K091326 · Activatek, Inc. · EGJ · May 27, 2009 · Physical Medicine
Device Facts
Record ID
K091326
Device Name
ACTIVAPATCH
Applicant
Activatek, Inc.
Product Code
EGJ · Physical Medicine
Decision Date
May 27, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5525
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ActivaTek ActivaPatch Iontophoresis System is intended to be used for the administration of soluble salts into the body for medical purposes and as an alternative to hypodermic injection.
Device Story
ActivaPatch Iontophoresis System is a self-contained, disposable, single-use patch for transdermal drug delivery. Device comprises electronic module, LED indicator, active/return electrodes, and adhesive foam. Operator applies patch to skin and pulls activation tab. Device measures skin impedance via voltage monitoring; LED signals impedance status (high, moderate, or optimal). Upon reaching optimal impedance, device initiates treatment timer. System delivers 40 mA*min (half dose) followed by 40 mA*min (second half dose) for total 80 mA*min. LED provides visual feedback on delivery phase and completion. If impedance rises (e.g., patch detachment), device suspends timer and reverts to measurement phase. Device automatically shuts off upon completion. Used in clinical settings; operated by healthcare providers. Benefits patient by providing non-invasive alternative to hypodermic injection for soluble salt administration.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Disposable, single-use iontophoresis patch. Dimensions: 5.7" x 3.4" x 0.2". Energy source: 6V internal power. Features electronic module with LED indicator for status monitoring. Impedance-based control logic. Adhesive foam backing for skin attachment.
Indications for Use
Indicated for the administration of soluble salts into the body for medical purposes as an alternative to hypodermic injection.
Regulatory Classification
Identification
An iontophoresis device is a device that is intended to use a direct current to introduce ions of soluble salts or other drugs into the body and induce sweating for use in the diagnosis of cystic fibrosis or for other uses if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug. When used in the diagnosis of cystic fibrosis, the sweat is collected and its composition and weight are determined. Additionally, an iontophoresis device intended for any other purposes is a prescription device that is intended to use a current to introduce ions of drugs or non-drug solutions into the body for medical purposes other than those specified in paragraph (a) of this section, meaning that the device is not intended for use in diagnosis of cystic fibrosis, or a specific drug is not specified in the labeling of the iontophoresis device.
Special Controls
*Classification.* Class II (special controls). The device is classified as class II. The special controls for this device are:(i) The following performance testing must be conducted:
(A) Testing using a drug approved for iontophoretic delivery, or a solution if identified in the labeling, to demonstrate safe use of the device as intended;
(B) Testing of the ability of the device to maintain a safe pH level; and
(C) If used in the ear, testing of the device to demonstrate mechanical safety.
(ii) Labeling must include adequate instructions for use, including sufficient information for the health care provider to determine the device characteristics that affect delivery of the drug or solution and to select appropriate drug or solution dosing information for administration by iontophoresis. This includes the following:
(A) A description and/or graphical representation of the electrical output;
(B) A description of the electrode materials and pH buffer;
(C) When intended for general drug delivery, language referring the user to drug labeling approved for iontophoretic delivery to determine if the drug they intend to deliver is specifically approved for use with that type of device and to obtain relevant dosing information; and
(D) A detailed summary of the device-related and procedure-related complications pertinent to use of the device, and appropriate warnings and contraindications, including the following warning:
*Warning:* Potential systemic adverse effects may result from use of this device. Drugs or solutions delivered with this device have the potential to reach the blood stream and cause systemic effects. Carefully read all labeling of the drug or solution used with this device to understand all potential adverse effects and to ensure appropriate dosing information. If systemic manifestations occur, refer to the drug or solution labeling for appropriate action.(iii) Appropriate analysis/testing must demonstrate electromagnetic compatibility, electrical safety, thermal safety, and mechanical safety.
(iv) Appropriate software verification, validation, and hazard analysis must be performed.
(v) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
(vi) The elements of the device that may contact the patient must be assessed for sterility, for devices labeled as sterile.
(vii) Performance data must support the shelf life of the elements of the device that may be affected by aging by demonstrating continued package integrity and device functionality over the stated shelf life.
Predicate Devices
Trivarion Buffered Iontophoresis Electrode System (K061522)
{0}------------------------------------------------
K091326
## Section 6 510(K) Summary
# MAY 27 2009
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.87 and 807.92. Summary preparation date is May 5, 2009.
## A. Contact Information [21 CFR 807.92(a)(1)].
ActivaTek Inc. 2734 S. 3600 West, Unit F West Valley, UT 84120
Tel: 1-801-969-0883 Fax: 1-800-680-5520 Contact person: Jamal Yanaki
#### B. Device Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Trade Name: ActivaPatch Iontophoresis System
Device Common Name: Iontophoresis Patch Electrode
Classification Name: Iontophoresis, other uses
Product Code: EGJ
Panel: Physical Medicine
Device Classification: Class III
### C. Predicate Devices [21 CFR 807.92(a)(3)]
The ActivaTek ActivaPatch Iontophoresis System uses the same materials as the following predicate device: Trivarion Buffered Iontophoresis Electrode System K061522 manufactured by ActivaTek, Inc.
The ActivaTek ActivaPatch Iontophoresis System uses similar technology and has equivalent physical output characteristics as the following predicate device: Action Patch Iontophoresis System K030395 manufactured by EMPI, Inc.
