LIGHTCYCLER MRSA ADVANCED TEST FOR USE WITH THE LIGHTCYCLER 2.0 INSTRUMENT, MODEL 05205727
Applicant
Roche Molecular Systems, Inc.
Product Code
NQX · Microbiology
Decision Date
Jul 6, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Indications for Use
The LightCycler® MRSA Advanced Test is a qualitative in vitro diagnostic test for the direct detection of nasal colonization with methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA infections in healthcare settings. The test is performed on the LightCycler® 2.0 Instrument with nasal swab specimens from patients suspected of colonization, uses swab extraction and mechanical lysis for specimen preparation followed by polymerase chain reaction (PCR) for the amplification of MRSA DNA, and fluorogenic target specific hybridization probes for the detection of the amplified DNA. The LightCycler® MRSA Advanced Test is not intended to diagnose, guide or monitor treatment for MRSA infections. Concomitant cultures are necessary to recover organisms for epidemiology typing or for further susceptibility testing.
Device Story
The LightCycler® MRSA Advanced Test is an in vitro diagnostic assay for detecting MRSA nasal colonization. Input: nasal swab specimens. Process: mechanical lysis of bacterial cell walls using glass beads in a MagNA Lyser Instrument; PCR amplification of target DNA using hot-start Taq polymerase and AmpErase (uracil-N-glycosylase) to prevent amplicon contamination; detection via FRET hybridization probes. Output: automated melting peak analysis performed by the LightCycler® 2.0 Instrument to identify MRSA-specific amplicons. Used in clinical laboratories by trained personnel. Results are interpreted via LightCycler® Software and Micro Analysis Software to aid infection control programs in healthcare settings. Benefits include rapid identification of colonized patients to facilitate isolation and infection control measures.
Clinical Evidence
Multi-site prospective study (n=1,402 evaluable specimens) compared the device against direct chromogenic culture, broth culture, and a second FDA-cleared NAAT. Results vs. direct chromogenic culture: 95.2% positive percent agreement (91.1-97.8% CI) and 96.4% negative percent agreement (95.2-97.4% CI). Reproducibility studies across 3 sites, 2 operators, 5 days, and 3 lots showed high consistency (99-100% for positive/weak positive samples).
Technological Characteristics
Uses mechanical lysis (glass beads) for sample prep. Real-time PCR amplification with hot-start Taq polymerase and AmpErase. Detection via FRET hybridization probes. Platform: LightCycler® 2.0 Instrument. Software-based melting peak analysis. Qualitative result generation. Standard laboratory equipment required (centrifuge, MagNA Lyser).
Indications for Use
Indicated for the direct detection of nasal colonization with methicillin-resistant Staphylococcus aureus (MRSA) in patients suspected of colonization to aid in the prevention and control of MRSA infections in healthcare settings. Not for diagnosing, guiding, or monitoring treatment of MRSA infections.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k091409
B. Purpose for Submission:
Determine substantial equivalence for this 510(k) for MRSA from nasal specimens.
C. Measurand:
Detects Methicillin-resistant Staphylococcus aureus (MRSA) by targeting the right extremity (RE) of the Staphylococcal Cassette Chromosome mec (SCCmec)/orfX junction.
D. Type of Test:
Nucleic Acid Amplification Test, DNA, Methicillin-resistant Staphylococcus aureus (MRSA), qualitative
E. Applicant:
Roche Molecular Systems, Inc
F. Proprietary and Established Names:
LightCycler® MRSA Advanced Test
G. Regulatory Information:
1. Regulation section:
21CFR866. 1640 Antimicrobial susceptibility test powder
2. Classification:
Class II
3. Product code:
NQX - Nucleic Acid Amplification test, MRSA, Direct Specimen
OOI - Nuclei Acid Amplification System, Real Time
4. Panel:
83 - Microbiology
H. Intended Use:
1. Intended use:
The LightCycler® MRSA Advanced Test is a qualitative in vitro diagnostic test for the direct detection of nasal colonization with methicillin-resistant (MRSA) to
{1}
aid in the prevention and control of MRSA infections in healthcare settings. The test is performed on the LightCycler® 2.0 Instrument with nasal swab specimens from patients suspected of colonization, uses swab extraction and mechanical lysis for specimen preparation followed by polymerase chain reaction (PCR) for the amplification of MRSA DNA, and fluorogenic target specific hybridization probes for the detection of the amplified DNA.
