NFOCUS NEUROMEDICAL ECHO DISTAL ACCESS CATHETER

K090918 · Nfocus Neuromedical, Inc. · DQY · Jul 30, 2009 · Cardiovascular

Device Facts

Record IDK090918
Device NameNFOCUS NEUROMEDICAL ECHO DISTAL ACCESS CATHETER
ApplicantNfocus Neuromedical, Inc.
Product CodeDQY · Cardiovascular
Decision DateJul 30, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The Echo Distal Access Catheter is indicated for the introduction of interventional and/or diagnostic devices and fluid infusion into the peripheral, coronary, and neuro vasculature.

Device Story

Echo Distal Access Catheter is a single-stiffness percutaneous catheter designed for vascular access. It features an atraumatic distal tip, a luer hub, and a hydrophilic shaft coating to minimize friction. Used in peripheral, coronary, and neuro vasculature, the device facilitates the delivery of interventional or diagnostic tools and fluid infusion. It is compatible with 6F or greater introducer sheaths and requires a separate rotating hemostasis valve and guidewire. Operated by clinicians in a clinical setting, the catheter provides a pathway for therapeutic or diagnostic procedures, potentially improving access to target vascular sites.

Clinical Evidence

Bench testing and in vivo testing performed. Performance, sterilization, and biocompatibility testing verified the device performs as designed and is suitable for its intended use. No clinical trial data provided.

Technological Characteristics

Single-stiffness percutaneous catheter; 6F outer diameter; 0.070" inner diameter; 105cm working length. Materials: PTFE, Pebax, Stainless Steel, hydrophilic coating. Transition stiffness from hub to tip. Non-powered, mechanical device.

Indications for Use

Indicated for patients requiring introduction of interventional/diagnostic devices or fluid infusion into peripheral, coronary, or neuro vasculature.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for NFOCUS neuromedical. The logo consists of a stylized "N" made up of triangles, followed by the word "NFOCUS" in bold, black letters. Below the word "NFOCUS" is the word "neuromedical" in a smaller, italicized font. K0909118 JUL 30 2009 #### 5. 510(K) SUMMARY Date Prepared: Trade Name: Common Name: Classification: Applicant: Contact Person: Predicate Devices: March 30, 2009 Echo™ Distal Access Catheter Percutaneous Catheter Class II, 21 CFR 870.1250, Product Code DQY Nfocus™ Neuromedical 2191 E. Bayshore Road, Suite 100 Palo Alto, CA 94303 USA Tel: 650-845-3064 Fax: 650-813-1869 Bob Founds Director, Quality Assurance and Regulatory Affairs Envoy Guiding Catheter, manufactured by Cordis Corporation Neuron Intracranial Access System, manufactured by Penumbra, Inc. #### Device Description: The Echo " Distal Access Catheter is a single-stiffness catheter with an atraumatic tip at the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating to reduce friction during use. Device dimensions and configuration are shown on the product label. The Echo Distal Access Catheter is compatible with introducer sheaths having an inner diameter of 6F or greater. A rotating hemostasis valve with side-arm adapter and a compatible guidewire are required, but not provided with the Echo catheter. #### Intended Use: The Echo Distal Access Catheter is indicated for the introduction of interventional and/or diagnostic devices and fluid infusion into the peripheral, coronary, and neuro vasculature. The indicated use is substantially equivalent to that of the legally marketed predicate devices. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for NFOCUS neuromedical. The logo consists of a stylized letter N made up of three triangles, followed by the word "NFOCUS" in a bold, sans-serif font. Below the word "NFOCUS" is the word "neuromedical" in a smaller, italicized font. The logo is black and white. # Technological Characteristics of the Device Compared to the Predicate Device: The Echo Distal Access Catheter uses similar technology, has similar intended use, functions, materials and method of operation as the following predicate device(s): | Device | Nfocus Neuromedical<br>Echo Distal Access<br>Catheter (Subject<br>Device) | Cordis Endovascular<br>Systems ENVOY<br>Guiding Catheter | Penumbra Inc., Neuron<br>Intracranial Access<br>System | |----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The Echo Distal Access<br>Catheter is indicated for<br>the introduction of<br>interventional and/or<br>diagnostic devices and<br>fluid infusion into the<br>peripheral, coronary,<br>and neuro vasculature. | The Envoy Guiding<br>Catheter is intended for<br>use in the peripheral,<br>coronary, and<br>neurovasculature for the<br>introduction of<br>interventional /<br>diagnostic devices. | The Neuron Intracranial<br>Access System is<br>indicated for the<br>introduction of<br>interventional devices<br>into the peripheral,<br>coronary, and neuro<br>vasculature. | | Design: | | | | | Outer Diameter (OD) | 6 French | 6 French | 6 French | | Inside Diameter (ID) | 0.070" | 0.070" | 0.070" | | Working Length | 105cm | 90-100cm | 95-105cm | | Catheter shaft | Transition in stiffness<br>from hub to tip | Transition in stiffness<br>from hub to tip | Transition in stiffness<br>from hub to tip | | Materials | Predominantly PTFE,<br>Pebax, Stainless Steel,<br>Hydrophilic coating | Predominantly PTFE,<br>Nylon/polyurethane,<br>Stainless Steel, | Stainless Steel,<br>platinum, hydrophilic<br>coating | Note: Materials in the predicate devices are not known with certainty. Material equivalence is demonstrated by in vivo performance tests and biocompatibility tests. ### Summary of Studies: In vitro bench testing and in vivo testing have been performed on the device materials and finished devices. Performance, sterilization and biocompatibility testing verified that the Echo Distal Access Catheter performs as designed and is suitable for its intended use. ### Conclusion: The data presented in this submission demonstrate that the Nfocus Neuromedical Echo Distal Access Catheter is substantially equivalent to the predicate devices identified in regards to device design, materials, and intended use. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (DHHS). The seal features an abstract eagle design, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" encircling the top portion of the seal. The eagle is stylized with thick, black lines, and the text is in a simple, sans-serif font. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Nfocus™ Neuromedical c/o Mr. Bob Founds Director, Quality Assurance and Regulatory Affairs 2191 E. Bay Shore Road Suite 100 Palo Alto, Ca 94303 THE 30 2009 K090918 Re: > Trade/Device Name: EchoTM Distal Access Catheter Common Name: Catheter, Percutaneous Regulation Number: 21 CFR 870.1250 Regulatory Class: II Product Code: DQY Dated: July 16, 2009 Received: July 21, 2009 Dear Mr. Founds: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Bob Founds forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, R.Z. MP Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for NFocus Neuromedical. The logo consists of a stylized letter N made up of three triangles, followed by the word "NFOCUS" in bold, sans-serif font. Below the word "NFOCUS" is the word "neuromedical" in a smaller, italicized font. #### INDICATIONS FOR USE STATEMENT 4. 510(K) Number (if known): KO90919 Device Name: Nfocus™ Neuromedical Echo™ Distal Access Catheter Indications for Use: The Nfocus Neuromedical Echo Distal Access Catheter Vessel is indicated for the introduction of interventional and/or diagnostic devices and fluid infusion into the peripheral, coronary, and neuro vasculature. Prescription Use X (21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Off (Division Sign-Off) Division of Caralovascular Devices Kogog 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...