SMITH & NEPHEW RF CANNULATE

K090955 · Smith & Nephew, Inc. · GXI · Jul 16, 2009 · Neurology

Device Facts

Record IDK090955
Device NameSMITH & NEPHEW RF CANNULATE
ApplicantSmith & Nephew, Inc.
Product CodeGXI · Neurology
Decision DateJul 16, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Smith & Nephew RF Cannulae are intended to facilitate placement of Smith & Nephew RF Denervation Probes. The Smith & Nephew RF Cannulae are indicated for use in RF heat lesion procedures for the relief of pain.

Device Story

Smith & Nephew RF Cannulae are surgical access tools designed to facilitate the placement of RF Denervation Probes during radiofrequency heat lesion procedures. The cannulae are provided in various lengths, gauges, and tip configurations (sharp or blunt; straight or curved) to accommodate different anatomical sites. The device is used by clinicians in a clinical setting to provide a pathway for the RF probe to reach the target tissue. By enabling precise probe placement, the cannulae assist in the delivery of RF energy to perform nerve ablation, which is intended to provide pain relief for the patient.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Cannulae available in 18, 20, and 22 gauge sizes with varying lengths (5cm, 10cm, 15cm) and tip configurations (sharp/blunt, straight/curved). Designed for manual insertion to guide RF probes. No energy source or software components.

Indications for Use

Indicated for use in RF heat lesion procedures for the relief of pain in patients requiring denervation probe placement.

