A comprehensive review of clinical literature was provided to demonstrate that the use of branched polyester vascular grafts for debranching and hybrid procedures is standard clinical practice.
Literature review; Clinical practice; Hybrid procedures; Debranching
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Comprehensive review of the literature; Literature review
Patients undergoing aortic repair via debranching and hybrid procedures
Not applicable for this study
Clinical performance and safety of branched polyester vascular grafts in debranching and hybrid procedures
Indications for Use
Gelweave™ branched vascular grafts: "Repair or replacement of damaged and diseased vessels of the abdomen and thoracic aorta in cases of aneurysm, dissection or coarctation". Gelweave Siena™ vascular grafts with and without radiopaque markers: "Repair or replacement of damaged and diseased vessels of the thoracic aorta in cases of aneurysm, dissection or coarctation". "Gelweave™ branched vascular grafts, including Siena™ grafts can also be used for debranching i.e. reconstruction of the aortic vessels and associated Hybrid procedures".
Device Story
Polyester vascular grafts with gelatin sealant; used in open heart surgery for repair/replacement of diseased/damaged aorta and associated vessels. Branched design accommodates side vessel reconstruction and cardiopulmonary bypass attachment. Used in hospital operating rooms by qualified surgeons. Expanded indication allows use in 'debranching' and 'Hybrid procedures' (open surgical debranching combined with endovascular aortic repair). Device provides structural replacement for damaged vasculature; hybrid use allows less invasive treatment options for high-risk patients.
Clinical Evidence
No clinical testing performed. Evidence consists of a comprehensive literature review supporting the use of standard branched polyester vascular grafts for debranching and hybrid procedures. Bench testing demonstrated that insertion and manipulation of a 26F endovascular delivery catheter through a 10mm side branch did not cause structural damage or blood leakage under 120 mmHg pressure.
Technological Characteristics
Vascular graft prosthesis; materials: polyester and gelatin sealant. Sterilization: ethylene oxide. Form factor: branched and unbranched tubular grafts, some with radiopaque markers. Intended for single use. No changes to design, materials, or manufacturing compared to predicates.
Indications for Use
Indicated for patients requiring repair or replacement of damaged/diseased vessels of the abdomen and thoracic aorta due to aneurysm, dissection, or coarctation, including use in debranching and hybrid procedures.
Regulatory Classification
Identification
A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”
Gelweave Siena™ grafts without radiopaque markers (K040829)
Gelweave Siena™ grafts with radiopaque markers (K060142)
Submission Summary (Full Text)
{0}------------------------------------------------
K090987
VASQUTEK
**PERUNO**
JUN 19 2009
JUN 19 2009
VASCUTEK Ltd. a TERUMO Company Newmains Avenue Inchinnan Renfrewshire, PA4 9RR Scotland
Tel: {+44) (0)141 812 5555 Fax: (+44) (0)141 812 7650 www.vascutek.com
## 510(k) SUMMARY
# VASCUTEK GELWEAVE™ BRANCHED VASCULAR GRAFTS (K952293) AND GELWEAVE SIENA " VASCULAR GRAFTS WITH (K060142) AND WITHOUT (K040829) RADIOPAQUE MARKERS
#### Date prepared: 2nd of June 2009
Prepared in accordance with CFR21 Part 807, Sec. 807.92: Content and format of a 510(k) summary.
Page 1 of 5
Prosthesis, vascular graft Common/Usual Name :
Proprietary Name(s) : Vascutek Gelweave 10 branched vascular grafts and Gelweave Siena M vascular grafts with and without radiopaque markers
Prosthesis, vascular graft, of 6mm and greater diameter Classification Name :
Classification :
The Food and Drug Administration has classified these devices as Class II devices under classification 21 CFR 870.3450. Classification, DSY.
Applicant and : 510(k) submitter/ contact
Karen Kelso Regulatory Affairs Manager Vascutek Ltd. Newmains Avenue Inchinnan Industrial Estate Renfrewshire, Scotland PA4 9RR, United Kingdom
Phone: +44 (141) 814 5527 Fax: +44 (141) 814 3431
{1}------------------------------------------------
# Device Predicates:
| Predicate Device Name | FDA Clearance Date | 510(k) |
|-------------------------------------------------------|--------------------|---------|
| Gelweave™ grafts | 19th December 1995 | K952293 |
| Gelweave Siena™ grafts without<br>radioopaque markers | 20th May 2004 | K040829 |
| Gelweave Siena™ grafts with radiopaque<br>markers | 6th February 2006 | K060142 |
* NOTE: Gelweave Branched Grafts - Documentation submitted to Dorothy Abel, FDA, dated 21° November 2001, regarding Vascutek's decision not to submit a 510k for branched Gelweave grafts. This decision ultilised FDA Guidance document (K97-1) "Deciding when to submit a 510k for an existing device".
