← Product Code [MQV](/productcode/MQV) · K091031

# PIONEER SURGICAL NANOSS BVF-E (K091031)

_Pioneer Surgical Technology · MQV · Nov 17, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K091031

## Device Facts

- **Applicant:** Pioneer Surgical Technology
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Nov 17, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Pioneer FortrOss Bone Void Filler is indicated for bony voids or gaps that are not intrinsic to the stability of bony structure. These defects may be surgically created osseous defects or defects created from traumatic injury to the bone. The product is indicated to be used in the posterolateral spine in conjunction with bone marrow aspirate and autograft bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

## Device Story

Pioneer FortrOss Bone Void Filler is a porous calcium phosphate material combined with a porcine gelatin carrier. It functions as an osteoconductive scaffold to fill bony voids or gaps not intrinsic to structural stability. Used in the posterolateral spine, the device is mixed with bone marrow aspirate and autograft bone by a surgeon. The material is designed to resorb over time and be replaced by natural bone during the healing process. It serves as a bone graft substitute to facilitate bone repair in surgical or traumatic defect sites.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Porous calcium phosphate material with porcine gelatin carrier. Osteoconductive scaffold. Resorbable. Form factor is a bone void filler for surgical implantation.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- Pioneer Surgical NanOss BVF-E ([K081558](/device/K081558.md))
- Actifuse™ Bone Graft Substitute ([K082575](/device/K082575.md))
- Formagraft™ Collagen Bone Graft Matrix ([K050798](/device/K050798.md))
- Vitoss® Scaffold Foam ([K032288](/device/K032288.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Pioneer FortrOss Bone Void Filler

## 510(k) Summary

NOV 1 7 2009

# Pioneer Surgical Technology Pioneer FortrOss Bone Void Filler

### ADMINISTRATIVE INFORMATION

| Manufacturer Name:         | Pioneer Surgical Technology<br>375 River Park Circle<br>Marquette, MI 49855                                                     |                      |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------|
|                            | Telephone:                                                                                                                      | +1 (906) 226-4812    |
|                            | Fax:                                                                                                                            | +1 (906) 226-4459    |
| Official Contact:          | Jonathan Gilbert                                                                                                                |                      |
| Representative/Consultant: | David J. Collette, MD<br>Floyd G. Larson<br>PaxMed International, LLC<br>11234 El Camino Real, Suite 200<br>San Diego, CA 92130 |                      |
|                            | Telephone:                                                                                                                      | +1 (858) 792-1235    |
|                            | Fax:                                                                                                                            | +1 (858) 792-1236    |
|                            | email:                                                                                                                          | dcollette@paxmed.com |

### DEVICE NAME AND CLASSIFICATION

| Trade/Proprietary Name:     | Pioneer FortrOss Bone Void Filler                                                     |
|-----------------------------|---------------------------------------------------------------------------------------|
| Common Name:                | Bone Void Filler                                                                      |
| Classification Regulations: | Filler, Bone Void, Calcium Compound<br>21 CFR 880.3045<br>Class II (special controls) |
| Product Code:               | MQV                                                                                   |
| Classification Panel:       | Orthopaedic and Rehabilitation Devices Panel                                          |
| Reviewing Branch:           | Restorative Devices Branch                                                            |

email:

flarson@paxmed.com

### INTENDED USE

Pioneer FortrOss Bone Void Filler is indicated for bony voids or gaps that are not intrinsic to the stability of bony structure. These defects may be surgically created osseous defects or defects created from traumatic injury to the bone. The product is indicated to be used in the posterolateral spine in conjunction with bone marrow aspirate and autograft bone. The product

pg 1 of 2

{1}------------------------------------------------

510(k) Summary

・

Pioneer FortrOss Bone Void Filler

provides a bone void filler that resorbs and is replaced with bone during the healing process.

#### DEVICE DESCRIPTION

Pioneer FortrOss Bone Void Filler is a porous calcium phosphate material mixed with a porcine gelatin carrier. The product is an osteoconductive scaffold mixed with a gelatin carrier for use in repairing bony defects in spinal indications.

#### EQUIVALENCE TO MARKETED DEVICES

Pioneer FortrOss Bone Void Filler is identical in composition and performance to Pioneer BVF-E cleared (K081558). Predicate devices with similar characteristics, which may include ceramic materials, porcine gelatin carrier, porous structure, and presentation are cited below.

Devices to Which Substantial Equivalence is Claimed:

K081558 - Pioneer Surgical NanOss BVF-E K082575 -- Actifuse™ Bone Graft Substitute K050798 - Formagraft™ Collagen Bone Graft Matrix K032288 -- Vitoss® Scaffold Foam

The comparisons and testing conducted on FortrOss Bone Void Filler demonstrate that the device is substantially equivalent to other bone void fillers currently in commercial distribution.

pg 2 of 2

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with flowing lines, positioned to the right. Encircling the figure is text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". The text is arranged in a circular fashion around the upper left portion of the logo.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Pioneer Surgical Technology % PaxMed International, LLC David J. Collette, M.D. 11234 El Camino Real, Suite 200 San Diego, CA 92130

NOV 1 7 2009

Re: K091031

Trade/Device Name: Pioneer FortrOss Bone Void Filler Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: October 29, 2009 Received: October 30, 2009

Dear Dr. Collette:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - David J. Collette, M.D.

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Summary

Pioneer FortrOss Bone Void Filler

### Indications for Use

510(k) Number: K091031

Device Name: Pioneer FortrOss Bone Void Filler

Indications for Use:

Pioneer FortrOss Bone Void Filler is indicated for bony voids or gaps that are not intrinsic to the stability of bony structure. These defects may be surgically created osseous defects or defects created from traumatic injury to the bone. The product is indicated to be used in the posterolateral spine in conjunction with bone marrow aspirate and autograft bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

Prescription Use x (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

H.H.Hilly FOR M.MELKERSON

Page 1 of 1

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K091031

---

**Source:** [https://fda-staging.innolitics.com/device/K091031](https://fda-staging.innolitics.com/device/K091031)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com