BD PROBETEC CHLAMYDIA TRACHOMATIS (CT) Q AMPLIFIED DNA ASSAY
K091724 · Becton, Dickinson & CO · MKZ · Nov 13, 2009 · Microbiology
Device Facts
Record ID
K091724
Device Name
BD PROBETEC CHLAMYDIA TRACHOMATIS (CT) Q AMPLIFIED DNA ASSAY
Applicant
Becton, Dickinson & CO
Product Code
MKZ · Microbiology
Decision Date
Nov 13, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3120
Device Class
Class 1
Indications for Use
The BD ProbeTec™ Chlamydia trachomatis (CT) Q⁺ Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Chlamydia trachomatis DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in BD SurePath™ Preservative Fluid using an aliquot that is removed prior to processing for the BD SurePath Pap test. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease.
Device Story
The BD ProbeTec GC Q* Amplified DNA Assay is an in vitro diagnostic test for N. gonorrhoeae DNA. It utilizes Strand Displacement Amplification (SDA) technology. The process involves automated DNA extraction from clinical specimens via chemical lysis, binding to para-magnetic particles, washing, and elution. The BD Viper System performs the extraction and amplification. Reagents are provided in disposable microwells; the system rehydrates reagents, transfers the mixture to an amplification microwell, and incubates it in a thermally-controlled fluorescent reader. The system monitors fluorescence from a target-specific probe and an Extraction Control (EC) probe. An automated algorithm calculates peak fluorescence (MaxRFU) and compares it to a threshold to determine results (positive, negative, or EC failure). The device is used in clinical laboratory settings by trained personnel. It aids clinicians in diagnosing gonococcal urogenital disease, facilitating appropriate patient management.
Clinical Evidence
Clinical study evaluated 1715 female subjects across 11 sites. Sensitivity and specificity were determined by comparing BD SurePath specimen results to a Patient Infected Status (PIS) algorithm based on three reference NAAT methods. Overall sensitivity was 100.0% (51/51) and specificity was 99.9% (1662/1664). Performance was consistent across symptomatic and asymptomatic populations. Reproducibility was assessed using simulated specimens at various concentrations, showing high concordance and consistent MaxRFU values across multiple sites and runs.
Technological Characteristics
Uses Strand Displacement Amplification (SDA) with fluorescently-labeled detector probes. Reagents are dried in disposable microwells (Priming and Amplification). Employs para-magnetic particles for DNA extraction. Operates on the BD Viper System, which provides automated extraction, thermal control, and fluorescent detection. Connectivity is integrated within the BD Viper System platform. Software-based automated signal processing and result interpretation.
Indications for Use
Indicated for symptomatic and asymptomatic individuals to aid in the diagnosis of gonococcal urogenital disease. Specimens include female endocervical swabs, male urethral swabs, patient-collected vaginal swabs, male/female urine, and gynecological specimens in BD SurePath Preservative Fluid.
Regulatory Classification
Identification
Chlamydia serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to chlamydia in serum. Additionally, some of these reagents consist of chlamydia antisera conjugated with a fluorescent dye used to identify chlamydia directly from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genus Chlamydia and provides epidemiological information on these diseases. Chlamydia are the causative agents of psittacosis (a form of pneumonia), lymphogranuloma venereum (a venereal disease), and trachoma (a chronic disease of the eye and eyelid).
Predicate Devices
BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay (K081825)
Gen-Probe APTIMA Assay for Neisseria gonorrhoeae (AGC) (K062440)
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K091724
B. Purpose for Submission:
To obtain a Substantial equivalence determination for the addition of pre-aliquot samples from gynecological specimens stored in SurePath™ Preservative Fluid as an additional sample type to be tested on the BD Viper™ System in extracted mode previously cleared (K081824).
C. Measurand:
Chlamydia trachomatis DNA
D. Type of Test:
Qualitative Determination of Chlamydia trachomatis DNA using the Strand Displacement Nucleic Acid Amplification technology
E. Applicant:
Becton, Dickinson and Company
F. Proprietary and Established Names:
BD ProbeTec™ Chlamydia trachomatis (CT) Q⁺ Amplified DNA Assay
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| MKZ | I | 866.3120 | Microbiology |
H. Intended Use:
1. Intended use:
The BD ProbeTec™ Chlamydia trachomatis (CT) Q⁺ Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Chlamydia trachomatis DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in BD SurePath™
{1}
Page 2 of 25
Preservative Fluid using an aliquot that is removed prior to processing for the BD SurePath Pap test. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease.
2. Indications for use: Same as indications for use
3. Special condition for use statement(s):
For Prescription Use Only
4. Special instrument Requirements:
BD Viper System with automated nucleic acid extraction mode
I. Device Description:
The BD ProbeTecCT Q® Amplified DNA Assay is based on the simultaneous amplification and detection of target DNA using amplification primers and a fluorescently-labeled detector probe. The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, prewarmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermally-controlled fluorescent readers. The presence or absence of C. trachomatis DNA is determined by calculating the peak fluorescence (Maximum Relative Fluorescent Units (MaxRFU)) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.
In addition to the fluorescent probe used to detect amplified C. trachomatis target DNA, a second fluorescently-labeled oligonucleotide is incorporated in each reaction. The Extraction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the C. trachomatis-specific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is re-hydrated upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper instrument and an automated algorithm is applied to both the EC and C. trachomatis-specific signals to report specimen results as positive, negative, or EC failure.
J. Substantial Equivalence Information:
{2}
Page 3 of 25
1. **Predicate device names:**
BD ProbeTec™ Chlamydia trachomatis (CT) Q⁺ Amplified DNA Assay
Gen-Probe ACT Assay
2. **Predicate K number(s):**
K081824
K053446
3. **Comparison with predicate:**
Device Comparison: CTQ Assay Specimen Collection for the BD Viper System in Extracted Mode
| | BD ProbeTec CTQ Assay,PreservCyt Solution Specimens (Device) | BD ProbeTec CTQ Assay, Swab and Urine Specimens (K081824) | Gen-Probe ACT (K053446) |
| --- | --- | --- | --- |
| Specimen Types | • Same as K081825
• Gynecological specimen in SurePath Preservative Fluid | • Endocervical swab (females)
• Vaginal self-collected swab (in a clinical setting) (females)
• Urethral swab (males)
• Neat urine (female and male)
• UPT urine (female and male) | • Endocervical swab (females)
• Vaginal swab (females)
• Urethral swab (males)
• Neat urine (female and male)
• UTT urine (female and male)
• Gynecological specimen in PreservCyt Solution |
| Specimen Collection and Transport Accessories | • Same as K081824
• Liquid Based Cytology Specimen (LBC) Dilution Tube | • Endocervical kit
• Urethral kit
• Vaginal kit
• UPT
• Neat urine (Qx Sample Tube) | • Unisex swab kit
• Vaginal swab kit
• Urine collection kit
• Specimen transfer kit (for gynecological specimen in PreservCyt Solution) |
Device Comparison: Specimen Collection
| | BD ProbeTec CTQ Assay, SurePath Preservative Fluid (Device) | Gen-Probe ACT (K053446) |
| --- | --- | --- |
| Specimen Collection | • Gynecological specimen collected and placed in SurePath Preservative Fluid (per TriPath’s instruction for use).
• Sample for CTQ/GCQ testing is drawn from original cytology specimen vial before the specimen is processed for cytology testing. | • Gynecological specimen collected and placed in PreservCyt Solution (per Cytyc’s instruction for use).
