BARD 3DMAX LIGHT MESH, MODELS: 0117310, 0117311, 0117312, 0117320, 0117321, 011732

K091659 · C.R. Bard, Inc. · FTL · Aug 3, 2009 · General, Plastic Surgery

Device Facts

Record IDK091659
Device NameBARD 3DMAX LIGHT MESH, MODELS: 0117310, 0117311, 0117312, 0117320, 0117321, 011732
ApplicantC.R. Bard, Inc.
Product CodeFTL · General, Plastic Surgery
Decision DateAug 3, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Bard® 3DMAX™ Light Mesh is a sterile, single use device indicated for use in the reinforcement of soft tissue where weakness exists, such as the repair of hernia defects.

Device Story

Bard® 3DMAX™ Light Mesh is a sterile, single-use, anatomically shaped surgical mesh designed for laparoscopic inguinal hernia repair. Constructed from knitted polypropylene monofilaments (0.0048-inch diameter), the device features a preformed shape with sealed edges to facilitate easier positioning compared to flat mesh sheets. It includes an 'M'-shaped orientation marker and an arrow, both composed of polypropylene monofilament dyed with Phthalocyaninato(2-) copper, to assist surgeons in identifying the medial edge during placement. The mesh is designed to be stretched in two directions to accommodate and reinforce tissue defects. Used by surgeons in an operating room setting, the device provides a structural scaffold for tissue reinforcement. The anatomical design and orientation markers aim to improve surgical efficiency and accuracy during hernia repair procedures, potentially reducing operative time and improving patient outcomes.

Clinical Evidence

Bench testing only. No clinical data provided. Laboratory testing compared the subject device to predicates (Soft Mesh, 3DMAX™ Mesh, Mersilene Mesh) evaluating thickness, pore size, density, stiffness, tensile strength, percent elongation, suture pullout strength, burst strength, and tear resistance per FDA guidance. Biocompatibility testing for polypropylene and the copper-dyed orientation marker yielded acceptable results.

Technological Characteristics

Knitted polypropylene monofilaments (0.0048-inch diameter); anatomically shaped with sealed edges; medial orientation marker (polypropylene with Phthalocyaninato(2-) copper colorant); sterile, single-use; non-absorbable synthetic mesh.

