Alpha-bsm Bone Substitute Material is an implantable synthetic calcium phosphate bone graft material nano-that forms a crystalline matix that resorbs and is replaced with new bone during the healing process. It is indicated for use in filling and/or augmentation of bone voids, gaps or defects that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Alpha-bsm Bone Substitute Material is intended to be used in bony voids or gaps to fill and/or augment dental intraosseous, introral and maxillofacial defects. These defects include, but are not limited to, periodontal/infrabony defects; alveolar ridge augmentation (osteotomy, apicoectomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); sinus lifts; cystic defects: craniofacial augmentation.
Device Story
Synthetic, biocompatible bone graft substitutes (Alpha-bsm, Beta-bsm, Gamma-bsm, EquivaBone) used for filling/augmenting non-load-bearing bone defects. Powder component mixed with solution to form paste or putty; applied manually or via syringe to treatment site. Hardens at body temperature into poorly crystalline hydroxyapatite (PCHA); osteoconductive; resorbed and replaced by natural bone. EquivaBone includes demineralized bone matrix (DBM). Used in dental/maxillofacial procedures by clinicians. Benefits patient by facilitating bone healing in voids/gaps.
Clinical Evidence
Bench testing only. Performance data consistent with Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices (April 28, 2005) submitted. EquivaBone DBM component assayed for osteoinductive potential in athymic nude mouse model.
Technological Characteristics
Synthetic calcium phosphate bone graft material. EquivaBone also contains sodium carboxymethyl cellulose and demineralized bone matrix (DBM). Hardens at body temperature to poorly crystalline hydroxyapatite (PCHA). Delivery via silicone bulb, syringe-to-syringe, or manual application. Sterile, single-use kits.
Indications for Use
Indicated for filling/augmentation of non-load-bearing bone voids, gaps, or defects (surgically created or traumatic) in dental, intraoral, and maxillofacial regions. Includes periodontal/infrabony defects, alveolar ridge augmentation, dental extraction sites, sinus lifts, cystic defects, and craniofacial augmentation.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Beta-bsm Injectable Bone Substitute Material (K090242)
Gamma-bsm Moldable Bone Substitute Material (K090242)
EquivaBone Osteoinductive Bone Graft Substitute (K080329)
EquivaBone Osteoinductive Bone Graft Substitute (K090310)
Submission Summary (Full Text)
{0}------------------------------------------------
K091729
### 5. 510(k) Summaries
DEC 2 3 2009
- 510(k) Summary Alpha-bsm 5.1
### ETEX Corporation Submitter: 38 Sidney Street Cambridge, MA 02139 Registration No .: 1225112 Owner/Operator No .: 9014709
Contact Person: Christopher Klaczyk Regulatory Affairs Manager Office: (617) 577-7270 x160 Mobile: (617) 710-8091 Fax: (617) 577-7170 E-Mail: cklaczyk@etexcorp.com
Date Prepared: June 9, 2009
Product Code(s): LYC (21 CFR 872.3930)
Device Class: II (21 CFR 872.3930)
Classification Panel: Dental
Classification Name: Bone Grafting Material, Synthetic (21 CFR 872.3930)
Proprietary Name: Alpha-bsm Bone Substitute Material
Predicate Device(s): a-BSM Bone Substitute Material (ETEX Corporation, K962548 LYC Dental indication) a-BSM Bone Substitute Material (ETEX Corporation, K980223 LZK Maxillofacial indication) a-BSM Bone Substitute Material (ETEX Corporation, K983009 GXP Cranioplasty indication) a-BSM Bone Substitute Material (ETEX Corporation, K032307 GXP Cranioplasty indication) a-BSM Bone Substitute Material (ETEX Corporation, K072636 MQV Extremities, Pelvis and Spine indication) GRAFTON® DBM (Osteotech, K051188, NUN)
Device Description: Alpha-bsm Bone Substitute Material is a synthetic, biocompatible bone graft substitute material. At the time of use, the powder component is combined with a specified volume of mixing solution and mixed to form a paste. Mixing is facilitated by a silicone bulb mixing system. The
Page 5-1 .
