AG Industries filters are replacement filters intended for use in oxygen concentrator machines to help remove contaminants, including air borne bacteria and other particulate debris from an air stream. When used with oxygen concentrator machines, the replacement filters may be used in the home, nursing home, hospital, patient care facility, etc.
Device Story
Device consists of replacement filters for oxygen concentrators; functions as intake or final filtration stage to remove airborne bacteria and particulate debris from air stream. Air is drawn into compressor, passes through filter, then through sieve beds for oxygen concentration; final filter may be used before delivery to patient. Used in home, nursing home, hospital, or patient care facilities. Filters utilize glass microfiber media within styrene butadiene or polypropylene housings. Output is filtered air; benefits patient by reducing exposure to contaminants. No active electronic components or software.
Clinical Evidence
No clinical data. Bench testing only: Bacterial Filtration Efficiency (BFE), Dioctyl Phthalate (DOP) aerosol testing, and biocompatibility testing (ISO Guinea Pig Maximization, Acute Systemic Injection, MEM Elution, and Material Mediated Rabbit Pyrogen tests).
Technological Characteristics
Materials: Glass microfiber filter media; Styrene Butadiene or Polypropylene housings. Principle: Mechanical filtration of air stream. Form factor: Various configurations (hose barb, NPT, ISO male/female connectors) for oxygen concentrator integration. Connectivity: None (passive). Sterilization: Not specified.
Indications for Use
Indicated for use as replacement filters in oxygen concentrator machines to remove airborne bacteria and particulate debris from the air stream. Suitable for use in home, nursing home, hospital, and patient care facility settings.
Regulatory Classification
Identification
A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.
{0}------------------------------------------------
KC091363
## 510(K) SUMMARY
# FEB - 4 2010
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMVDA 1990 and 21 CFR §807.92.
510(k) number:
# 1. Submitter's Identification:
AG Industries 3637 Scarlet Oak Blvd. St Louis, MO 63122 USA Submitter Phone: (866) 222-8988 Submitter Fax: (970) 494-2052 Submitter Contact: Michael Amann Title: Vice-President Date Summary Prepared: 1-5-2009
## 2. Name of the Device:
Please reference the following table for Proprietary and Common names of the devices included in this submission.
| Proprietary Name | Common Name |
|-----------------------------------|-------------------|
| AG1038831 | Compressor Filter |
| LL201 | Compressor Filter |
| LL202 | Compressor Filter |
| LL205 | Compressor Filter |
| HCF100 | Compressor Filter |
| BF910 (Platinum) | Compressor Filter |
| BF900 (Non-Platinum) | Compressor Filter |
| BF950 (Platinum and Non-Platinum) | Compressor Filter |
| BF200C | Bacterial Filter |
| BF100 | Bacterial Filter |
| BF500 | Bacterial Filter |
| BF600 | Bacterial Filter |
#### 3. Predicate Device Information:
#### Comparison to Predicate Devices:
The above devices are compared to similar devices marketed by Porous Media which were cleared under K061426. The features of each are exhibited in the following tables.
{1}------------------------------------------------
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| Model | AG1038831 | DBF 32 |
| Intended Use | The filters are intended to help<br>remove air-borne contaminants,<br>including air borne bacteria and<br>other particulate debris, from the<br>air stream of a respiratory device. | Same |
| Filtration Efficiency | 99.9+% BFE | 99.999%BFE |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Styrene Butadiene or approved<br>equivalent | Polystyrene or approved equivalent |
| Air Flow Resistance | 1.3 cm H20 @ 10 Lpm | 3.0 cm H2O @ 100 scfh |
| Maximum Flow Rate | 100 LPM | Same |
| Connection | 12.72 mm to fit machine | 22 mm ISO male/female |
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|------------------------------------------|------------------------------------|
| Model | LL205 | DBF 27 |
| Intended Use | Same as AG1038831 Above | Same |
| Filtration Efficiency | 99.99+% BFE | 99.999% BFE |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Styrene Butadiene or approved equivalent | Polystyrene or approved equivalent |
| Air Flow Resistance | .78 cm H20 @ 10 Lpm | 4.0 cm H2O @ 100 scfh |
| Maximum Flow Rate | 100 LPM | 100 LPM |
| Connection | Male 22 mm ISO | Same |
.