The LightCycler® MRSA Advanced Test is not intended to diagnose, guide or monitor treatment for MRSA infections. Concomitant cultures are necessary to recover organisms for epidemiology typing or for further susceptibility testing.
2. Indication(s) for use:
The LightCycler® MRSA Advanced Test is a qualitative in vitro diagnostic test for the direct detection of nasal colonization with methicillin-resistant (MRSA) to aid in the prevention and control of MRSA infections in healthcare settings. The test is performed on the LightCycler® 2.0 Instrument with nasal swab specimens from patients suspected of colonization, uses swab extraction and mechanical lysis for specimen preparation followed by polymerase chain reaction (PCR) for the amplification of MRSA DNA, and fluorogenic target specific hybridization probes for the detection of the amplified DNA.
The LightCycler® MRSA Advanced Test is not intended to diagnose, guide or monitor treatment for MRSA infections. Concomitant cultures are necessary to recover organisms for epidemiology typing or for further susceptibility testing.
3. Special conditions for use statement(s):
Prescription Use
4. Special instrument requirements:
To be used with the LightCycler 2.0 Instrument
I. Device Description:
Specimen collection swab heads are cut into lysis tubes and subjected to heating to inactivate specimens. Subsequently lysis tubes are transferred in the MagNA Lyser Instrument where mechanical lysis of bacterial cell walls occurs, resulting in crude lysate preparations. After a brief centrifugation step to spin down glass beads and swab fibers, the processed specimen are subjected to PCR analysis using the LightCycler® MRSA Advanced Test.
The processed samples and the amplification mixture containing hot-start Taq polymerase are placed in LightCycler® Capillaries (20μL) in which PCR amplification will occur. Each LightCycler® MRSA Advanced Test reaction contains an internal control, which is designed to control for specimen inhibition, and to monitor reagent integrity. Also present in each LightCycler® MRSA Advanced Test is the AmpErase (uracil-N-glycosylase) enzyme. It recognizes and catalyzes the destruction of DNA strands containing deoxyuridine, but not DNA containing
2
{2}
deoxythymidine. Since amplicons produced with the LightCycler® MRSA Advanced Test contain deoxyuridine, potential amplicon contaminants are eliminated during a prolonged heating step performed prior to the start of PCR amplification.
A target sequence in a plasmid is simultaneously amplified in the Positive Control. The Positive Control is intended to monitor for reagent failure and is included into each run. Each run also includes a Negative Control used to detect reagent or environmental contamination by MRSA DNA.
MRSA and Internal Control amplicons are detected by fluorescence using a specific pair of hybridization probes. The probes attach to a specific internal sequence in the amplified fragment and are positioned in a closed proximity to one another. Upon excitation, these bound probes emit a fluorescence signal of a specific wavelength using a process called Fluorescence Resonance Energy Transfer (FRET). The emitted light is measured by the LightCycler® 2.0 Instrument. MRSA or internal control specific amplicons are detected in parallel in two different detection channels and thus can be differentiated.
After completion of the real-time PCR process, a melting peak analysis is performed by the LightCycler® 2.0 Instrument. Single stranded DNA amplicons with bound hybridization probes are subjected to increasing temperatures. When the PCR products reach a specific temperature one of the two bound hybridization probes melts off, resulting in a loss of fluorescence signal. The decrease in the fluorescence signal occurs at a specific temperature and results in melting peaks which are used to identify and distinguish MRSA- and IC-specific amplicons.
After a visual identification of the melting peaks is performed using Tm Bars in the LightCycler® Software, test results are transferred to a dedicated interpretation tool (the Micro Analysis Software) and a report is generated.
