K091025 · bioMerieux, Inc. · JSO · Mar 2, 2010 · Microbiology
Device Facts
Record ID
K091025
Device Name
CHROMID VRE AGAR, MODEL: REF 43 851
Applicant
bioMerieux, Inc.
Product Code
JSO · Microbiology
Decision Date
Mar 2, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1700
Device Class
Class 2
Indications for Use
chromID™ VRE agar is a selective and differential chromogenic medium containing 8 µg/mL of vancomycin, for the qualitative detection of Enterococcus faecium and E. faecalis showing acquired vancomycin resistance (VRE) in stool specimens. chromID™ VRE agar can be used as an aid to identify, prevent and control VRE colonization in healthcare settings. chromID™ VRE agar is not intended to diagnose VRE infection nor to guide or monitor treatment for infections. Subculture to non-selective media (e.g. trypticase soy agar with 5% sheep blood) is needed for further identification, susceptibility testing and epidemiological typing.
Device Story
chromID™ VRE agar is a selective, differential chromogenic culture medium used in clinical laboratories for the qualitative detection of vancomycin-resistant Enterococcus faecium and E. faecalis from stool specimens. The medium contains a nutritive base, vancomycin (8 µg/mL), and two chromogenic substrates. It functions by inhibiting non-resistant enterococci and most other bacteria while allowing VRE to grow. The principle of operation relies on enzymatic activity: β-galactosidase-producing VRE E. faecium colonies appear violet, while α-glucosidase-producing VRE E. faecalis colonies appear blue-to-green after 24-48 hours of aerobic incubation. Healthcare providers visually inspect the plates for these characteristic colors to identify potential VRE colonization. This information assists in infection control and epidemiological surveillance. The device is intended for professional use in clinical settings; it does not diagnose active infection or guide treatment. Subculture to non-selective media is required for further identification and susceptibility testing.
Clinical Evidence
Performance evaluated at four laboratories using 1299 stool samples compared to conventional reference methods (Gram stain, catalase, VITEK 2, 16S-500 sequencing, agar dilution). At 24h, sensitivity was 97.1% and specificity 99.7%. At 48h, sensitivity was 96.9% and specificity 99.7%. Reproducibility was 100% after 48h. Interference, cross-reactivity, and recovery studies were also performed.
Technological Characteristics
Selective/differential chromogenic medium; nutritive base with peptones, 8 µg/mL vancomycin, and chromogenic substrates. Form factor: agar plate. Manual reading of colony color (violet for E. faecium, blue-to-green for E. faecalis). Aerobic incubation at 35-37°C in the dark.
Indications for Use
Indicated for qualitative detection of acquired vancomycin-resistant Enterococcus faecium and E. faecalis in stool specimens to aid in VRE colonization identification, prevention, and control in healthcare settings. Not for diagnosing VRE infection or guiding/monitoring treatment.
Regulatory Classification
Identification
A culture medium for antimicrobial susceptibility tests is a device intended for medical purposes that consists of any medium capable of supporting the growth of many of the bacterial pathogens that are subject to antimicrobial susceptibility tests. The medium should be free of components known to be antagonistic to the common agents for which susceptibility tests are performed in the treatment of disease.
Predicate Devices
Remel Bile Esculin Azide Agar w/ 6ug/ml Vancomycin (K972359)
Submission Summary (Full Text)
{0}
See doc
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K091025
B. Purpose for Submission:
To obtain Premarket Notification clearance of chromID™ VRE Agar for the qualitative detection of Enterococcus faecium and E.faecalis showing acquired vancomycin resistance (VRE) in stool specimens.
C. Measurand:
Vancomycin resistant E. faecium and E. faecalis
D. Type of Test:
Detection of vancomycin resistant Enterococcus faecium and Enterococcus faecalis using a selective and differential chromogenic medium.
E. Applicant:
bioMérieux, Inc.