{1}------------------------------------------------
#### D. Device Description [21 CFR 807.92(a)(4)]
The ActivaTek ActivaPatch Iontophoresis System (Iontophoresis Patch) consists of a self-contained, disposable single-use iontophoresis patch, instructions for use, and an alcohol prep pad containing 70% isopropyl alcohol.
The Iontophoresis Patch consists of an electronic module, an LED indicator, an active electrode, and a return electrode. These elements are incorporated under an adhesive foam covering which adheres the device to the skin. Its dimensions are 5.7 inches (length), 3.4 inches (width), and 0.2 inches (thickness).
The Iontophoresis Patch is designed to deliver a calibrated and fixed half dose of 40 mAmp*minutes and a full dose of 80 mAmp*minutes.
The ActivaPatch Iontophoresis System sequence of operation is signaled by its LED.
- a) The operator applies the ActivaPatch to the skin.
- b) The operator then pulls the tab located on the outer surface of the patch.
- c) In the Initial Skin Impedance Measurement Phase, the ActivaPatch measures voltage across the active and return electrodes to indicate the level of skin impedance. For high skin impedance, the LED flashes once every 4 seconds. This indicates that the skin impedance is too high for iontophoretic delivery. The treatment timer is not started. For moderate skin impedances, the LED flashes pulses four times every 30 seconds. The patch continues to monitor the skin impedance but does not turn on the treatment timer. When the skin impedance becomes optimal, the LED blinks once every 30 seconds. This indicates that the impedance is appropriate to start an iontophoretic treatment session and the ActivaPatch starts the treatment timer.
- d) In the First Half Dose Delivery Phase, the ActivaPatch LED continues to blink once every 30 seconds. This phase lasts 80 minutes and delivers an ionic dosage of 40 mA*minutes. At the end of the First Half Dose Delivery, the LED blinks short-long-short sequences every 30 seconds for 3 minutes. This signals to the user that the transition to the Second Half Dose Delivery Phase has begun.
- e) During the Second Half Dose Delivery Phase, the ActivaPatch LED blinks two pulses every 30 seconds. During this final phase, another 40 mA*minute dose is delivered. At the end of the delivery, the power in the device is automatically shut off and the LED does not blink.
- f) If at any time the skin impedance increases to high levels (e.g., the electrode falls off of the skin site) the ActivaPatch goes into the Skin Measurement Phase and the LED flashes pulses once every 4 seconds. The treatment timer is suspended and restarts when the skin impedance falls to optimal levels.
{2}------------------------------------------------
### E. Device Specification [21 CFR 807.92(a)(6)]
The ActivaPatch Iontophoresis System delivers a calibrated half dose of 40 mAmp and a full dose of 80 mAmps. It operates with a voltage of 6V.
#### F. Indications for Use [21 CFR 807.92(a)(5)]
The ActivaTek ActivaPatch Iontophoresis System is intended to be used for the administration of soluble salts into the body for medical purposes and as an alternative to hypodermic injection.
#### G. Conclusion [21 CFR 807.92(a)(3)]
Technologically, the ActivaPatch Iontophoresis System was found to be substantially equivalent to the currently cleared EMPI Action Patch (K030395) and ActivaTek Trivarion Buffered Iontophoresis Electrode (K061522). The indications are identical to the previously cleared EMPI Action Patch (K030395). We believe that there are no new questions of safety or efficacy raised by the introduction of the ActivaTek ActivaPatch Iontophoresis System.
6-3
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three stripes on its wing, representing the department's commitment to health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
MAY 27 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ActivaTek, Inc. % Mr. Jamal Yanaki President & CEO 2734 S. 3600 W. Unit F Salt Lake City, Utah 84119
Re: K091326
Trade/Device Name: ActivaTek ActivaPatch Iontophoresis System Regulation Number: 21 CFR 890.5525 Regulation Name: Iontophoresis device Regulatory Class: III Product Code: EGJ Dated: May 3, 2009 Received: May 9, 2009
Dear Mr. Yanaki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act), as long as you comply with all of the Act's requirements relating to drugs labeled or promoted with the device as described below. You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 - Mr. Jamal Yanaki
Our substantially equivalent decision does not apply to the drugs that you will label or promote for use with your device. Therefore, you may neither label nor promote your device for use with specific drugs, nor package drugs with your device prior to FDA having approved the drugs for iontophoretic administration. For information on the requirements for marketing new drugs, you may contact:
> Director Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland
As you are aware, there are concerns relating to the fact that no drug is currently labeled for administration via an iontophoresis device. The Agency currently is evaluating this public health concern regarding the safety and effectiveness of this route of administration of drugs, and in the near future will inform manufacturers of certain additional steps the Agency believes are necessary to address this concern.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (240) 276-0120.
{5}------------------------------------------------
Page 3 - Mr. Jamal Yanaki
Also, please not the regulation entitled, "Misbranding by reference to premarket notification"(21) CFR Part 807.97). You may obtain other general information on your responsibilities under the , Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
For
Mark N. Melkerson
Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Section 5 Indications for Use
510(k) Number (if known):_ 1< (3 9 1326
Device Name: ActivaTek ActivaPatch Iontophoresis System
Indications for Use:
The ActivaTek ActivaPatch Iontophoresis System is intended to be used for the administration of soluble salts into the body for medical purposes and as an alternative to hypodermic injection.
Prescription Use X (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
***_*****_**. (Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K091326
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.