• LBC specimen is first processed for cytology and then an aliquot is drawn from the remaining specimen in the vial for CT/GC testing. |
{3}
Page 4 of 25
Device Comparison: Specimen Processing
| | BD ProbeTec CTQ Assay, SurePath Preservative Fluid (Device) | BD ProbeTec CTQ Assay, Swab and Urine Specimens (K081824) |
| --- | --- | --- |
| Specimen Processing | • Same as K081824 without the 15 minute pre-warm step for LBC Dilution Tube specimens
• Use of LBC Specimen Rack to prevent pre-warming of LBC specimens | Pre-warm specimens (swabs and urines) for 15 minutes before running BD Viper System |
K. Standard/Guidance Document Referenced (if applicable): Not Applicable
L. Test Principle:
The BD ProbeTecCT Q® Amplified DNA Assay is based on the simultaneous amplification and detection of target DNA using amplification primers and a fluorescently-labeled detector probe (8, 9). The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, pre-warmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermally-controlled fluorescent readers. The presence or absence of C. trachomatis DNA is determined by calculating the peak fluorescence (Maximum Relative Fluorescent Units (MaxRFU)) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.
In addition to the fluorescent probe used to detect amplified C. trachomatis target DNA, a second fluorescently-labeled oligonucleotide is incorporated in each reaction. The Extraction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the C. trachomatis-specific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is re-hydrated upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper instrument and an automated algorithm is applied to both the EC and C. trachomatis-specific signals to report specimen results as positive, negative, or EC failure.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
{4}
Page 5 of 25
a. Precision/Reproducibility
Reproducibility of the BD Viper System using the BD ProbeTec CT Qx Assay was evaluated at three clinical sites on one BD Viper System per site. A panel of simulated specimens was tested that comprised CT and GC organisms seeded into swab diluent for the BD ProbeTec CT Qx Assay. Simulated endocervical and urethral specimens contained a clean endocervical swab whereas the simulated urine and vaginal swab specimens did not. Uninoculated swab diluent for the BD ProbeTec CT Qx Assay was used for the CT negative samples. Nine replicates of each panel member were tested every day for five days on each BD Viper System. The data are summarized on page 20 Table 9.
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Control Set for the BD ProbeTec CT/GC Qx Amplified DNA Assays: 24 CT/GC Qx Positive Control Tubes containing approximately 2400 copies each of pCTB4 and pGCint3 linearized plasmids in carrier nucleic acid, and 24 CT/GC Qx Negative Controls Tubes containing carrier nucleic acid alone. The concentrations of the pCTB4 and pGCint3 plasmids are determined by UV spectrophotometry.
d. Detection limit:
CT QX Assay Analytical Sensitivity:
The Limits of Detection (LODs) for the CT QX Assay with C. trachomatis serovar H in urine and swab specimens when extracted on the BD Viper System were determined to be < 15 CT elementary bodies (EB) per mL for neat and Qx UPT urine and < 30 CT EB per mL for expressed vaginal, endocervical swab, and BD SurePath specimens. A correlation of EB to Inclusion-forming units (IFU) suggests that the CT QX assay LODs with serovar H in urine and swab specimens correspond to < 1 IFU per mL (15). The CT QX Assay on the BD Viper System in extracted mode was able to detect 16 isolates representing 15 CT serovars (A, B, Ba, C, D, E (2)*, F, G, H, I, J, K, LGV1, LGV2, and LGV3) with ≥ 95% proportion positive at a concentration of 15 EB per mL in CT/GC QX Swab Diluent.
* Testing with CT serovar E included both the type strain and the nvCT strain, a new variant with a deletion in the cryptic plasmid.
e. Analytical specificity:
DNA from 141 organisms listed in Table 1 was extracted on the BD Viper System and tested with the BD ProbeTec CT Qx Amplified DNA Assay. All potential cross-reactive species were tested at ≥ 1×10⁸ cells/mL except where noted. The CT Qx Assay did not cross-react with any of the organisms tested.
5
{5}
Page 6 of 25
Table 1: Potential Cross-reacting Microorganisms.
| Acinetobacter calcoaceticus | Epstein Barr Virus *** | Peptostreptococcus productus | Neisseria elongata subsp. nitroreduscens (2) |
| --- | --- | --- | --- |
| Acinetobacter lwoffii | Escherichia coli | Plesiomonas shigelloides | Neisseria elongata |
| Actinomyces israelii | Flavobacterium meningosepticum | Propionibacterium acnes | Neisseria flava (4) |
| Adenovirus*** | Gardnerella vaginalis | Providencia stuartii | Neisseria flavescens (4) |
| Aeromonas hydrophilia | Gemella haemolysans | Pseudomonas aeruginosa | Neisseria gonorrhoeae |
| Alcaligenes faecalis* | Haemophilus influenzae | Salmonella minnesota | Neisseria lactamica (7) |
| Bacillus subtilis* | Herpes Simplex Virus ** | Salmonella typhimurium | Neisseria meningitidis (12) |
| Bacteroides fragilis | Human papillomavirus (16 and 18)*** | Staphylococcus aureus | Neisseria mucosa (5) |
| Candida albicans* | Kingella kingae | Staphylococcus epidermidis | Neisseria perflava (8) |
| Candida glabrata* | Klebsiella pneumoniae | Streptococcus agalactiae | Neisseria polysaccharea (2) |
| Candida tropicalis* | Lactobacillus acidophilus* | Streptococcus mitis | Neisseria sicca (5) |
| Chlamydia pneumoniae*** | Lactobacillus brevis | Streptococcus mutans | Neisseria subflava (15) |
| Chlamydia psittaci* | Lactobacillus jensenii* | Streptococcus pneumoniae* | Neisseria weaverii (3) |
| Citrobacter freundii | Listeria monocytogenes | Streptococcus pyogenes | |
| Clostridium perfringens | Mobiluncus mulieris | Streptomyces griseus** | |
| Corynebacterium renale | Moraxella lacunata* | Trichomonas vaginalis** | |
| Cryptococcus neoformans* | Moraxella osloensis | Veillonella parvula | |
| Cytomegalovirus** | Morganella morganii | Vibrio parahaemolyticus | |
| Edwardsiella tarda | Mycobacterium gordonae | Yersinia enterocolitica | |
| Enterobacter cloacae | Mycobacterium smegmatis | Branhamella catarrhalis (5) | |
| Enterococcus faecalis | Peptostreptococcus anaerobius | Neisseria cinerea (2) | |
| Enterococcus faecium | Peptostreptococcus asaccharolyticus | Neisseria elongata subsp. Glycolytica | |
(n) number of strains tested in the BD ProbeTec CT Q⁴ Assay
* Tested at > 1×10⁷ cells or EB per mL; **Tested at > 1×10⁶ cells or viral particles per mL; ***Tested at ≥ 1×10⁶ genomic equivalents per mL; *** tested at ≥ 1×10⁵ TCID₅₀/mL
f. Assay cut-off:
{6}
The presence or absence of $C$ trachomatis DNA is determined by calculating the peak fluorescence (MaxRFU) over the course of the amplification process and by comparing this measurement to a predetermined threshold value. The magnitude of the MaxRFU score is not indicative of the level of organism in the specimen. If the $C$ trachomatis-specific signal is greater than or equal to a threshold of 125 MaxRFU, the EC fluorescence is ignored by the algorithm. If the $C$ trachomatis-specific signal is less than a threshold of 125 MaxRFU, the EC fluorescence is utilized by the algorithm in the interpretation of the result. If assay control results are not as expected, patient results are not reported.
# CT $\mathbf{Q}^{\mathrm{x}}$ Interfering Substances
The performance of the BD ProbeTec CT $\mathbf{Q}^{\mathrm{x}}$ Assay on the BD Viper System in extracted mode was evaluated in the presence of potential interfering substances which may be encountered in swab, urine and/or BD SurePath specimens. Potential interfering substances were spiked into $\mathbf{Q}^{\mathrm{x}}$ UPT urine and vaginal swab specimen matrices as well as BD SurePath specimens in LBC Specimen Dilution Tubes, in both the presence and the absence of CT elementary bodies (30 CT EB/mL in urine matrix and 90 CT EB/mL in swab/LBC Specimen Dilution Tube matrix). Results are summarized in Table 2.