Indications for Use

Indicated for reinforcement of soft tissue where weakness exists, such as hernia defect repair in patients requiring surgical intervention.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K091659 page 43 ## SECTION 7.0 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS #### A. Submitter Information AUG 0 8 0009 Submitter's Name: Address: Davol Inc. Subsidiary of C.R. Bard, Inc 100 Crossings Boulevard Warwick, R1 02886 - Telephone: Fax: Contact Person: Date of Preparation: (401) 825-8582 (401) 825-8765 Gail Dow June 3, 2009 #### B. Device Name Trade Name: Common/Usual Name: Classification Name: Bard® 3DMAX™ Light Mesh Surgical Mesh Surgical Mesh, Polymeric #### C. Predicate Device Name Bard® Soft Mesh (Davol Inc.) Trade name: Trade name: Bard® 3DMAX™ Mesh (Davol Inc.) Mersilene Mesh (Ethicon, Inc.) Trade name: #### D. Device Description The Proposed Bard® 3DMAX™ Light Mesh, is designed to fit the inguinal anatomy. The device is anatomically shaped and preformed with sealed edges that allow for easier positioning of the device than a traditional flat sheet of mesh in a laparoscopic inguinal hernia repair. The device is constructed of knitted polypropylene monofilaments 0.0048 inches in diameter. The knit construction allows the mesh to be stretched in both directions in order to accommodate and reinforce tissue defects. The device contains an orientation marker in an "M" shape to designate the medial edge of the mesh and an arrow pointing to the medial aspect. Both the arrow and "M" will aid the user during positioning and placement of the device. The orientation marker is composed of polypropylene monofilament dyed with Phthalocyaninato(2-) copper colorant. # PREMARKET NOTIFICATION FOR THE BARD® 3DMAX™ Light Mesh ### -CONFIDENTIAL- {1}------------------------------------------------ #### E. Intended Use The Bard® 3DMAX™ Light Mesh is a sterile, single use device indicated for use in the reinforcement of soft tissue where weakness exists, such as the repair of hernia defects. #### ட் Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use Bard® 3DMAX™ Light Mesh, the currently marketed Bard® 3DMAX™ Mesh, Bard® Soft Mesh and Ethicon Mersilene Mesh devices are all indicated for repair of hernia defects. The intended use for the 3DMAX™ Light Mesh is the same as the Predicate devices, Soft Mesh and 3DMAX™ Mesh, with the exception of the indication for repair of chest wall defects. The Predicate Bard® Soft Mesh is a major component of the Proposed 3DMAX™ Light Mesh and both are made from the same polypropylene monofilaments with a diameter of approximately 0.0048 inches. The Proposed 3DMAX™ Light Mesh has similar physical attributes and performance characteristics as the Predicates 3DMAX™ Mesh, Soft Mesh, and Mersilene Mesh. Furthermore, it has similar materials and manufacturing methods as the Predicate 3DMAX™ Mesh. The key differences in the Proposed 3DMAX™ Light Mesh compared to the currently marketed 3DMAX™ Mesh are the diameter of the polypropylene monofilament and the knit pattern of the mesh. In order to create a thin lightweight mesh and reduce the amount of material used in the mesh compared to traditional monofilament polypropylene meshes, the Proposed 3DMAX™ Light Mesh is constructed from monofilaments with a diameter of approximately 0.0048 inches. However, the weights (grams/square inch) of Mersilene Mesh (0.0274) and Soft Mesh (0.0282) are comparable to the 3DMAX™ Light Mesh (0.0272). In addition, the Proposed 3DMAX™ Light Mesh has larger pore knit construction than the Predicate Bard 3D Max™ Mesh. The 3DMAX™ Light Mesh, and the Predicate 3DMAX™ Mesh are similar as both are designed to fit the inquinal anatomy. Unlike the Predicates. Mersilene Mesh and Soft Mesh, which are flat mesh sheets, the Proposed 3DMAX™ Light Mesh is anatomically shaped and preformed with sealed edges that allow for easier positioning in a laparoscopic inguinal hernia repair. The edge seal design will be the same as the currently marketed 3DMAX™ Mesh, however a layer of polypropylene mesh will be added during the edge seal lamination process to provide sufficient material to create a smooth, non-porous edge. Unlike the Predicates, Mersilene Mesh and Soft Mesh, the Proposed 3DMAX™ Light Mesh and the currently marketed 3DMAX™ Mesh contain a medial marker composed of polypropylene monofilament and Phthalocyaninato(2-) copper colorant. ### PREMARKET NOTIFICATION FOR THE BARD® 3DMAX™ Light Mesh #### -CONFIDENTIAL- {2}------------------------------------------------ K091659 Page 3/3 #### G. Performance Data Laboratory bench testing was performed to assess the effects of the new characteristics of the Proposed 3DMAX™ Light Mesh. These tests compared the Proposed Product against the Predicate Products Soft Mesh, 3DMAX™ Mesh and Mersilene Mesh. In accordance with FDA's "Guidance for the Preparation of a Premarket Notification for a Surqical Mesh" (March 2. 1999), tests included physical and performance evaluations of the products in terms of thickness, pore size, density, stiffness, tensile strength, percent elongation, suture pullout strength, burst strength and tear resistance. Laboratory testing discussed in this submission demonstrates that the material chosen and the design utilized in manufacturing 3DMAX™ Light Mesh is substantially equivalent to the referenced Predicate Products. The results of this testing are in Section 15. Additionally, biocompatibility testing covering the polypropylene material used in the Proposed 3DMAX™ Light Mesh, as well as the orientation marker composed of polypropylene monofilament and Phthalocyaninato(2-) copper colorant have received acceptable results. Results are summarized in Section 14.0 and copies of the test reports are in Exhibit 4. The results show that the 3DMAX™ Light Mesh is substantially equivalent to the currently marketed Predicate devices. Therefore, based on laboratory testing and biocompatibility data, the Proposed 3DMAX™ Light Mesh, is safe and effective for its intended use. ## PREMARKET NOTIFICATION FOR THE BARD® 3DMAX™ Light Mesh ### -CONFIDENTIAL- {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 0 3 2009 C.R. Bard, Inc. % Ms. Gail Dow Regulatory Affairs Associate 100 Crossings Boulevard Warwick, Rhode Island 02886 Re: K091659 Trade/Device Name: Bard® 3DMax"" Light Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: June 3, 2009 Received: June 9, 2009 Dear Ms. Dow: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Ms. Gail Dow forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, sincerely yours, Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATION FOR USE STATEMENT 510(k) Number (if known): ---- Not Known Bard® 3DMAx™ Light Mesh Device Name: The Bard® 3DMAX™ Light Mesh will be indicated for use in Indications for Use: the reinforcement of soft tissue where weakness exists, such as the repair of hernia defects. Over-The-Counter AND/OR Prescription Use _____________________________________________________________________________________________________________________________________________________________ X Use (21 CFR 801 (Part 21 CFR 801 Subpart D) Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Keane for mxm (Division Sign Off) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K091659
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...