BLACKROCK NEUROPORT BIOPOTENTIAL SIGNAL PROCESSING SYSTEM
Applicant
I2s Micro Implantable Systems, LLC
Product Code
GWL · Neurology
Decision Date
May 28, 2009
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 882.1835
Device Class
Class 2
Indications for Use
The Blackrock NeuroPort Biopotential Signal Processing System supports recording, processing and display of biopotential signals from user supplied electrodes. Biopotential signals include: Electrocorticography (ECoG), electroencephalography (EEG), electromyography (EMG), electrocardiography (ECG), electrooculography (EOG) and Evoked Potential (EP).
Device Story
System acquires, processes, and displays biopotential signals (ECoG, EEG, EMG, ECG, EOG, EP) from user-supplied electrodes or NeuroPort Array. High input impedance amplifier (1000 Megohm) captures signals; 16-bit A/D conversion at sampling rates up to 30,000 Hz. Device supports up to 128 channels (256 via cascading). Used in clinical/research settings by clinicians/researchers. Output displayed on PC for interpretation; no physiologic alarms provided. Passive signal splitter allows simultaneous connection to user equipment. System facilitates signal analysis via third-party software (e.g., Persyst Insight, NeuroPlex, Matlab, Spike 2). Benefits include flexible, high-channel-count signal acquisition for neurophysiological monitoring and research.
Clinical Evidence
Bench testing only. Documentation included software testing, system testing, and standards compliance (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26) to demonstrate substantial equivalence.
Technological Characteristics
Physiologic signal amplifier; 1000 Megohm input impedance; 16-bit A/D conversion; up to 30,000 Hz sampling rate; >90 dB CMRR. Connectivity via interface cables and passive signal splitter. Supports cascading for 256 channels. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-26. Power: 110 VAC.
Indications for Use
Indicated for recording, processing, and display of biopotential signals (ECoG, EEG, EMG, ECG, EOG, EP) in research institutions, clinics, hospitals, operating rooms, and epilepsy evaluation environments using user-supplied electrodes or the NeuroPort Array.
Regulatory Classification
Identification
A physiological signal amplifier is a general purpose device used to electrically amplify signals derived from various physiological sources (e.g., the electroencephalogram).
Predicate Devices
NeuroPort Neural Signal Processing (NSP) System (K042626)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Blackrock Microsystems. The logo is a stylized figure eight with the company name below it. The figure eight is black and white, and the company name is in a sans-serif font.
# 510(k) Summary
Device Name: Blackrock NeuroPort Biopotential Signal Processing System
510(k) Sponsor: I2S Micro Implantable Systems, LLC, d/b/a Blackrock Microsystems 391 Chipeta Way, Suite G Salt Lake City, UT 84108
510(k) Contact: Chris DeCaria Director of Product Development Blackrock Microsystems 391 Chipeta Way, Suite G Salt Lake City, UT 84108
Summary Date: May 15, 2009
Trade Name: Blackrock NeuroPort Biopotential Signal Processing System
Common Name: Bio-potential Signal Acquisition System
Classification Name: CFR 882.1835 Physiologic Signal Amplifier, Product Code: GWL CFR 882.1845 Physiological Signal Conditioners, Product Code: GWK
Predicate Devices:
510{k) Number: K042626
Trade Name: NeuroPort™ Neural Signal Processing (NSP) System
510(k) Number: K060803 Trade Name: g.USBamp
510(k) Number: K040113 Trade Name: Sandman SD20 Amplifier
#### 1.0 Description of Device
The Blackrock NeuroPort Biopotential Signal Processing System (System), when connected to customer supplier bio-potential electrodes or the NeuroPort Electrode Array (K042384), supports recording and display of bio-potential signals. Bio-potential signals that can be recorded by the System include:
- 1. Electrooculography (EOG),
- Electrocorticography (ECoG), 2.
Confidential
Page 1-2
{1}------------------------------------------------
- electroencephalography (EEG), 3.
- 4. electromyography (EMG),
- న్న electroencephalography (ECG) and
- evoked potential (EP). 6.
