SILVER PU ANTIBACTERIAL FOAM DRESSING - NON ADHERENT AND ISLAND
K091013 · Advanced Medical Solutions Limited · FRO · Jul 30, 2009 · SU
Device Facts
| Record ID | K091013 |
| Device Name | SILVER PU ANTIBACTERIAL FOAM DRESSING - NON ADHERENT AND ISLAND |
| Applicant | Advanced Medical Solutions Limited |
| Product Code | FRO · SU |
| Decision Date | Jul 30, 2009 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
Silver PU Antibacterial Foam Dressings are indicated for the management of light to moderate exuding partial and full thickness wounds, such as: Decubitus (pressure) ulcers, Diabetic ulcers, Leg ulcers, Graft and donor sites, Lacerations and abrasions, 1st and 2nd degree burns, Trauma wounds (dermal lesions, trauma injuries or incisions), Post-operative surgical wounds. The Silver PU Antibacterial Foam Dressings are indicated for external use only.
Device Story
Silver PU Antibacterial Foam Dressings are sterile, absorbent wound dressings; available in non-adherent and island (adhesive) variants. Construction includes hydrophilic polyurethane foam laminated to silver-coated nylon mesh; island variant adds polyurethane waterproof membrane with acrylic adhesive. Device provides bacterial barrier and maintains moist wound environment; elemental silver (min 0.26mg/cm²) protects dressing from bacterial contamination and reduces wound odor. Used in clinical or home settings for management of exuding wounds; applied by healthcare professionals or patients. Antibacterial effect lasts up to 7 days. Benefits include infection protection, odor reduction, and support for moist wound healing; suitable for use under compression.
Clinical Evidence
No clinical trials performed. Evidence consists of in vitro antimicrobial testing against S. aureus (including MRSA), S. epidermidis (including MRSE), S. pyogenes, E. faecalis (VRE), and E. coli. Biocompatibility testing conducted per BS EN ISO 10993-1. Sterilization validation performed per harmonized standards. Animal and bench testing support performance claims.
Technological Characteristics
Materials: Polyurethane film/foam, silver-coated nylon mesh, acrylic adhesive. Energy source: None. Form factor: Various sizes/shapes, non-adherent or island variants. Connectivity: None. Sterilization: Validated per harmonized standards. Biocompatibility: BS EN ISO 10993-1 compliant. Manufacturing: BS EN ISO 14971 risk analysis implemented.
Indications for Use
Indicated for management of light to moderate exuding partial and full thickness wounds in patients with decubitus ulcers, diabetic ulcers, leg ulcers, graft/donor sites, lacerations, abrasions, 1st/2nd degree burns, trauma wounds, and post-operative surgical wounds. For external use only.
Predicate Devices
Submission Summary (Full Text)
{0}------------------------------------------------
K091013
page 1 of 2
## JUL 30 2009
## Section 5 - Traditional 510(k) Notification :- Summary
This Traditional 510(k) notification is to provide substantial equivalence for Advanced Medical Solutions Limited's Silver PU Antibacterial Foam Dressings, which are substantially equivalent to currently marketed devices intended for wound care.
| Submitted by:- | Advanced Medical Solutions Limited<br>Road Three<br>Winsford Industrial Estate<br>Winsford, Cheshire<br>CW7 3PD<br>United Kingdom |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:- | Mrs Claire Ryan<br>Regulatory Affairs Manager<br>Telephone: + 44(0)1606 545569<br>Fax: + 44(0)1606 863600<br>Email: claire.ryan@admedsol.com |
| Date prepared:- | 3rd April 2009 |
| Common Name:- | Silver PU Antibacterial Foam Dressings |
| Trade Names:- | Not yet defined |
| Classification Name:- | Dressing, Wound, Drug |
| Classification:- | Unclassified |
| Product Code:- | FRO |
#### Legally marketed device(s) for substantial equivalence comparison :-
Allevyn Ag Dressings, 510(k) # K063835, manufactured by Smith & Nephew. Mepilex Ag Dressing, 510(k) # K071354 / K061554, manufactured by Mölnlycke Health Care.
#### Device Description :-
The Silver PU Antibacterial Foam Dressings is available in two variants:
Silver PU Antimicrobial Foam Dressing - Non-Adherent variant consists of a low friction waterproof polyurethane film, which also provides a bacterial barrier, laminated to hydrophilic absorbent polyurethane foam with a silver coated nylon mesh wound contact layer.
