← Product Code [NAD](/productcode/NAD) · K091133

# APLION TOPICAL CARE SYSTEM (K091133)

_Aplion Medical,Llc · NAD · May 20, 2009 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K091133

## Device Facts

- **Applicant:** Aplion Medical,Llc
- **Product Code:** [NAD](/productcode/NAD.md)
- **Decision Date:** May 20, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4020
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Aplion Topical Care System is an occlusive wound dressing which is intended to provide a moist wound healing environment to facilitate the normal wound healing process. It also permits the introduction of topical wound treatment solutions and suspensions.

## Device Story

The Aplion Topical Care System is an occlusive wound dressing designed to cover wounds and maintain a moist healing environment. The system comprises the dressing itself, a tubeset connection, and a fluid-delivery component. It allows for the delivery of topical wound treatment solutions and suspensions directly to the wound site. The device is intended for use by healthcare providers to facilitate the normal wound healing process.

## Clinical Evidence

No clinical data. Biocompatibility testing was conducted according to ISO 10993-1 and 510(k) Memorandum G95-1.

## Technological Characteristics

Synthetic polymeric material (e.g., polyurethane); includes tubeset connection and fluid-delivery component. Biocompatibility testing per ISO 10993-1.

## Regulatory Identification

An occlusive wound dressing is a nonresorbable, sterile or non-sterile device intended to cover a wound, to provide or support a moist wound environment, and to allow the exchange of gases such as oxygen and water vapor through the device. It consists of a piece of synthetic polymeric material, such as polyurethane, with or without an adhesive backing. This classification does not include an occlusive wound dressing that contains added drugs such as antimicrobial agents, added biologics such as growth factors, or is composed of materials derived from animal sources.

## Predicate Devices

- Wound Cell Transparent Dressing ([K020781](/device/K020781.md))
- Dermastream ([K060046](/device/K060046.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K091133
page 1 of 2

MAY 20 2009

# 510(k) SUMMARY Occlusive Wound Dressing Date of Summary: 04/17/09

#### A. General Provisions Submitter's Name: Aplion Medical, LLC. Submitter's Address: 2425 South 900 West Salt Lake City, UT 84119 Contact Person: Curtis Jensen Quality Manager, Ceramatec, Inc. Classification Name: Occlusive Wound Dressing 21 CFR 878.4020 Aplion Topical Care System Proprietary Name: Common Name: Occlusive Wound Dressing

## B. Name of Predicate Device(s)

- Occlusive Wound Dressing (Product Code NAD): K020781 . Kinetic Concepts, Inc., Wound Cell Transparent Dressing
- . Occlusive Wound Dressing (Product Code - NAD): K060046 Enzysurge, Ltd., Dermastream

## C. Device Description

An occlusive wound dressing is a nonresorbable, sterile or non-sterile device intended to cover a wound, to provide or support a moist wound environment, and to allow the exchange of gasses such as oxygen and water vapor through the device. It consists of a piece of synthetic polymeric material, such as polyurethane, with or without an adhesive backing (21 CFR 878.4020).

The Aplion Topical Care System ("Aplion System") consists of an occlusive wound dressing, a tubeset connection, and a small fluid-delivery component.

## D. Indications for Use

The Aplion Topical Care System is an occlusive wound dressing which is intended to provide a moist wound healing environment to facilitate the normal wound healing process. It also permits the introduction of topical wound treatment solutions and suspensions.

{1}------------------------------------------------

K09//33
Page 2 of 2

### E. Safety and Biocompatibility

#### Safety and Biocompatibility Summary

As a Class I, 510(k) exempt device, FDA review of the Aplion System is not necessary; nonetheless the materials used in the Aplion System were chosen for their biocompatibility, function, and suitability for the intended use of the device. Biocompatibility testing of the entire system was completed according to ISO 10993-1 and 510(k) Memorandum G95-1. Details on the individual materials and the biocompatibility test results are included in this submission.

The Aplion System has been tested to and complies with all applicable Safety and Biocompatibility Standards.

# F. Conclusion

The Aplion System, subject of this 510(k), is substantially equivalent to the Wound Cell Transparent Dressing (Kinetic Concepts; K020781) and the Dermastream (Enzysurge; K060046) products previously cleared by FDA.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA", which is arranged in a circular fashion around the left side of the logo. The logo is simple and monochromatic.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 20 2009

Aplion Medical, LLC % Ceramatec, Inc. Mr. Curtis Jensen Quality Manager 2425 South 900 West Salt Lake City, Utah 84119

Re: K091133

Trade/Device Name: Aplion Topical Care System Regulation Number: 21 CFR 878.4020 Regulation Name: Occlusive wound dressing Regulatory Class: I Product Code: NAD Dated: April 17, 2009 Received: April 20, 2009

Dear Mr. Jensen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at

{3}------------------------------------------------

Page 2 - Mr. Curtis Jensen

(240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at (240) 276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

page 1 of 1

510(k) Number (if known): K091133

Device Name: Aplion Topical Care System

Indications for Use: The Aplion Topical Care System is an occlusive wound dressing which is intended to provide a moist wound healing environment to facilitate the normal wound healing process. It also permits the introduction of topical wound treatment solutions and suspensions.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

4-1

Over-The-Counter Use (21 CFR 801 Subpart C)

(Please do not write below this line – Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David Kramer
for MXM

(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices

| 510(k) Number | K091133 |
|---------------|---------|
|---------------|---------|

4-1

---

**Source:** [https://fda-staging.innolitics.com/device/K091133](https://fda-staging.innolitics.com/device/K091133)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).