NOVABONE DENTAL PUTTY - BIOACTIVE SYNTHETIC BONE GRAFT
Applicant
Novabone Products, LLC
Product Code
LYC · Dental
Decision Date
Jun 15, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of NovaBone Dental Putty is to provide a safe, biocompatible synthetic bone graft material for use in oral, dental intraosseous, and craniofacial defects. It is used alone in a manner comparable to autogenous bone graft chips or allograft bone particulate (Demineralized Freeze Dried Bone) or may be mixed with either (typically 1:1 ratio v/v) as a bone graft extender. Typical uses include: - Periodontal/Infrabony defects - Ridge Augmentation (sinusotomy, osteotomy, cystectomy) - Extraction (ridge maintenance/augmentation, implant preparation/ sites placement) - Sinus lifts - Cystic cavities - Cranio-facial augmentation For larger defects, a mixture of NovaBone Dental Putty with an equal volume of allograft or autograft bone and bone marrow may improve new bone formation.
Device Story
NovaBone Dental Putty is an osteoconductive, synthetic bone void filler; composed of calcium-phosphorus-sodium-silicate (Bioglass) particulate mixed with a synthetic binder. Supplied sterile in PET-G trays or disposable syringes; used by clinicians in dental/oral/craniofacial surgery. Upon implantation, binder absorbs, allowing tissue infiltration between bioactive glass particles; particles act as a scaffold for new bone growth, eventually being replaced by host bone. Device provides no mechanical strength; requires stabilization for load-bearing sites. Modification adds syringe delivery format to existing tray-in-pouch offering.
Clinical Evidence
No clinical data provided. Substantial equivalence based on identical composition, formulation, and technological characteristics to legally marketed predicate devices. Bench testing only.
Technological Characteristics
Osteoconductive bone void filler; bioactive glass (45S5 Bioglass) particulate with synthetic binder. Pliable, cohesive putty form. Supplied sterile in PET-G trays or disposable plastic syringes. Biocompatible and resorbable. Non-structural scaffold.
Indications for Use
Indicated for patients requiring bone void filling or augmentation in oral, dental intraosseous, and craniofacial defects, including periodontal/infrabony defects, alveolar ridge augmentation, dental extraction sites, sinus lifts, and cystic cavities. Contraindicated for load-bearing defects requiring mechanical support; not for use to gain screw purchase or stabilize screw placement.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
NovaBone Putty - Bioactive Synthetic Bone Graft (K082672)
Submission Summary (Full Text)
{0}------------------------------------------------
K091484
Image /page/0/Picture/1 description: The image shows the logo for NOVABONE SYNTHETIC BONE GRAFT PRODUCTS. The logo is in black and white and features the word "NOVABONE" in large, bold letters. Below the word "NOVABONE" is the phrase "SYNTHETIC BONE GRAFT PRODUCTS" in smaller letters. The logo is simple and professional, and it effectively communicates the company's focus on bone graft products.
JUN 1 5 2009
### Special 510(k): Device Modification Summary 05/15/09 NovaBone Dental Putty - Bioactive Synthetic Bone Graft .
#### Submitter Information: 1.
| Name: | NovaBone Products, LLC |
|------------|----------------------------------------------------|
| Address: | 13709 Progress Boulevard, #33<br>Alachua, FL 32615 |
| Telephone: | (386) 462-7660 |
| Facsimile: | (386) 418-1636 |
| Contact: | David M. Gaisser |
#### 2. Name of Device:
| Trade Name: | NovaBone Dental Putty – Bioactive Synthetic Bone Graft |
|--------------------|------------------------------------------------------------------------------|
| Common Name: | Osteoconductive Bone Void Filler<br>Synthetic Resorbable Bone Graft Material |
| Regulation Number: | 21 CFR 888.3045 |
| Regulation Name: | Bone Void Filler |
#### Legally Marketed Predicate Device: 3.
| Predicate #1: | NovaBone Dental Putty – Bioactive Synthetic Graft<br>[K063549] |
|---------------|----------------------------------------------------------------|
| Predicate #2: | PerioGlas - Bone Graft Particulate<br>[K040278] |
| Predicate #2: | NovaBone Putty - Bioactive Synthetic Bone Graft<br>[K082672] |
#### 4. Device Description
NovaBone Dental Putty is an osteoconductive, bone void filler device. The device is intended for dental intraosseous, oral, and cranio-/maxillofacial bony defects and is supplied sterile. It is composed of a calcium-phosphorussodium-silicate (Bioglass) particulate mixed with a synthetic binder that acts as a temporary binding agent for the particulate. The particulate and binder are provided premixed as a pliable cohesive material. The mixed device is supplied sterile, packaged either in a PET-G tray or in a disposable plastic syringe. On implantation, the binder is absorbed to permit tissue infiltration between the Bioglass particles. The particles are slowly absorbed and replaced by new bone tissue during the healing process. The device is intended for dental intraosseous, oral, and cranio-/maxillofacial bony defects. It is supplied sterile.