{1}------------------------------------------------
resulting paste can be administered to the treatment site by injection or manual application. The material can be shaped into a desired form in-situ prior to implantation. After the paste is applied to the treatment site, it hardens at body temperature and converts to an apatitic calcium phosphate material. The end product, poorly crystalline hydroxyapatite (PCHA), is of low crystalline order with a similar chemical and crystalline structure to that of natural bone minerals. Alpha-bsm Bone Substitute Material is an osteoconductive material that is resorbed and replaced by natural bone over time.
# Intended Use:
Alpha-bsm Bone Substitute Material is an implantable synthetic calcium phosphate bone graft material nano-that forms a crystalline matix that resorbs and is replaced with new bone during the healing process. It is indicated for use in filling and/or augmentation of bone voids, gaps or defects that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
Alpha-bsm Bone Substitute Material is intended to be used in bony voids or gaps to fill and/or augment dental intraosseous, introral and maxillofacial defects. These defects include, but are not limited to, periodontal/infrabony defects; alveolar ridge augmentation (osteotomy, apicoectomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); sinus lifts; cystic defects: craniofacial augmentation.
Materials: Synthetic calcium phosphate
Performance Data: Testing consistent with Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices (dated April 28, 2005) has been submitted.
{2}------------------------------------------------
K091729
#### 510(k) Summary - Beta-bsm 5.2
| Submitter: | ETEX Corporation |
|------------|-----------------------------|
| | 38 Sidney Street |
| | Cambridge, MA 02139 |
| | Registration No.: 1225112 |
| | Owner/Operator No.: 9014709 |
Contact Person: Christopher Klaczyk Regulatory Affairs Manager Office: (617) 577-7270 x160 Mobile: (617) 710-8091 Fax: (617) 577-7170 E-Mail: cklaczyk@etexcorp.com
Date Prepared: June 9, 2009
Product Code(s): LYC (21 CFR 872.3930)
Device Class: II (21 CFR 872.3930)
Classification Panel: Dental
Classification Name: Bone Grafting Material, Synthetic (21 CFR 872.3930)
Proprietarv Name: Beta-bsm Injectable Bone Substitute Material
Predicate Device(s): a-BSM Bone Substitute Material (ETEX Corporation, K962548 LYC Dental indication) a-BSM Bone Substitute Material (ETEX Corporation, K980223 LZK Maxillofacial indication) a-BSM Bone Substitute Material (ETEX Corporation, K983009 GXP Cranioplasty indication) a-BSM Bone Substitute Material (ETEX Corporation, K032307 GXP Cranioplasty indication) a-BSM Bone Substitute Material (ETEX Corporation, K072636 MOV Extremities, Pelvis and Spine indication) Beta-bsm Injectable Bone Substitute Material (ETEX Corporation, K090242) GRAFTON® DBM (Osteotech, K051188, NUN)
Device Description: Beta-bsm Injectable Bone Substitute Material is a synthetic, biocompatible bone graft substitute material. At the time of use, the powder component is combined with a specified volume of mixing solution and mixed to form a
{3}------------------------------------------------
paste. Mixing is facilitated by a syringe-to-syringe mixing system. The resulting paste can be administered to the treatment site by injection or manual application. The material can be shaped into a desired form in-situ prior to implantation. After the paste is applied to the treatment site, it hardens at body temperature and converts to an apatitic calcium phosphate material. The end product, poorly crystalline hydroxyapatite (PCHA), is of low crystalline order with a similar chemical and crystalline structure to that of natural bone minerals. Beta-bsm Injectable Bone Substitute Material is an osteoconductive material that is resorbed and replaced by natural bone over time.