Silver PU Antimicrobial Foam Dressing - Island variant consists of a polyurethane waterproof membrane coated with pressure sensitive acrylic adhesive, which also provides a bacterial barrier, with a centrally located hydrophilic absorbent polyurethane foam with a silver coated nylon mesh wound contact layer.
Each Silver PU Antibacterial Foam Dressing contains a minimum of 0.26mg/cm² of elemental silver, the antibacterial action of the silver protects the dressing from bacterial contamination and provides an effective barrier to bacterial penetration.
{1}------------------------------------------------
091013
Page 272
The antibacterial properties of the dressing are effective for up to 7 days, as demonstrated in vitro, against a broad spectrum of microorganisms associated with wound infection, such as Staphylococcus aureus, including MRSA, Staphylococcus epidermidis, including MRSE, Strentococcus pvogenes, Enterococcus faecalis (VRE) and Escherichia coli. Silver Antibacterial Foam Dressings may reduce odour caused by micro-organisms in the wound. Odour reduction results from the antibacterial effect in the dressing.
Silver PU Antibacterial Foam Dressings are soft, highly absorbent, conformable and create the ideal environment for moist wound healing.
Silver PU Antibacterial Foam Dressings are suitable for use under compression bandaging.
Silver PU Antibacterial Foam Dressings are available in various sizes and shapes. The dressings are packaged in individual pouches, then into shelf cartons.
#### Indications for use:
Silver PU Antibacterial Foam Dressings are indicated for the management of light to moderate exuding partial and full thickness wounds, such as decubitus (pressure) ulcers, diabetic ulcers, leg ulcers, graft and donor sites, lacerations and abrasions, 18 and 210 degree burns, trauma (dermal lesions, trauma injuries or incisions) and post-operative surgical wounds. Silver PU Antibacterial Foam Dressings are indicated for external use only.
#### Manufacturing :-
Silver PU Antibacterial Foam Dressings will be manufactured according to the product specification and under good manufacturing practices (GMP). A risk analysis has been performed in accordance with BS EN ISO 14971 to identify possible failure mode during manufacturing. Manufacturing controls have been developed and implemented to address the identified risk factors based on the criticality of the failure mode.
Advanced Medical Solutions Ltd, Silver PU Antibacterial Foam Dressings meets all the established specifications prior to release to ensure the device is safe, effective and correctly labelled for its intended use.
#### Testing :-
Performance data for the Silver PU Antibacterial Foam Dressings have been established using antimicrobial, animal and bench testing. The biocompatibility of Advanced Medical Solutions Limited Silver PU Antibacterial Foam Dressings have been demonstrated to be in compliance with the requirements of BS EN ISO 10993-1(Biological Evaluation of Medical Devices). Sterilisation validation has been performed in compliance with harmonised standards.
#### Statement of Substantial Equivalence:-
The indication for use, performance testing and antimicrobial activity for the Silver PU Antibacterial Foam Dressings are substantially equivalent to the predicate devices; Allevyn Ag Dressings, 510(k) # K063835, manufactured by Smith & Nephew and Mepilex Ag Dressing, 510(k) # K071354 / K061554. manufactured by Moinlycke Health Care. The biocompatibility and performance testing for the Silver PU Antibacterial Foam Dressings have demonstrated that the device is safe and effective for the indications of use.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three legs and a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the border of the circle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Advanced Medical Solutions Limited % Ms. Claire Ryan Road Three Winsford Industrial Estate Winsford, Cheshire CW7 3PD United Kingdom
JUL 3 0 2009
Re: K091013
Trade/Device Name: Silver PU Antibacterial Foam Dressings Regulatory Class: Unclassified Product Code: FRO Dated: July 17, 2009 Received: July 22, 2009
## Dear Ms. Ryan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Ms. Claire Ryan
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
. Sincerely yours,
incerely yours,
ark N. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 000020
## Indications for Use
510(k) Number (if known): KO91013
Device Name: Silver PU Antibacterial Foam Dressings
Indications for Use:
Silver PU Antibacterial Foam Dressings are indicated for the management of light to moderate exuding partial and full thickness wounds, such as: '.
- Decubitus (pressure) ulcers �
- Diabetic ulcers 0
- Leg ulcers .
- Graft and donor sites �
- Lacerations and abrasions 9
- 1st and 2nd degree burns �
- Trauma wounds (dermal lesions, trauma injuries or incisions) e
- Post-operative surgical wounds
The Silver PU Antibacterial Foam Dressings are indicated for external use only
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
David Kione for MKM
Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K091013