VASOVIEW 6 PRO ENDOSCOPIC VESSEL HARVESTING SYSTEM, MODEL VH-2500
K091733 · Maquet Cardiovascular, LLC · GEI · Aug 28, 2009 · General, Plastic Surgery
Device Facts
Record ID
K091733
Device Name
VASOVIEW 6 PRO ENDOSCOPIC VESSEL HARVESTING SYSTEM, MODEL VH-2500
Applicant
Maquet Cardiovascular, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 28, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VASOVIEW 6 PRO System is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissection of tissue including dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extraperitoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection/vessel harvesting along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or radial artery for use in coronary artery bypass grafting. Thoracoscopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels and other tissues of the chest wall. The VASOSHIELD Pressure Controlling Syringe is indicated for controlling pressure during the preparation and irrigation of blood vessels prior to use as a bypass graft.
Device Story
The VasoView 6 Pro is an endoscopic vessel harvesting system used in minimally invasive surgery. The system includes a harvesting cannula with four lumens housing a 7mm endoscope, C-Ring wire, endoscope washer tube, and a Bipolar BiSector for tissue ligation and division. Accessories include an extended-length endoscope, dissection tip for blunt tissue isolation, and a shortport blunt tip trocar for instrument access. The system also includes the VasoShield pressure-controlling syringe, which features a pressure relief valve and setting ring to manage irrigation pressure for bypass grafts. Operated by surgeons in an OR setting, the device enables visualization and mechanical/electrosurgical dissection of vessels. The output is the harvested vessel, prepared for use as a bypass graft. The device benefits patients by facilitating minimally invasive access, potentially reducing surgical trauma compared to open harvesting techniques.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and verification results demonstrating performance equivalent to predicate devices.
Technological Characteristics
Electrosurgical cutting and coagulation system. Components: 7mm endoscope, C-Ring wire, Bipolar BiSector, blunt tip trocar, and pressure-controlled syringe. Connectivity: None. Energy source: Electrosurgical (bipolar). Sterilization: Not specified. Materials: Not specified.
Indications for Use
Indicated for patients undergoing minimally invasive endoscopic surgery for arterial bypass, including coronary artery bypass grafting (CABG) and peripheral artery bypass. Used for blunt dissection of blood vessels (saphenous vein, radial artery), ducts, and structures in the extraperitoneal, subcutaneous extremity, and thoracic spaces. Also indicated for cutting tissue and coagulation. VASOSHIELD syringe indicated for pressure-controlled preparation/irrigation of bypass grafts.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
K091733
Maquet Cardiovascular LLC
VasoView 6 Pro Endoscopic Vessel Harvesting System
System
Special 510(k)
AUG 2 8 2009
## APPENDIX A: 510(K) SUMMARY
| Submitter | MAQUET Cardiovascular LLC |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's<br>Address | 40 Barbour Pond Drive<br>Wayne, NJ 07470 |
| Telephone | (408) 635-3987 |
| Fax | (408) 635-3907 |
| Contact Person | Artie Kaushik |
| Date Prepared | June 10, 2009 |
| Device Trade<br>Name | VASOVIEW 6 PRO Endoscopic Vessel Harvesting System |
| Device<br>Common Name | Electrosurgical cutting and coagulation device and accessories |
| Device<br>Classification<br>Name | Electrosurgical cutting and coagulation device and accessories |
| Device<br>Classification | Class II |
| Summary of<br>substantial<br>equivalence | The design, materials, method of delivery, and intended use features of the<br>VASOVIEW 6 PRO Endoscopic Vessel Harvesting System are substantially<br>equivalent with regard to those features in the predicate device: the VV6<br>(K041981, August 20, 2004) and VASOSHIELD Pressure Controlling Syringe<br>(K082725, December 17, 2008) |
| Device<br>description | The VASOVIEW 6 PRO Endoscopic Vessel Harvesting System consists of the<br>VASOVIEW 6 PRO Harvesting Cannula which is designed for use in<br>conjunction with the 7mm Endoscope. The Harvesting Cannula has four lumens<br>which house the Endoscope, C-Ring wire, endoscope washer tube and Bipolar<br>BiSector for ligation and division of vessel branched. The system includes the<br>following accessories- (1) The 7mm Extended Length Endoscope and Dissection<br>Tip for blunt dissection of tissue and isolation of structures in the cavity. (2)<br>Shortport Blunt Tip Trocar (BTT) to provide a port for insertion of endoscoic<br>instruments into an incision site. (3) VASOSHIELD Pressure Controlling<br>Syringe is a fully assembled 60 mL syringe with a pressure relief valve that is<br>controlled by a pressure setting ring. |
{1}------------------------------------------------
The VASOVIEW 6 PRO System is indicated for use in minimally invasive Indications for surgery allowing access for vessel harvesting, and is primarily indicated for Use patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissection of tissue including dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extraperitoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection/vessel harvesting along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or radial artery for use in coronary artery bypass grafting. Thoracoscopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels and other tissues of the chest wall. The VASOSHIELD Pressure Controlling Syringe is indicated for controlling pressure during the preparation and irrigation of blood vessels prior to use as a bypass graft. The VASOVIEW 6 PRO Endoscopic Vessel Harvesting System incorporates the Technological same fundamental scientific technology as the predicate devices. characteristics The results of the verification testing demonstrate that the VASOVIEW 6 PRO Performance meet the established acceptance criteria and performs in a manner equivalent to data the predicate device. No new safety or effectiveness issues were raised during the testing program. The results of the verification testing for the VASOSHIELD Pressure Controlling Syringe also demonstrate that the syringe meets established acceptance criteria and performs in a manner equivalent to its predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with healthcare, with three overlapping lines forming the staff and a serpent winding around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Maquet Cardiovascular, LLC % Ms. Artie Kaushik Associate, Regulatory Affairs 170 Baytech Drive San Jose, California 95134
## AUG 2 8 2009
Re: K091733
Trade/Device Name: VASOVIEW 6 PRO Endoscopic Vessel Harvesting System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI, GBX Dated: July 31, 2009 Received: August 4, 2009
Dear Ms. Kaushik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Ms. Artie Kaushik
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## APPENDIX B: INDICATIONS FOR USE STATEMENT
510(k) number (if known) The 510(k) number has not been issued yet.
Device name
VASOVIEW 6 PRO Endoscopic Vessel Harvesting System
Indications for Use
The VASOVIEW 6 PRO System is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissection of tissue including dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extraperitoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection/vessel harvesting along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or radial artery for use in coronary artery bypass grafting. Thoracoscopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels and other tissues of the chest wall.
The VASOSHIELD Pressure Controlling Syringe is indicated for controlling pressure during the preparation and irrigation of blood vessels prior to use as a bypass graft.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The-Counter Use ﮐﮯ OR Prescription Use (Per 21 CFR 801.109)
K091733
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.