EZ-MIO and EZ-IO: The EZ-MIO manual driver and EZ-IO power driver provide intraosseous access in the distal tibia of adults when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours. Vidaport Intraosseous Infusion System: The VidaPort provides intraosseous access in the proximal tibia, when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours. The device is for use in adult patients only. EZ-IO Humeral Head: The Humeral Head EZ-IO provides intraosseous access in the Humeral Head, when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
Device Story
Intraosseous (IO) access system comprising manual/power drivers and needles; provides vascular access in distal tibia, proximal tibia, or humeral head. Used in emergent, urgent, or medically necessary clinical settings when IV access is unattainable. Operates by inserting needle into bone marrow space to facilitate fluid/medication delivery. Intended for adult patients; duration of use up to 24 hours. Healthcare providers use device to establish life-saving access; output is established IO route. Benefits include rapid, reliable access for resuscitation or therapy when peripheral veins are inaccessible.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on identical technological characteristics and design to predicate devices.
Technological Characteristics
Hypodermic single lumen needle system. Includes manual and power-driven insertion mechanisms. Materials, design, and biocompatibility are identical to previously cleared predicates. Device is intended for short-term (up to 24 hours) intraosseous access.
Indications for Use
Indicated for adult patients requiring intraosseous access in the distal tibia, proximal tibia, or humeral head when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
# 5. 510(k) Summary
Image /page/0/Picture/2 description: The image contains two lines of text. The first line appears to be handwritten and reads "KO91140". The second line is a stylized, block-lettered word "vidacare". The font used for "vidacare" has a unique, almost circuit-like appearance.
722 Isom Road. San Antonio, TX 78216 Tel (210) 375 8500 Fax (210) 375 8537 Toll Free (within US) 866 479 8500 www.vidacare.com
OCT 1 4 2009
### SUMMARY
722 Isom Road
210-375-8500
VidaCare Corporation
San Antonio, TX 78216
Submitter's name: Address:
Phone: Fax number:
Name of contact person:
210-375-8537 Grace Holland Requlatory Specialists, Inc 3722 Ave. Sausalito Irvine, CA 92606 Phone: 949-262-0411
Fax: 949-552-2821
| Date the summary was prepared: | Original, April 17, 2009<br>Revised, October 10, 2009 |
|--------------------------------|-------------------------------------------------------|
|--------------------------------|-------------------------------------------------------|
EZ-MIO Distal Tibia, EZ-IO Distal Tibia Name of the devices: Vidaport Intraosseous Infusions System EZ-IO, Humeral Head
Intraosseous Infusion System Common or usual name: Hypodermic single lumen needle Classification name:
The legally marketed devices to which we are claiming equivalence [807.92(a) (3)]:
| | 510(k) Number | Trade or Proprietary or Model Name | Manufacturer |
|---|---------------|---------------------------------------|----------------|
| 1 | K062956 | EZ-MIO, EZ-IO, Distal Tibia | Vidacare Corp. |
| 2 | K032885 | Vidaport Intraosseous Infusion System | Vidacare Corp. |
| 3 | K052408 | EZ-IO, Humeral Head | Vidacare Corp. |
{1}------------------------------------------------
Indications for Use:
## EZ-MIO and EZ-IO:
The EZ-MIO manual driver and EZ-IO power driver provide intraosseous access in the distal tibia of adults when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
## Vidaport Intraosseous Infusion System:
The VidaPort provides intraosseous access in the proximal tibia, when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours. The device is for use in adult patients only.
## EZ-IO Humeral Head:
The Humeral Head EZ-IO provides intraosseous access in the Humeral Head, when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
Summary of the technological characteristics of our device compared to the predicate devices:
This submission extends the indications for use to include usage in emergent, urgent, or medically necessary cases for up to 24 hours. There have been no changes to the design or components of the devices cleared under 510(k) K062956, K032885 and K052408, and therefore the comparison of technological characteristics listed below are identical.
Target Population Driver Design Features Needle Design Technique Sterility Biocompatibility Where Used
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. In the center of the seal is an emblem that features a stylized representation of an eagle with its wings spread. The eagle is facing to the right and has three lines extending from its wings.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-0609 Silver Spring, MD 20993-0002
Vidacare Corporation C/O Ms. Grace Holland Regulatory Specialist Regulatory Specialist, Incorporated 3722 Avenue Sausalito Irvine, California 92606
## OCT 1 4 2009
Re: K091140
Trade/Device Name: Vidacare® Needle for EZ-M10 and EZ-IO, Vidaport Intraosseous Infusion System, and EZ-IO Humeral Head
Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle
Regulatory Class: II
Product Code: FMI
Dated: September 25, 2009
Received: September 28, 2009
Dear Ms. Holland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH /CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Susan Lanier
Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
#### 4. Indications for Use Statement
#### Indications for Use
. 199
510(k) Number (if known): K091140
### Device Name: Vidacare® Needle for EZ-MIO and EZ-IO, Vidaport Intraosseous Infusion System, and EZ-IO Humeral Head
Indications for Use:
#### EZ-MIO and EZ-IO:
The EZ-MIO manual driver and EZ-IO power driver provide intraosseous access in the distal tibia of adults when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
#### Vidaport Intraosseous Infusion System:
The VidaPort provides intraosseous access in the proximal tibia, when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours. The device is for use in adult patients only.
#### EZ-IO Humeral Head:
The Humeral Head EZ-IO provides intraosseous access in the Humeral Head, when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
X Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
-------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anthony D. Arata
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Page 1 of 1
REGULATORY SPECIALISTS, INC
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.