K090831 · Tanthera, Inc. · LEJ · May 29, 2009 · General, Plastic Surgery
Device Facts
Record ID
K090831
Device Name
TANTHERA-T35
Applicant
Tanthera, Inc.
Product Code
LEJ · General, Plastic Surgery
Decision Date
May 29, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4635
Device Class
Class 2
Indications for Use
Intended to provide ultraviolet light to tan the skin.
Device Story
TanThera-T35 is a clam-shell tanning chamber using UVA1 (340-400 nm) LED technology to induce skin tanning. Device features upper canopy and lower bench with acrylic platform for supine positioning. System utilizes multiple LED modules mounted on copper tube heat sinks with chilled-water liquid cooling. Operation is managed by T-Max timing system, temperature control, and flow sensors; session terminates automatically if operating parameters are exceeded or cooling fails. User operates device manually, with ability to terminate session via switch. Requires use of FDA-cleared eye protection. Benefits include tanning response via LED-based UVA1 emission.
Clinical Evidence
No clinical data. Substantial equivalence is based on structural, operational, and spectral comparisons to predicate devices and adherence to 21 CFR 1040.20 performance standards for sunlamp products.
Technological Characteristics
Clam-shell tanning chamber; UVA1 (340-400 nm) LED source (Nichia NCSU033A); copper tube heat sinks; chilled-water liquid cooling system; acrylic platform; T-Max timing system; low-voltage DC power (13V/12V/-12V). Complies with 21 CFR 1040.20.
Indications for Use
Indicated for individuals seeking to tan the skin via ultraviolet light exposure. Contraindications include individuals not wearing FDA-cleared eye protection.
Regulatory Classification
Identification
A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.
Special Controls
*Classification.* Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following:
(i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and
(ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly.
(2) Demonstrate that device is mechanically safe to prevent user injury.
(3) Demonstrate software verification, validation, and hazard analysis.
(4) Demonstrate that device is biocompatible.
(5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment.
(6)
*Labeling* —(i)*Sunlamp products.* (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement: Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions.
(ii)
*Sunlamp products and UV lamps intended for use in sunlamp products.* Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.”
(B) “Contraindication: This product must not be used if skin lesions or open wounds are present.”
(C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.”
(D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.”
(c)
*Performance standard.* Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.
{0}------------------------------------------------
K090831
Center for Devices and Radiological Health March 19, 2009
# SECTION 15 - 510(k) SUMMARY
MAY 2 9 2009
The 510(k) Summary is contained on the following pages.
## 510(k) Summary
For
# TanThera-T35 UVA1 Light Emitting Diode Tanning Chamber
Date Prepared: March 19, 2009
#### Section 15.1 - Sponsor Information
Contact Person: Peter D. Fiset, CEO Phone: 518.281.7383 Email: PeterFiset@Opthera.com
Christopher W. Macomber, COO Phone: 518.265.2882 Email: ChrisMacomber@Opthera.com
Opthera, Inc. 5 Upper Loudon Road Loudonville, NY 12211 USA Fax: 518.436.0485
## Section 15.2 - Outside Regulatory Counsel
Foley & Lardner LLP 3000 K St., NW Suite 500 Washington, DC 20007
Contact Person: Nathan A. Beaver (202) 295-4039 (telephone) (202) 672-5399(facsimile) nbeaver@foley.com (email)
{1}------------------------------------------------
90831
Center for Devices and Radiological Health March 19, 2009
# Section 15.3 - Device Name
·Proprietary Name: TanThera-T35
·Common/Usual Name: UVA1 Light Emitting Diode Tanning Chamber
. Classification Names and Numbers: Ultraviolet lamp for tanning, General Hospital, LEJ.
# Section 15.3 - Device Description and Intended Use
Indications for Use: Intended to provide ultraviolet light to tan the skin.
The TanThera-T35 ("T35") is a tanning chamber employing UVA1 (340 nm - 400 nm) light-emitting diode ("LED") technology as the ultraviolet light source in order to provide a tanning response to human skin. It is designed in a clam-shell format with an upper ("canopy") and lower ("bench") radiant surface in order to provide a two-sided tan with a person in supine position on an acrylic platform.