J. Substantial Equivalence Information:
1. Predicate device name(s): BD GeneOhm MRSA Assay
1. Predicate 510(k) number(s): K033415 K042357
{3}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| | Device | Predicate |
| Item | LightCycler® MRSA Advanced Test | BD GeneOhm™ MRSA Assay |
| Intended Use | Direct and qualitative detection of MRSA | Same |
| Sample Type | Nasal swab | Same |
| Lysis | Mechanical lysis using glass beads | Same |
| Mode of detection | Nucleic acid amplification (DNA) utilizing real-time PCR | Same |
| Internal Assay Controls | Internal assay control to detect inhibitory specimens and to confirm integrity of reagent in negative samples | Same |
| External Controls | Provided: positive control to monitor for reagent integrity | Same |
| DNA Target Sequence | Sequence incorporating the insertion site of the SCCmec in the S. aureus orfX gene | Similar but not identical |
| Workflow | Multiple manual steps including pipetting, fluid transfer, vortexing, centrifugation and sample heating | Same |
| | | |
| Differences | | |
| Assay platform | LightCycler | SmartCycler |
| Detection Chemistry | Paired target-specific hybridization probes using fluorescence resonance energy transfer (FRET) | Target-specific hybridization probes (molecular beacon technology) |
| Result analysis | Melting peak analysis | Amplification curves analysis |
| Time to result | 2 hours | 60- 75 minutes |
| | | |
K. Standard/Guidance Document Referenced (if applicable):
Not applicable
L. Test Principle:
The LightCycler® MRSA Advanced Test relies on three major processes:
1) Specimen preparation by mechanical lysis of the bacterial cell walls
{4}
2) PCR amplification of target DNA and detection by specific hybridization probes
3) Automated result generation after melting peak analysis.
## Specimen Preparation
The mechanical lysis of the nasal swab specimens is performed by using the LightCycler® Advanced Lysis Kit and the MagNA Lyser Instrument. Swab heads are cut into lysis tubes and subjected to heating to inactivate all bacteria. Subsequently lysis tubes are transferred in the MagNA Lyser Instrument where mechanical lysis of bacterial cell walls occurs, resulting in crude lysate preparations. After a brief centrifugation step to spin down glass beads and swab fibers, the processed specimen are subjected to PCR analysis using the LightCycler® MRSA Advanced Test.
## PCR Amplification
### Target Selection
The primers and probes in the LightCycler® MRSA Advanced Test detect a proprietary sequence indicative of the integration of the SCCmec cassette into the S. aureus chromosome, indicating the presence of MRSA DNA.
### Amplification
The processed samples and the amplification mixture containing hot-start Taq polymerase are placed in LightCycler® Capillaries (20μL) in which PCR amplification will occur. Each LightCycler® MRSA Advanced Test reaction contains an internal control, which is designed to control for specimen inhibition, and to monitor reagent integrity. The AmpErase (uracil-N-glycosylase) enzyme included in the LightCycler® MRSA Advanced Test recognizes and catalyzes the destruction of DNA strands containing deoxyuridine, but not DNA containing deoxythymidine. Since amplicons produced with the LightCycler® MRSA Advanced Test contain deoxyuridine, potential amplicon contaminants are eliminated during a prolonged heating step performed prior to the start of PCR amplification.
A target sequence in a plasmid is simultaneously amplified in the Positive Control. The Positive Control is intended to monitor for reagent failure and is included into each run. Each run also includes a Negative Control used to detect reagent or environmental contamination by MRSA DNA.
## Specific detection of PCR products by Hybridization Probes
MRSA and Internal Control amplicons are detected by fluorescence using a specific pair of hybridization probes. The probes attach to a specific internal sequence in the amplified fragment and are positioned in a closed proximity to one another. Upon excitation, these bound probes emit a fluorescence signal of a specific wavelength using a process called Fluorescence Resonance Energy Transfer (FRET). The emitted light is measured by the LightCycler® 2.0 Instrument. MRSA or internal control specific amplicons are detected in parallel in two different detection channels and thus can be differentiated.
