ELECYS ANTI-CCP CALCHECK

K091601 · Roche Diagnostics Corp. · JJX · Sep 25, 2009 · Clinical Chemistry

Device Facts

Record IDK091601
Device NameELECYS ANTI-CCP CALCHECK
ApplicantRoche Diagnostics Corp.
Product CodeJJX · Clinical Chemistry
Decision DateSep 25, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

For use in the verification of the calibration established by the Elecsys Anti-CCP reagent on the indicated Elecsys and cobas e immunoassay analyzers. For In Vitro Diagnostic Use; For Prescription Use Only; To be used with the Elecsys Anti-CCP reagent only

Device Story

Elecsys Anti-CCP CalCheck is a lyophilized quality control product containing human antibodies to cyclic citrullinated peptides (CCP) in a human serum matrix. It is used in clinical laboratory settings to verify the calibration of the Elecsys Anti-CCP reagent on specific Roche immunoassay analyzers (Elecsys 2010, MODULAR ANALYTICS E170, cobas e 411, cobas e 601). The product is reconstituted with distilled or deionized water by laboratory personnel. By testing these known-value samples, clinicians verify that the analyzer's calibration remains accurate, ensuring the reliability of patient Anti-CCP test results used in the diagnosis or monitoring of autoimmune conditions.

Clinical Evidence

No clinical data. Bench testing only. Value assignment was performed using multiple analyzer platforms (Elecsys 2010/cobas e411 and MODULAR ANALYTICS E140/cobas e 601) with target values established as the median of independent series of analyses. Stability was validated via accelerated studies (24-month shelf-life) and real-time studies (4-hour open-vial stability).

Technological Characteristics

Lyophilized human serum matrix containing human antibodies to CCP and preservatives. Three-level concentration format. Reconstituted with 1.0 mL distilled/deionized water. Traceable to a commercially available FDA-cleared second-generation Anti-CCP assay. Storage: 2-8 °C unopened; 20-25 °C for 4 hours reconstituted.

Indications for Use

Indicated for use as a calibration verification material for the Elecsys Anti-CCP reagent on Elecsys and cobas e immunoassay analyzers. No specific patient population is described as this is a laboratory quality control product.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k091601 B. Purpose for Submission: New device C. Measurand: Human IgG autoantibodies to cyclic citrullinated peptides (CCP) D. Type of Test: Calibration verification material for the Elecsys Anti-CCP reagent E. Applicant: Roche Diagnostics Corporation F. Proprietary and Established Names: Elecsys Anti-CCP CalCheck G. Regulatory Information: 1. Regulation section: 21 CFR § 862.1660 Quality control material (assayed and unassayed) 2. Classification: Class I (reserved) 3. Product code: JJX, Single (specified) analyte controls (assayed and unassayed) 4. Panel: Clinical Chemistry (75) H. Intended Use: 1. Intended use(s): See indication(s) for use below. 2. Indication(s) for use: For use in the verification of the calibration established by the Elecsys Anti-CCP reagent on the indicated Elecsys and cobas e immunoassay analyzers. 3. Special conditions for use statement(s): For In Vitro Diagnostic Use; For Prescription Use Only; To be used with the Elecsys Anti-CCP reagent only 4. Special instrument requirements: Elecsys 2010, MODULAR ANALYTICS E170, cobas e 411 and cobas e 601 analyzers {1} 2 I. Device Description: The Elecsys Anti-CCP CalCheck is a lyophilized product consisting of human antibodies to CCP (derived from human serum) in a human serum matrix containing preservatives. During manufacture, the analyte is spiked into the matrix at the desired concentration levels. Each Elecsys Anti-CCP CalCheck is calibrated against master calibrators standardized against a commercially available FDA-cleared second-generation Anti-CCP assay. All human source materials used in the preparation of the proposed device were tested using FDA-approved methods and shown to be free from Hepatitis B Surface Antigen, and antibodies to Hepatitis C and HIV. J. Substantial Equivalence Information: 1. Predicate device name(s): Roche Diagnostics, Elecsys C-Peptide CalCheck 2. Predicate 510(k) number(s): k040157 3. Comparison with predicate: | Item | Device | Predicate (k040157) | | --- | --- | --- | | Similarities | | | | Intended use | For use in the verification of the calibration established by the Elecsys Anti-CCP reagent on the indicated Elecsys and cobas e immunoassay analyzers. | For use in the verification of the calibration established by the Elecsys C-Peptide reagent on the Elecsys immunoassay systems. | | Levels | Three | same | | Format | Lyophilized | Same | | Handling | Reconstitute with exactly 1.0 mL distilled or deionized water and allow to stand closed for 15 minutes, then mix gently by inversion. | Same | | Stability | Unopened: Store at 2 to 8 °C until expiration date. Reconstituted: 20 to 25 °C; 4 hours | Same | | Differences | | | | Matrix | Human serum | Equine serum | K. Standard/Guidance Document Referenced (if applicable): - Guidance for Industry and FDA Staff – Assayed and Unassayed Quality Control Material L. Test Principle: Not Applicable {2} 3 M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not Applicable b. Linearity/assay reportable range: Not Applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: There is no recognized standard for CCP. Therefore, the CCP CalCheck material is traceable to a commercially available FDA-cleared second-generation Anti-CCP assay. Value assignment: Each of the three CalCheck levels is value assigned using a minimum of three Elecsys 2010/cobas e411 analyzers and three MODULAR ANALYTICS E140/cobas e 601 analyzers. A minimum of six independent series of analyses are performed on each instrument platform. Each sample is tested in duplicate. The target value is then calculated as the median of the determined values. The target values are given in the table below: | Anti-CCP CalCheck Level | Anti-CCP Target Values (U/mL) | | --- | --- | | Check 1 (low) | 20 | | Check 2 (medium) | 300 | | Check 3 (high) | 375 | Stability: The unopened stability of the CalChecks was determined using accelerated studies. Based on the accelerated studies, the unopened shelf-life was estimated to be 24 months. The sponsor is currently performing real-time studies to support the unopened stability claim predicted from accelerated studies. The open vial stability was determined using real-time studies. The data supports the claim that the reconstituted product is stable up to 4 hours when stored at 20 to 25 °C. d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: {3} a. Method comparison with predicate device: Not Applicable b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: Target values and acceptable ranges for pilot lot are listed below: For the Elecsys 2010/cobas e 411 analyzers | Level | Value | Range | Units | | --- | --- | --- | --- | | Check 1 | 21.2 | 17.0 – 25.4 | U/mL | | Check 2 | 261 | 209 – 313 | U/mL | | Check 3 | 347 | 260 - 434 | U/mL | For the MODULAR ANALYTICS E170/cobas e 601 analyzers | Level | Value | Range | Units | | --- | --- | --- | --- | | Check 1 | 18.7 | 15.0 – 22.4 | U/mL | | Check 2 | 267 | 214 – 320 | U/mL | | Check 3 | 350 | 263 - 438 | U/mL | # N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. # O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...