Table 2: CT Q ${}^{x}$ Interfering Substances.
| Interpretation | Swab | Urine | SurePath |
| --- | --- | --- | --- |
| No Interference Observed | Blood (≤ 60%) Seminal Fluid Mucus Over The Counter vaginal products and contraceptives Hemorrhoidal cream Prescription vaginal treatments Leukocytes (1x10^6 cells/mL) 1x10^6 cells/mL Neisseria gonorrhoeae | Blood (≤1%) Seminal fluid Mucus Antibiotics Analgesics Phenazopyridine Over The Counter deodorant sprays and powders Hormones Leukocytes Albumin <1 mg/mL Glucose Acidic urine (pH 4.0) Alkaline urine (pH 9.0) Bilirubin 1x10^6 cells/mL Neisseria gonorrhoeae Organisms associated with Urinary Tract Infections | Blood (≤ 1%) Seminal Fluid Mucus Over The Counter vaginal products and contraceptives Hemorrhoidal cream Prescription vaginal treatments Leukocytes (1x10^6 cells/mL) 1x10^6 cells/mL Neisseria gonorrhoeae |
| May cause extraction control (EC) failures | Blood (> 60%) | Not applicable | Not applicable |
| May cause False Negative results | Not applicable | Not applicable | Not applicable |
{7}
Page 8 of 25
2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Not Applicable
3. Clinical studies:
Swab and Urine Specimen Clinical Study
Clinician-collected endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female Qx UPT and neat urine specimens were collected from 1059 female subjects and 479 male subjects attending OB/GYN, sexually transmitted disease (STD) and family planning clinics at seven geographically diverse clinical sites in North America. Subjects were classified as symptomatic if they reported symptoms such as dysuria, urethral discharge, coital pain/difficulty/bleeding, testicular or scrotum pain/swelling, abnormal vaginal discharge, or pelvic/uterine/adnexal pain. Subjects were classified as asymptomatic if they did not report symptoms. Sixty five female subjects and 7 male subjects were excluded from the data analysis due to age requirement violations, antibiotic treatment in the last 21 days, opting to withdraw from the study after initially consenting, failure to obtain paired swab and urine specimens, urine quantity less than 20 mL, or transport and storage errors related to specimen collection. Therefore, the final data analysis included 994 compliant female subjects and 472 compliant male subjects. Five specimens were collected from each of the 994 eligible female subjects. A urine specimen was collected and split into Qx UPT, neat urine and the two reference urine specimen collection devices followed by a vaginal swab specimen and three randomized endocervical swab specimens. Up to four specimens were collected from each of the 472 eligible male subjects. Up to three randomized urethral swab specimens were collected followed by a urine specimen that was split into Qx UPT, neat urine and the two reference urine specimen collection devices. BD ProbeTec CT Qx assay results were generated from the Qx UPT and neat urine specimens, the vaginal swab specimen, one endocervical swab specimen and one male urethral swab specimen. The remaining two endocervical swab specimens, up to two male urethral swab specimens, and the two reference urine specimens for each male and female subject were tested using two reference methods: the BD ProbeTec ET CT/AC assay and another commercially available NAAT (Nucleic Acid Amplification Test). Specimen testing was conducted either at the site of specimen collection or at a designated BD Viper testing site. All performance calculations were based on the total number of BD ProbeTec CT Qx assays results for endocervical, vaginal and male urethral swab specimens, and male and female Qx UPT and neat urine specimens compared to a patient infected status (PIS) algorithm for each gender. In the algorithm, the designation of a subject as being infected with CT or not was based on endocervical swab and urine specimen results from the commercially available BD ProbeTec ET CT/AC assay and the other commercially available NAAT. Subjects were considered infected with CT if two of the four endocervical swab and urine
{8}
Page 9 of 25
specimens (or two of the three or four urethral swab and urine specimens) tested positive in the BD ProbeTec ET CT/AC assay and the other reference NAAT (one specimen testing positive in each NAAT). Subjects were considered non-infected if less than two reference NAAT results were positive. A total of 5388 BD ProbeTec CT Qx assay results was used to calculate sensitivity and specificity. Performance of the assay with endocervical swabs, patient collected vaginal swab specimens (in a clinical setting), female UPT and neat urine was assessed in the clinical study. Separate performance was calculated for specimens collected from pregnant females. For the latter, sensitivity compared to patient infected status for FS was 62.5% (5/8): the test and reference NAAT swab specimens were negative; the test and reference NAAT urine specimens were positive yielding a PIS positive result. FV sensitivity was 75% (6/8): the test and reference NAAT swab specimens were negative; the test and reference NAAT urine specimens were positive yielding a PIS positive result. FNU and FUPT sensitivity were 100% (8/8). Specificity was 94.7% (18/19) for FS, FV, FNU, and FUPT separately.
a. Clinical Sensitivity:
Sensitivity compared to patient infected status for FS was 62.5% (5/8): the test and reference NAAT swab specimens were negative; the test and reference NAAT urine specimens were positive yielding a PIS positive result. FV sensitivity was 75% (6/8): the test and reference NAAT swab specimens were negative; the test and reference NAAT urine specimens were positive yielding a PIS positive result. FNU and FUPT sensitivity were 100% (8/8).
Neat and Q^x UPT Urine Stability
Pools of CT negative male and female urine specimens were used in analytical experiments to support the urine storage and transport stability claims. For neat urine, pools were co-spiked with CT serovar H and GC strain ATCC 19424 at 45 EB per mL and 150 cells per mL, respectively. Neat urine specimens were stored at either 2-8°C for 1, 3 or 7 days; or at 30°C for 8, 24 or 30 h; or at -20°C for 180 days. At each time point, samples were removed from storage and tested with the BD ProbeTec CT Q^x Assay on the BD Viper System in extracted mode. Thirty-two assay replicates were generated for each condition (sample type/temperature/duration). The expected results were obtained with the CT Q^x assay under all conditions tested.
For Q^x UPT urine, pooled specimens were co-spiked with CT serovar H and GC strain ATCC 19424 at 45 EB per mL and 150 cells per mL, respectively. The spiked urine specimen pools were then stored at either 2-8°C for 24 h or 30°C for 8 h prior to transfer into Q^x UPT tubes. The Q^x UPT specimen pools were then stored either at 2-8°C for 14, 21 or 30 days; or at 30°C for 14, 21 or 30 days; or at -20°C for 180 days. At each time point Q^x UPT specimens were removed from storage and tested with the BD ProbeTec CT Q^x Assay on the BD Viper System in extracted mode. Thirty-two assay replicates were generated for each condition (sample type/temperature/duration). The expected results were obtained with the CT Q^x assay under all conditions tested.
9
{9}
Page 10 of 25
# Vaginal Dry and Expressed Swab Stability
Pools of CT negative vaginal swab matrix were used in analytical experiments to support the storage and transport stability claims for dry vaginal swab specimens. Pools were co-spiked with CT serovar H and GC strain ATCC 19424 to achieve 90 EB per mL and 300 cells per mL, respectively, when seeded onto swabs and expressed in CT/GC Q⁸ Swab Diluent. Seeded dry swabs were stored at 2-8°C for 3, 7, or 14 days; or at 30°C for 3, 7 or 14 days; or at -20°C for 30, 60, or 180 days. At each time point, dry swabs were removed from storage and expressed into 2 mL of CT/GC Q⁸ Swab Diluent and evaluated with the BD ProbeTec CT Q⁸ Assay on the BD Viper System in extracted mode. Thirty-two assay replicates were generated for each condition (sample type/temperature/duration). The expected results were obtained with the CT Q⁸ assay under all conditions tested.