{2}------------------------------------------------
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|---------------------------------------------|------------------------------------|
| Model | LL201 | DBF 24 |
| Intended Use | Same as AG1038831 Above | Same |
| Filtration Efficiency | 99.999+% BFE | 99.999% |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Styrene Butadiene or approved<br>equivalent | Polystyrene or approved equivalent |
| Air Flow Resistance | 12.02 cm H20 @ 85 Lpm | 4.0 cm H2O @ 100 scfh |
| Maximum Flow Rate | 100 LPM | Same |
| Connection | Male 22 mm ISO | Same |
:
. . .
·
.
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|---------------------------------------------|------------------------------------|
| Model | LL202 | DBF 24 |
| Intended Use | Same as AG1038831 Above | Same |
| Filtration Efficiency | 99.9+% BFE | 99.999% |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Styrene Butadiene or approved<br>equivalent | Polystyrene or approved equivalent |
| Air Flow Resistance | .39 cm H20 @ 10 Lpm | 4.0 cm H2O @ 100 scfh |
| Maximum Flow Rate | 100 LPM | Same |
| Connection | Male 22 mm ISO | Same |
:
:
:
{3}------------------------------------------------
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|---------------------------------------------|------------------------------------|
| Model | HCF100 | DBF 32 |
| Intended Use | Same as AG1038831 Above | Same |
| Filtration Efficiency | 99.999+% BFE | 99.999% |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Styrene Butadiene or approved<br>equivalent | Polystyrene or approved equivalent |
| Air Flow Resistance | .33 cm H20 @ 10.5 Lpm | 3.0 cm H2O @ 100 scfh |
| Maximum Flow Rate | 100 LPM | Same |
| Connection | 22mm ISO Male/Female | Same |
.
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|---------------------------------------------|----------------------------------------------------|
| Model | BF910 (Platinum) | DBF 25 |
| Intended Use | Same as AG1038831 Above | Same |
| Filtration Efficiency | 99.99+% BFE | 99.999% |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Styrene Butadiene or approved<br>equivalent | Polystyrene or approved equivalent |
| Air Flow Resistance | .4 cm H20 @ 5 Lpm | 18 in H2O (Platinum 10)<br>7 in H2O (non-Platinum) |
| Maximum Flow Rate | 100 LPM | Same |
| Connection | 3/8 " FNPT and 22mm OD | 3/8 " FNPT |
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
1
:
{4}------------------------------------------------
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|---------------------------------------------|----------------------------------------------------|
| Model | BF900 (Non-Platinum) | DBF 25 |
| Intended Use | Same as AG1038831 Above | Same |
| Filtration Efficiency | 99.999+% BFE | 99.999% |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Styrene Butadiene or approved<br>equivalent | Polystyrene or approved equivalent |
| Air Flow Resistance | 1.6 cm H20 @ 84.9 LPM (Non-Platinum) | 18 in H2O (Platinum 10)<br>7 in H2O (non-Platinum) |
| Maximum Flow Rate | 100 LPM | Same |
| Connection | 3/8" FNPT and 22mm OD | 3/8" FNPT |
.
.
-
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|---------------------------------------------|----------------------------------------------------|
| Model | BF950 ( Platinum & Non<br>Platinum) | DBF 25 |
| Intended Use | Same as AG1038831 Above | Same |
| Filtration Efficiency | 99.999+% BFE | 99.999% |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Styrene Butadiene or approved<br>equivalent | Polystyrene or approved equivalent |
| Air Flow Resistance | .45 cm H2O @ 10.3 LPM | 18 in H2O (Platinum 10)<br>7 in H2O (non-Platinum) |
| Maximum Flow Rate | 100 LPM | Same |
| Connection | 3/8" FNPT and 22mm OD | 3/8" FNPT |
.
{5}------------------------------------------------
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|-----------------------------------------|------------------------------------|
| Model | BF200C | DDF4711M03Y |
| Intended Use | Same as AG1038831 Above | Same |
| Filtration Efficiency | 99.999+% BFE | 99.999% |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Polypropylene or approved<br>equivalent | Polystyrene or approved equivalent |
| Connection | 3/8" Hose Barb | 3/8" Hose Barb |
·
.