5
{5}
After completion of the real-time PCR process, a melting peak analysis is performed automatically by the LightCycler® 2.0 Instrument. Single stranded DNA amplicons with bound hybridization probes are subjected to increasing temperatures. When the PCR products reach a specific temperature one of the two bound hybridization probes melts off, resulting in a loss of fluorescence signal. The decrease in the fluorescence signal occurs at a specific temperature and results in melting peaks which are used to identify and distinguish MRSA- and IC-specific amplicons.
## Automated result generation after melting peak analysis
After a visual identification of the melting peaks is performed using TM Bars in the LightCycler® Software, test results are transferred to a dedicated interpretation tool (the Micro Analysis Software) and a report is generated.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Reproducibility study was performed using a 12 member panel with varying concentrations of MRSA and methicillin sensitive Staphylococcus epidermidis (MSSE) at a constant level. Two operators at each of three sites performed each per day for five days on three reagent lots; 4 specimens x 3 replicates x 5 days x 3 sites x 3 lots x 2 operators.
The panel included a negative sample, below the LOD (20 CFU/swab expected to yield a positivity rate of between 30 to 70%), weak positive (300 CFU/swab), and positive (800 CFU/swab). The negative panel member yield negative results from a low of 99% to 100%. The below LOD panel member positivity rate ranged from a low of 42% to 47%, the weak positive panel member positivity rate ranged from a low of 99% to 100%, and the positive panel member positivity rate ranged from a low of 99% to 100% depending on the site.
## Summary of Reproducibility Results
{6}
7
| MRSA
CFU/swab | Sample
Type | Lot to Lot | | | Site to Site | | | Day to Day | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Lot | No./No.
Tested | % | Site | No./No.
Tested | % | Day | No./No.
Tested | % |
| 0 | Negative | 1 | 90/90 | 100% | 1 | 90/90 | 100% | 1 | 53/54** | 98% |
| | | 2 | 90/90 | 100% | 2 | 89/90** | 99% | 2 | 54/54 | 100% |
| | | 3 | 89/90** | 99% | 3 | 90/90 | 100% | 3 | 54/54 | 100% |
| | | | 4 | 54/54 | 100% | | | | | |
| | | | | | | | | 5 | 54/54 | 100% |
| 20* | Below
LOD | 1 | 52/90 | 58% | 1 | 38/90 | 42% | 1 | 25/54 | 46% |
| | | 2 | 30/90 | 33% | 2 | 41/90 | 46% | 2 | 26/54 | 48% |
| | | 3 | 39/90 | 43% | 3 | 42/90 | 47% | 3 | 22/54 | 41% |
| | | | 4 | 25/54 | 46% | | | | | |
| | | | | | | | | 5 | 23/54 | 43% |
| 300 | Weak
Positive | 1 | 90/90 | 100% | 1 | 89/90 | 99% | 1 | 53/54 | 98% |
| | | 2 | 89/90 | 99% | 2 | 90/90 | 100% | 2 | 54/54 | 100% |
| | | 3 | 90/90 | 100% | 3 | 90/90 | 100% | 3 | 54/54 | 100% |
| | | | 4 | 54/54 | 100% | | | | | |
| | | | | | | | | 5 | 54/54 | 100% |
| 800 | Positive | 1 | 90/90 | 100% | 1 | 90/90 | 100% | 1 | 53/54** | 98% |
| | | 2 | 90/90 | 100% | 2 | 89/90** | 99% | 2 | 54/54 | 100% |
| | | 3 | 89/90** | 99% | 3 | 90/90 | 100% | 3 | 54/54 | 100% |
| | | | 4 | 54/54 | 100% | | | | | |
| | | | | | | | | 5 | 54/54 | 100% |
* Concentration that produces approximately 30%–70% positive results.
** There was an apparent transposition of one positive and one negative sample positioned adjacent to each other in the same run. Excluding these samples from the analysis results for lot to lot, site to site, and day to day reproducibility would be 100%.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The LightCycler® MRSA Advanced Test includes an Internal Control (IC); at least one Positive Control (PC) and one Negative Control (NC, required but not included in the kit) must be processed with each run. The valid run rate was >95% and was acceptable.