The System is not a monitoring system. No physiologic alarms are provided. The acquisition and display of bio-potential signals is for the interpretation and use of the clinician.
#### 1.2 Clinical Application
The System is used in research institution, clinic, hospital, operating room and epilepsy evaluation environments to acquire (record) bio-potential signals from user supplied electrodes or the NeuroPort Array. All of these electrodes require a high input impedance amplifier to acquire and display the bio-potential signal from the human body the electrodes contact.
## 2.0 Intended use of Device
The Blackrock NeuroPort Biopotential Signal Processing System supports recording, processing and display of biopotential signals from user supplied electrodes. Biopotential signals include: Electrocorticography (ECoG), electroencephalography (EEG), electromyography (EMG), electrocardiography (ECG), electrooculography (EOG) and Evoked Potential (EP).
### 3.0 Technological Characteristics
The only hardware modifications to the predicated NeuroPort System are:
- 1. The addition of interface cables to support connection of user supplied electrodes to the amplifier. These interface cables are verified by the continuity of the single pathway, commercial electrode input connector to the amplifier output connector.
- 2. A passive signal splitter box allows the user supplied electrodes to connect to the System Amplifier and the user's own equipment. This passive box will be verified by continuity of the signal pathways between the user supplied electrode input connector and the two signal connector outputs (System Amplifier output connector and User defined equipment connector).
- Sync to an external EEG instrument, such as the X-CEL TEK device is addressed in 3. the software testing, see Section 6.0.
# Confidential
{2}------------------------------------------------
- 4. The syncing of two Systems to be capable of recording up to 256 input channels is supported in the software testing, see Section 6.
## 3.1 Comparison to Predicates
| Feature | System<br>Under Review | Predicate<br>NeuroPort<br>Instrument<br>(K060523) | Predicate<br>g.USBamp<br>(K060803) | Predicate<br>Sandman SD20<br>Amplifier<br>(K0040113) | Comments |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use,<br>Indications<br>for Use | The Blackrock NeuroPort<br>Biopotential Signal<br>Processing System<br>supports recording,<br>processing and display of<br>biopotential signals from<br>user supplied electrodes.<br>Biopotential signals<br>include, but are not limited<br>to: Electrocorticography<br>(ECOG),<br>electroencephalography<br>(EEG), electromyography<br>(EMG),<br>electrocardiography<br>(ECG), electrooculography<br>(EOG) and Evoked<br>Potential (EP). | The intended use of<br>the Cyberkinetics<br>Neurotechnology<br>Systems, Inc.<br>NeuroPort Neural<br>Signal Processor<br>System is for<br>temporary (< 30<br>days) recording and<br>monitoring of brain<br>electrical activity. | Measuring,<br>recording and<br>analysis of<br>electrical activity of<br>the brain and/or<br>through the<br>attachment of<br>multiple electrodes<br>at various locations<br>to aid in morning<br>and diagnosis as<br>routinely found in<br>clinical settings of<br>EEG. | The SD20<br>Amplifier is<br>intended to be use<br>by or under the<br>directions of a<br>physician for<br>acquisition of EEG,<br>polygraphy and<br>polysomnography<br>signals and<br>transmission of<br>these signals to a<br>PC during<br>recording of<br>neurophysiology<br>examinations. | Equivalent<br>The Blackrock<br>NeuroPort<br>Biopotential<br>Signal<br>Processing<br>System under<br>review acquires,<br>conditions and<br>supports analysis<br>of the same types<br>of biopotential<br>signals as the<br>predicate<br>devices.<br>Polygraphy and<br>polysmnography<br>includes EEG,<br>EMG, ECG,<br>EOG,<br>biopotential<br>signals. |