#### Intended Use 5.
NovaBone Dental Putty is indicated to be packed into bony voids or gaps to fill and/or augment oral, dental intraosseous, and craniofacial defects. These defects
NOVABONE PRODUCTS, LLC 13709 PROGRESS BOULEVARD, #33 · ALACHUA, FL 32615 · (386) 462-7660 · FAX (386) 418-1636 www.novabone.com
{1}------------------------------------------------
### Special 510(k): Design Modification NovaBone Products, LLC NovaBone Dental Putty - Bioactive Synthetic Bone Graft
include: periodontal/infrabony defects; alveolar ridge augmentation may (sinusotomy, osteotomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); sinus lifts; cystic defects; craniofacial augmentation. NovaBone Dental Putty may be used alone in a manner comparable to autogenous bone graft chips or allograft bone particulate (demineralized freeze dried bone), or it may be mixed with either as a bone graft extender.
#### Technological Characteristics 6.
The technological characteristics of the NovaBone Dental Putty device are similar to those of the predicates. The device and the predicates are designed as osteoconductive space-filling devices to be gently packed into defect sites and used as non-structural scaffolds for the body's natural healing and bone regeneration process. The device indications are unchanged from those of the predicate devices. The device is intended to be used alone, or in combination with autogenous or allograft bone.
The NovaBone Dental Putty device of this submission is identical in composition and formulation to the NovaBone Dental Putty and NovaBone Putty predicate devices. The primary component of NovaBone Dental Putty also is identical to the bioactive glass (4555 Bioglass) particulate found in the PerioGlas predicate. This synthetic material is both biocompatible and osteoconductive. The NovaBone Dental Putty includes a synthetic binder as an inert carrier for ease of handling and delivery, forming a premixed cohesive material. The binder is biocompatible and is absorbed after implantation, opening space between the bioactive glass particles for cell infiltration and bone formation. The bioactive glass particulate remains for a longer post-implantation period, acting as a scaffold for bone ingrowth. This particulate is absorbed and replaced by new bone tissue.
The syringe package format as described in this submission is a product line extension, increasing the device package format availability. Similar syringe packages have been cleared for the NovaBone Putty [K082672] and PerioGlas [K040278] predicates.
#### 7. Warnings and Precautions
NovaBone Dental Putty does not possess sufficient mechanical strength to support load-bearing defects prior to hard tissue ingrowth. In cases where load support is required, standard internal or external stabilization techniques must be followed to obtain rigid stabilization in all planes.
NovaBone Dental Putty is intended for use by clinicians familiar with bone grafting and internal/external fixation techniques. NovaBone Dental Putty must not be used to gain screw purchase or to stabilize screw placement.
{2}------------------------------------------------
Confidential
Special 510(k): Design Modification NovaBone Products, LLC NovaBone Dental Putty -- Bioactive Synthetic Bone Graft
#### Complications 8.
Possible complications are the same as to be expected of autogenous bone grafting procedures. These may include: superficial wound infection, deep wound infection, deep wound infection with osteomyelitis, delayed union, loss of reduction, failure of fusion, loss of bone graft, graft protrusion and / or dislodgement, and general complications that may arise from anesthesia and / or surgery. Complications specific to oral/dental use are those as may be typically observed for similar bone grafting procedures and may include: tooth sensitivity, gingival recession, flap sloughing, resorption or ankylosis of the treated root, abscess formation.
#### Conclusion 9. .
The NovaBone Dental Putty device modification subject to this submission is to include a filled-syringe package format to augment the current tray-in-pouch format. This device modification does not result in a change in technological characteristics of the device. NovaBone Putty continues to be safe and effective as a non-structural osteoconductive bone void filler for osseous defects.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/11 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The seal is black and white.
### Public Health Service
# JUN 1 5 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David M. Gaisser Vice President, Operations/Regulatory Affairs/Quality Assurance NovaBone Products, LLC 13709 Progress Boulevard, #33 Alachua, Florida 32615
Re: K091484
Trade/Device Name: NovaBone Dental Putty - Bioactive Synthetic Bone Graft Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: May 15, 2009 Received: May 19, 2009
Dear Mr. Gaisser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Mr. Gaisser
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Rumrer
Susan Runner, D.D.S. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Special 510(k): Design Modification NovaBone Products, LLC NovaBone Dental Putty - Bioactive Synthetic Bone Graft Confidential
# STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known):
K0914884
Device Name: _ NovaBone Dental Putty - Bioactive Synthetic Bone Graft
Indications For Use:
The intended use of NovaBone Dental Putty is to provide a safe, biocompatible synthetic bone graft material for use in oral, dental intraosseous, and craniofacial defects. It is used alone in a manner comparable to autogenous bone graft chips or allograft bone particulate (Demineralized Freeze Dried Bone) or may be mixed with either (typically 1:1 ratio v/v) as a bone graft extender. Typical uses include:
- · Periodontal/Infrabony defects
- · Ridge Augmentation (sinusotomy, osteotomy, cystectomy)
- · Extraction (ridge maintenance/augmentation, implant preparation/ sites placement)
- · Sinus lifts
- · Cystic cavities
- · Cranio-facial augmentation
For larger defects, a mixture of NovaBone Dental Putty with an equal volume of allograft or autograft bone and bone marrow may improve new bone formation.
Prescription Use XX
OR (Per 21 CFR 801.109)
Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kein Huly for MSA
(Division Sign-Off) Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
510(k) Number: K091484
10 of 84
1-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.