K091722 · Biomet, Inc. · FMF · Mar 12, 2010 · General Hospital
Device Facts
Record ID
K091722
Device Name
COAXIAL APPLICATOR SYSTEM
Applicant
Biomet, Inc.
Product Code
FMF · General Hospital
Decision Date
Mar 12, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The CoAxial™ Applicator System is designed to facilitate pre-mixing of allograft, or synthetic bone graft materials for application to an orthopedic surgical site, with autologous blood, plasma, platelet-rich plasma, or other specific blood component(s) as deemed necessary by the clinical use requirements. In addition, the CoAxial™ Applicator System is intended for the application of autologous blood, plasma, platelet-rich plasma, or other specific blood component(s), as deemed necessary by the surgeon's determination of the clinical use requirements to facilitate the preparation of soft tissue autograft or allograft material prior to the application of the graft material to a repair site.
Device Story
Dual-chamber piston syringe system; 12 ml outer chamber and 1.2 ml inner chamber; facilitates simultaneous dispensing of two fluids at 10:1 ratio. Used in orthopedic surgical settings by surgeons to pre-mix bone graft materials (allograft, autograft, synthetic) with autologous blood components (plasma, PRP). Device enables controlled application of graft mixtures to repair sites. Benefits include efficient, consistent mixing and delivery of graft materials, potentially improving handling and application at surgical sites.
Clinical Evidence
No clinical data provided. Bench testing included torque testing (separation force, unscrewing torque, assembly, stress cracking), volume accuracy, assembly time, and mixing thoroughness. Bone graft handling testing compared 11:1, 10:1, and 9:1 ratios, showing no statistically significant differences in cohesive strength.
Technological Characteristics
Dual-syringe piston system; 12 ml outer and 1.2 ml inner chambers; 10:1 fluid dispensing ratio. Manual operation. Sterile, single-use. No electronic components or software.
Indications for Use
Indicated for patients requiring orthopedic surgical site bone graft application or soft tissue graft preparation. Used by surgeons to mix/apply autologous blood, plasma, or platelet-rich plasma with bone graft or soft tissue graft materials.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
SmartJet™ Bone Grafting Liquid Applicator (K011032)
{0}------------------------------------------------
K0917222
## MAR ] 2 2010
# 510(k) Summary
### Applicant/Sponsor: Biomet Biologics, LLC 56 East Bell Drive P.O. Box 587 Warsaw, IN 46581 Phone: 800-348-9500 www.biomet.com
Contact Person: Lonnie Witham Summary Preparation Date: March 11, 2010
Device Trade Name: CoAxial™ Applicator System Common Name: Piston Syringe
Classification Name: Piston Syringe FMF (21 CFR 880.5860)
| 510(k) Number | Predicate Device | Manufacturer |
|---------------|-------------------------------------------|--------------------------------------------------|
| K011032 | SmartJet™ Bone Grafting Liquid Applicator | Harvest Technologies Corp.<br>Plymouth, MA 02360 |
| K020252 | SmartJet™ Liquid Grafting Applicator | Harvest Technologies Corp.<br>Plymouth, MA 02360 |
| K883338 | Surgical Sealant Applicator (FibriJet®) | Micromedics, Inc.<br>St. Paul, MN 55107 |
### Legally Marketed Predicate Devices To Which Substantial Equivalence Is Claimed:
#### Device Description:
The CoAxial™ Applicator System consists of the CoAxial™ Spray Kit and CoAxial™ Applicator Tips. Within the CoAxial™ Spray Kit is a sterile, single-use, syringe type device (CoAxial™ Applicator), connector and Spray Tip. The CoAxial™ Applicator is comprised of two separate chambers to facilitate pre-mixing of bone graft materials with blood or blood products.
#### Statement of Intended Use:
The CoAxial™ Applicator System is designed to facilitate pre-mixing of allograft, or synthetic bone graft materials for application to an orthopedic surgical site, with autologous blood, plasma, platelet-rich plasma, or other specific blood component(s) as deemed necessary by the clinical use requirements
In addition, the CoAxial™ Applicator System is intended for the application of autologous blood, plasma, platelet-rich plasma, or other specific blood component(s), as deemed necessary by the surgeon's determination of the clinical use requirements to facilitate the preparation of
{1}------------------------------------------------
soft tissue autograft or allograft material prior to the application of the graft material to a repair site.