Beta-bsm Iniectable Bone Substitute Material is an implantable synthetic calcium phosphate bone graft material that forms a nano-crystalline matrix that resorbs and is replaced with new bone during the healing process. It is indicated for use in filling and/or augmentation of bone voids, gaps or defects that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
> Beta-bsm Injectable Bone Substitute Material is intended to be used in bony voids or gaps to fill and/or augment dental intraosseous, intraoral and maxillofacial defects. These defects include, but are not limited to, periodontal/infrabony defects: alveolar ridge a augmentation (osteotomy. apicoectomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); sinus lifts; cystic defects; craniofacial augmentation.
Materials: Synthetic calcium phosphate
Performance Data: Testing consistent with Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices (dated April 28, 2005) has been submitted.
### Intended Use:
{4}------------------------------------------------
K091729
#### 5.3 510(k) Summary - Gamma-bsm
| Submitter: | ETEX Corporation<br>38 Sidney Street<br>Cambridge, MA 02139<br>Registration No.: 1225112<br>Owner/Operator No.: 9014709 |
|------------|-------------------------------------------------------------------------------------------------------------------------|
|------------|-------------------------------------------------------------------------------------------------------------------------|
Contact Person: Christopher Klaczyk Regulatory Affairs Manager Office: (617) 577-7270 x160 Mobile: (617) 710-8091 Fax: (617) 577-7170 E-Mail: cklaczyk@etexcorp.com
Date Prepared: June 9, 2009
Product Code(s): LYC (21 CFR 872.3930)
Device Class: II (21 CFR 872.3930)
Classification Panel: Dental
Classification Name: Bone Grafting Material, Synthetic (21 CFR 872.3930)
Proprietary Name: Gamma-bsm Moldable Bone Substitute Material
Predicate Device(s): a-BSM Bone Substitute Material (ETEX Corporation, K962548 LYC Dental indication) a-BSM Bone Substitute Material (ETEX Corporation, K980223 LZK Maxillofacial indication) a-BSM Bone Substitute Material (ETEX Corporation, K983009 GXP Cranioplasty indication) a-BSM Bone Substitute Material (ETEX Corporation, K032307 GXP Cranioplasty indication) a-BSM Bone Substitute. Material (ETEX Corporation. K072636 MQV Extremities, Pelvis and Spine indication) Gamma-bsm Moldable Bone Substitute Material (ETEX Corporation, K090242) GRAFTON® DBM (Osteotech, K051188, NUN)
Device Description: Gamma-bsm Moldable Bone Substitute Material is a synthetic, biocompatible bone graft substitute material. At the time of use, the powder component is combined with a specified volume of mixing solution and mixed to form a
{5}------------------------------------------------
putty. The resulting putty is administered to the treatment site by manual application. The material can be shaped into a desired form in-situ prior to implantation. After the putty is applied to the treatment site, it hardens at body temperature and converts to an apatitic calcium phosphate product, poorly crystalline material. The end hydroxyapatite (PCHA), is of low crystalline order with a similar chemical and crystalline structure to that of natural Gamma-bsm Moldable Bone Substitute bone minerals. Material is an osteoconductive material that is resorbed and replaced by natural bone over time.
Intended Use: Gamma-bsm Moldable Bone Substitute Material is an implantable synthetic calcium phosphate bone graft material that forms a nano-crystalline matrix that resorbs and is replaced with new bone during the healing process. It is indicated for use in filling and/or augmentation of bone voids, gaps or defects that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
> Gamma-bsm Moldable Bone Substitute Material is intended to be used in bony voids or gaps to fill and/or augment dental intraosseous, intraoral and maxillofacial defects, These defects include, but are not limited to, periodontal/infrabony alveolar defects: ---------------------------------------------------------------------------------------------------------------------------------------------------------------------(osteotomy. apicoectomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); sinus lifts; cystic defects; craniofacial augmentation.
