SILK BLACK BRAIDED SURGICAL SUTURES, NYLON MONOFILAMENT BLUE AND BLACK SURGICAL SUTURES
K091574 · Supplidores Unidos Internacionales, S.A. · GAT · Oct 9, 2009 · General, Plastic Surgery
Device Facts
Record ID
K091574
Device Name
SILK BLACK BRAIDED SURGICAL SUTURES, NYLON MONOFILAMENT BLUE AND BLACK SURGICAL SUTURES
Applicant
Supplidores Unidos Internacionales, S.A.
Product Code
GAT · General, Plastic Surgery
Decision Date
Oct 9, 2009
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
General soft tissue approximation and/or ligation including use in Cardiovascular, Opthalmic, and Neurological procedures.
Device Story
AMERICAN SUTURE™ non-absorbable surgical sutures are sterile, single-use devices used for soft tissue approximation, wound closure, and ligation. Materials include silk (braided, silicone-coated), nylon (monofilament), polypropylene (monofilament), and polyester (braided, silicone-coated). Sutures are provided with or without attached needles. Used by clinicians in surgical settings (cardiovascular, ophthalmic, neurological). Device provides mechanical tissue support; performance verified against USP monographs for diameter, tensile strength, and needle attachment. Benefits include reliable wound closure and tissue approximation.
Clinical Evidence
Bench testing only. Performance verified via mechanical testing (diameter, tensile strength, needle attachment) per USP 30 monographs. Biocompatibility testing performed per AAMI/ANSI/ISO 10993-5 (cytotoxicity) and 10993-10 (irritation/hypersensitivity). Sterility validated per AAMI/ANSI/ISO 11137-1. Additional testing conducted per Mexican norms (FEUM) for intracutaneous irritation, systemic injection, and implantation.
Indicated for general soft tissue approximation and/or ligation, including cardiovascular, ophthalmic, and neurological procedures. Prescription use only.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
**510(k) Number:** K091574
page 1/3
# 510(k) SUMMARY (As Required by 21 CFR 807. 92)
| Submitted by: | Jorge Lay<br>Principal<br>Suplidores Unidos Internacionales S.A.<br>Avenida Miguel Hidalgo 31B<br>Colonia Lago de Guadalupe<br>Cuautitlan Estado de Mexico<br>Mexico<br>Tel: +52 (55) 58 77 42 41<br>Fax: (305) 551-9139 | OCT - 9 2009 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Date of Summary: | May 1 <sup>st</sup> , 2009 | |
| Device Name: | AMERICAN SUTURE <sup>TM</sup> Non Absorbable Surgical Sutures with/without Needles (Silk, Nylon, Polypropylene, and Polyester) | |
| Common Name: | Surgical Sutures | |
| Classification Name: | Sutures | |
| Class: | 2 | |
| Product Code: | GAT, GAW, GAP, and GAR | |
| Regulation Number: | 21 CFR 878-5000, 5010, 5020, and 5030 | |
| Predicative Device: | CP Medical Non-absorbable Surgical Sutures | |
| 510 K Number | Product Description |
|--------------|-----------------------------------------------|
| K001184 | Silk Non-absorbable Surgical Sutures |
| K001173 | Nylon Non-absorbable Surgical Sutures |
| K001185 | Polypropylene Non-absorbable Surgical Sutures |
Modifications:
K001172
There are no modifications to the device design that affect safety and effectiveness for its intended use.
Polyester Non-absorbalbe surgical Sutures
{1}------------------------------------------------
K091574 page 2/3
AMERICAN SUTURE™ Non-Absorbable Surgical Sutures with/without Needle ⁄ Device Description AMERIOAN OOTORE - Non-Pyrogenic devices used for safe and reliable soft tissue approximation, wound closure, and ligation.
For safe and reliable soft tissue approximation, wound closure and ligation. Intended Use:
AMERICAN SUTURE™ Non-Absorbable Surgical Sutures with/without Needle Technological AMERICAN OO FORE - Non Abberbast of the legally marketed CP Medical Non-absorbable Surgical Sutures.
# Technical Comparison to Predicative Devices
| American ™ Non Absorbable<br>Surgical Sutures | CP Medical Non Absorbable<br>Surgical Sutures |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| American Surgical Sutures are indicated for soft<br>tissue approximation and ligation | CP Medical non Absorbable Sutures are indicated<br>soft tissue approximation and ligation. |
| American Silk sutures are Intended for Use in<br>gastrointestinal, neurological, and cardiovascular<br>surgery | CP Medical sutures are Intended for Use in<br>gastrointestinal, neurological, and cardiovascular<br>surgery |
| American Sutures are made of Silk, Nylon,<br>Polypropylene, and Polyester materials. Sterile, Single<br>Use, Non Absorbable. With or without Needles<br>attached | CP Medical Sutures made of Silk, Nylon,<br>Polypropylene and Polyester materials.<br>Sterile, Single Use, Non Absorbable. With or<br>without Needles attached. |
| American Silk sutures are made from pure Silk<br>monofilaments, braided then coated with<br>Silicone | CP Medical Silk Sutures are braided from pure<br>Silk Monofilaments, braided then coated with<br>Silicone |
| American Nylon Sutures are made from<br>monofilaments of a polyamide material and dyed<br>colored Black or Blue | CP Medical sutures are made from monofilaments<br>polyamide material and dyed colored Black or Blue |
| American Polypropylene sutures are made from<br>Monofilaments of synthetic Polyolefin | CP Medical Polypropylene sutures are made from<br>monofilaments of synthetic Polypropylene 1 |
| American Polyester Sutures are made from<br>monofilaments of synthetic Polyethylene, braided<br>then coated with Silicone | CP Medical Polyester sutures are made form<br>monofilaments of synthetic Polyethylene, braided,<br>then coated with Silicone |
| American Surgical Sutures meet or exceed the<br>performance requirements for Non-Absorbable<br>surgical sutures as defined in the Official monograph<br>of the United States Pharmacopeia:<br>Diameter <861><br>Tensile Strength <881><br>Needle Attachment <871><br>Suture Length Requirement | CP Medical Sutures meet or exceed the<br>performance requirements for Non-Absorbable<br>surgical sutures as defined in the Official<br>monograph of the United States Pharmacopeia:<br>Diameter <861><br>Tensile Strength <881><br>Needle Attachment <871><br>Suture Length Requirement |
| American surgical Sutures are offered with or without<br>Needles | CP Medical sutures are offered with or without<br>Needles |
: ....