Pools of CT negative vaginal swab matrix were used in analytical experiments to support the storage and transport stability claims for expressed vaginal swab specimens. Pools were spiked with CT serovar H and GC strain ATCC 19424 to achieve 90 EB per mL and 300 cells per mL, respectively. The spiked swab matrix was stored at 2-8°C for 7, 14 or 30 days; or at 30°C for 7, 14 or 30 days; or at -20°C for 30, 60, or 180 days. At each time point, samples were removed from storage and tested with the BD ProbeTec CT Q⁸ Assay on the BD Viper System in extracted mode. Thirty-two assay replicates were generated for each condition (sample type/temperature/duration). The expected results were obtained with the CT Q⁸ assay under all conditions tested.
# Endocervical and Urethral Swab Specimen Stability
Pools of CT negative endocervical swab matrix were used in analytical experiments to support the storage and transport stability claims for endocervical and urethral swab specimens. Pools of swab matrix were spiked with CT serovar H and GC strain ATCC 19424 at 90 EB per mL and 300 cells per mL, respectively. The pools were dispensed in 2 mL volumes into BD sample tubes to simulate "wet" endocervical specimens and stored at either 2-8°C for 7, 14 or 30 days; or at 30°C for 7, 14 or 30 days; or at -20°C for 30, 60, or 180 days. At each time point, samples were removed from storage and tested with the BD ProbeTec CT Q⁸ Assay on the BD Viper System in extracted mode. Thirty-two assay replicates were generated for each condition (sample type/temperature/duration). The expected results were obtained with the CT Q⁸ assay under all conditions tested.
# Post Pre-warm Specimen Stability
Pools of male and female CT negative neat urine were used in analytical experiments to support the storage stability claims for pre-warmed neat and Q⁴ UPT urine specimens. Pooled specimens were spiked with CT serovar H and GC strain ATCC 19424 at 45 EB per mL and 150 cells per mL, respectively and either added to Q⁴ UPT tubes or left untreated as neat urine. Both specimen types were pre-warmed at 114°C for 15 min, and cooled for 15 min. After the pre-warm process, specimen tubes were stored at either 2-8°C for 1, 3 or 7 days; or at 30°C for 1, 3 or 7 days; or at -20°C for 30 or 180 days. At each time point samples were removed from storage and tested
10
{10}
Page 11 of 25
with the BD ProbeTec CT Qx Assay on the BD Viper System in extracted mode. Thirty-two assay replicates were generated for each condition (sample type/temperature/duration). The expected results were obtained with the CT Qx assay under all conditions tested.
Pools of CT negative vaginal and endocervical swab specimen matrices in CT/GC Qx Swab Diluent were used in analytical experiments to support the storage stability claims for pre-warmed expressed vaginal, endocervical, and male urethral swab specimens. For both types of matrix, pooled specimens were spiked with CT serovar H and GC strain ATCC 19424 at 90 EB per mL and 300 cells per mL, respectively and aliquotted into 2 mL volumes in BD specimen tubes. The tubes were pre-warmed at 114°C for 15 min and cooled for 15 min. After the pre-warm process, the specimen tubes were stored either at 2-8°C for 3 or 7 days; or at 30°C for 3 or 7 days; or at -20°C for 30 or 180 days. At each time point, samples were removed from storage and tested with the BD ProbeTec CT Qx Assay on the BD Viper System in extracted mode. Thirty-two assay replicates were generated for each condition (sample type/temperature/duration). The expected results were obtained with the CT Qx assay under all conditions tested.
## BD SurePath Specimen Stability
Pools of CT and GC negative BD SurePath clinical specimens were used in analytical experiments to support the storage and stability claims. Pools were co-spiked with CT serovar H and GC strain ATCC 19424 to achieve 90 EB per mL and 300 cells per mL, respectively. The pools were dispensed in 10 mL volumes in BD SurePath vials and stored at either 2-8°C or 30°C. After 30 days, 0.5 mL from each vial was removed and added to an LBC Specimen Dilution Tube. The specimens in the LBC Specimen Dilution Tube were then stored at 2-8°C for 30 or 90 days; or at 30°C for 30 or 90 days; or at -20°C for 90 days. At each time point, samples were removed from storage and tested with the BD ProbeTec CT Qx Assay on the BD Viper System in extracted mode. Twenty-four assay replicates were generated for each condition (temperature/duration). The expected results were obtained with the CT Qx assay under all conditions tested.
Hypothetical positive and negative predictive values (PPV & NPV) for the CT Qx Assay with swab and urine specimens are shown in Table 3A. Hypothetical positive and negative predictive values (PPV & NPV) for the CT Qx Assay from the multi-center clinical trial for BD SurePath specimens are shown in Table 3B. Hypothetical positive and negative predictive values (PPV & NPV) for the CT Qx Assay from the multi-center clinical trial for PreservCyt specimens are shown in Table 3C. These calculations are based on hypothetical prevalence and overall sensitivity and specificity (compared to the patient infected status) of 94.5% and 98.9% for swab and urine specimens, of 95.0% and 99.7% for BD SurePath specimens, and of 94.1% and 99.8% for PreservCyt specimens.
11
{11}
Page 12 of 25
Table 3A: CT Hypothetical Positive and Negative Predictive Values (Swabs/Urines) Compared to Patient Infected Status.
| Prevalence (%) | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) |
| --- | --- | --- | --- | --- |
| 2 | 94.5 | 98.9 | 64.1 | 99.9 |
| 5 | 94.5 | 98.9 | 82.1 | 99.7 |
| 10 | 94.5 | 98.9 | 90.7 | 99.4 |
| 20 | 94.5 | 98.9 | 95.6 | 98.6 |
| 30 | 94.5 | 98.9 | 97.4 | 97.7 |
| 40 | 94.5 | 98.9 | 98.3 | 96.4 |
| 50 | 94.5 | 98.9 | 98.9 | 94.7 |
Table 3B: CT Hypothetical Positive and Negative Predictive Values (BD SurePath) Compared to Patient Infected Status.
| Prevalence (%) | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) |
| --- | --- | --- | --- | --- |
| 2 | 95.0 | 99.7 | 86.6 | 99.9 |
| 5 | 95.0 | 99.7 | 94.3 | 99.7 |
| 10 | 95.0 | 99.7 | 97.2 | 99.4 |
| 20 | 95.0 | 99.7 | 98.8 | 98.8 |
| 30 | 95.0 | 99.7 | 99.3 | 97.9 |
| 40 | 95.0 | 99.7 | 99.5 | 96.8 |
| 50 | 95.0 | 99.7 | 99.7 | 95.2 |
Table 3C: CT Hypothetical Positive and Negative Predictive Values (PreservCyt) Compared to Patient Infected Status.
| Prevalence (%) | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) |
| --- | --- | --- | --- | --- |
| 2 | 94.1 | 98.8 | 90.6 | 99.9 |
| 5 | 94.1 | 98.8 | 96.1 | 99.7 |
| 10 | 94.1 | 98.8 | 98.1 | 99.3 |
| 20 | 94.1 | 98.8 | 99.2 | 98.5 |
| 30 | 94.1 | 98.8 | 99.5 | 97.5 |
| 40 | 94.1 | 98.8 | 99.7 | 96.2 |
| 50 | 94.1 | 98.8 | 99.8 | 94.4 |
12
{12}
A total of 5388 CT Q $^x$ Assay results from swab and urine specimens was evaluated from seven geographically diverse clinical sites. A frequency distribution of the initial MaxRFU values for the CT Q $^x$ assay is shown in Figure A and B. The distribution of MaxRFU values from CT Q $^x$ true positive, true negative, false positive and false negative specimens (i.e., from those specimens that yielded results which were discordant with the patient infected status (PIS)) is shown in Table 4A.