#### Materials: Synthetic calcium phosphate
- Performance Data: Testing consistent with Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices (dated April 28, 2005) has been submitted.
{6}------------------------------------------------
K0911729 Traditional 510(k) Submission –Bone Grafting Material
#### 5.4 510(k) Summary - EquivaBone
| Submitter: | ETEX Corporation<br>38 Sidney Street<br>Cambridge, MA 02139<br>Registration No.: 1225112<br>Owner/Operator No.: 9014709 |
|------------|-------------------------------------------------------------------------------------------------------------------------|
|------------|-------------------------------------------------------------------------------------------------------------------------|
Contact Person: Christopher Klaczyk Regulatory Affairs Manager Office: (617) 577-7270 x160 Mobile: (617) 710-8091 Fax: (617) 577-7170 E-Mail: cklaczyk@etexcorp.com
Date Prepared: June 9, 2009
Product Code(s): NUN (21 CFR 872.3930)
Device Class: II (21 CFR 872.3930)
Classification Panel: Dental
Classification Name: Bone Grafting Material, Human Source (21 CFR 872.3930)
Proprietary Name: EquivaBone Osteoinductive Bone Graft Substitute
Predicate Device(s): a-BSM Bone Substitute Material (ETEX Corporation, K962548 LYC Dental indication) a-BSM Bone Substitute Material (ETEX Corporation, K980223 LZK Maxillofacial indication) a-BSM Bone Substitute Material (ETEX Corporation, K983009 GXP Cranioplasty indication) a-BSM Bone Substitute Material (ETEX Corporation, K032307 GXP Cranioplasty indication) a-BSM Bone Substitute Material (ETEX Corporation, K072636 MQV Extremities, Pelvis and Spine indication) EquivaBone Osteoinductive Bone Graft Substitute (ETEX Corporation, K080329 and K090310) GRAFTON® DBM (Osteotech, K051188, NUN)
Device Description: EquivaBone is a biocompatible bone graft substitute material consisting of synthetic calcium phosphate, carboxymethyl cellulose (CMC) and human demineralized bone matrix (DBM). It is supplied in a single use kit as
{7}------------------------------------------------
sterile powders and hydration solution that are mixed together at the time of use in the operating room to form flowable putty which is implanted manually or can be extruded through a syringe. After implantation the product hardens at body temperature and resorbs and remodels during the healing process. Each lot of DBM contained within EquivaBone is assayed for osteoinductive potential in an athymic nude mouse model. This may or may not be predictive of EquivaBone osteoinductivity in humans.
Intended Use: EquivaBone Osteoinductive Bone Graft Substitute is an implantable synthetic calcium phosphate bone graft material that forms a nano-crystalline matrix combined with demineralized bone matrix that resorbs and is replaced with new bone during the healing process. It is indicated for use in filling and/or augmentation of bone voids, gaps or defects that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
> Osteoinductive Bone Graft Substitute is EquivaBone intended to be used in bony voids or gaps to fill and/or augment dental intraosseous. intraoral and maxillofacial These defects include, but are not limited to. defects. periodontal/infrabony defects; alveolar ridge augmentation (osteotomy, apicoectomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); sinus lifts; cystic defects; craniofacial augmentation.
Materials: Synthetic calcium phosphate, sodium carboxymethyl cellulose, Demineralized Bone Matrix (DBM)
Performance Data: Testing consistent with Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices (dated April 28, 2005) has been submitted.
