PHILIPS MX 16 SLICE CT SYSTEM PHASE11, MODEL 9896 056 56511

K091195 · Philips Medical Systems (Cleveland), Inc. · JAK · May 27, 2009 · Radiology

Device Facts

Record IDK091195
Device NamePHILIPS MX 16 SLICE CT SYSTEM PHASE11, MODEL 9896 056 56511
ApplicantPhilips Medical Systems (Cleveland), Inc.
Product CodeJAK · Radiology
Decision DateMay 27, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Philips MX 16-slice CT system can be used as a Head and Whole Body Computed Tomography X-ray System featuring a continuously rotating x-ray tube and detector array with multislice capability up to 16 slices simultaneously. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients.

Device Story

The MX 16-slice CT system is a whole-body diagnostic imaging device; utilizes a continuously rotating X-ray tube and detector array to acquire transmission data; performs multi-slice acquisition (up to 16 slices simultaneously). Data is reconstructed by computer into cross-sectional images at various angles and planes. Operated by trained radiology staff in clinical settings. Output consists of cross-sectional images displayed on equipment monitors for physician interpretation. Device aids in clinical diagnosis by providing anatomical visualization. Benefits include non-invasive diagnostic capability for a wide range of patient conditions.

Clinical Evidence

No clinical data provided. Safety and effectiveness are demonstrated through bench testing, adherence to recognized international standards (IEC 60601-1, ISO 13485), and compliance with FDA performance standards for radiation-emitting products.

Technological Characteristics

Whole-body CT scanner; rotating X-ray tube and detector array; 16-slice multi-slice capability. Complies with IEC 60601-1 (electrical/mechanical safety) and 21 CFR Subchapter J (radiation safety). Software-based image reconstruction. System includes gantry, patient support, and display equipment.

Indications for Use

Indicated for head and whole body computed tomography imaging in all patients.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K091195 MAY 27 2009 # 510(k) Summary Philips Medical Systems (Cleveland) Inc. "Brilliance Volume" This summary of this 510(k) provides safety and effectiveness information submitted in accordance with the requirements of 21 CFR 807.92. #### 1. Submitter Philips Medical Systems (Cleveland), Inc. 595 Miner Road Cleveland, OH 44143 (440) 483-3000 # Contact: | | Michael A. Chilbert, Ph.D., P.E. | |------------|----------------------------------| | Telephone: | 440 483 3284 | | FAX: | 440 483 4918 | | Email: | Michael.chilbert@philips.com | Date of Summary: February 25, 2009 #### 2. Device Name and Classification | Device Name: | MX 16-slice CT System, phase II | |----------------------|----------------------------------| | Classification Name: | Computed Tomography X-Ray System | The FDA has classified the Computed Tomography X-Ray System and its accessories as Class II in 21 CFR 892.1750 (Product Code 90 JAK) #### 3. Predicate Device Information In the opinion of Philips Medical Systems Inc., the MX 16-slice CT System, phase II is of comparable type and substantially equivalent to the legally marketed devices currently in commercial distribution, namely: the Brilliance CT 16-slice in K012009 (a.k.a. MX8000 IDT CT Scanner). It expands the capabilities of the MX 16-slice CT system (K083498). #### Device Description 4. The MX 16-slice CT System, phase II is a Whole Body Computed Tomography X-Rav System featuring a continuously rotating X-ray tube and detectors gantry and multi-slice capability. The acquired x-ray transmission data is reconstructed by computer into crosssectional images of the body taken at different angles and planes. This device also includes signal analysis and display equipment, patient and equipment supports. components and accessories. {1}------------------------------------------------ ### 5. Intended Use of the device The MX 16-slice CT system can be used as a Whole Body Computed Tomography X-ray System featuring a continuously rotating x-ray tube and detector array with multislice capability up to 16 slices simultaneously. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients. ### 6. Comparison to Predicate Devices In the opinion of Philips Medical Systems (Cleveland), Inc., the "Philips MX 16-slice" CT scanner is of comparable type and substantially equivalent to the legally marketed devices because it has the similar technological characteristics and sub-assemblies as the current commercial distribution of Brilliance CT 16-slice (K012009). The safety and effectiveness of the "Philips MX 16-slice" is assured by adherence to Good Manufacturing Practices (GMP) 21 CFR 820 and to International Standards ISO 13485:2003. Potential hazards are identified in a hazard analysis and controlled in the following manner: Software safety is assured by the company procedures that conform to accepted practices. Quality assurance procedures are adhered to, and meeting the specifications and functional requirements is demonstrated via testing. Electrical and Mechanical safety is assured by adherence to IEC 60601-1 Standards. Radiation safety is assured by compliance with 21 CFR. Subchapter J Performance Standards. Based on the above considerations, it is Philips's opinion that the MX 16-slice CT Scanner is substantially equivalent in safety and effectiveness to the predicate device: Brilliance CT 16-slice with 510(k) Number K012009. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract shapes that resemble human figures or birds in flight. The overall design is simple and conveys a sense of government authority and public service. ### Public Health Service 27 2009 MAY Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Philips Medical Systems (Cleveland), Inc. % Mr. Ned Devine Senior Staff Engineer Underwriters Laboratories, Inc. 333 Pfingsten Road NORTHBROOK IL 60062 Re: K091195 Trade/Device Name: Philips MX 16-slice CT system Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: May 8, 2009 Received: May 12, 2009 Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. 21 CFR 876.xxx 21 CFR 884.xxx 21 CFR 892.xxx Other (Gastroenterology/Renal/Urology) (Obstetrics/Gynecology) (Radiology) (240) 276-0115 (240) 276-0115 (240) 276-0120 (240) 276-0100 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, [signature] Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure Page 2 {4}------------------------------------------------ 510(k) Number (if known): K091195 Page 1 of 1 Philips MX 16-slice CT system Device Name: Indications for Use: The Philips MX 16-slice CT system can be used as a Head and Whole Body Computed Tomography X-ray System featuring a continuously rotating x-ray tube and detector array with multislice capability up to 16 slices simultaneously. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Hels Remm (Division Sign Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...