A total of 1714 CT Q $^x$ Assay results from BD SurePath specimens was evaluated from eleven geographically diverse clinical sites. A frequency distribution of the initial MaxRFU values for the CT Q $^x$ assay is shown in Figure B. The distribution of MaxRFU values from CT Q $^x$ true positive, true negative, false positive and false negative specimens (i.e., from those specimens that yielded results which were discordant with the patient infected status (PIS)) is shown in Table 4B.

Figure A: Frequency Distribution of MaxRFU for the CT Q $^x$ Assay (Swabs and Urines).
{13}
Page 14 of 25

Figure B: Frequency Distribution of MaxRFU for the CT Q $^x$ Assay (BD SurePath Specimens).
Table 4A: CT Q $^x$ MaxRFU Ranges for False Negative, False Positive, True Negative and True Positive Results (Swabs/Urines).
| | MaxRFU range | 0-49 | 50-99 | 100-124 | 125-149 | 150-199 | 200-249 | 250-349 | 350-499 | 500-799 | ≥800 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| n | | 4590 | 26 | 1 | 1 | 3 | 3 | 0 | 5 | 4 | 755 |
| FN | FNU | 8 | 0 | 0 | | | | | | | |
| | FS | 10 | 0 | 0 | | | | | | | |
| | FUPT | 8 | 0 | 0 | | | | | | | |
| | FV | 4 | 0 | 0 | | | | | | | |
| | MNU | 2 | 0 | 0 | | | | | | | |
| | MS | 8 | 0 | 0 | | | | | | | |
| | MUPT | 2 | 0 | 0 | | | | | | | |
| | Total | 42 | 0 | 0 | | | | | | | |
| FP | FNU | | | | 0 | 1 | 0 | 0 | 0 | 0 | 4 |
| | FS | | | | 0 | 0 | 0 | 0 | 2 | 2 | 11 |
| | FUPT | | | | 0 | 0 | 2 | 0 | 0 | 0 | 5 |
| | FV | | | | 0 | 1 | 0 | 0 | 0 | 0 | 6 |
| | MNU | | | | 0 | 0 | 0 | 0 | 0 | 1 | 2 |
| | MS | | | | 0 | 1 | 0 | 0 | 0 | 0 | 5 |
| | MUPT | | | | 1 | 0 | 1 | 0 | 0 | 0 | 5 |
| | Total | | | | 1 | 3 | 3 | 0 | 2 | 3 | 38 |
| TN | FNU | 868 | 5 | 0 | | | | | | | |
| | FS | 857 | 6 | 0 | | | | | | | |
| | FUPT | 866 | 5 | 0 | | | | | | | |
| | FV | 866 | 4 | 1 | | | | | | | |
| | MNU | 368 | 0 | 0 | | | | | | | |
{14}
Page 15 of 25
| | MaxRFU range | 0-49 | 50-99 | 100-124 | 125-149 | 150-199 | 200-249 | 250-349 | 350-499 | 500-799 | ≥800 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| n | | 4590 | 26 | 1 | 1 | 3 | 3 | 0 | 5 | 4 | 755 |
| | MS | 364 | 1 | 0 | | | | | | | |
| | MUPT | 359 | 5 | 0 | | | | | | | |
| | Total | 4548 | 26 | 1 | | | | | | | |
| TP | FNU | | | | 0 | 0 | 0 | 0 | 2 | 1 | 104 |
| | FS | | | | 0 | 0 | 0 | 0 | 1 | 0 | 104 |
| | FUPT | | | | 0 | 0 | 0 | 0 | 0 | 0 | 107 |
| | FV | | | | 0 | 0 | 0 | 0 | 0 | 0 | 111 |
| | MNU | | | | 0 | 0 | 0 | 0 | 0 | 0 | 99 |
| | MS | | | | 0 | 0 | 0 | 0 | 0 | 0 | 93 |
| | MUPT | | | | 0 | 0 | 0 | 0 | 0 | 0 | 99 |
| | Total | | | | 0 | 0 | 0 | 0 | 3 | 1 | 717 |
Table 4B: CT Q $^s$ MaxRFU Ranges for False Negative, False Positive, True Negative and True Positive Results (BD SurePath Specimens).
| MaxRFU Range | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 0-49 | 50-99 | 100-124 | 125-149 | 150-199 | 200-249 | 250-349 | 350-499 | 500-799 | ≥800 |
| FN | 7 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| FP | 0 | 0 | 0 | 0 | 0 | 1 | 1 | 1 | 0 | 2 |
| TN | 1562 | 6 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| TP | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 133 |
| Total | 1569 | 6 | 1 | 0 | 0 | 1 | 1 | 1 | 0 | 135 |
# Controls
During the swab/urine clinical evaluation, there were no CT Q $^s$ positive control failures from 253 CT Q $^s$ plate runs. For the CT Q $^s$ negative control, a failure was observed in 1 of 253 CT Q $^s$ plate runs. During the BD SurePath specimen clinical evaluation, there was one CT Q $^s$ positive control failure from 120 CT Q $^s$ plates that were run. The CT/GC Q $^s$ positive and negative control MaxRFU values observed in the clinical trials are shown in Table 5.
{15}
Page 16 of 25
Table 5: Distribution of MaxRFU Results for the CT Q $^{\text{x}}$ Assay Negative and Positive Controls
| Control | Statistic | Swab and Urine Specimen Clinical Study | BD SurePath specimen Clinical Study |
| --- | --- | --- | --- |
| CT Qx Negative Control | N | 252 | 119 |
| MaxRFU | Maximum | 41 | 11 |
| | 95th Percentile | 4 | 0 |
| | Median | 0 | 0 |
| | Mean | 1 | 0 |
| | 5th Percentile | 0 | 0 |
| | Minimum | 0 | 0 |
| | | | |
| CT Qx Positive Control | N | 253 | 119 |
| MaxRFU | Maximum | 2378 | 2173 |
| | 95th Percentile | 2184 | 2128 |
| | Median | 1968 | 1942 |
| | Mean | 1939 | 1864 |
| | 5th Percentile | 1597 | 1340 |
| | Minimum | 629 | 168 |
# Swab and Urine Specimen Clinical Study
Clinician-collected endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female $\mathrm{Q}^{\mathrm{x}}$ UPT and neat urine specimens were collected from 1059 female subjects and 479 male subjects attending OB/GYN, sexually transmitted disease (STD) and family planning clinics at seven geographically diverse clinical sites in North America. Subjects were classified as symptomatic if they reported symptoms such as dysuria, urethral discharge, coital pain/difficulty/bleeding, testicular or scrotum pain/swelling, abnormal vaginal discharge, or pelvic/uterine/adnexal pain. Subjects were classified as asymptomatic if they did not report symptoms. Sixty five female subjects and 7 male subjects were excluded from the data analysis due to age requirement violations, antibiotic treatment in the last 21 days, opting to withdraw from the study after initially consenting, failure to obtain paired swab and urine specimens, urine quantity less than $20~\mathrm{mL}$ , or transport and storage errors related to specimen collection.
{16}
Page 17 of 25
Therefore, the final data analysis included 994 compliant female subjects and 472 compliant male subjects.
Five specimens were collected from each of the 994 eligible female subjects. A urine specimen was collected and split into Q⁴ UPT, neat urine and the two reference urine specimen collection devices followed by a vaginal swab specimen and three randomized endocervical swab specimens. Up to four specimens were collected from each of the 472 eligible male subjects. Up to three randomized urethral swab specimens were collected followed by a urine specimen that was split into Q⁴ UPT, neat urine and the two reference urine specimen collection devices. BD ProbeTec CT Q⁸ assay results were generated from the Q⁴ UPT and neat urine specimens, the vaginal swab specimen, one endocervical swab specimen and one male urethral swab specimen. The remaining two endocervical swab specimens, up to two male urethral swab specimens, and the two reference urine specimens for each male and female subject were tested using two reference methods: the BD ProbeTec ET CT/AC assay and another commercially available NAAT (Nucleic Acid Amplification Test). Specimen testing was conducted either at the site of specimen collection or at a designated BD Viper testing site.
