← Product Code [EZD](/productcode/EZD) · K091306

# AMSURE PVC INTERMITTENT CATHETER (K091306)

_Amsino International, Inc. · EZD · Aug 21, 2009 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K091306

## Device Facts

- **Applicant:** Amsino International, Inc.
- **Product Code:** [EZD](/productcode/EZD.md)
- **Decision Date:** Aug 21, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The AMSure™ PVC Intermittent Catheter is intended for use in the drainage of urine from the bladder.

## Device Story

The AMSure™ PVC Intermittent Catheter is a sterile, single-use medical device consisting of PVC tubing and a funnel. It is designed for the drainage of urine from the bladder. The device is used in clinical or home settings to facilitate bladder emptying. It is a manual device operated by healthcare professionals or patients. The device functions as a conduit for urine drainage; it does not involve electronic components, software, or automated processing.

## Clinical Evidence

Bench testing only. Performance and biocompatibility testing demonstrated safety and effectiveness for the intended use.

## Technological Characteristics

Materials: PVC tubing and funnel. Form factor: Intermittent catheter. Energy source: None (manual). Sterilization: Sterile, single-use. No software or electronic components.

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- AMSure™ Urethral Catheterization Tray ([K030712](/device/K030712.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### AMSINO

Special 510(k) Submission ----------------------------------------------------------------------------------------------------------------------------------------------------

K09/306

# 510(K) Summary of Safety and Effectiveness AMSelf™ PVC Intermittent Catheter

| Company:                 | Amsino International, Inc.<br>855 Towne Center Drive<br>Pomona,CA 91767<br>(909)626-5888                                                                                                          | AUG 21 2009 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Contact:                 | Holiven Ji<br>Manager of Regulatory Affairs                                                                                                                                                       |             |
| Date Prepared:           | April 22, 2009                                                                                                                                                                                    |             |
| Classification Name:     | Tray, Catheterization, Sterile Urethral, with or without Catheter<br>(876.5130)                                                                                                                   |             |
| Common Name:             | Urethral Catheter                                                                                                                                                                                 |             |
| Proprietary Name:        | <i>AMSure</i> ™ PVC Intermittent Catheter                                                                                                                                                         |             |
| Product Code:            | FCM                                                                                                                                                                                               |             |
| Medical Specialty:       | Gastroenterology/Urology                                                                                                                                                                          |             |
| Device Class:            | Class II                                                                                                                                                                                          |             |
| Unmodified Device:       | <i>AMSure</i> ™ Urethral Catheterization Tray (K030712)                                                                                                                                           |             |
| Device Description:      | The <i>AMSure</i> ™ PVC Intermittent Catheter is a sterile, single-<br>use patient device, compromising a PVC tubing and funnel.                                                                  |             |
| Intended Use:            | The <i>AMSure</i> ™ PVC Intermittent Catheter is intended for use in<br>the drainage of urine from the bladder.                                                                                   |             |
| Comparison to Predicate: | The <i>AMSure</i> ™ PVC Intermittent Catheter is a configuration<br>and label modification of the <i>AMSure</i> ™ Urethral<br>Catheterization Tray (K030712) and is intended for the same<br>use. |             |
| Non-Clinical Testing:    | Performance and biocompatibility testing has demonstrated<br>the safety and effectiveness of the <i>AMSure</i> ™ PVC<br>Intermittent Catheter for its intended use.                               |             |

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo and the word "DEPARTMENT". The logo is a stylized image of an eagle with three lines representing its wings. The word "DEPARTMENT" is in a bold, sans-serif font and is underlined. The image is in black and white.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

### AUG 2 1 2009

Mr. Holiven Ji Manager of International Regulatory Affairs Amsino International, Incorporated 855 Towne Center Drive POMONA CA 91767

Re: K091306

Trade/Device Name: AMSure" PVC Intermittent Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZD Dated: July 20, 2009 Received: July 28, 2009

Dear Mr, Ji:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical

{2}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Lammy M. Mouck

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use Statement

510(k) Number:

K091/306

(if known)

:

510(k) Number --

Device Name: AMSure PVC Intermittent Catheter

Indications The AMSure PVC Intermittent Catheter is intended for use in for Use: the drainage of urine from the bladder.

Prescription Use **√** AND/OR Over-The-Counter Use

(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                                           | <div> <img alt="Signature" src="signature.png"/> </div> |
|---------------------------------------------------------------|---------------------------------------------------------|
| Division of Reproductive, Abdominal, and Radiological Devices | K091306                                                 |

Page - 4

---

**Source:** [https://fda-staging.innolitics.com/device/K091306](https://fda-staging.innolitics.com/device/K091306)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).