#### Device Description:
The design and all other properties of the Vascutek Gelweave™ branched vascular grafts and Gelweave Siena™ vascular grafts with and without radiopaque markers remain unchanged. The Vascutek Gelweave™ branched vascular grafts and Gelweave Siena™ vascular grafts with and without radiopaque markers function as vascular grafts manufactured from standard polyester and gelatin sealant technologies.
Additional images and information are provided in the catalogue number table presented in Table 3 of this 510(k) submission.
Vascutek Gelweave™ vascular grafts were cleared by FDA on the 19th of December 1995 (K952293). The cleared indications for use were "Repair or replacement of damaged and diseased vessels of the abdomen and thoracic aorta in cases of aneurysm, dissection or coarctation".
The branched versions of the Gelweave™ vascular grafts were later made available to also accommodate reconstruction of the associated side vessels and allow intra-operative attachment to cardiopulmonary bypass. This is information clearly detailed in the product Instructions for Use.
In 2004 Vascutek received 510(k) clearance for the Siena™ grafts without radiopaque markers, followed by Siena™ grafts with radiopaque markers in 2006. The intention of the radiopaque markers was to aid in vivo visualisation. The branches present on the Gelweave grafts, like all other Gelweave™ branched grafts was to accommodate Siena™M reconstruction of the associated branch vessels and also intra-operative aftachment to cardiopulmonary bypass. The cleared indications for use were; "Repair or replacement of damaged and diseased vessels of the thoracic aorta in cases of aneurysm, dissection or coarchation". This is clearly detailed in the 510(k) cleared product Instructions for Use.
As stated, the presence of side branches allows accommodation of aortic side vessel reconstruction as well as the ability to attach the patient to cardiopulmonary bypass. It is evident from the available literature and clinical opinion that it is also considered standard practice to use branched polyester vascular grafts to undertake debranching and associated Hybrid procedures. These procedures allow reconstruction of the aortic side vessels at the same time as repair of the aortic aneurysm by using standard polyester vascular grafts in combination with endovascular devices, usually in a single procedure.
Vascutek are therefore requesting FDA clearance to make reference to debranching and vasociated Hybrid procedures for the Vascutek Gelweave™ branched vascular grafts and Gelweave Siena " yascular grafts with and without radiopague markers in the product labelling, specifically the indications for use, product Instructions for Use and associated marketing literature.
{2}------------------------------------------------
There are no other changes to the products i.e. there are no changes to the design, materials, packaging, sterilisation, shelf life or any other parameters. The only change is a request for an expanded indication for use. The grafts remain as finished products, provided sterile for single use only.
#### Intended Use:
The intended use for the Vascutek Gelweave™ branched vascular grafts and Gelweave Siena™ vascular grafts with and without radiopaque markers is currently;
## Gelweave " branched vascular grafts:
"Repair or replacement of damaged and diseased vessels of the abdomen and thoracic aorta in cases of aneurysm, dissection or coarctation".
# Gelweave Siena™ vascular grafts with and without radiopaque markers:
"Repair or replacement of damaged and diseased vessels of the thoracic aorta in cases of aneurysm, dissection or coarctation".
The subject of this 510(k) is to request FDA clearance to expand this intended use as follows:
"Gelweave" branched vascular grafts, including Siena™ grafts can also be used for debranching i.e. reconstruction of the aortic vessels and associated Hybrid procedures".
Hybrid procedures are defined as a treatment combination employing open surgical debranching with endovascular aortic repair.
As stated, it is evident from the literature and clinical opinion that it is considered standard practice to use branched polyester vascular grafts to undertake debranching and associated Hybrid procedures. These procedures do not change the overall intended use of the grafts, which are still used in open heart surgery to repair or replace a diseased and damaged aorta and associated vessels.
The difference is therefore not critical to the intended therapeutic, diagnostic, prosthetic or surgical use of the devices. The difference also does not affect the safety and effectiveness of the device when used as labelled.
# Principles of Operation and Technology:
Vascutek Gelweave™ branched vascular grafts and Gelweave Siena™ vascular grafts with and without radiopaque markers are used to repair or replace a diseased and damaged in aorta and associated vessels.
Vascutek Gelweave™ branched vascular grafts and Gelweave Siena™ vascular grafts with and without radiopaque markers are intended only for use in the hospital operating room by suitably qualified surgeons.
Vascutek Gelweave™ pranched vascular grafts and Gelweave Siena™ vascular grafts with and without radiopaque markers provided sterile by ethylene oxide for single use only.
There devices have the same technological characteristics as the predicates i.e. there are no
changes to the Vascutek Gelweave" branched vascular grafts and Gelweave Siena™ vascular grafts with and without radiopaque markers other than the request for an expanded indication.
{3}------------------------------------------------
Performance: The performance of the Vascutek Gelweave™ branched vascular grafts and Gelweave Siena™ vascular grafts with and without radiopaque markers remains unchanged. The grafts are still used in open heart surgery to repair or replace a diseased and damaged aorta and associated vessels.