F. Proprietary and Established Names:
chromID™ VRE Agar
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1700
2. Classification:
Class II
3. Product code:
JSO Culture media, Antimicrobial susceptibility test, excluding Mueller Hinton Agar
{1}
4. Panel:
Microbiology
H. Intended Use:
1. Intended use:
chromID™ VRE agar is a selective and differential chromogenic medium containing 8 µg/mL of vancomycin, for the qualitative detection of E. faecium and E. faecalis showing acquired vancomycin resistance (VRE) in stool specimens. chromID™ VRE agar can be used as an aid to identify, prevent and control VRE colonization in healthcare settings. chromID™ VRE is not intended to diagnose VRE infection nor to guide or monitor treatment for infections. Subculture to non-selective media (e.g. trypticase soy agar with 5% sheep blood) is needed for further identification, susceptibility testing and epidemiological typing.
2. Indications for use:
chromID™ VRE agar is a selective and differential chromogenic medium containing 8 µg/mL of vancomycin, for the qualitative detection of E. faecium and E. faecalis showing acquired vancomycin resistance (VRE) in stool specimens. chromID™ VRE agar can be used as an aid to identify, prevent and control VRE colonization in healthcare settings. chromID™ VRE is not intended to diagnose VRE infection nor to guide or monitor treatment for infections. Subculture to non-selective media (e.g. trypticase soy agar with 5% sheep blood) is needed for further identification, susceptibility testing and epidemiological typing.
3. Special conditions for use statement:
Prescription use
4. Special instrument requirements:
Not applicable
I. Device Description:
chromID™ VRE agar is translucent and a light tan color. After the plates are inoculated and incubated, VRE colonies will have either a violet color for E. faecium or a blue-to-green color for E. faecalis. Non-VRE colonies will be colorless or have colors other than violet or blue-to-green.
See doc
{2}
J. Substantial Equivalence Information:
1. Predicate device name:
Remel Bile Esculin Azide Agar w/ 6μg/mL Vancomycin
2. Predicate K number:
K972359
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | chromID™ VRE agar is a selective and differential chromogenic medium containing 8μg/mL of vancomycin, for the qualitative detection of E. faecium and E. faecalis showing acquired vancomycin resistance (VRE) in stool specimens. | Remel Bile Esculin Azide Agar w/ 6μg/mL vancomycin is a solid medium recommended for use in qualitative procedures as a screening method for primary isolation and presumptive identification of vancomycin resistant enterococci (VRE) from surveillance cultures. |
| Reading | Manual | Manual |
| Inoculum | Direct Specimen | Direct Specimen |
| Specimen | Stool samples | Urine, stool |
| Differences | | |
| --- | --- | --- |
| Item | | Predicate |
| Detection Method | Two chromogenic substrates provide for the direct detection of E. faecium and E. faecalis through characteristic colony color
• E. faecium: violet color for β-galactosidase-producing strains
• E. faecalis: blue-to-green color for α-glucosidase-producing strains | Organisms positive for esculin hydrolysis hydrolyze the glycoside esculin to esculetin and dextrose. The esculetin reacts with ferric ammonium citrate to form a dark brown or black complex in the medium. |
| Incubation | 35 - 37°C in aerobic | 33 - 37°C in aerobic conditions or |
See doc
{3}
| Differences | | |
| --- | --- | --- |
| Item | | Predicate |
| | conditions, in the dark | in 5 – 10% CO₂ |
K. Standard/Guidance Document referenced (if applicable):
Not applicable
L. Test Principle:
chromID™ VRE agar (patents pending) consists of a rich nutritive base including a variety of peptones. It also contains chromogenic substrates and a mixture of antibiotics including vancomycin (8 µg/mL) which enable
- the specific and selective growth of VRE,
- the direct detection of *E. faecium* and *E. faecalis* through the characteristic color of colonies:
- *E. faecium*: violet color for β-galactosidase producing strains
- *E. faecalis*: blue-to-green color for α-glucosidase producing strains
The selective mixture inhibits enterococcal strains that do not express acquired vancomycin resistance; enterococcal strains that express intrinsic vancomycin resistance (i.e. vanC phenotype: *E. gallinarum* and *E. casseliflavus*); and most Gram-negative and Gram-positive bacteria.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Reproducibility was demonstrated at three sites using a set of ten organisms which were tested in triplicate each day for three days at three different sites. The study included vancomycin susceptible and vancomycin resistant *E. faecalis* and *E. faecium* (both with vanA and vanB). The quality control reference strains were also included in the study.
