K091285 · Avicenna Laser Technology, Inc. · ILY · Jul 28, 2009 · Physical Medicine
Device Facts
Record ID
K091285
Device Name
AL T LASER, MODEL AVI HPLL-12
Applicant
Avicenna Laser Technology, Inc.
Product Code
ILY · Physical Medicine
Decision Date
Jul 28, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ALT Laser, Model AVI HPLL-12 is intended to emit energy in the infrared spectrum to provide topical heating for use when heat is indicated in the temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis, and promoting relaxation of the muscle tissue.
Device Story
Device emits 980nm infrared and 650nm red visible laser light in overlapping patterns; 650nm light serves as visual indicator for invisible 980nm beam. System includes safety interlock: if 650nm aiming beam is absent, 980nm infrared output is locked out and operator is alerted. Used in clinical settings by qualified users; requires protective eyewear for all present. Provides topical heating to treat pain and muscle stiffness. Benefits include non-invasive thermal therapy for musculoskeletal conditions.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Infrared lamp (21 CFR 890.5500). Emits 980nm infrared and 650nm visible red laser light. Features safety interlock system for beam monitoring. Class IV laser product requiring controlled access and protective eyewear.
Indications for Use
Indicated for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis, and promoting muscle relaxation. Contraindicated in pregnant females (abdominal/lumbosacral), children (epiphyseal lines), patients with pacemakers (thorax/over pacemaker), over thyroid/ovaries/testicles, and patients taking heat/light-sensitive drugs (e.g., certain steroids).
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
# Section 5: 510k) Summary
The Summary of Safety and Effectivenes on the ALT Laser, Model AVI awailson available and represented the time the time the alterations of the conclusions or recommendations set forth.
| Applicant | Avicenna Laser Technology, Inc.<br>Bruce Coren, DVM, MS.<br>9822 East Washington Street, Suite 6<br>Chagrin Falls, Ohio 44023<br>JUL 2 8 2009 |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone | (561) 722 - 1153 |
| Facsimile | (561) 820 - 9646 |
| Date | April 24, 2009 |
| Name | ALT Laser, Model AVI HPLL-12 |
| Classification | Infrared Lamp, 21 CFR 890.5500 |
| Predicate: | ALT Laser, Model VTR75, K031612 market clearance date December 11,<br>2003 |
| Description | The system will produce a 980nm infrared and a 650 nm red visible laser<br>light in overlapping patterns. The two types of light will be mixed such<br>that the visible 650 nm light becomes a reasonable indicator of the<br>invisible infrared 980nm light. The 650 nm visible aiming laser light will<br>be sensed and if the light is absent, it will lock out the infrared 980nm light<br>and alert the operator. |
| Intended Use | ALT Laser, Model AVI HPLL-12 is intended to emit energy in the infrared<br>spectrum to provide topical heating for use when heat is indicated in the<br>temporary relief of minor muscle and joint pain, muscle spasm, pain and<br>stiffness associated with arthritis, and promoting relaxation of the muscle<br>tissue. |
| Contraindications | Do not apply infrared light to abdominal or lumbosacral points in<br>pregnant females. Do not apply infrared light to the epiphyseal lines in children. Do not apply infrared light to the thorax or over the pacemaker itself<br>in patients with pacemakers. Do not apply infrared light over the thyroid gland, ovaries and<br>testicles. Do not apply infrared light to patients who are taking drugs that have<br>heat or light sensitive contraindications, such as but not limited to<br>certain types of steroids |
| Warning | Do not use in the presence of flammable solvents or anesthetics. Never look directly into the laser light source or at scattered laser light<br>from any reflective surfaces. Never sight down the beam into the<br>source. Direct eye contact with the output beam from the laser will cause<br>serious damage and possible blindness. Avoid direct exposure to the laser light. The intensity of the beam can<br>easily cause flesh burns or ignite clothing. Never look directly into the beam or at a specular reflection even<br>while wearing protective eyewear. |
| Caution | Always wear laser safety eyewear, which is optically dense at the wavelength of operation. Limit access to the laser to qualified users who are familiar with laser safety practices and who are aware of the dangers involved. As a precaution against accidental exposure to the output beam or its reflection, those using the system should wear laser safety glasses as required by the wavelength being generated. Use the laser in an enclosed room. Laser light will remain collimated over long distances and therefore prevents a potential hazard if not confined. Post warning signs in the area of the laser beam to alert those present. Advise all those using the laser of these precautions. It is good practice to operate the laser in a room with controlled and restricted access. Always wear laser safety eyewear, which is optically dense at the wavelength of operation. Use safety interlocks to restrict access to laser areas and prevent unauthorized use of the laser. Post warning signs for class IV laser products. |
| Substantial<br>Equivalency<br>Information | The light at infrared wavelengths is therapeutic in nature and although the output of infrared energy varies between devices this difference only impacts the amount of time to attain the therapeutic thermal effect of the infrared energy. Therefore, the operational differences and output between the predicate device ALT Laser VTR 75 and ALT Laser, Model AVI HPLL-12 does not present a significant effect in the therapeutic outcome. |
| Technological<br>Characteristics | The intended use and technological characteristic of the two devices do not vary significantly. The safety and effectiveness of the Avicenna Laser Technology, Inc.'s ALT Laser, Model AVI HPLL-12 infrared device is comparable to that of the Avicenna's ALT Laser Model VTR 75 device. |
{1}------------------------------------------------
## ku91280 510(k) SUMMARY, continue
The Summary of Safety and Effectiveness on the ALT Laser, Model AVI HPLL-12 reflects data available and represented at the time the submission was prepared, but caution should be exercised in interpreting the data. The results of future studies and or tests may require alterations of the conclusions or recommendations set forth.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 8 2009
Avicenna Laser Technology, Inc. % Bruce R. Coren, DVM, MS President 9822 East Washington Street, Suite 6 Chagrin Falls, Ohio 44023
Re: K091285
Trade/Device Name: ALT Laser, Model AVI HPLL-12 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: ILY Dated: June 19, 2009 Received: July 6, 2009
Dear Dr. Coren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
### Page 2 - Bruce R. Coren, DVM, MS
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrl/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800).638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours
Chlare Buchup
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Section 4: Indication for Use Summary
510(k) Number (if known): K091285
Device Name: _ _ _ _ _ _ _ _ _ _ _ ALT Laser, Model AVI HPLL-12
Indications For Use:
ALT Laser, Model AVI HPLL-12 is intended to emit energy in the infrared spectrum to provide topical heating for use when heat is indicated in the temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis, and promoting relaxation of the muscle tissue.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
DuilR.Oale for mken
(Division Sign-Off) Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K091285
Prescription Use X (Per 21 CFR 801.109)
OR
Over-The-Counter-Use _
- 11A-
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.