## Summary of Technological Characteristics:
The CoAxial™ Applicator is a dual-syringe system, similar to all three predicate devices. It is comprised of an inner, smaller syringe and an outer, larger syringe, instead of two separate syringes. While the design differs from the predicates, it is designed to function in the same manner. Two fluids can be held by separate syringes and all applicators allow for dispensing the two liquids simultaneously.
The dimensions of the two chambers that comprise the CoAxial™ Applicator are 12 ml (outer) and 1.2 ml (inner), which are similar to the 10 ml and 1 ml volumes held by the two syringes in both the Harvest SmartJet™ Bone Grafting Liquid Applicator predicates. The 10:1 ratio of dispelled fluids is the same for the SmartJet™ systems. The FibriJet® Applicator has a ratio of 】1:1.
#### Summary of Performance Data:
Torque testing (separation force, unscrewing torque, ease of assembly, resistance of overriding, stress cracking) and volume accuracy testing was performed on the CoAxial™ Applicator. Results either met the acceptance criteria set forth within the applicable standard, or if there were no set acceptance criteria, the results were acceptable per the test protocol.
Assembly time and mixing thoroughness were evaluated against the FibriJet® Applicator, manufactured by Micromedics. The CoAxial™ Applicator System demonstrated ease of use advantage over the FibriJet® Applicator, with a three-fold decrease in assembly time using an assembly test subject who was unfamiliar with either device assembly. The results of the mixing analysis were comparable for the two devices.
The CoAxial™ Applicator System is intended to be used for the application of liquids in a 10:1 ratio, and the instructions for use clearly define how to achieve this application. Bone graft handling testing was performed to compare the effect that an 11:1, 10:1, and 9:1 ratio of PRP/PP to autologous serum has on bone graft cohesive strength. No statistically significant differences between the groups were observed.
#### Non-Clinical Testing:
Non-clinical laboratory testing was performed to determine substantial equivalence. Testing indicated that the Fat Concentration System is substantially equivalent to currently marketed predicate devices. The results indicated that the device was functional within its intended use.
#### Clinical Testing:
None provided as a basis for substantial equivalence.
{2}------------------------------------------------
#### Adverse Events:
A search of FDA's MAUDE database for the period of Jan. 1, 2008 thru Jan. 31, 2010 revealed that no adverse events were reported under K011032, K020252, and K883338.
#### Substantial Equivalence:
The CoAxial™ Applicator System is shown to be substantially equivalent to the previously cleared devices, the Harvest SmartJet™ Bone Grafting Liquid Applicator- K011032, SmartJer™ Liquid Grafting Applicator - K020252, and the Micromedics FibriJet® Applicator - K883338. The design, sizes, and materials are similar to the aforementioned predicates and there are no new performance characteristics that introduce new safety and/or effectiveness issues.
All trademarks are property of Biomet, Inc.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Lonnie Witham Regulatory Affairs Consultant Biomet, Biologics, LLC 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581-0587
MAR 1 2 2010
Re: K091722
Trade/Device Name: CoAxial Applicator System Regulation Number: 21CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: March 3, 2010 Received: March 4, 2010
Dear Mr. Witham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 Mr. Witham
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): NA _________________________________________________________________________________________________________________________________________________
Device Name: CoAxial Applicator System
Indications for Use:
The CoAxial Applicator System is designed to facilitate pre-mixinq of allograft, autograft, or synthetic bone graft materials for application to an orthopedic surgical site, with autologous blood, plasma, platelet-rich plasma, or other specific blood component(s) as deemed necessary by the clinical use requirements.
In addition, the CoAxial Applicator System is intended for the application of autologous blood, plasma, platelet-rich plasma, or other specific blood component(s), as deemed necessary by the surgeon's determination of the clinical use requirements to facilitate the preparation of soft tissue autograft or allograft material prior to the application of the graft material to a repair site.
Prescription Use YES (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CBER, Office of Blood Research and Review
Division Sign-Off) vision of Anesthesiology, General Hospital ection Control, Dental Devices
10(k) Number:
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.