Regulatory Classification

Identification

A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness Smith & Nephew RF Cannulae 090955 JUL 1 6 2009 Submitted By: Date: Contact Person: Proprietary Name: Common Name: Classification Name and Reference: Smith & Nephew, Inc. 1450 Brooks Road Memphis, TN 38116 April 03, 2009 Mason W. Robbins, RPCV Regulatory Affairs Specialist Tel: (901) 399-6021 Fax: (901) 399-1557 Smith & Nephew RF Cannulae Probe. Radiofrequency Lesion 21 CFR 882.4725, Radiofrequency lesion probe Device Product Code and Panel Code: GXI / Neurology / 84 #### Device Description: Smith & Nephew RF Cannulae are intended to facilitate the placement of Smith & Nephew RF Denervation Probes for the use in RF heat lesion procedures. The RF Cannulae are offered in a variety of lengths, gauges and tip configurations to accommodate various anatomical locations and differences in patient anatomy. ## Intended Use: Smith & Nephew RF Cannulae are intended to facilitate placement of Smith & Nephew RF Denervation Probes. The Smith & Nephew RF Cannulae are indicated for use in RF heat lesion procedures for the relief of pain. ## Technological Characteristics: The proposed Smith & Nephew RF Cannulae are technologically similar to the predicate Smith & Nephew RF Cannulae previously cleared under K063467. Several design modifications were made to the proposed RF Cannulae including changes to improve cannula insertion performance. ### Substantial Equivalence Information: The proposed RF Cannulae are substantially equivalent to the predicate device listed below based on similarities in design features, overall indications for use and material composition. - . Smithy & Nephew RF Cannulae (K063467) {1}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is depicted in a simple, line-art style, and the text is in a sans-serif font. The overall design is clean and professional, reflecting the department's role in public health and welfare. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUL 1 6 2009 Smith & Nephew, Inc. c/o Mason W. Robbins, RPCV Regulatory Affairs Specialist 1450 Brooks Rd. Memphis, TN 38116 Re: K090955 Trade/Device Name: Smith & Nephew RF Cannulae (See Enclosed list for Models) Regulation Number: 21 CFR 882.4725 Regulation Name: Radiofrequency Lesion Probe Regulatory Class: II Product Code: GXI Dated: June 25, 2009 Received: June 30, 2009 Dear Mr. Robbins: . We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Malvina B. Eydelman, M. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ cc: HFZ- 401 DMC HFZ- 404 510(k) Staff HFZ- 460 (DONED - Marsha L. Burke Nicholas) D.O. OC Numbers: | Division of Enforcement A | 240-276-0115 | |------------------------------------------------------------------------------------|--------------| | Dental, ENT and Ophthalmic Devices Branch | 240-276-0115 | | OB/GYN, Gastro. & Urology Devices Branch | 240-276-0115 | | General Hospital Devices Branch | 240-276-0115 | | General Surgery Devices Branch | 240-276-0115 | | Division of Enforcement B | 240-276-0120 | | Cardiovascular & Neurological Devices Branch | 240-276-0120 | | Orthopedic, Physical Medicine & Anesthesiology Devices and<br>Radiological Devices | 240-276-0120 | Drafted: Marsha L. Burke Nicholas, July 7, 2009 Edited: Marisol Lendor, July 7, 2009 Final: Marisol Lendor, July 7, 2009 Typed: Marisol Lendor, July 7, 2009 # IMAGE COPY | Carles of Children Court Children Controller<br>A natur w | a long the britain range of the comments in and meet<br>Bearing September 1 La Sales<br>change and control and the later and and | Career of Concession Company Company Comments of Concession Come of Concession Comparison of Concession Comers | Contrastrum, between belief | and the state with the been and the first and and the production of the country of the country of the county of the country of the country of the country of the county of the<br>Wellier on 16, A 10 to be province the 1<br>1 | .<br>Company of Children | A Comments of the Carder<br>Barren - - Verselling - Marin - An Photo Party & Production - Party Party & | -------------------------------------------<br>1 NETWYF 200<br>Carlos Career of Children<br>Port Port F F REAL<br>And Charles of the Same<br>. | Carlos Control Concession Company Controller Company Continued Company Company Company Company Company Company Company Company Company Comments of Children Comments of Childr | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | <br>100 Amount<br>And And of a contract and consisted and second an any and a more of the comments of the commend and the commend of the commend of the commend of the comments of the comments | | I | <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | An also a comment of the | MA & READ THE THERE F<br> | ------------------------------------- | 1 | | | Comments of A Comments of A | | --------------------------------------------------------------------------------------------- | Company of the American Company of Concelling Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company | Acres of the Artistics of Acres of American<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | | Acres Clause | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | A 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -<br>NA MARKE A HOLE LE BOOK & BUILD BE ALL BE | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>C C C C C CLANDING CHART | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | THE LEAST PATTER<br> | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | 1 | {4}------------------------------------------------ # Enclosure 1 . and the control control control control controllers. : 100 million and the states of the states and the contraction of the comments of the count . . . . . . . and the comments of the comments of : : ## Smith & Nephew RF Cannulae Models : . : : . . | Catalogue Number | Description | |------------------|------------------------------------------| | 7210273 | 22 Gauge RF Cannula | | 72200031 | 22 gauge, 5cm, 4mm sharp, curved tip | | 72200032 | 22 gauge, 5cm, 4mm sharp, curved tip | | 72200033 | 22 gauge, 10cm, 10mm sharp, curved tip | | | 22 gauge, 10cm, 10mm sharp, straight tip | | 7210274 | 20 Gauge RF Cannula | | 7210275 | 20 gauge, 10cm, 10mm blunt, curved tip | | 7210276 | 20 gauge, 10cm, 10mm sharp, curved tip | | 7210277 | 20 gauge, 10cm, 5mm sharp, straight tip | | 7210278 | 20 gauge, 15cm, 10mm blunt, curved tip | | 7210279 | 20 gauge, 15cm, 10mm sharp, curved tip | | 72200805 | 20 gauge, 15cm, 5mm sharp, straight tip | | 72200806 | 18 Gauge RF Cannula | | | 18 Gauge, 10cm, 10 mm sharp, curved tip | | | 18 Gauge, 15cm, 10 mm sharp, curved tip | {5}------------------------------------------------ # Premarket Notification Indications for Use Statement 510(k) Number (if known): K090955 Smith & Nephew RF Cannulae Device Name: Indications for Use: Smith & Nephew RF Cannulae are intended to facilitate placement of Smith & Nephew RF Denervation Probes. The Smith & Nephew RF Cannulae are indicated for use in RF heat lesion procedures for the relief of pain. Prescription Use (Part 21 CFR 801.109) AND/OR Over-the-Counter Use (Optional Format 1-2-96) # (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE OŇ ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) MLB Nicholas Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K090955 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...