Bench Testing: For verification purposes testing was undertaken to demonstrate that there was no change in the sealing properties of the Gelweave™ branched vascular grafts as a result of introduction and manipulation of an endovascular delivery catheter via a side branch. A 26 F introducer was inserted into the 10mm side branch of 21 Gelweave grafts. The introducer and side branch were tied to allow movement back and forth within the side branch. After this process, the graft was pressurised with the citrated horse blood to 120 mm/Hg. During this pressurisation process, the delivery catheter was deliberately pushed into the main body of the Gelweave graft, i.e. it was forced onto the graft body wall. The force used to push the delivery catheter into the graft wall far exceeded that which a surgeon would use in the operating room. The delivery catheter was pushed onto the graft body wall 10 times over a 2 minute period. This is again in excess of what would be undertaken in the operating room during this procedure. Any blood lost from the graft due to this manual action of the delivery catheter was collected on a pre-weighed absorbent tissue. This was reweighed after the test and the total weight of blood lost from the graft calculated. The results have shown that the insertion of a 26 F endovascular delivery catheter through a 10mm side branch of a standard Gelweave graft does not alter the sealing properties of the ronft i.e. does not cause damage to the structure of the graft, which would result in blood leakage.
Clinical Information: No clinical testing has been undertaken, however a comprehensive review of the literature in relation to clinical use of standard branched polyester vascular grafts in relation to performance of debranching and associated hybrid procedures is provided. These procedures are considered standard clinical practice in light of current endovascular procedures and allow surgeons to undertake repair of aortic vessels and aneurysm in either a single or two-stage procedure.
Debranching and hybrid procedures are considered an alternative treatment in the light of current endovascular procedures as they allow the surgeon to undertake repair of aortic vessels and aortic aneurysms in a less invasive approach; the gold standard of aortic arch aneurysm and thoracoabdominal aneurysm, still remains conventional open surgical repair. Since open surgical repair is a rather invasive approach and the surgical insult is often intolerable for high risk elderly patients, other alternative treatments options have emerged. As discussed, these new treatment options include hybrid procedures and also branched endovascular stent grafts.
Gonclusion: In conclusion, the Vascutek Gelweave™ branched vascular grafts and Gelweave Siena™ vascular grafts with and without radiopague markers are substantially equivalent to the predicate devices in commercial distribution and results of non-clinical tests and supporting information from the literature demonstrates that the devices continue to be safe, effective, and perform as well as the predicate devices.
## Substantial Equivalence:
Substantial Lquivalence: " branched vascular grafts and Gelweave Siena™ vascular grafts with and without radiopaque markers are substantially equivalent in overall intended use, design and materials, performance, principles of operation and technology to the predicate This is shown comprehensively in Table 2, Section 10 of this pre-market devices. notification.
{4}------------------------------------------------
There are no changes to the Vascutek Gelweave™ branched vascular grafts and Gelweave There and the one of a crown of thout radiopaque markers other than the request for an
expanded indication.
1. 1. 1. 1.
Karai Leto
Signature
.............................................................................................................................................................................. Date
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a stylized caduceus, a symbol often associated with healthcare, with a double helix-like design.
Public Health Service
JUN 19 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Vascutek, Ltd. c/o Ms. Karen Kelso Regulatory Affairs Manager Newmains Avenue Inchinnan Industrial Estate Renfrewshire, Scotland PA4 9RR, United Kingdom
Rc: K090987
Trade Nmae: Vascutek Gelweave™ branched vascular grafts and
Gelweave SienaTM vascular grafts with and without radiopaque markers Regulation Number: 21 CFR 870.3450 Regulation Name: Vascular Graft Prosthesis Regulatory Class: Class II (two) Product Code: DSY Dated: June 2, 2009
Received: June 9, 2009
# Dear Ms. Kelso:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Karen Kelso
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to -http://www.fda.gov/cdrh/mdr/. _______________________________________________________________________________________________________________________________________________
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet addresshttp://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Dmina R. Vachner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K. 09.0987
Device Name:
Gelweave™ Branched Vascular Grafts Vascutek including Gelweave Siena™ Vascular Grafts with and without Radioapaque Markers
Indications For Use:
Gelweave "M branched vascular graffs:
"Repair of replacentent of damaged and diseased vessels of the abdomen and thoracic aorta in cases of anewrysm, dissection or courclation".
Sicua TM grafts vascular with and without Gelweave radiopaque markers:
"Repair or replacement of damaged and diseased vessels of the thoracie aorta in cases of anew ysm, dissection or coarctation".
"Gelweave" brunched vascular grafts, including Siena" grafts can also be used for debranching i.e. reconstruction of the aortic vessels and associated Hybrid procediires".
Prescription Use (Part 21 CFR 80L Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
地边流!
Donna R. Richter
(Division Sign Off)
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number K09
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.