All performance calculations were based on the total number of BD ProbeTec CT Q⁴ assays results for endocervical, vaginal and male urethral swab specimens, and male and female Q⁴ UPT and neat urine specimens compared to a patient infected status (PIS) algorithm for each gender. In the algorithm, the designation of a subject as being infected with CT or not was based on endocervical swab and urine specimen results from the commercially available BD ProbeTec ET CT/GC/AC assay and the other commercially available NAAT. Subjects were considered infected with CT if two of the four endocervical swab and urine specimens (or two of the three or four urethral swab and urine specimens) tested positive in the BD ProbeTec ET CT/AC assay and the other reference NAAT (one specimen testing positive in each NAAT). Subjects were considered non-infected if less than two reference NAAT results were positive. A total of 5388 BD ProbeTec CT Q⁴ assay results was used to calculate sensitivity and specificity.
Performance of the assay with endocervical swabs, patient collected vaginal swab specimens (in a clinical setting), female UPT and neat urine was assessed in the clinical study. Separate performance was calculated for specimens collected from pregnant females. Sensitivity compared to patient infected status for FS was 62.5% (5/8): the test and reference NAAT swab specimens were negative; the test and reference NAAT urine specimens were positive yielding a PIS positive result. FV sensitivity was 75% (6/8): the test and reference NAAT swab specimens were negative; the test and reference NAAT urine specimens were positive yielding a PIS positive result. FNU and FUPT sensitivity were 100% (8/8). Specificity was 94.7% (18/19) for FS, FV, FNU, and FUPT separately.
17
{17}
Page 18 of 25
# BD SurePath Specimen Clinical Study
Endocervical swab specimens and BD SurePath specimens were collected from 1728 compliant female subjects attending family planning, OB/GYN, and sexually transmitted disease clinics at eleven geographically diverse clinical sites in North America. Subjects were classified as symptomatic if they reported symptoms such as dysuria, coital pain/difficulty/bleeding, abnormal vaginal discharge, or pelvic/uterine/adnexal pain. Subjects were classified as asymptomatic if they did not report symptoms.
Three randomized endocervical swab specimens and a BD SurePath specimen were collected from each female subject. The three reference endocervical swabs were tested with the BD ProbeTec ET CT/GC/AC assay, the BD ProbeTec CT Q® assay, and another commercially available NAAT (Nucleic Acid Amplification Test). Sensitivity and specificity for BD SurePath specimens were calculated by comparing results to a patient infected status (PIS) algorithm. The designation of positive or negative PIS was based on the endocervical swab specimen results from the three reference methods. At least two positive reference results were required to establish a subject as PIS-positive. At least two negative reference results were required to establish a subject as PIS-negative.
The distribution of cervical sampling devices used in the clinical study according to clinical collection site is summarized in Table 6.
Table 7 summarizes the number of results from symptomatic and asymptomatic subjects designated as infected or non-infected with CT according to the PIS algorithm.
Table 8 summarizes the CT Q® Assay Performance for BD SurePath Specimens Compared to Patient Infected Status (by symptomatic status).
Table 6: Summary of Cervical Sampling Devices Used in the BD SurePath specimen Clinical Study
| Cervical Sampling Device Used | Clinical Collection Site Number | | | | | | | | | | | Total |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | |
| Broom-Type Device | 54 | 50 | 511 | 18 | 374 | 0 | 127 | 0 | 0 | 71 | 0 | 1205 |
| Spatula/Cytobrush | 0 | 25 | 0 | 0 | 182 | 112 | 32 | 24 | 103 | 8 | 37 | 523 |
{18}
Page 19 of 25
Table 7: CT Q $^x$ Assay Performance for Swabs and Urines Compared to Patient Infected Status (by specimen type and symptomatic status).
| Specimen Type | Symptomatic Status | N | Sensitivity | 95% C.I. | Specificity | 95% C.I. | PPV% | NPV% | Error Initial/Final |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| FS | A | 450 | 93.0% (53/57) | (83.0% - 98.1%) | 98.0% (385/393) | (96.0% - 99.1%) | 86.9 | 99.0 | 2/0 |
| | S | 543 | 89.7% (52/58) | (78.8% - 96.1%) | 98.6% (478/485) | (97.0% - 99.4%) | 88.1 | 98.8 | 1/1 |
| | Total | 993 | 91.3% (105/115) | (84.6% - 95.8%) | 98.3% (863/878) | (97.2% - 99.0%) | 87.5 | 98.9 | 3/1 |
| FV | A | 449 | 98.2% (56/57) | (90.6% - 100.0%) | 99.5% (390/392) | (98.2% - 99.9%) | 96.6 | 99.7 | 0/0 |
| | S | 544 | 94.8% (55/58) | (85.6% - 98.9%) | 99.0% (481/486) | (97.6% - 99.7%) | 91.7 | 99.4 | 0/0 |
| | Total | 993 | 96.5% (111/115) | (91.3% - 99.0%) | 99.2% (871/878) | (98.4% - 99.7%) | 94.1 | 99.5 | 0/0 |
| FNU | A | 450 | 93.0% (53/57) | (83.0% - 98.1%) | 100.0% (393/393) | (99.1% - 100.0%) | 100.0 | 99.0 | 0/0 |
| | S | 543 | 93.1% (54/58) | (83.3% - 98.1%) | 99.0% (480/485) | (97.6% - 99.7%) | 91.5 | 99.2 | 0/0 |
| | Total | 993 | 93.0% (107/115) | (86.8% - 96.9%) | 99.4% (873/878) | (98.7% - 99.8%) | 95.5 | 99.1 | 0/0 |
| FUPT | A | 450 | 94.7% (54/57) | (85.4% - 98.9%) | 99.5% (391/393) | (98.2% - 99.9%) | 96.4 | 99.2 | 0/0 |
| | S | 543 | 91.4% (53/58) | (81.0% - 97.1%) | 99.0% (480/485) | (97.6% - 99.7%) | 91.4 | 99.0 | 0/0 |
| | Total | 993 | 93.0% (107/115) | (86.8% - 96.9%) | 99.2% (871/878) | (98.4% - 99.7%) | 93.9 | 99.1 | 0/0 |
| MS | A | 215 | 88.6% (31/35) | (73.3% - 96.8%) | 98.9% (178/180) | (96.0% - 99.9%) | 93.9 | 97.8 | 1/0 |
| | S | 257 | 93.9% (62/66) | (85.2% - 98.3%) | 97.9% (187/191) | (94.7% - 99.4%) | 93.9 | 97.9 | 1/0 |
| | Total | 472 | 92.1% (93/101) | (85.0% - 96.5%) | 98.4% (365/371) | (96.5% - 99.4%) | 93.9 | 97.9 | 2/0 |
| MNU | A | 215 | 100.0% (35/35) | (90.0% - 100.0%) | 98.9% (178/180) | (96.0% - 99.9%) | 94.6 | 100.0 | 0/0 |
| | S | 257 | 97.0% (64/66) | (89.5% - 99.6%) | 99.5% (190/191) | (97.1% - 100.0%) | 98.5 | 99.0 | 0/0 |
| | Total | 472 | 98.0% (99/101) | (93.0% - 99.8%) | 99.2% (368/371) | (97.7% - 99.8%) | 97.1 | 99.5 | 0/0 |
{19}
Page 20 of 25
| Specimen Type | Symptomatic Status | N | Sensitivity | 95% C.I. | Specificity | 95% C.I. | PPV% | NPV% | Error Initial/Final |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| MUPT | A | 215 | 100.0% (35/35) | (90.0% - 100.0%) | 98.9% (178/180) | (96.0% - 99.9%) | 94.6 | 100.0 | 0/0 |
| | S | 257 | 97.0% (64/66) | (89.5% - 99.6%) | 97.4% (186/191) | (94.0% - 99.1%) | 92.8 | 98.9 | 0/0 |
| | Total | 472 | 98.0% (99/101) | (93.0% - 99.8%) | 98.1% (364/371) | (96.2% - 99.2%) | 93.4 | 99.5 | 0/0 |
| Total | | 5388 | 94.5% (721/763) | (92.6% - 96.0%) | 98.9% (4575/4625) | (98.6% - 99.2%) | 93.5 | 99.1 | 5/1 |
Note: For female subjects, infections localized to the endocervix or urethra have been reported in the literature (16-20). Analyses were performed on the endocervical swab specimens and female UPT and neat urine specimens to further characterize the ten negative female endocervical swabs (105/115) and the eight negative female UPT and neat urine specimens (107/115).
