# MODIFICATION TO SYNTEC-TAICHUN NON-STERILE BONE PLATE AND SCREW IMPLANTS (K090839)

_Syntec Scientific Corp. · HRS · Apr 24, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K090839

## Device Facts

- **Applicant:** Syntec Scientific Corp.
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Apr 24, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The bone plates and screws are provided non-sterile. The devices are intended to treat fractures of various bones, including the clavicle, scapula, pelvis, long bone (humerus, ulna, radius, femur, tibia, and fibula), and small bone (metacarpals, metatarsals, and phalanges).

## Device Story

Syntec-Taichung Bone Plates and Screws are metallic orthopedic fixation implants used to stabilize bone fractures. Plates are manufactured from 316LS stainless steel (ASTM F138) or pure titanium. Devices are provided non-sterile and require surgical implantation by a physician. Plates range from 1.5mm to 6.5mm in thickness, 6.5mm to 17.5mm in width, and 24.0mm to 394.0mm in length (2 to 22 holes). The system is designed to be used with screws previously cleared under K983495. The implants provide mechanical stabilization to facilitate bone healing.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Materials: 316LS Stainless Steel (ASTM F138) or Pure Titanium. Dimensions: 1.5-6.5mm thickness, 6.5-17.5mm width, 24.0-394.0mm length. Form factor: Bone plates with 2-22 holes. Non-sterile. Mechanical fixation.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Syntec-Taichung Bone Plates and Screw Implant ([K983495](/device/K983495.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K090839

### 510(K) SUMMARY

| Submitter's name          | Syntec Scientific Corporation                                                                                                         | APR 24 2005 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Address                   | 2, Kung San Rd,<br>Chuan Shing Industrial Zone,<br>Shen Kang,<br>Chang Hua, Taiwan.<br>Telephone :886-4-7987099<br>Fax: 886-4-7987077 |             |
| Contact person            | Carol Chang                                                                                                                           |             |
| Name of the device        | Syntec-Taichung Bone Plates and Screw<br>Implant                                                                                      |             |
| Trade or proprietary name | Syntec-Taichung Bone Plates and Screw<br>Implant                                                                                      |             |
| Classification name       | Plate, Fixation, Bone                                                                                                                 |             |
| Prode Code                | HRS                                                                                                                                   |             |
| Regulation Number         | 888.3030                                                                                                                              |             |
| Class                     | II                                                                                                                                    |             |
| Predicate device:         | Syntec-Taichung Bone Plates and Screw<br>Implant                                                                                      |             |

### Description of the Device:

The plate is manufactured from commercially 316LS Stainless Steel which in conformance with ASTM F138 and an alternative material which is Pure Titanium. The plates designed are from 1.5mm to 6.5mm thickness, from 6.5mm to 17.5mm width, and are available variously in length from 24.0mm (2 holes) to 394.0mm (22 holes). It should assembly with the same screw that was the predicate device in concurrence by FDA, (K983495).

#### Indications for Use:

The bone plates and screws are provided non-sterile. The devices are intended to treat fractures of various bones, including the clavicle, scapula, pelvis, long bone (humerus, ulna, radius, femur, tibia, and fibula), and small bone (metacarpals, metatarsals, and phalanges).

### Substantially Equivalence:

The predicate device information provided supports the substantially equivalence.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol. The text is in all capital letters and is evenly spaced around the circle.

AD

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Syntec Scientific Corporation % Ms. Carol Chang Regulatory Affairs Specialist 3F1. 96 Chung Hsaio E. Rd. Sec 3 Taipei China (Taiwan) 106

APR 2 4 2009

Re: K090839

Trade/Device Name: Syntec-Taichung Bone Plates and Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS, HWC Dated: March 27, 2009 Received: March 27, 2009

Dear Ms. Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{2}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

For Pdo/m

De

Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

## 510(k) Number (if known): K090839

## Device Name: Syntec-Taichung Non-Sterile Bone Plate and Screw Implant

## Indications for Use:

The bone plates and screws are provided non-sterile. The devices are intended to treat fractures of various bones, including the clavicle, scapula, pelvis, long bone (humerus, ulna, radius, femur, tibia, and fibula), and small bone (metacarpals, metatarsals, and phalanges).

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number

---

**Source:** [https://fda-staging.innolitics.com/device/K090839](https://fda-staging.innolitics.com/device/K090839)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).