A MRSA strain representative of RE2 types is used for this external positive control. MRSA strains representing RE3 and RE7 types may be used as additional external positive controls to monitor assay primers and probes not directly controlled in the assay. External controls may be used in accordance with local, state, and federal accrediting organization, as applicable.
d. Detection limit:
**Limit of Detection Study**
The Limit of Detection (LoD) study was determined using three MRSA types
{7}
(RE2, RE3, and RE7) covered by the primers and probes in the assay and for the three swab types.
Cultures of these strains were quantified, diluted to values spanning the range of 100 to 400 colony forming units (CFU) per swab, and absorbed onto swabs previously soaked into various transport media. All dilutions around the LOD value were tested in replicates of at least 30. The limit of detection obtained for each strain type and swab type tested represents the lowest number of CFU/swab at which a positive result will be obtained with at least 95% confidence. Results indicate that the LightCycler® MRSA Advanced Test will produce a positive result with 95% confidence for a swab containing 240 CFU.
Summary of Results for three swab types tested
| | MRSA types | | |
| --- | --- | --- | --- |
| Swab Types | RE2 | RE3 | RE7 |
| Liquid Stuart transport media | 240 cfu/swab | 140 cfu/swab | 240 cfu/swab |
| Amies gel without charcoal transport media | 180 cfu/swab | 224 cfu/swab | 240 cfu/swab |
| Amies gel with charcoal transport media | 240 cfu/swab | 224 cfu/swab | 240 cfu/swab |
## Analytical Inclusivity
A total of 137 well characterized MRSA strains were obtained through the Network on Antimicrobial Resistance in S. aureus at a concentration of 10⁵ CFU/mL.
| USA type | Number of isolates tested | Isolates with positive results by the LightCycler® MRSA Advanced Test |
| --- | --- | --- |
| 100 | 54* | 53 |
| 200 | 5 | 5 |
| 300 | 37 | 37 |
| 400 | 4 | 4 |
| 500 | 7 | 7 |
| 600 | 3 | 3 |
| 700 | 4 | 4 |
| 800 | 7 | 7 |
| 1000 | 7 | 7 |
| 1100 | 4 | 4 |
| Not typable/ unknown | 5 | 5 |
| Total | 137 | 136 (99.3%) |
*Of the 54 USA100 strains, one strain (NRS642) was negative by the LightCyclerMRSA Advanced Test
An additional 1,643 isolates were collected in Europe (65%) and in the U.S.
{8}
(35%). There were 1,561 (95%) correctly identified as MRSA; 82 were negative by the LightCycler® MRSA Advanced Test.
e. Analytical specificity:
## Cross Reactivity
The study included pathogenic microorganisms and contaminants potentially present in nasal microflora. Microorganisms tested were 13 viral, 66 bacterial and 7 fungal species at concentrations of $10^{6}$ to $10^{7}\mathrm{CFU/mL}$, or as genomic DNA at concentrations $10~\mathrm{pg/PCR}$. Human DNA at concentration of $5\mathrm{ng}/\mathrm{reaction}$ was also tested. *Rothia mucilaginosa* revealed a weak positive melting peak in the target channel 610. The results were negative for MRSA $98.9\%$ of the organisms tested.
In addition, 100 methicillin sensitive *S. aureus* (MSSA), 117 methicillin resistant Coagulase negative *Staph* (CoNS) and 104 methicillin sensitive CoNS at concentrations $10^{4} - 10^{5}$ CFU/reaction were tested. Results were all negative by the LightCycler® MRSA Advanced Test.
Ten *Staphylococcus aureus* isolates characterized as Borderline Oxacillin Resistant *S. aureus* (BORSA) were included in the evaluation of BORSA study. They have an oxacillin MIC of $\geq 4\mu\mathrm{g/mL}$ and were negative by an assay for the detection of PBP2' of *S. aureus*. The BORSA isolates were tested at a concentration range of $10,000 - 100,000$ CFU/lysis tube. All the results were negative.