- Of the 115 female subjects defined as positive by the PIS algorithm, ten had infections localized to the urethra as indicated by the urine reference result (i.e, BD ProbeTec ET CT/AC assay and other NAAT endocervical swab specimens were negative, BD ProbeTec ET CT/AC assay and other NAAT urine specimens were positive.) The BD ProbeTec CT Qx assay was negative for nine of the ten endocervical swab specimens from these subjects.
- Of the 115 female subjects defined as positive by the PIS algorithm, three had infections localized to the endocervix as indicated by the endocervical reference result (i.e, BD ProbeTec ET CT/AC assay and other NAAT urine specimens were negative, BD ProbeTec ET CT/AC assay and other NAAT endocervical swab specimens were positive.) The BD ProbeTec CT Qx assay was negative for UPT and neat urine for these three subjects.
Table 8: CT Qᵀ Assay Performance for BD SurePath Specimens Compared to Patient Infected Status (by symptomatic status).
| | | Performance Compared to Patient Infected Status | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Symptomatic Status | N | Sensitivity | 95% C.I. | Specificity | 95% C.I. | PPV % | NPV % | Error Initial/Final |
| A | 115
6 | 92.7%
(76/82) | (84.8% - 97.3%) | 99.7%
(1071/1074) | (99.2% - 99.9%) | 95.9 | 99.4 | 2/1 |
| S | 558 | 98.3%
(57/58) | (90.8% - 100.0%) | 99.6%
(498/500) | (98.6% - 100.0%) | 96.6 | 99.8 | 0/0 |
| Total | 171
4¹ | 95.0%
(133/140) | (90.0% - 98.0%) | 99.7%
(1569/1574) | (99.3% - 99.9%) | 96.6 | 99.6 | 2/1 |
## Reproducibility
Reproducibility of the BD Viper System using the BD ProbeTec CT Qx Assay was evaluated at three clinical sites on one BD Viper System per site. A panel of simulated specimens was tested that comprised CT and GC organisms seeded into swab diluent for the BD ProbeTec CT Qx Assay. Simulated endocervical and urethral specimens contained a clean endocervical swab whereas the
¹ Of the 1728 compliant female subjects, 14 subjects did not have a CT Qᵀ assay result for the BD SurePath specimen, therefore the final data analysis included 1714 compliant female subjects.
20
{20}
simulated urine and vaginal swab specimens did not. Uninoculated swab diluent for the BD ProbeTec CT Qx Assay was used for the CT negative samples. Nine replicates of each panel member were tested every day for five days on each BD Viper System. The data are summarized in Table 9.
Table 9: Summary of Reproducibility Data for Swabs and Urines on the BD Viper System for the CT Qx Assay.
| | | | | | | Within Run | | Between Runs Within Site | | Between Site | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Specimen Type | CT EB's/mL | GC Cells/mL | % Correct | 95% CI | MaxRF UMean | SD | %CV | SD | %CV | SD | %CV |
| Endocervical / Urethra1 | 0 | 0 | 98.5% (133/135) | (94.8-99.8%) | 29.9 | 233.0 | 778.5 | 0.0 | 0.0 | 33.9 | 113.4 |
| | 30 | 0 | 100.0% (135/135) | (97.3-100.0%) | 2011.2 | 114.1 | 5.7 | 0.0 | 0.0 | 14.8 | 0.7 |
| | 0 | 100 | 100.0% (135/135) | (97.3-100.0%) | 1.4 | 6.0 | 442.7 | 1.0 | 76.9 | 0.0 | 0.0 |
| | 30 | 250 | 100.0% (135/135) | (97.3-100.0%) | 1991.9 | 118.0 | 5.9 | 17.6 | 0.9 | 10.4 | 0.5 |
| | 75 | 100 | 100.0% (135/135) | (97.3-100.0%) | 1954.8 | 169.4 | 8.7 | 0.0 | 0.0 | 0.0 | 0.0 |
| Urine/Vaginal | 0 | 0 | 100.0% (135/135) | (97.3-100.0%) | 0.9 | 5.0 | 542.4 | 0.0 | 0.0 | 0.0 | 0.0 |
| | 30 | 0 | 100.0% (135/135) | (97.3-100.0%) | 1999.8 | 131.8 | 6.6 | 34.2 | 1.7 | 0.0 | 0.0 |
| | 0 | 100 | 100.0% (135/135) | (97.3-100.0%) | 0.8 | 3.4 | 442.4 | 0.0 | 0.0 | 0.0 | 0.0 |
| | 30 | 250 | 100.0% (135/135) | (97.3-100.0%) | 1995.2 | 125.8 | 6.3 | 33.1 | 1.7 | 52.9 | 2.7 |
| | 75 | 100 | 100.0% (135/135) | (97.3-100.0%) | 2014.4 | 109.5 | 5.4 | 0.0 | 0.0 | 0.0 | 0.0 |
A second study was conducted internally to characterize the reproducibility of test results (i.e., proportion positive or negative) at target levels below the analytical Limit of Detection (LOD) of the BD ProbeTec CT Qx Assay. A panel of simulated specimens was tested that comprised CT and GC organisms seeded into swab diluent at two different levels (1:10, 1:100) each of which was below the analytical LOD for the respective organism. These levels were selected to fall within the dynamic range of the analytical LOD curve for this assay. Fifteen replicates of each panel member were tested every day for five days across three BD Viper Systems. The data are summarized in Table 10.
{21}
Page 22 of 25
Table 10: Characterization of System Reproducibility at Target Levels below the Analytical Limit of Detection for the CT Q $^{\text{x}}$ Assay for Swabs and Urines.
| Specimen Type | Dilution of Analytical LOD | % Positive | 95% CI (Positive) | Max RFU Mean (Positive) | % Negative | 95% CI (Negative) | Max RFU Mean (Negative) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Endocervical/Urethral | 1:10 | 70.2 (158/225) | (63.8, 76.1) | 1794.2 | 29.8 (67/225) | (23.9, 36.2) | 2.6 |
| Endocervical/Urethral | 1:100 | 10.2 (23/225) | (6.6,14.9) | 1643.8 | 89.8 (202/225) | (85.1, 93.4) | 1.6 |
| Urine/Vaginal | 1:10 | 64.4 (145/225) | (57.8, 70.7) | 1733.9 | 35.6 (80/225) | (29.3, 42.2) | 4.6 |
| Urine/Vaginal | 1:100 | 10.7 (24/225) | (7.0, 15.5) | 1666.6 | 89.3 (201/225) | (84.5, 93.0) | 2.4 |
A reproducibility study of the BD Viper System using the BD ProbeTec CT Q $^{\text{x}}$ Assay was also conducted for Liquid Based Cytology (LBC) specimens at three clinical sites on one BD Viper System per site. A panel of simulated specimens comprising CT and GC organisms seeded into LBC Specimen Dilution Tubes containing LBC medium was tested with the BD ProbeTec CT Q $^{\text{x}}$ Assay. Uninoculated LBC Specimen Dilution Tubes containing LBC medium were used for the CT negative samples. Nine replicates of each panel member were tested every day for five days on each BD Viper System. The data are summarized in Table 13. Two additional target levels were included in the panels to characterize the reproducibility of test results (i.e., proportion positive or negative) at target levels below the analytical Limit of Detection (LOD) of the BD ProbeTec CT Q $^{\text{x}}$ Assay. These additional specimens comprised CT and GC organisms seeded into LBC Specimen Dilution Tubes containing LBC medium at dilutions of 1:10 and 1:100 of the respective analytical LODs of each analyte. These levels were selected to fall within the dynamic range of the analytical LOD curves for the BD ProbeTec CT Q $^{\text{x}}$ and GC Q $^{\text{x}}$ assays. Nine replicates of each panel member were tested every day for five days across the three BD Viper Systems. The data are summarized in Tables 11 and 12.