¹ Polyolefin and Polypropylene are both synthetic polymers widely used in the manufacturing of surgical sutures.
{2}------------------------------------------------
K091524
page 3/3
Testing:
AMERICAN SUTURE™ Non-Absorbable Surgical Sutures with/without Needle have AMENTO IN OO to refformance and safety testing to verify mechanical properties and functioning, as well as biocompatibility and sterility.
The American Sutures™ Surgical sutures intended to be introduced have been tested The Amenoun Guturoo - Ourgromance with the following test methods:
## FDA Recognized Standards
- ecogmized Standards
AAMI / ANSI / ISO 10993-5:1999, Biological evaluation of medical devices -- Part 5: . Tests for In Vitro Cytotoxicity. (Biocompatibility)
- Fests for In Ville Ofter 3-10:2002, Biological evaluation of medical devices Part 10: ● Tests for irritation and delayed-type hypersensitivity. (Biocompatibility)
- Tests for Inteation and dolarsa type tify on and routine control of ETO Sterilization
- AAMI / ANSI / ISO 11137-1:2000 Validation and routine control of GAMMA Sterilization
## Currently FDA recognized USP Edition Monographs
USP 30 <861>:2007 Sutures Diameter USP 30 <871> 2007 Sutures Needle Attachment USP 30 <881>:2007 Sutures Tensile Strength
### Mexican Norms.
Federal Mexican Code for Medical Devices - F.E.U.M. 2006 (Mex. Pharmacopoeia) FEUM- MGA-DM-3171, pages 220-221- Intracutaneous Irritation Test FEUM-MGA-DM-3083, pages 217-220 - Systemic Injection Test FEUM-MGA-DM-3081, pages 215-216 Implantation Test
Official Mexican Norms NOM-067-SSA-1993 for Medical Sutures NOM -B-1-70 Analysis Methods for Steels and Smelting (Needles) NOM -- B-119-78 Determination of Rockwell Hardness for Metals (Needles) NOM-BB-46-76 Determination of Corrosion Resistance for Hypodermic Needles
State States
. : routine : 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized depiction of an eagle or bird-like figure with three flowing lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Suplidores unidos Internacionales S.A. % VHP Consulting Mr. Victor Pereira P.O. Box 97-0132 Coconut Creek, Florida 33067-0132
OCT - 9 2009
### Re: K091574
Trade/Device Name: AMERICAN SUTURE™ Surgical Sutures – SILK BLACK – Braided AMERICAN SUTURE™ Surgical Sutures - NYLON BLACK and BLUE - Braided
> AMERICAN SUTURE™ Surgical Sutures – POLYPROPYLENE AMERICAN SUTURE™ Surgical Sutures – POLESTER - Braided
Regulation Number: 21 CFR 878.5000
Regulation Name: Nonabsorbable poly(ethylene terephthalate) surgical suture Regulatory Class: Class II Product Code: GAT Dated: August 1, 2009
Received: August 11, 2009
Dear Mr. Pereira:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 - Mr. Victor Pereira
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours
Sincerely yours,
Mark N. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number: K091574
Device Name: AMERICAN SUTURE™ Surgical Sutures- SILK BLACK - Braided
INDICATIONS FOR USE:
General soft tissue approximation and/or ligation including use in Cardiovascular, Opthalmic, and Neurological procedures.
Prescription Use_X (Part 21 CFR 801 Subpart D) AND/OR
Over- The Counter Use: (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Krave for MKM
(Division Sign-Off)
Division of Surgical, Orthopedic. and Restorative Devices
510(k) Number K091574
{6}------------------------------------------------
## 510(k) Number: K091574
Device Name: AMERICAN SUTURE™ Surgical Sutures- NYLON BLACK and BLUE - Braided
## INDICATIONS FOR USE:
General soft tissue approximation and/or ligation including use in Cardiovascular, Opthalmic, and Neurological procedures.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over- The Counter Use: (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
DEN Darind Krane for arxи
(ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K091574
{7}------------------------------------------------
510(k) Number: K091574
Device Name AMERICAN SUTURE™ Surgical Sutures- POLYPROPYLENE
#### INDICATIONS FOR USE:
General soft tissue approximation and/or ligation including use in Cardiovascular, Opthalmic, and Neurological procedures.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over- The Counter Use: (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Keane for MXul
(Division S Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K091574
{8}------------------------------------------------
510(k) Number: K091574
Device Name: AMERICAN SUTURE™ Surgical Sutures- POLESTER - Braided
## INDICATIONS FOR USE:
General soft tissue approximation and/or ligation including use in Cardiovascular, Opthalmic, and Neurological procedures.
X Prescription Use_ (Part 21 CFR 801 Subpart D) AND/OR
Over- The Counter Use: (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Keane for mxm
(Division Sign Off)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K091524
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.