Preliminary results at 24 hours demonstrated 80% reproducibility. After incubation for the full 48 hours, overall reproducibility was 100%.
See doc
{4}
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The recommended quality control (QC) organisms, *E. faecium* (vanA) ATCC 700221 and *E. faecalis* (vanB) ATCC 51299 as positive controls, and *E. faecalis* ATCC 29212 as negative control were used. Quality control data was compiled across all three sites and overall QC results were acceptable.
QC Data Summary at 48 hours
| Organism | Expected Result on chromID™ VRE | chromID™ VRE Positive | chromID™ VRE Negative |
| --- | --- | --- | --- |
| E.faecalis ATCC 29212 | No Growth (Negative) | 1 | 96 |
| E. faecalis ATCC 51299 | Growth of Blue-to-Green colored Colonies (Positive) | 97 | 0 |
| E.faecium ATCC 700221 | Growth of Violet colored Colonies (Positive) | 97 | 0 |
In one instance, the negative control, *E.faecalis* ATCC 29212 exhibited breakthrough growth on the chromID™ VRE agar medium. A note was added to the table in the Quality Control section of the package insert recommending the use of a light organism inoculum in order to mitigate erroneous quality control results.
d. Detection limit:
Not applicable
e. Analytical specificity:
Cross-reactivity Study
A Cross-reactivity study was performed using at total of 99 strains representing various organism groups.
One strain of *Enterococcus raffinosus* produced colorless colonies after 24 hours of incubation which subsequently developed violet pigmentation after 48 hours of incubation.
One strain of *Citrobacter freundii* produced violet *Klebsiella pneumoniae* pigmented colonies after 24 hours incubation. Two strains of produced violet colored colonies after 24 and 48 hours of incubation.
See doc
{5}
Two strains of *P. aeruginosa*, five strains of *P. fluorescens*, and one strain each of *S. maltophilia* and *A. baumanii* grew producing colorless and/or atypically pigmented colonies.
One strain each of *C. albicans*, *C. tropicalis*, *C. glabrata*, *Geotrichum candidum*, *Saccromyces cerevisiae*, *Aspergillus niger*, and *Penicillium abeanum*, all grew producing colorless or blue colonies on the chromID™ VRE medium.
The study results are reflected in the Limitation and Cross Reactivity sections of the device package insert.
## Interference Study
Commonly used medicinal substances namely, Preparation H Ointment, Boudreaux’s Butt Paste, Glycerin, Dulcolax, Vaseline and Tucks Pads, as well as human blood and physiological saline were found to decrease growth of VRE isolates on the chromID™ VRE medium in comparison to the control. However, the chromogenic reaction remained unaffected.
Testing of both Preparation H Cream and Miconazole 7 resulted inhibition or delayed growth of VRE isolates on the chromID™ VRE medium at 24 and 48 hours incubation. However, the chromogenic reaction remained unaffected.
Cary Blair transport medium was tested and found to have no detrimental effect on the growth, recovery or chromogenic reaction of VRE isolates on the chromID™ VRE medium.