## Interference Study
### Exogenous Interferent
The following exogenous substances, which are components of decongestants and substances used to relieve nasal dryness and/or irritation, have been shown not to interfere with the detection of MRSA by the LightCycler® MRSA Advanced Test. Ointments and sprays were spiked to MRSA negative and MRSA positive Liquid Stewart swabs.
Gels from the Amies gel swab container with and without charcoal were also included. No interferences observed.
{9}
10
# Exogenous Substances Tested For Interference With The LightCycler® MRSA Advanced Test
| Trade name | Active ingredient |
| --- | --- |
| InfectoPyoderm® | Mupirocin 20 mg/g (antibiotic) |
| Turixin® | Mupirocin calcium 21.5 mg/g (antibiotic) |
| Bepanthen® nasal ointment | Dexpanthenol 50 mg/g (pro vitamine B5) |
| Accumed 12-hour Nasal Spray | Oxymetazoline hydrochloride 0.05% (decongestant) |
| Otriven® | Xylometazoline 0.1% (decongestant) |
## Endogenous Interferent
Blood was found to inhibit the LightCycler® MRSA Advanced Test. The IC amplification is inhibited at the lowest blood volume tested (5 μL) while the MRSA amplification is affected at higher levels (> 10 μL). However, the control concept in the MRSA Advanced Test insures that inhibitory samples are invalidated when tested, thus preventing false negative results from being reported. The data also indicated that excessive amounts of nasal mucus may be inhibitory and result in an invalid test result.
## Carry over/contamination
Samples with very high concentration of MRSA (i.e. 10⁵, and 10⁶ CFU/swab) were alternated with MRSA negative samples through sample preparation, PCR preparation and amplification in the LightCycler 2.0 Instrument. Data demonstrated cross-contamination at MRSA concentration higher than (> 10⁵ CFU/swab).
## f. Assay cut-off:
Multiple LightCycler® runs on multiple instruments were analyzed to determine the average Tm for the Positive Control/Target signals in channel (610 nm) and valid Internal Control signals in channel (670 nm). The baseline (assay cut-off) was determined by the analysis of negative samples. The baseline values were then confirmed in an additional study.
## 2. Comparison studies:
### a. Method comparison with predicate device:
A total of 1,620 nasal swab specimens were collected from subjects at five sites across the United States and tested with the LightCycler® MRSA Advanced Test.
{10}
Of the 1,620 specimen tested, 1,402 specimens were eligible to be included in statistical analyses. Performance characteristics were determined by comparing the LightCycler® MRSA Advanced Test with a second FDA-cleared nucleic acid amplification test (NAAT), direct culture, and broth culture.
A double-headed nasal swab was collected from each subject. One swab head was directly streaked onto a chromogenic agar plate with cefoxitin, and then processed according to the package insert for testing with the LightCycler® MRSA Advanced Test. The second swab head was directly streaked onto a separate chromogenic plate with cefoxitin, and then processed for testing with the second FDA-cleared NAAT test (according to the package insert). Thereafter, the second swab head was transferred into Trypticase Soy Broth and incubated for 48 hours at 35-37°C and then sub-cultured onto a chromogenic plate with cefoxitin (broth culture). Chromogenic culture plates were incubated at 35-37°C for 20-48 hours. Presumptive MRSA colonies from all culture plates were confirmed by coagulase testing and Gram staining if found after 44-48 hours of incubation. Each participating site performed all tests.
Samples that grew MRSA on direct chromogenic culture from either swab head A and/or swab head B were considered MRSA positive. The following are the performance summary tables of the LightCycler® MRSA Advanced Test with CHROMagar, another NAAT, and enrichment culture:
Comparison of the LightCycler MRSA Advanced Test with Direct Chromogenic Culture
| | Direct chromogenic culture | | |
| --- | --- | --- | --- |
| LightCycler® MRSA Advanced Test | Positive | Negative | Total |
| Positive | 178 | 44 | 222 |
{11}
12
| | Direct chromogenic culture | | |
| --- | --- | --- | --- |
| LightCycler® MRSA Advanced Test | Positive | Negative | Total |
| Negative | 9 | 1,171 | 1,180 |
| Total | 187 | 1,215 | 1,402 |
| Positive Percent Agreement (95% exact CI*) | 95.2% (91.1%, 97.8%) | | |
| Negative Percent Agreement (95% exact CI) | | 96.4% (95.2%, 97.4%) | |
Note: Included in this summary are 1,402 evaluable specimens that had valid LightCycler MRSA Advanced Test and direct culture results.
