable sharps containers meet or exceed OSHA recommendations for sharps containers.
#### INTENDED USE:
Sharpsmart™ S2 disposable sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps. This device is intended for "Over-the-Counter" use.
S2 model is a 1 quart size. The container has a volume of 1.6quarts quarts, fill capacity of 1.16 quarts and an empty weight of 0.95 lb.
S2 model Color - Red with translucent lid or yellow with translucent lid.
S2 model outer dimensions of 7.50 "h x 4.50"w x 5.50 "d. (190mm x 115mm x 140mm ).
{2}------------------------------------------------
# PREDICATE PRODUCT COMPARISON TABLE:
| Manufacturer | Daniels Sharpsmart, Inc | Solutions, Inc |
|---------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Sharpsmart Disposable<br>Sharps Container | Sharps Away Disposable<br>Container |
| K Number | K091736 | K072667 |
| Indication for Use | Sharpsmart Disposable<br>Containers are intended to<br>be used for the safe disposal<br>of hazardous sharps | Sharps Away Disposable<br>Containers are intended to<br>be used for the safe disposal<br>of hazardous sharps |
| Target population | Healthcare professional | Healthcare professional |
| Where used | Healthcare facilities | Healthcare facilities |
| Material | Polypropylene | Polypropylene |
| Sharp Access | Sharps inserted through the<br>top in a vertical position<br>with sharp side down<br>through a hole formed with<br>flaps | Sharps inserted through the<br>top in a vertical position<br>with sharp side down<br>through a hole formed with<br>flaps |
| Sharps closure | Flaps are closed and locked<br>in place for removal | Flaps are closed and locked<br>in place for removal |
| Impact Resistance | Yes | Yes |
| Puncture Resistance | Yes | Yes |
| Leak Resistance | Yes | Yes |
| Single Use | Yes | Yes |
| Non-sterile | Yes | Yes |
## SUBSTANTIAL EQUIVALENCE DISCUSSION OF SIMILARITIES AND DIFFERENCES:
The Sharpsmart S2 Disposable Sharps Containers are similar to the Solution, Inc Sharps Away Disposable Containers in:
Non-Clinical test comparisons are viewed through: Intended Use, Materials and design.
Clinical tests are discussed in the Performance testing section.
Daniels submits the following information to demonstrate that the Sharpsmart S2 Disposable Sharps Container shares indications, design principles, materials and properties with the following predicate devices, which have been determined by FDA to be white the 2010 wais alent to pre-amendment devices: Sharps Away Disposable Containers (K072667) from Solutions, Inc.
Intended Use Comparison
{3}------------------------------------------------
The indications for use of the Sharpsmart S2 Disposable Sharps Container are not new indications in that they are the same as or are included in those for the predicate devices. Sharpsmart and the predicate devices are containers intended to be used for the disposal of contaminated medical sharps in health care facilities.
### Design and Materials Comparison
The design and functional characteristics of the Sharpsmart and the predicate devices are similar. They are constructed from polymeric materials. The Disposable Sharps Containers and Sharps Away containers are intended to be disposed once used. All of the devices conform to national or international standards for puncture resistance, impact resistance and leakage. They have means to prevent contact between the user and the contents, and are designed with features to easily and safely determine if they are full. They do not have features to bend, break or shear needles.
### Performance Testing Comparison
Test methods: UN Recommendations on the Transport of Dangerous Goods. 15th revised Ed. 6.1.5.3; AS/NZS 4031:1992
| Puncture Resistance -Passed | Handle Testing - Passed |
|-----------------------------|-------------------------|
| Leak Resistance - Passed | Drop Test - Passed |
| Impact Testing - Passed | Stacking Test - Passed |
Vibration - Passed
#### Conclusion:
The Sharpsmart S2 Disposable Sharps Containers introduces no new questions concerning the safety or effectiveness of the Sharpsmart Disposable Sharps Containers and is thus substantially equivalent to the predicate device.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Daniels Corporation PTY, Limited C/O Ms. Andrea Arredondo Compliance Manger Daniels Sharpsmart, Incorporated 2133-126 Upton Drive, #436 Virginia Beach, Virginia 23454
OCT 3 0 2009
Re: K091736
Trade/Device Name: Sharpsmart™ S2 Disposable Sharps Contianer Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: MMK Dated: October 2, 2009 Received: October 2, 2009
Dear Ms. Arredondo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2- Ms. Arredondo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRH Offices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Clinton D. Van
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use:
510(k) Number (if known): K091736 -
Device Name: Sharpsmart™ S2 disposable sharps container
Model: S2
Indications For Use: Sharpsmart™ S2 disposable sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps. This device is intended for "Over-the-Counter" use.
Physical Attributes: S2 model is a 1 quart size. The container has a volume of 1.6quarts quarts, fill capacity of 1.16 quarts and an empty weight of 0.95 lb.
S2 model Color - Red with translucent lid or yellow with translucent lid.
S2 model outer dimensions of 7.50 "h x 4.50"w x 5.50 "d. (190mm x 115mm x 140mm ).
Prescription Use __
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ X
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K 0 917 3 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.