## Collection Swab Study
The influence of collection swabs on the recovery of VRE was also evaluated. Rayon and nylon flocked swabs were tested dry and coupled with Amies transport media. After 1 and 4 hours of contact there was found to be no significant difference in the sensitivity of detection of VRE between dry swabs and those in Amies transport medium. After 18 hours of contact at room temperature, the nylon flocked swab with Amies transport media resulted in the best recovery for VRE, followed by nylon flocked dry swabs and rayon swab with Amies transport media. After 18 hours at 2-8°C, the nylon flocked swabs with and without Amies transport medium resulted in the best recovery of VRE, followed by the rayon swabs with and without Amies transport medium.
See doc
{6}
f. Analytical Sensitivity
Recovery Study
A recovery study was performed to determine the lowest number of colony forming units (CFU) of vancomycin resistant enterococci (VRE) that will grow on the chromID™ VRE medium. One strain of *E. faecalis* (vanA) bioMerieux stock no. 05.02.228; and one strain of *E. faecium* (vanA) ATCC 700221; were evaluated. Ten-fold serial dilutions were prepared in saline and then plated in duplicate onto chromID™ VRE medium. After 48 hours incubation, the number of CFU was counted on each plate. The number of CFU obtained on the two plates from each dilution was averaged. The percent recovery on chromID™ VRE was determined to be equivalent to 100 CFU/mL of sample.
g. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The performance of the chromID™ VRE agar was evaluated at four geographically diverse laboratory sites. A total of 1299 stool specimens were evaluated. Stool specimens were inoculated on chromID™ VRE agar medium and the Bile Esculin Azide Agar with 6μg/ml of Vancomycin (BEAV). Both plates were observed for growth at 24 and 48 hours. Colonies with violet or blue-to-green pigment (chromID™ VRE) or brown to black pigment diffusing into the medium (BEAV), were identified by a combination of conventional reference methods to include Gram stain, catalase, VITEK 2 GP, and 16S-500 sequencing. Vancomycin resistance was confirmed by agar dilution.
See doc
{7}
Percent agreement of the chromID™ VRE compared to conventional reference methods is presented in Table 1.
| Table 1. chromID™ VRE vs Conventional Methods | Positive % Agreement | Negative % Agreement |
| --- | --- | --- |
| chromID™ VRE
agar @24h
incubation | 97.1% (334/344)
(95% CI = 94.7%, 98.6%) | 99.7% (952/955)
(95% CI = 99.1%, 99.9%) |
| chromID™ VRE
agar @48h
incubation | 96.9% (344/355)
(95% CI = 94.5%, 98.4%) | 99.7% (941/944)
(95% CI = 99.1%, 99.3%) |
Performance data for the chromID™ VRE agar compared to the VITEK® 2 GP Identification, for both isolates of E.faecalis and E.faecium is presented in Tables 2 and 3.
In Table 2, of the nine isolates that were not identified as E.faecium by VITEK®2, one isolate was negative on the chromID™ VRE agar at 24 hours but was false positive after 48 hours. The other eight isolates were false positive on chromID™ VRE at 24 and 48 hours.
| Table 2. chromID™ VRE vs VITEK® 2 Identification
E. faecium | Positive % Agreement | Negative % Agreement |
| --- | --- | --- |
| chromID™ VRE
agar @ 24h
incubation | 94.3% (316/335)
(95% CI = 91.3%, 96.6%) | 11.1% (1/9)*
(95% CI = 0.3%, 42.3%) |
| chromID™ VRE
agar @ 48h
incubation | 96.7% (324/335)
(95% CI = 94.2%, 98.4%) | 0.0% (0/9)*
(95% CI = 0.0%, 33.6%) |
See doc
{8}
In Table 3, all $E$ .faecalis isolates were identified by VITEK® 2 however four isolates did not produce the characteristic blue-to-green color at 24 hours. At 48 hours incubation, three of the four isolates developed the characteristic blue-to-green color.
| Table 3. chromID™ VRE vs VITEK® 2 Identification E. faecalis | Positive % Agreement | Negative % Agreement |
| --- | --- | --- |
| chromID™ VRE agar @ 24h incubation | 87.1% (27/31) (95% CI = 70.2%, 96.4%) | 0/0* |
| chromID™ VRE agar @ 48h incubation | 96.8% (30/31) (95% CI = 83.3%, 99.9%) | 0/0* |
Performance data for the chromID™ VRE agar compared to the vancomycin MIC, for both isolates of $E$ faecalis and $E$ faecium is presented in Tables 4 and 5.