{8}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Mr. Christopher Klaczyk Regulatory Affairs Manager ETEX Corporation 38 Sidney Street Cambridge, Massachusetts 02139
Re: K091729
Trade/Device Name: EquivaBone Osteoinductive Bone Graft Substitute Gamma-bsm Moldable Bone Substitute Material Beta-bsm Injectable Bone Substitute Material Alpha-bsm Bone Substitute Material
DEC 2 3 2009
Regulation Number: 21CFR 872.3930
Regulation Name: Bone Grafting Material
Regulatory Class: II
Product Code: LYC, NUN
Dated: December 9, 2009
Received: December 10, 2009
Dear Mr. Klaczyk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
### Page 2- Mr. Klaczyk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
. Anton V. Maier
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{10}------------------------------------------------
#### 4.4 Indications For Use - EquivaBone
510(k) Number (if known): K091729
Device Name:
EquivaBone Osteoinductive Bone Graft Substitute
Indications for Use:
EquivaBone Osteoinductive Bone Graft Substitute is an implantable synthetic calcium phosphate bone graft material that forms a nano-crystalline matrix combined with demineralized bone matrix that resorbs and is replaced with new bone during the healing process. It is indicated for use in filling and/or augmentation of bone voids, gaps or defects that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
EquivaBone Osteoinductive Bone Graft Substitute is intended to be used in bony voids or gaps to fill and/or augment dental intraosseous, intraoral and maxillofacial defects. These defects include, but are not limited to, periodontal/infrabony defects; alveolar ridge augmentation (osteotomy, apicoectomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); sinus lifts; cystic defects; craniofacial augmentation.
Prescription Use (Part 21 CFR 801 Subpart D)
. AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Th
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K091729
{11}------------------------------------------------
K091729
### 4. Indications For Use
#### Indications For Use - Alpha-bsm 4.1
K091729 510(k) Number (if known):
Device Name:
Alpha-bsm Bone Substitute Material
Indications for Use:
Alpha-bsm Bone Substitute Material is an implantable synthetic calcium phosphate bone graft material that forms a nano-crystalline matrix that resorbs and is replaced with new bone during the healing process. It is indicated for use in filling and/or augmentation of bone voids, gaps or defects that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
Alpha-bsm Bone Substitute Material is intended to be used in bony voids or gaps to fill and/or augment dental intraosseous, intraoral and maxillofacial defects. These defects include, but are not limited to, periodontal/infrabony defects; alveolar ridge augmentation (osteotomy, apicoectomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); sinus lifts; cystic defects; craniofacial augmentation.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Jh
Division Sian-Off Division of Anesthesialogy, General Ho Infection Control. Dental Devices
510(k) Number: K091729
{12}------------------------------------------------
K091729
#### 4.3 Indications For Use - Gamma-bsm
K091729 510(k) Number (if known):
Device Name:
Gamma-bsm Moldable Bone Substitute Material
Indications for Use:
Gamma-bsm Moldable Bone Substitute Material is an implantable synthetic calcium phosphate bone graft material that forms a nano-crystalline matrix that resorbs and is replaced with new bone during the healing process. It is indicated for use in filling and/or augmentation of bone voids, gaps or defects that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
Gamma-bsm Moldable Bone Substitute Material is intended to be used in bony voids or gaps to fill and/or augment dental intraosseous, intraoral and maxillofacial defects. These defects include, but are not limited to, periodontal/infrabony defects; alveolar ridge augmentation (osteotomy, apicoectorny, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); sinus lifts; cystic defects; craniofacial augmentation.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
h
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K091729
{13}------------------------------------------------
### Indications For Use - Beta-bsm 4.2
K09 j729
510(k) Number (if known): K091729
Device Name:
Beta-bsm Injectable Bone Substitute Material
Indications for Use:
Beta-bsm Injectable Bone Substitute Material is an implantable synthetic calcium phosphate bone graft material that forms a nano-crystalline matrix that resorbs and is replaced with new bone during the healing process. It is indicated for use in filling and/or augmentation of bone voids, gaps or defects that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
Beta-bsm Injectable Bone Substitute Material is intended to be used in bony voids or gaps to fill and/or augment dental intraosseous, intraoral and maxillofacial defects. These defects include, but are not limited to, periodontal/infrabony defects; alveolar ridge augmentation (osteotomy, apicoectorny, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); sinus lifts; cystic defects; craniofacial augmentation.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kinh
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.