← Product Code [HRS](/productcode/HRS) · K091614

# OSTEOMED FOOT PLATING SYSTEM (K091614)

_Osteomed LP · HRS · Sep 10, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K091614

## Device Facts

- **Applicant:** Osteomed LP
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Sep 10, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The OsteoMed Foot Plating System is indicated for use in trauma, general surgery, and reconstructive procedures of the foot, ankle or other bones appropriate for the device. The OsteoMed Foot Plating System implants are intended for single use only.

## Device Story

System consists of plates (compression, locking, elongated, compression-elongated holes), screws (locking, non-locking, cannulated), washers, and implantable K-wires. Used in trauma, general surgery, and reconstructive procedures of foot, ankle, or other bones. Operated by surgeons in clinical settings to provide rigid bone fixation. Instrumentation facilitates modification, insertion, and removal of implants. Implants are single-use only.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Materials: Titanium (ASTM F-67, F-136, F-1472) or Stainless Steel (ASTM F-138, F-139). Instrumentation: Stainless steel, anodized aluminum, medical grade polymers. Mechanical fixation system; no software or energy source.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Device Proprietary Name:

Device Common Name:

Classification Name:

Product Code:

Name of Submitter:

K091614 (pg. 1 of 1)

OsteoMed Foot Plate and Screw Rigid Fixation System

OsteoMed Foot Plating System

21 CFR § 888.3030: Single/multiple component metallic bone fixation appliances and accessories

HRS. HWC

OsteoMed L. P. 3885 Arapaho Road Addison, Texas 75001 Phone: (972) 677-4781 Fax: (972) 677-4778

Contact Person:

Date Prepared:

Alma Relja, RAC

June 1, 2009

Summary:

This submission describes the OsteoMed Foot Plating System consisting of various shape and sizes plates featuring compression, locking, elongated or compression elongated holes, and ulated locking, non-locking and cannulated screws, implantable K-wires, washers, and angulated instrumentation. The OsteoMed Foot Plating System is indicated for use in trauma, general surgery, and reconstructive procedures of the foot, ankle or other bones appropriate for the size of the device. All implants are intended for single use only. Surgical instrumentation is provided to facilitate modification, insertion, and removal of implants.

The system contains several modules based on the size of the device and application site such as fixation/reconstruction of small fragment bones, forefoot, mid-foot, rear-foot, ankle, , or other bones appropriate for the size of the device.

The implants are made of Titanium (ASTM F-67, ASTM F-136, or ASTM F-1472) or Stainless Steel (ASTM F-138 or ASTM F-139). The instrumentation is made from various grades of stainless steel, anodized aluminum, and/or medical grade polymers.

Equivalence for this system is based on similarities in intended use, material, design and operational principle to the predicate devices listed in this submission. Also, note that some sections of this system could have been letter to file based on the OsteoMed previously cleared submissions.

Due to the similarity of materials and design to both pre-enactment and post-enactment devices we believe that the OsteoMed Foot Plating System does not raise any new safety or effectiveness issues.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.

SEP 1 0 2009

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Osteomed L.P. % Ms. Alma Relja, RAC Regulatory Affairs Specialist 3885 Arapaho Road Addison, Texas 75001

Re: K091614

Trade/Device Name: OsteoMed Foot Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS. HWC Dated: September 3, 2009 Received: September 4, 2009

Dear Ms. Relja:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

## Page 2 - Ms. Alma Relja, RAC

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Harbarefnehn

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Eyaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known): 糸の91614 Device Name: OsteoMed Foot Plating System

Indications for Use:

The OsteoMed Foot Plating System is indicated for use in trauma, general surgery, and reconstructive procedures of the foot, ankle or other bones appropriate for the device.

The OsteoMed Foot Plating System implants are intended for single use only.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sputu for mxn
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K091614

Page 1 of 1

(Posted November 13, 2003)

---

**Source:** [https://fda-staging.innolitics.com/device/K091614](https://fda-staging.innolitics.com/device/K091614)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
