K091245 · Preservation Solutions, Inc. · KDN · Jun 17, 2009 · Gastroenterology, Urology
Device Facts
Record ID
K091245
Device Name
COSTORSOL
Applicant
Preservation Solutions, Inc.
Product Code
KDN · Gastroenterology, Urology
Decision Date
Jun 17, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5880
Device Class
Class 2
Indications for Use
CoStorSol® is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.
Device Story
CoStorSol® is a sterile, non-pyrogenic organ preservation solution; formulation includes soluble colloids, buffers, sodium/potassium salts, redox stabilizers, and phosphoric compounds to support tissue viability via ATP regeneration. Used by transplant clinicians for hypothermic flushing and cold storage of harvested kidney, liver, and pancreas organs. Supplied in 1-liter flexible bags; must be chilled to 2°–6°C prior to use. Acts as initial vasculature flush medium and subsequent cold storage medium. Facilitates organ transport and eventual transplantation. Removes requirement for point-of-use microfiltration compared to previous versions. Benefits patient by maintaining organ viability during the interval between donor removal and recipient transplantation.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993. Particulate matter testing conducted per USP <788> for large volume injections. Sterilization validated per ISO 17665 or USP <1211>.
Technological Characteristics
Sterile, non-pyrogenic aqueous solution containing colloids, buffers, salts, and phosphoric compounds. Supplied in 1-liter flexible administration pouches. Hypothermic storage (2°–6°C). Biocompatibility per ISO 10993; particulate limits per USP <788>; sterilization per ISO 17665/USP <1211>.
Indications for Use
Indicated for flushing and cold storage of kidney, liver, and pancreas organs during organ removal from donors, in preparation for storage, transport, and transplantation into recipients.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}------------------------------------------------
K09/245
182
### 510(k) SUMMARY
#### SUBMITTER INFORMATION
- A. Company Name: Preservation Solutions Inc.
JUN 1 7 2009
- B. Company Address: 980 Proctor Drive Elkhorn, Wisconsin 53121
- C. Company Phone: 262 723 6715
- D. Company Facsimile: 262 723 4013
- E. Contact Person: William Wagner Quality Assurance Director
## DEVICE IDENTIFICATION
- A. Device Trade Name: CoStorSol®
- B. Device Common Name: Organ Storage Solution
- C. Classification Name: Isolated kidney perfusion and transport system and accessories
- D. Class II (21 CRF 876.5880)
- E. Device Code: KDN
### IDENTIFICATION OF PREDICATE DEVICES
CoStorSol®, to be stored at ambient room temperature, is a transplant graft storage solution for harvested organs, which is substantially equivalent to the original CoStorSol® solution cleared to market under premarket notifications K073693 and K083453.
### DEVICE DESCRIPTION
Preservation Solutions, Inc. manufacturers CoStorSol® according to a "recipe" pioneered at the University of Wisconsin by Folkert O. Belzer, MD and James H. Southard, PhD. Indeed, the cold storage solution is often referred to as "Belzer UW" solution. The formulation includes soluble colloids, buffers, sodium and potassium salts, redox stabilizers, and phosphoric compounds to aid tissue viability by enabling regeneration of adenosine triphosphate (ATP).
CoStorSol® is a clear to light yellow, sterile, non-pyrogenic solution for hypothermic flushing and storage of organs. The solution is packaged in 1-liter bags, which must be chilled to between 2° and 6° C prior to use. The solution may be used without any point of use filtration.
Traditional 510(k) CoStorSol - No Time of Use Filtration
{1}------------------------------------------------
# INDICATIONS FOR USE
K091245
2 of 2
CoStorSol® is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.
# COMPARISON TO PREDICATE DEVICES
CoStorSol® solution was cleared under 510(k) K083453 with a specified shelf life of one (1) year when stored under refrigeration at temperatures no lower than 2°C, or at ambient indoor There has been no change in the formulation, packaging, or temperatures up to 25°C. indication for use for the product. The proposed CoStorSol® solution is identical to the predicate, but label instructions to pass the solution though a microfilter (Pall Blood Transfusion Filter) at the time of use have been removed. Preservation Solutions, Inc. has compiled data showing that unfiltered CoStorSol® is equivalent to filtered CoStorSol®. The directions to chill CoStorSol®, prior to its use for flushing or storing organs for transplant, remain unaltered. Only the instruction to connect a sterile microfilter to the spike port while draining the solution from its flexible container has been deleted from the labeling. The solution was, and is still designed for storage without freezing. CoStorSol® will continue to be supplied in 1-liter flexible bags for connection to standard administration sets for flushing of harvested organs. CoStorSol® is used as an initial vasculature flush medium for an organ, and then as a cold storage medium.
# BIOCOMPATIBILITY, STERILIZATION, PACKAGING, AND BENCH TESTING
Test results have shown CoStorSol® itself to be a biocompatible solution, supplied in flexible solution administration pouches, which have likewise been tested and shown to be biocompatible. The ISO 10993 series of standards were referenced during the planning and execution of all biocompatibility testing. Particulate matter does not exceed the limits set in USP Section <788>, for large volume injections. CoStorSol® is supplied sterile and non-pyrogenic in order to assure safety for transplant recipients. Sterilization processes were validated according to either ISO 17665 or USP Section <1211>, as appropriate.
### CONCLUSION
The above statements establish substantial equivalence between the predicate, CoStorSol® with instructions to pass the solution through a microfilter at the time of use, and the proposed product. CoStorSol® without instructions for final filtration.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircles the eagle. The seal is in black and white and has a simple, official design.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 1 7 2009
Preservations Solutions, Inc. % Mr. Neil Burris Principal Neil Burris and Associates 4250 Grove Street DENVER CO 80211
Re: K091245
Trade/Device Name: ÇoStorSo1® Regulation Number: 21 CFR 876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: II Product Code: KDN Dated: April 20, 2009 Received: April 28, 2009
# Dear Mr. Burris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
21 CFR 876.xxx 21 CFR 884.xxx 21 CFR 892.xxx Other
(Gastroenterology/Renal/Urology) (240) 276-0115 (Obstetrics/Gynecology) (240) 276-0115 (Radiology) (240) 276-0120 (240) 276-0100
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K09/245
1 of 1
### 5. Statement of Indication for Use
Device Name: CoStorSol®
Indications for Use
CoStorSol® is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.
Prescription Use XXXX (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Laura De Norris
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Traditional 510(k) CoStorSol - No Time of Use Filtration
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.