K091139 · Conmed Corporation · FYD · May 5, 2009 · General, Plastic Surgery
Device Facts
Record ID
K091139
Device Name
CONMED AER DEFENSE SMOKE EVACUATOR
Applicant
Conmed Corporation
Product Code
FYD · General, Plastic Surgery
Decision Date
May 5, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5070
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The CONMED® Aer Defense™ Smoke Evacuator is designed to remove smoke, aerosols and mitigate odors produced by surgical smoke during electrosurgical procedures.
Device Story
CONMED Aer Defense Smoke Evacuator is a high-suction, high-flow surgical smoke evacuation system. Device uses a two-stage thru-flow vacuum motor to draw surgical smoke from the surgical site through vacuum tubing into a replaceable, self-contained FilterOne ULPA-grade carbon filter. System operates in multiple modes: Normal, Laparoscopic (low flow), Footswitch, Continuous, ConMed ESU Automatic (adjusts flow based on ESU power usage), and Integrated Operating Room Interface. Used in OR settings by surgical staff. Output is the removal of smoke/aerosols/odors, improving visibility and air quality for the surgical team. Device benefits patients and staff by reducing exposure to surgical smoke byproducts. Verification testing confirms safety and performance equivalence to predicate systems.
Clinical Evidence
Bench testing only. No clinical data presented. Verification and validation testing included electrical safety (IEC 60601-1-1, IEC 60601-1-2, UL 60601-1), mechanical design verification, and software verification to confirm operational modes and performance requirements.
Technological Characteristics
Two-stage thru-flow vacuum motor; ULPA-grade carbon filter (99.9995% efficiency for 0.12um particles); suction flow rates up to 20 LPM (Laparoscopic) or 25 CFM (Normal); serial port for OR integration; electrical safety per IEC 60601-1-1, IEC 60601-1-2, and UL 60601-1.
Indications for Use
Indicated for removal of smoke, aerosols, and mitigation of odors produced by surgical smoke during electrosurgical procedures in patients undergoing surgery.
Regulatory Classification
Identification
Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
Porta Plumesafe™ 601 Smoke Evacuation System (K924732)
ConMed Integrated Service Manager SM40SE (K053464)
Submission Summary (Full Text)
{0}------------------------------------------------
K091139
| 4. | Summary of Safety and Effectiveness | | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Submitted by: | CONMED Electrosurgery Division<br>14603 East Fremont Avenue<br>Centennial, CO 80112 USA<br>Telephone: 303-269-8296<br>Facsimile: 303-699-9854 | | |
| Contact Person: | Damaris Velez | | |
| Date Prepared: | February 12, 2009 | | |
| Proprietary Name: | CONMED® Aer Defense™ Smoke Evacuator | | |
| Common Name: | Electrosurgical Surgical Smoke Evacuator | | |
| Classification Name: | Apparatus, Exhaust, Surgical<br>FYD | | |
| Predicate Device: | This product is similar in design, composition, and function to the<br>Porta Plumesafe™ 601 Smoke Evacuation System, K924732 and<br>the Smoke Evacuation section of the ConMed Integrated Service<br>Manager SM40SE, K053464. | | |
# Technological Characteristics
## Smoke Evacuator:
The CONMED Aer Defense™ Smoke Evacuator has been designed with a high suction, high flow rate, two-stage thru flow vacuum motor. Its motor is used to draw the surgical smoke from the surgical site, through the vacuum tubing and into the CONMED Aer Defense™ Smoke Evacuator's filter where the surgical smoke is processed by a series of filtration stages. A single disposable filter is used to simplify the installation and removal during filter changes.
## Filter:
The filter for the Aer Defense™ Smoke Evacuator, like the filters for the predicate devices, is a replaceable self-contained filter that is completely enclosed to help protect the health care personnel from potential contamination during filter changes. The FilterOne™ is an Ultra Low Penetration Air (ULPA) grade with carbon filter that has a filter efficiency of 99.9995% for particle sizes of 0.12um or greater. These characteristics are identical to the predicate filters.
The filter life for the FilterOne™ is based on the flow rate through the filter and is 35 hrs in Lap Mode, and 10 hrs at maximum flow in Normal Mode. The filter life for the
Aer Defense™ Smoke Evacuator
{1}------------------------------------------------
predicate devices is based on time of use and is 8 hrs maximum. Bench testing has shown that the filter life for the Aer Defense™ Smoke Evacuator is substantially equivalent to the predicate filters based on bench testing.
# Suction Flow Rate:
The maximum suction flow rate for the Aer Defense™ Smoke Evacuator ranges from a maximum 20 LPM in Laparoscopic Mode to 25 CFM in Normal Mode and is substantially equivalent to the flow range of 20 LPM in Normal Mode and 34 CFM in Boost Mode for the predicate devices, K924732 and K053464.
## Performance Characteristics:
Performance Testing: Design Verification & Validation bench testing was designed and conducted to show that the Aer Defense™ Smoke Evacuator operates as safe and effective as the predicate devices. K924732 and K053464.
Electrical Verification testing showed that the Aer Defense™ Smoke Evacuator met its operational mode and electrical design requirements as specified in the test protocol. Test results indicate that the Aer Defense™ Smoke Evacuator is substantially equivalent to the predicate devices, K924732 and K053464.