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|-----------------------------------------|------------------------------------|
| Model | BF100 | DDF4711M03Y |
| Intended Use | Same as AG1038831 Above | Same |
| Filtration Efficiency | 99.999+% | 99.999% |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Polypropylene or approved<br>equivalent | Polystyrene or approved equivalent |
| Connection | 3/16"- ¼" Stepped Bar | 1/8" Hose Barb |
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|-----------------------------------------|------------------------------------|
| Model | BF500 | DDF4700M03Y |
| Intended Use | Same as AG1038831 Above | Same |
| Filtration Efficiency | 99.999+% | 99.97% |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Polypropylene or approved<br>equivalent | Polystyrene or approved equivalent |
| Connection | 1/4" Straight Barb | Same |
ﺮ
,
{6}------------------------------------------------
| Feature | AG Industries | Porous Media<br>K061426 |
|-----------------------|-----------------------------------------|------------------------------------|
| Model | BF600 | DDF4711M03Y |
| Intended Use | Same as AG1038831 Above | Same |
| Filtration Efficiency | 99.999+% | 99.999% |
| Filter Material | Glass Microfiber | Same |
| Housing Material | Polypropylene or approved<br>equivalent | Polystyrene or approved equivalent |
| Connection | 1/8" Hose Barb | Same |
The following tests were performed to support substantial equivalence:
- . Bacterial Filtration Efficiency
- Dioctyl Phthalate (DOP) Aerosol Test
- . ISO Guinea Pig Maximization Test
- . ISO Acute Systemic Injection Test
- . MEM Elution Test
- . Material Mediated Rabbit Pyrogen Test
The above tests confirmed that the performance of the subject filters is substantially equivalent to the predicate device(s). The Biocompatibility tests confirmed that the following filters BF500, BF200C, BF100 and BF600, which contain materials not in previously cleared devices, present no biocompatibility issues.
# 4. Device Description:
# COMPRESSOR & BACTERIA FILTERS ( AG1038831, LL205, LL201, LL202, HCF100, BF910, BF900, BF950, BF200C, BF100, BF500, BF600):
Room air is drawn into the compressor of a respiratory device through the bacterial intake filter. From the compress, the air passes through the Compressor Filter, if one is installed on the machine and proceeds to the sieve beds. The sieve beds condition the air by removing nitrogen from the air stream, which results in higher concentration of oxygen. The air then passes through the final filter before being supplied to the patient.
# 5. Intended Use:
AG Industries filters are replacement filters intended for use in oxygen concentrator machines to help remove contaminants, including air borne bacteria and other particulate
{7}------------------------------------------------
debris from an air stream. When used with oxygen concentrator machines, the replacement filters may be used in the home, nursing home, hospital, patient care facility, etc.
## 6. Discussion of Clinical Tests Performed:
Not applicable
## 7. Conclusions:
The subject devices have the same intended use and similar characteristics as the predicate devices. No new questions of safety or effectiveness are raised by differences in technology or materials. Thus, the AG Industies COMPRESSOR & BACTERIA FILTERS ( AG1038831, LL205, LL201, LL202, HCF100, BF910, BF900, BF950, BF200C, BF100, BF500, BF600) are substantially equivalent to the predicate devices.
{8}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms or lines extending upwards, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular pattern around the figure. The text is in all caps and is separated by dots.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
# FEB - 4 2010
ndd AG Industries C/O Mr. Ian Gordon Senior Vice-President Emergo Group, Incorporated 1705 South Capital of Texas Highway, Suite 500 Austin, Texas 78746
Re: K091363
Trade/Device Name: Compressor & Bacteria Filters (AG1038831, LL205, LL201, LL202, HCF100, BF910, BF900, BF950, BF200C, BF100, BF500, BF600
Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: II Product Code: CAH Dated: January 21, 2010 Received: January 22, 2010
Dear Mr. Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
## Page 2 - Mr. Gordon
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/cdrh/comp/ for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
## INDICATIONS FOR USE
## 510(k) Number :
# Device Name: COMPRESSOR & BACTERIA FILTERS ( AG1038831, LL205, LL201, LL202, HCF100, BF910, BF900, BF950, BF200C, BF100, BF500, BF600):
## Indications for Use:
AG Industries filters are replacement filters intended for use in oxygen concentrator machines to help remove contaminants, including air borne bacteria and other particulate debris from an air stream. When used with oxygen concentrator machines, the replacement filters may be used in the home, nursing home, hospital, patient care facility, etc.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schultheiss
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
KO91363 510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.