| Typical<br>Biopotential<br>Signals<br>Recorded | Electroencephalography<br>(EEG)<br>Electrocorticography<br>(ECOG)<br>Electrocardiography (ECG)<br>Electromyography (EMG)<br>Electrooculography (EOG)<br>Evoked potential (EP) | EEG | EEG<br>ECoG<br>ECG<br>EMG<br>EOG | EEG<br>ECG<br>EMG<br>EOG<br>EP | Equivalent |
| Number of<br>Signal<br>Recording<br>Channels | Up to 128 with one<br>device;<br>Up to 256 by cascading<br>two devices | 96 | Up to 128 | Up to 128 | Equivalent |
| Amplifier<br>Input<br>Impedance | 1000 Megohm | 1000 Megohm | > 100 Megohm | Unknown | Same as<br>predicate<br>NeuroPort<br>Instrument. |
| A/D<br>Conversion | 16 Bit | 16 Bit | 24 Bit | 16 Bit | Equivalent |
| Sampling<br>Rate | Up to 30,000 Hz | Up to 30,000 Hz | Up to 38,400 Hz | Up to 32,000 Hz | Equivalent |
| CMRR | > 90 dB | > 90 dB | > 100 dB | > 100 dB | Equivalent |
·
{3}------------------------------------------------
| Feature | System<br>Under Review | Predicate<br>NeuroPort<br>Instrument<br>(K060523) | Predicate<br>g.USBamp<br>(K060803) | Predicate<br>Sandman SD20<br>Amplifier<br>(K0040113) | Comments |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Analysis<br>Software | Embedded, commercially<br>available and user<br>defined.<br>Examples:<br>Persyst Insight™<br>NeuroPlex™<br>Matlab™<br>Spike 2™<br>and others. | Embedded,<br>commercially<br>available and user<br>defined.<br>Examples:<br>Persyst Insight™<br>NeuroPlex™<br>Matlab™<br>Spike 2™<br>and others. | Embedded and<br>commercially<br>available.<br>Examples:<br>Matlab™<br>and others. | Embedded and<br>commercially<br>available | Equivalent<br>The<br>physician/clinician<br>may apply their<br>own analysis<br>software as a<br>research tool or<br>other<br>commercially<br>available<br>software<br>supporting<br>analysis. |
| Power | 110 VAC | 110 VAC | USB Port | Unknown | Safety of the<br>power source is<br>verified by<br>standards<br>compliance. |
| Alarms | No | No | No | Unknown | Same |
| Safety<br>Standards<br>Compliance | IEC 60601-1:1998<br>IEC 60601-1-2:2001<br>IEC 60601-2-26 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-26 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-26 | IEC 60601-1<br>IEC 60601-1-2 | Same |
## 4.0 Data Summary
Software documentation, system testing and standards compliance were provided to support the substantial equivalence, safety and effectiveness of the System.
## 5.0 Conclusions
The modifications to the Predicate NeuroPort System to create the Blackrock NeuroPort Biopotential Signal Processing System were evaluated and raise no new questions of safety or effectiveness.
Confidential
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, with stylized lines forming the wings and body.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
I2S Micro Implantable Systems, LLC c/o Gary Syring, Principal Consultant Quality & Regulatory Associates, Inc. 800 Levanger Lane Stoughton, WI 53589
MAY 2 8 2009
Re: K090957
Trade/Device Name: Blackrock NeuroPort Biopotential Signal Processing System Regulation Number: 21 CFR 882.1835 Regulation Name: Physiologic Signal Amplifier Regulatory Class: II Product code: GWL (and GWK) Dated: April 2, 2009
Received: April 6, 2009
# Dear Mr. Syring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Malvina B. Eglehnus, mD.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 690957
Indications for Use
510(k) Number (if known):
Device Name: Blackrock NeuroPort Biopotential Signal Processing System
Indications for Use:
The Blackrock NeuroPort Biopotential Signal Processing System supports recording, processing and display of biopotential signals from user supplied electrodes. Biopotential signals include: Electrocorticography (ECoG), electroencephalography (EEG), electromyography (EMG), electrocardiography (ECG), electrooculography (EOG) and Evoked Potential (EP).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
E.K. Rhm
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices
510(k) Number K090957
Page 1 of 1
Confidential
Page 3-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.