← Product Code [MBI](/productcode/MBI) · K091503

# MAGNUM MP FIXATION DEVICE (K091503)

_Arthrocare Corp. · MBI · Jun 17, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K091503

## Device Facts

- **Applicant:** Arthrocare Corp.
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Jun 17, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Magnum MP bone anchor with inserter is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon attachment Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction. ITB tenodesis: patellar ligament and tendon avulsions

## Device Story

Magnum MP is a bone anchor system with an inserter handle used in arthroscopic and orthopedic surgeries. It facilitates the fixation of soft tissue to bone. The device is operated by a surgeon in a clinical or OR setting. By securing soft tissue to bone, it aids in the repair and reconstruction of various tendons and ligaments, potentially improving patient recovery and joint stability. It is a mechanical fixation device.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Bone anchor system with inserter handle. Metallic bone fixation fastener (21 CFR 888.3040, Product code MBI).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Opus Magnum PI ([K070227](/device/K070227.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K091503

# 510(K) SUMMARY

ARTHROCARE CORPORATION MAGNUM M FIXATION DEVICE

## General Information

Submitter Name/Address:

JUN 17 2009

ArthroCare Corporation 680 Vaqueros Avenue Sunnyvale, CA 94085-3523

Laura N. Kasperowicz

May 19, 2009

Establishment Registration No .: 2951580

Contact Person:

Date Prepared:

Device Description

Trade Name: Device Model Name:

Magnum® MP Fixation Device

Sr. Manager, Regulatory Affairs

Generic/Common Name:

Bone Anchor, Fastener, Fixation, Soft Tissue

Classification Name:

Fastener, Fixation, Nondegradeable, Soft Tissue (Class II per 21 CFR 888.3040, Product code: MBI)

Predicate Devices Opus Magnum PI

K070227 (Cleared April 16, 2007)

### Product Description

The Magnum MP device is a bone anchor system with inserter handle designed for specific indications in arthroscopic and orthopedic procedures.

### Indications For Use

The Magnum MP bone anchor with inserter is indicated for use in fixation of soft tissue to bone. Examples of such procedures include:

Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, bicens tenodesis and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction and midfoot reconstruction

Foot: Hallux valgus reconstruction

Elbow: Tennis elbow repair, biceps tendon attachment

Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction. ITB tenodesis: patellar ligament and tendon avulsions

page vi

{1}------------------------------------------------

# 510(K) SUMMARY

### Substantial Equivalence

By definition, substantial equivalence means that a device has the same intended use and technical characteristics as the predicate device, or has the same intended use and different technological characteristics, but can be demonstrated to be as safe and effective as the predicate device. The Magnum MP design and technology is substantially equivalent to the existing Magnum PI Knotless Fixation Device cleared by the Food and Drug Administration [K070227]. The differences between the Magnum MP and the predicate device do not raise any questions regarding the safety and effectiveness of the implant. Furthermore, the materials are well characterized and have been used in predicate devices with similar indications. The proposed device, as designed, is as safe and effective as predicate devices.

#### Summary and Reason for 510k Notification

The purpose of this 510(k) is to notify the Food and Drug Administration of a proposed modification to an existing product. The proposed device, the Magnum MP Fixation Device is substantially equivalent to the Magnum PI Knotless Fixation Device originally cleared under K070227.

page vii

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services, USA. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" are arranged in a circular pattern around the eagle.

#### Public Health Service

JUN 17 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Arthrocare Corporation % Ms. Laura N. Kasperowicz Senior Manager, Regulatory Affairs 15285 Alton Parkway, Suite 200 Irvine, California 92618

Re: K091503

Trade/Device Name: Magnum MP Fixation Device Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: II Product Code: MBI Dated: May 19, 2009 Received: May 21, 2009

Dear Ms. Kasperowicz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing

{3}------------------------------------------------

Page 2 - Ms. Laura N. Kasperowicz

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Barbara Buehrig

Mark N. Melkerson Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# INDICATIONS FOR USE STATEMENT

510(k) Number:

# K091503

Device Name:

Magnum MP Fixation Device

Indications for Use:

The Magnum MP bone anchor with inserter is indicated for use in fixation of soft tissue to bone. Examples of such procedures include:

Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction and midfoot reconstruction

Foot: Hallux valgus reconstruction

Elbow: Tennis elbow repair, biceps tendon attachment

Knee: Extra-capsular repairs; reattachment of: medial coilateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions

Prescription Use (Part 21 CFR 801 Subpart D) X

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) NO

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Soneta

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K091503

---

**Source:** [https://fda-staging.innolitics.com/device/K091503](https://fda-staging.innolitics.com/device/K091503)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
