← Product Code [HWC](/productcode/HWC) · K090994

# LIGAFIX INTERFACE SCREW (K090994)

_Science For Biomaterials · HWC · May 6, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K090994

## Device Facts

- **Applicant:** Science For Biomaterials
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** May 6, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

LIGAFIX® / ComposiTCP® is a cannulated, sterile, single-use, resorbable interference bone screw made of a mixture of tri calcium phosphate (beta-TCP) and Poly Lactic Acid (PLA) designed for the interference fixation of grafts in anterior cruciate ligament reconstruction.

## Device Story

Resorbable cannulated interference bone screw; used for graft fixation in anterior cruciate ligament (ACL) reconstruction. Composed of beta-TCP/PLA composite; available in 60/40 and 30/70 ratios. Supplied sterile; single-use. Implanted by surgeons during orthopedic procedures. Provides mechanical fixation of graft within bone tunnel; material properties support sustained fixation during healing process. Resorbable nature eliminates need for secondary removal surgery.

## Clinical Evidence

No clinical data. Bench testing only: biological, mechanical, and biocompatibility tests confirmed biocompatibility and strength retention profiles compatible with the healing process.

## Technological Characteristics

Cannulated bone fixation screw; material: beta-TCP/PLA composite; ratios: 60/40 and 30/70; sterile; single-use; resorbable.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- LIGAFIX® INTERFERENCE SCREW ([K061262](/device/K061262.md))
- LIGAFIX® INTERFERENCE SCREW ([K070507](/device/K070507.md))
- LIGAFIX® / ComposiTCP® Interference screw ([K050407](/device/K050407.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "SPECIAL 510k" on the first line and "LIGAFIX® Interference screws" on the second line. The text is in a bold, sans-serif font. The word "SPECIAL" is in all caps, and the number "510k" is next to it. The word "LIGAFIX" is followed by the registered trademark symbol. The words "Interference screws" are below the word "LIGAFIX".

Image /page/0/Picture/1 description: The image shows the words "Science for Bio Materials" in a stylized font. The word "Bio" is subscripted. Underneath the words is a smaller line of text that is difficult to read. To the right of the words are three triangles pointing to the right, with each triangle having a different shade of gray. 

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

### 1. GENERAL INFORMATION

| Trade Name                            | LIGAFIX® Resorbable Interference Screw<br>ComposiTCP® Resorbable Interference Screw                                                                                                                                  |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name                           | Bone Fixation Screw                                                                                                                                                                                                  |
| Classification Name                   | Screw, Fixation, Bone                                                                                                                                                                                                |
| Class                                 | II                                                                                                                                                                                                                   |
| Product Code                          | HWC                                                                                                                                                                                                                  |
| CFR section                           | 21CFR 888.3040                                                                                                                                                                                                       |
| Device panel                          | Orthopedic                                                                                                                                                                                                           |
| Legally marketed<br>predicate devices | LIGAFIX® INTERFERENCE SCREW (K061262 and<br>K070507) manufactured by SCIENCE FOR<br>BIOMATERIALS                                                                                                                     |
| Submitter                             | SCIENCE FOR BIOMATERIALS<br>Sciences et Bio Matériaux<br>ZI du Monge<br>F 65100 LOURDES - FRANCE<br>Registration Number : 3004549189                                                                                 |
| Contact                               | Denis CLEMENT, CEO<br>Tel : +33 (0)5 62 42 21 01<br>Fax : +33 (0)5 62 42 21 00<br>e-mail : denis.clement@sbm-fr.com<br>Regulatory contact: Idée Consulting (FRANCE)<br>Isabelle DRUBAIX e-mail : idrubaix@nordnet.fr |

2009

### 2. DEVICE DESCRIPTION

LIGAFIX® / ComposiTCP® range of products consists of resorbable cannulated screws available in several models designed for the interference fixation of grafts in anterior cruciate ligament reconstruction.

LIGAFIX® / ComposiTCP® interference bone screw is made of a ceramic (beta-TCP) / polymer (Poly Lactic Acid -PLA) composite.

LIGAFIX® / ComposiTCP® interference screw is available in several sizes and in two beta-TCP/ PLA ratios 60/40 (LIGAFIX® 60/ ComposiTCP® 60) and 30/70 (LIGAFIX® 30/ ComposiTCP® 30)

LIGAFIX® / ComposiTCP® interference bone screws are supplied sterile and individually packaged in double heat sealed pouches.

Page 2 / 5

{1}------------------------------------------------

SPECIAL 510k

LIGAFIX® Interference screws

SCIENCE FOR BIO MATERIALS

K090994
(page 2 of 2)

#### 3. INTENDED USE

LIGAFIX® / ComposiTCP® is a cannulated, sterile, single-use, resorbable interference bone screw made of a mixture of tri calcium phosphate (beta-TCP) and Poly Lactic Acid (PLA) designed for the interference fixation of grafts in anterior cruciate ligament reconstruction.

### 4. PERFORMANCE DATA

Biological, mechanical and biocompatibility tests confirmed that LIGAFIX® / ComposiTCP® screws are highly biocompatible and presents the requisite strength to provide sustained fixation of the graft. LIGAFIX®/ ComposiTCP® screws strength retention profiles are compatible with the healing process.

#### 5, SUBSTANTIAL EQUIVALENCE

The modifications to LIGAFIX® / ComposiTCP® Interference screw (K050407, K061262 and K070507) consist of additional size of screw together with the deletion of a reference to surgical technique in the instructions for use.

The additional LIGAFIX® / ComposiTCP® Interference screws are substantially equivalent to their predicate devices LIGAFIX® / ComposiTCP® Interference screw (K050407, K061262 and K070507) in terms of intended use, material, design, mechanical properties and function.

Summary preparation date:

May 5, 2009

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

MAY - 6 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SCIENCE FOR BIOMATERIALS % Mr. Denis Clement CEO ZI du Monge F 65100 LOURDES - FRANCE

Re: K090994

Trade/Device Name: LIGAFIX® RESORBABLE INTEREFERNCE SCREW ComposiTCP® RESORBABLE INTERFERENCE SCREW Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: March 23, 2009 Received: April 7, 2009

Dear Mr. Clement :

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2- Mr. Denis Clement

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Daniel Keane

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

LIGAFIX® Interference screws

SCIENCE FOR BioMATERIALS

K090994 510(k) Number (if known):

LIGAFIX® RESORBABLE INTERFERENCE SCREW Device Name: ComposiTCP® RESORBABLE INTERFERENCE SCREW

#### Indications for Use:

"LIGAFIX® interference screws are designed for the interference fixation of grafts in anterior cruciate ligament reconstruction."

) Prescription Use AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## (Please do not write below this line-continue on another Page of NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sousta

\$
\int\$ (D
D
or

(Division Sign-On)
Division of Surgical, Orthopedic,
and Restorative Devices

510(k) Number K090994

Page 5 / 5

---

**Source:** [https://fda-staging.innolitics.com/device/K090994](https://fda-staging.innolitics.com/device/K090994)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).