In Table 4, three of four isolates were false positive as they produced violet colonies on chromID™ VRE but were confirmed to be vancomycin susceptible $E$ . faecium by MIC. The fourth isolate did not produce violet pigmented colonies.
| Table 4. chromID™ VRE vs Vancomycin MIC E. faecium | Positive % Agreement | Negative % Agreement |
| --- | --- | --- |
| chromID™ VRE agar @ 24h incubation | 94.4% (321/340) (95% CI = 91.4%, 96.6%) | 25% (1/4)* (95% CI = 0.6%, 80.6%) |
| chromID™ VRE agar @ 48h incubation | 97.1% (330/340) (95% CI = 94.7%, 98.6%) | 25% (1/4)* (95% CI = 0.6%, 80.6%) |
See doc
{9}
In Table 5, one sample produced blue-to-green colonies on chromID™ VRE and was vancomycin susceptible by MIC.
| Table 5. chromID™ VRE vs Vancomycin E. faecalis | Positive % Agreement | Negative % Agreement |
| --- | --- | --- |
| chromID™ VRE
agar @ 24h
incubation | 86.7% (26/30)
(95% CI = 69.3%, 96.3%) | 0.0% (0/1)*
(95% CI = 0.0%, 97.5%) |
| chromID™ VRE
agar @ 48h
incubation | 96.7% (29/30)
(95% CI = 82.8%, 99.9%) | 0.0% (0/1)*
(95% CI = 0.0%, 97.5%) |
A challenge set of 75 well characterized isolates were tested at one external site. The challenge isolates included vancomycin resistant and vancomycin susceptible *E. faecalis* and *E. faecium*, *E. gallinarum* and *E. casseliflavus*, and other gram positive microorganisms commonly isolated from the stools. Results of the challenge study were compared to the expected result as determined by the organism identification based on API® results or known reference strain and vancomycin resistance as determined by the vancomycin agar dilution method.
One challenge isolate *E. faecalis* 13029 vanB produced blue-purple colonies after 48 hours incubation. This was due to a rare situation of activity of both enzymes (α-glucosidase and β-galactosidase) occurring in one strain. Both enzymes reacted with the chromogen in the chromID™ VRE agar medium and this combination of activity produced the blue-purple color variation.
A second study was performed to evaluate the detection of enterococcal strains with low level vancomycin resistance. Forty-nine enterococcal strains characterized by intermediate or low levels of resistance to vancomycin were evaluated. The strains tested were from several international collections such as ATCC, LMG (Gent-Laboratorium voor Microbiologie, Belgium), NCTC (National Collection of Type Culture, United Kingdom) and the bioMérieux stock collection. *E. faecium* and *E. faecalis* isolates with a low level of vancomycin resistance were detected if their vancomycin MIC was greater than or equal to 4 μg/mL. For a majority of the isolates, the characteristic colony coloration was observed after 48 hours incubation. All the isolates of *E. casseliflavus* in this study were inhibited on the chromID™ VRE medium. Most isolates of *E. gallinarum* in this study were inhibited on the chromID™ VRE medium with the exception of two isolates, the first with the acquired vanA gene resulted in growth of violet colonies, a second with an MIC of greater than 64 μg/mL resulted in growth of colorless colonies.
See doc
{10}
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Vancomycin resistant *E. faecalis* presents as blue-to green color colonies.
Vancomycin resistant *E. faecium* presents as violet color colonies.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
See doc
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.