Table 11: Summary of Reproducibility Data for LBC Specimens on the BD Viper System for the CT Q $^{\text{y}}$ Assay.
| | | | | | Within Run | | Between Runs Within Site | | Between Site | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| CT EBs/mL | GC Cells/mL | % Correct | 95% CI | Mean MaxRFU | SD | %CV | SD | %CV | SD | %CV |
| 0 | 0 | 100.0% (135/135) | (97.3% - 100.0%) | 1.30 | 4.66 | 357.64 | 0.85 | 65.29 | 0.20 | 15.12 |
| 30 | 0 | 100.0% (135/135) | (97.3% - 100.0%) | 2021.95 | 225.94 | 11.17 | 16.58 | 0.82 | 21.52 | 1.06 |
| 0 | 100 | 100.0% (135/135) | (97.3% - 100.0%) | 1.35 | 3.63 | 268.97 | 0.00 | 0.00 | 0.87 | 64.48 |
| 30 | 250 | 100.0% (135/135) | (97.3% - 100.0%) | 2028.41 | 155.45 | 7.66 | 9.93 | 0.49 | 0.00 | 0.00 |
| 75 | 100 | 100.0% (135/135) | (97.3% - 100.0%) | 1964.40 | 170.91 | 8.70 | 44.37 | 2.26 | 8.70 | 0.44 |
{22}
Page 23 of 25
Table 12: Characterization of System Reproducibility at Target Levels below the Analytical Limit of Detection for the CT Q $^{\text{x}}$ Assay for LBC Specimens.
| Dilution of Analytica 1 LOD | % Positive | 95% CI (Positive) | MaxRFU Mean (Positive) | % Negative | 95% CI (Negative) | MaxRFU Mean (Negative) |
| --- | --- | --- | --- | --- | --- | --- |
| 1:10 | 50.4 (68/135) | (41.6 - 59.1) | 1935.9 | 49.6 (67/135) | (40.9 - 58.4) | 11.5 |
| 1:100 | 7.4 (10/135) | (3.6 - 13.2) | 1835.7 | 92.6 (125/135) | (86.8 - 96.4) | 9.4 |
# System Cross Contamination and Carryover
An internal study was conducted to evaluate the risk of producing a false positive result in either the same run on the BD Viper System in extracted mode (within run cross-contamination) or in a subsequent run (between run carryover). Testing was conducted using negative and positive samples on three BD Viper Systems. Negative samples consisted of CT/GC Q $^x$ Swab Diluent/LBC Specimen Dilution Tube with PreservCyt Solution. Positive samples consisted of a representative analyte (at $10^{5}$ CT EB/mL) spiked into CT/GC Q $^x$ Swab Diluent/LBC Specimen Dilution Tube with PreservCyt Solution. The overall rate of cross-contamination (i.e., with alternating columns of positive and negative samples and a prevalence of $50\%$ ) was $0.41\%$ (9/2208) for the CT/GC Qx Swab diluent and $0.45\%$ (5/1104) for the LBC Specimen Dilution Tube with PreservCyt Solution. The overall rate of carryover contamination (i.e., carryover between successive runs when the prevalence was $50\%$ in the previous run) was $0.36\%$ (8/2208) for the CT/GC Q $^x$ Swab diluent and $0.54\%$ (6/1104) for the LBC Specimen Dilution Tube with PreservCyt Solution. Cross-contamination and carryover rates across the three BD Viper Systems are summarized in Tables 13 and 14.
Table 13: Cross Contamination and Carryover Contamination (Swab/Urine).
| Assay Dispense Mode Selected | BD Viper System | Cross-Contamination | | | Carryover Contamination | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | n | Positive Results | Percent Positive | n | Positive Results | Percent Positive |
| Dual Assay | 1 | 736 | 5 | 0.68 | 736 | 1 | 0.14 |
| | 2 | 736 | 0 | 0.00 | 736 | 3 | 0.41 |
| | 3 | 736 | 4 | 0.54 | 736 | 4 | 0.54 |
| | Overall | 2208 | 9 | 0.41 | 2208 | 8 | 0.36 |
| Single Assay | 1 | 190 | 0 | 0.00 | 186 | 0 | 0.00 |
| | 2 | 188 | 1 | 0.53 | 186 | 1 | 0.54 |
| | 3 | 188 | 0 | 0.00 | 186 | 0 | 0.00 |
| | Overall | 566 | 1 | 0.18 | 568 | 1 | 0.18 |
{23}
Page 24 of 25
Table 14: Cross Contamination and Carryover Contamination (LBC Medium).
| Medium Type | BD Viper System | Cross-Contamination | | | Carryover Contamination | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | n | Positive Results | Percent Positive | n | Positive Results | Percent Positive |
| PreservCyt | 1 | 368 | 1 | 0.27 | 368 | 1 | 0.27 |
| | 2 | 368 | 3 | 0.82 | 368 | 0 | 0.00 |
| | 3 | 368 | 1 | 0.27 | 368 | 5 | 0.45 |
| | Overall | 1104 | 5 | 0.45 | 1104 | 6 | 0.54 |
## INTERPRETATION OF TABLES
## Symbols and Abbreviations
## Symbols
$(+)$ positive
(-) negative
number
$\%$ percentage
## Abbreviations
A Asymptomatic
CI Confidence Interval
CT Chlamydia trachomatis
CV Coefficient of Variation
EC Extraction Control
FN False Negative
FNU Female Neat Urine
FP False Positive
FS Female endocervical swab
FUPT Female urine in $\mathbf{Q}^{\mathrm{x}}$ UPT
FV Female vaginal swab
GC Neisseria gonorrhoeae
HIV Human Immunodeficiency Virus
I Indeterminate
IFU Inclusion Forming Units
LBC Liquid Based Cytology
LE Liquid level error
LOD Limit of Detection
MaxRFU Maximum relative fluorescent units
MNU Male Neat Urine
MS Male urethral swab
MUPT Male urine in $\mathbf{Q}^{\mathrm{x}}$ UPT
number
NA non-applicable
NAAT Nucleic Acid Amplification Test
NPA Negative Percent Agreement
NPV Negative Predictive Value
OB/GYN Obstetrics/Gynecology
PA Percent Agreement
PBS Phosphate Buffered Saline
PIS Patient Infected Status
PPA Positive Percent Agreement
PPV Positive Predictive Value
{24}
Page 25 of 25
QC Quality Control
S Symptomatic
SD Standard Deviation
SDA Strand Displacement Amplification
STD Sexually Transmitted Disease
TN True Negative
TP True Positive
UPT Urine Preservative Transport
N. Instrument: Same as described in K081824.
O. System Descriptions: Same as described in K081824
P. Other Supportive Device and Instrument Information: NA
Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
25
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.