Mechanical Verification testing has shown that the Aer Defense™ Smoke Evacuator has met its mechanical design requirements as specified in the protocol. Test results indicate that the Aer Defense™ Smoke Evacuator is substantially equivalent to the predicate devices, K924732 and K053464.
Software Verification testing has shown that the Aer Defense™ Smoke Evacuator has met the operational mode and software design requirements as specified in the protocol. Test results indicate that the Aer Defense™ Smoke Evacuator is substantially equivalent to the predicate devices. K924732 and K053464.
Validation testing has shown that the Aer Defense™ Smoke Evacuator has met its operational mode and validation requirements. Test results indicate that the Aer Defense™ Smoke Evacuator is substantially equivalent to the predicate devices, K924732 and K053464.
These acceptable test results for the non-clinical bench testing indicate that the Aer Defense™ Smoke Evacuator operation is substantially equivalent to the predicate devices, K924732 and K053464.
#### Electrical Safety:
Electrical Safety testing that was conducted for the Aer Defense™ Smoke Evacuator was done according to IEC 60601-1-1, IEC 60601-1-2, and UL 60601-1. This testing
Section 4 Page 2 of 7
{2}------------------------------------------------
indicates that the electrical safety for the Aer Defense™ Smoke Evacuator is substantially equivalent to the predicate devices, K924732 and K053464.
# Modes of Operation:
Normal Mode - Normal mode is the default activation mode. The smoke evacuator will run in this mode when none of the available modes have been selected and illuminated on the control panel (Laparoscopic, Continuous, Automatic, or Footswitch). While in Normal Mode, the smoke evacuator will start suction upon activation from an ESU. The Normal Mode is substantially equivalent to the Normal Mode and Boost Mode in predicate devices, K924732 and K053464.
Laparoscopic Mode - This mode is designed to provide a low flow suction control range for use in minimally invasive procedures. Laparoscopic mode is controlled by Normal, ConMed ESU Automatic, Footswitch, or Continuous activation modes. The Laparoscopic Mode is substantially equivalent to the Laparoscopic Mode and the Normal Mode in predicate devices K053464 and K924732 respectively.
Footswitch Mode - This mode starts suction upon activation of the Pneumatic Footswitch only. The smoke evacuator can be turned on or off by depressing the footswitch once to turn suction on and once again to turn suction off. This mode functions the same as the Footswitch Mode for the predicate devices, K924732 and K053464.
Continuous Activation Mode - This mode runs the smoke evacuator at the selected power level continuously, without activation of the Handpiece, until the mode has been deselected, another mode is selected, or the unit is turned off. This operating mode is not available in the predicate devices.
Electrical and Software Verification bench testing has shown this mode to be as safe and effective as the normal operation of the device and is substantially equivalent to the predicate devices, K924732 and K053464.
ConMed ESU Automatic Mode - This mode starts suction upon Handpiece activation from either a ConMed System 5000 or System 2450 ESU. The smoke evacuator will also employ automatic adjustments of the flow level based on the ConMed ESU power usage. This operating mode is not available in the predicate devices.
Electrical and Software Verification bench testing has shown this mode to be as safe and effective as the normal operation of the device and is substantially equivalent to the predicate devices, K924732 and K053464.
Integrated Operating Room Interface Mode - This mode allows the smoke evacuator front panel controls to be set through the devices serial port by an integrated operating room control center. This operating mode is not available in the predicate devices,
{3}------------------------------------------------
Electrical and Software Verification bench testing has shown this mode to be as safe and effective as the normal operation of the device and is substantially equivalent to the predicate devices, K924732 and K053464.
## Biocompatibility:
The CONMED® Aer Defense™ Smoke Evacuator and the predicate devices do not contact the patient and are substantially equivalent regarding biocompatibility.
## Indication for Use:
The CONMED® Aer Defense™ Smoke Evacuator is designed to remove smoke, aerosols and mitigate odors produced by surgical smoke during electrosurgical procedures. This indication is substantially equivalent to the indications of the predicate devices. K924732 and K053464.
## Summary:
The Electrical, Mechanical, Software, and Safety testing indicate that the CONMED Aer Defense Smoke Evacuator is substantially equivalent to the predicate devices identified in K924732 and K053464.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services, USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the top half of the circle. In the center of the seal is an emblem that resembles an eagle or other bird-like figure. The emblem is stylized and abstract, with flowing lines that suggest movement or flight.
MAY - 5 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Underwriters Laboratories, Incorporated C/O Mr. Jeff D. Rongero Responsible Third Party Official Conmed Corporation 12 Laboratory Drive Research Triangle Park, North Carolina 27709
Re: K091139
Trade/Device Name: Conmed Aer Defense Smoke Evacuator Regulation Number: 21 CFR 878.5070 Regulatory Class: Il Product Code: FYD Dated: April 17, 2009 Received: April 20, 2009
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2- Mr. Rongero
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0100. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Runner
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
| 510(k) Number (if known): | K091139 |
|---------------------------|---------|
|---------------------------|---------|
| Device Name: | CONMED <sup>®</sup> Aer Defense <sup>™</sup> Smoke Evacuator |
|--------------|--------------------------------------------------------------|
|--------------|--------------------------------------------------------------|
Indications for Use:
The CONMED Aer Defense™ Smoke Evacuator is designed to remove smoke, aerosols and mitigate odors produced by surgical smoke during electrosurgical procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Suan Dionnes
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page of
Kagillag 510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.