$^\circ \mathrm{CI} =$ Confidence Interval
Comparison of LightCycler MRSA Advanced Test with the second FDA cleared NAAT
| | The Second FDA-Cleared NAAT | | |
| --- | --- | --- | --- |
| LightCycler® MRSA Advanced Test | Positive | Negative | Total |
| Positive | 195 | 20 | 215 |
| Negative | 77 | 1,093 | 1,170 |
| Total | 272 | 1,113 | 1,385 |
| Positive Percent Agreement (95% exact CI) | 71.7% (65.9%, 77.0%) | | |
| Negative Percent Agreement (95% exact CI) | | 98.2% (97.2%, 98.9%) | |
Note: Included in this summary are 1,385 specimen with valid results for the MRSA Advanced Test and second FDA-cleared NAAT test which had additional concordant direct culture results from swab heads A and B
$^\circ \mathrm{CI} =$ Confidence Interval
Comparison of LightCycler MRSA Advanced Test with Broth Culture
{12}
| | Broth Culture | | |
| --- | --- | --- | --- |
| LightCycler® MRSA Advanced Test | Positive | Negative | Total |
| Positive | 184 | 38 | 222 |
| Negative | 21 | 1,152 | 1,173 |
| Total | 205 | 1,190 | 1,395 |
| Positive Percent Agreement (95% exact CI) | 89.8% (84.8%-93.5%) | | |
| Negative Percent Agreement (95% exact CI) | | 96.8% (95.6%, 97.7%) | |
Note: A total of 1395/1402 evaluable specimens that had valid LightCycler MRSA Advanced Test and enrichment culture results are included in this summary table. Enrichment culture results were missing /invalid for 7/1402 evaluable specimens.
$^\circ \mathrm{CI} =$ Confidence Interval
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
{13}
The overall number of positive MRSA samples by direct culture was 187 (13.3%). The study population in the clinical trial was grouped into subjects in hospital non-intensive care units (1,007; 71.8%), hospital intensive care units (67; 4.8%) nursing homes/extended care facilities (255; 18.2%), and medical facility staff (73; 5.2%).
N. Instrument Name:
Roche Molecular Systems LightCycler 2.0
O. System Descriptions:
1. Modes of Operation:
The instrument operates in batch mode with samples in glass capillaries covered by plastic stoppers.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
The LightCycler 2.0 instrument is operated through a separate PC connected by cable. The LightCycler software operates the instrument through the PC. The complete LightCycler Version 4.1 software documentation was reviewed in this submission as we consider the 2.0 a new instrument.
The MRSA assay uses an assay specific software module (MAS) to interact with LC 4.1 software and to calculate results. The complete MAS version 1.2 software documentation was reviewed in this submission.
In addition, the LightCycler 2.0 instrument falls under 862.2570 Instrumentation for clinical multiplex test systems regulation and its special controls. The appropriate product code from this regulation is OOI Real Time Nucleic Acid Amplification System.
The LightCycler 2.0 instrument also has RUO functionalities that can be used by end users to develop their own assays. Therefore, we applied the recently devised Combination RUO/IVD Instrumentation Policy. The company supplied all required software documentation and the IVD only user manual for review.
3. Specimen Identification:
14
{14}
Specimens are identified by their position in the sample carousel.
4. Specimen Sampling and Handling:
Specimen preparation is performed off-line, pipetted into glass capillaries, and manually loaded into a sample carousel for addition to the LightCycler 2.0.
5. Calibration:
No calibration needed.
6. Quality Control:
The use of quality control material is described in each specific assay package insert that uses the LightCycler 2.0 instrument.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not Applicable
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.