The i_Open 0.5T system is an open, whole body scanner. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the head, body, or extremities. The images produced by the i Open 0.5T system reflect the special distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Device Story
i_Open 0.5T is an open, whole-body permanent magnet MRI system (0.5 Tesla). Device uses magnetic resonance to excite protons in patient tissues; inputs include RF signals and magnetic field gradients. System transforms these inputs into 2D/3D images (T1, T2, proton density, MRA, MRCP) via integrated console software. Operated by trained clinicians in clinical settings; images are reviewed by physicians to assist in diagnosis. Benefits include non-invasive visualization of internal structures and organs.
Clinical Evidence
Bench testing only. Performance evaluated against FDA guidance for MR Diagnostic Devices, including NEMA standards for signal-to-noise ratio, image uniformity, geometric distortion, slice profile, and spatial resolution. No clinical data presented.
Technological Characteristics
0.5 Tesla permanent magnet MRI system; magnet material: Nd-B-Fe. Imaging capabilities: SE, FSE, STIR, FLAIR, Rewinded-GRE, Spoiled-GRE, TOF. Connectivity: DICOM printing/storage. Software: Windows XP-based interactive control, reconstruction, and post-processing. Conforms to NEMA and IEC standards.
Indications for Use
Indicated for diagnostic imaging of head, body, spine, and extremities in patients requiring MRI. Produces transverse, sagittal, coronal, and oblique images based on proton density, T1, T2, and flow. Contraindications not specified.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K090873
Page 1 of 3
# 510(k) Summary
APR 1 0 2009
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 ad 21CFR 807.92.
## I. General Information
### Establishment:
| Name: | Beijing Wandong Medical Equipment Co., Ltd |
|----------|-----------------------------------------------------------------------|
| Address: | No.7, SanJianFang NanLi, ChaoYang District,<br>Beijing, CHINA, 100024 |
| Phone: | +86 10 84575844 |
| Fax: | +86 10 84575842 |
Registration Number: 3004859018
| Contact Person: | Mr. Wang Weimin<br>Manager, | |
|-----------------|-----------------------------|-----------------|
| | Phone: | +86 10 84575844 |
| | Fax: | +86 10 84575842 |
| | E_Mail: | wmw@263.net.cn |
Date of Summary Preparation: Dec 17, 2008
Type of submission: traditional
### Device Name:
- ●Trade Name: i Open 0.5T
### ●Classification Number:
Magnetic Resonance Diagnostic Device, CFR 892.1000 90-LNH
- ●Classification: Class II
## ●Performance Standards:
None established under Section 514 the Food, Drug, and Cosmetic Act.
{1}------------------------------------------------
## II. Safety and Effectiveness Information.
#### . Device Description:
See Part C <Device Description> document.
#### ● Intended Use:
The i_Open 0.5T system is an open, whole body scanner. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the head, body, or extremities. The images produced by the i Open 0.5T system reflect the special distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
| Anatomical Region: | Head, Body, Spine, Extremities |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nucleus excited: | Proton |
| Diagnostic uses: | T1,T2 weighted<br>Proton density weighted<br>MIP-MRA<br>Water suppress imaging<br>fat suppress imaging<br>MRCP |
| Imaging capabilities: | 2D Spin Echo (SE)<br>2D,3D Fast Spin Echo(FSE)<br>2D Short Tau Inversion Recovery (STIR)<br>2D Fluid Attenuated Inversion Recovery (FLAIR)<br>2D,3D Rewinded Gradient Echo (2D, 3D Rewinded-GRE)<br>2D, 3D Spoiled Gradient Echo (2D,3D Spoiled-GRE)<br>2D,3D Time of Flight Angiography (TOF) |
#### . Technological Characteristics (comparison with predicate device):
The i_Open 0.5T is a 0.5 Tesla permanent MRI system. The magnet is mainly made of Nd-B-Fe material. The system software based on Windows XP is an interactive program integrated with scanning control, image reconstruction, reviewing, post-processing, DICOM printing.
#### ● Predicated Device:
K974212: Hitachi AIRIS II K001334: AIRIS II Version 4.1 Software
{2}------------------------------------------------
K090873
page 3 of 3
#### . Statement of Substantial Equivalence:
The i Open 0.5T is of comparable type and substantially equivalent to Hitachi AIRIS II (K974212) and AIRIS II version 4.1 Software (K001334) in that they are similar in technology and intended uses. Both of these systems are open-permanent-magnet MRI Imaging System, use Gradient Subsystem to provide controlled and uniform gradient magnet fields in the X, Y and Z directions, and use RF Subsystem to complete the function of RF signal transmitting/receiving and processing. Image reconstruction is controlled by console that has an interactive user interface, and the system produces 2D and 3D image that can be filmed or electronically stored for future review. Both of these systems have the traditional MRI units.
#### . General Safety and Effectiveness Concerns:
Operation of the i_Open 0.5T is substantially equivalent to the commercially available AIRIS II. The following are the safety parameter with action levels:
- . Maximum Static Field
- Rate of Change of Magnetic Field .
- RF Power Deposition .
- Acoustic Noise Levels
and performance levels:
- Specification Volume .
- Signal to Noise
- Image Uniformity
- . Geometric Distortion
- Slice Profile, Thickness and Gap
- High Contrast Spatial Resolution
specified by the FDA guidance document for MR Diagnostic Devices that will be evaluated. The i Open 0.5T will conform to the FDA recognized NEMA Standards for the measurement of performance and safety parameters and the international IEC standard for safety issues with Magnetic Resonance Imaging Devices. This will assure that the performance of this device can be considered safe and effective with respect to currently available system.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
## APR 1 0 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Beijing Wandong Medical Equipment Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25" Street NW BUFFALO MN 55313
Re: K090873 .
Trade/Device Name: i_Open 0.5 T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: March 30, 2009 Received: March 31, 2009
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
[signature]
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): _ < 0 90873
Device Name: i Open 0.5T
Indications for Use:
The i Open 0.5T system is an open, whole body scanner. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the head, body, or extremities. The images produced by the i Open 0.5T system reflect the special distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
| Anatomical Region: | Head, Body, Spine, Extremities |
|-----------------------|----------------------------------------------------|
| Nucleus excited: | Proton |
| Diagnostic uses: | T1,T2 weighted |
| | Proton density weighted |
| | MIP-MRA |
| | Water suppress imaging |
| | fat suppress imaging |
| | MRCP |
| Imaging capabilities: | |
| | 2D Spin Echo (SE) |
| | 2D,3D Fast Spin Echo(FSE) |
| | 2D Short Tau Inversion Recovery (STIR) |
| | 2D Fluid Attenuated Inversion Recovery (FLAIR) |
| | 2D,3D Rewinded Gradient Echo (2D, 3D Rewinded-GRE) |
| | 2D, 3D Spoiled Gradient Echo (2D,3D Spoiled-GRE) |
| | 2D,3D Time of Flight Angiography (TOF) |
Prescription Use
Yes AND/OR Over-The-Counter Use ______
(Part 21 CFR 801 (21 CFR 801 Subpart
Subpart D) C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number ________ K090873
・・・・・・・
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.