The principle parts of the T35 include multiple T35 LED Modules, forty-four (44) copper tube heat sink with a linear array of thirty-eight (38) T35 LED Modules each, a T-Max ® Timing System, a temperature control system, and a chilled-water liquid cooling system. The copper tube arrays with temperature equalizing internal counter-current flow dams, are arranged with twenty-six (26) tube arrays on the canopy and eighteen (18) tube arrays on the bench. The tube arrays are mounted behind a diffuse reflector that has cutouts for each of the T35 LED Modules. There is an acrylic shield between the tanner and the arrays to prevent tampering and contact with T35 LED Modules as well as to provide a minimum distance from the emitters to the tanner. The liquid cooling system has temperature and flow controls that sense and control water flow to maintain the LED modules at an operating temperature range. The bed is designed to terminate operation if the operating parameters of the T35 LED Modules are not met. Should the T35 LED Module temperature increase due to low coolant flow or high coolant inlet temperatures then the session terminates to prevent overheating and possible early burnout of the LED modules. A failure of the cooling system decreases LED lifetime but does not pose a significant risk to the tanner.
The T35-LED Module is comprised of Nichia Chip-Type LEDs (Model NCSU033A), a power connector, a fuse, and a 700 mA nominal current regulator. The T35-LED Module (Model 3N33X1) is clearly marked with a recommended lifetime of 10,000 hours. The T35 is powered by low-voltage DC power (13 ± 0.5 VDC, 12 ± 0.1 VDC, - 12 ± 0.1 VDC) and low-voltage DC-DC converters located in the bench of the tanning chamber with redundant fans for cooling. T35 LED Modules are replaceable by TanThera certified technicians only.
The T35 has a nominal peak wavelength output of 365nm and a maximum tanning session of 35 minutes. The total tanning chamber output less than 1250 W of UVA 1 light. An industry de-facto standard T-Max® timer regulates the session. The T35
{2}------------------------------------------------
Center for Devices and Radiological Health March 19, 2009
090831
incorporates a manual session termination switch that allows a tanner to discontinue a session at any time.
Tanners are instructed and required to wear FDA-cleared eye protection and to follow the tanning instructions as provided in the User Manual.
# Section 15.4 - Predicate Device
+Suntana SunSystem SunBed and SunBrella, FDA 510(k) Reference Number K800744, Output Range: UVA-only
. Currently marketed UVA1 bulb shown to effectively tan: Philips TL10 ·Substantial Equivalence Comparison:
The TanThera-T35 is substantially equivalent to currently marketed sunlamp products. The predicate scenario established to demonstrate substantial equivalence of the T35 to current sunlamp products utilized the FDA 510(k) approved Suntana SunSystem, SunBed and SunBrella for structural and operational equivalence, and the Philips TL10 UVA1 lamp for spectral irradiance equivalence for effective tanning.
The T35 represents a change in technology from mercury vapor based lamps to lightemitting diodes. With the exception of the change of emitter technology, no significant structural or performance changes are present in the T35 when compared to the Suntana SunSystem, and the Philips TL10 UVA1 lamp, nor are there significant changes when compared to other currently marketed sunlamp products. The spectral output, mode of operation, and general operating principles are similar to or the same as the listed predicates and currently available sunlamp products. Although there are some differences in the production of UV light compared to other sunlamp products, the differences do not raise any questions in regard to safety or efficacy. Therefore, the TanThera-T35 LED tanning chamber, when used in accordance with the user manual, is structurally substantially equivalent to the Suntana SunSystem and SunBrella tanning bed, and spectrally substantially equivalent to the Philips TL10 UVA1 bulb.
# Section 15.5 - Performance Characteristic Summary
The TanThera-T35 design references applicable parts of the FDA document "Sunlamp products and ultraviolet lamps intended for use in sunlamp products" (21 CFR 1040.20) April 1, 2008, which outlines performance standards for sunlamps and sunlamp products, and the FDA document "GUIDE FOR PREPARING PRODUCT REPORTS ON SUNLAMPS AND SUNLAMP PRODUCTS (21 CFR 1002)", September 1995 (Address corrections Aug. 2008).
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling an abstract human figure with outstretched arms, accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The symbol is black, and the text is also black.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 2 9 2009
TanThera, Inc. % Foley & Lardner LLP Mr. Nathan A Beaver Washington Harbour 3000 K Street, Northwest, Suite 500 Washington, District of Columbia 20007
Re: K090831
Trade/Device Name: TanThera -- T35 Regulation Number: 21 CFR 878.4635 Regulation Name: Ultraviolet lamp for tanning Regulatory Class: I Product Code: LEJ Dated: March 26, 2009 Received: March 26, 2009
Dear Mr. Beaver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at
{4}------------------------------------------------
Page 2 - Mr. Nathan A Beaver
(240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at (240) 276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sincerely yours,
Mark N. Mellon
Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (ff known):
ко90831
· Device Name: · · ··························································································································································
Indications For Use:
TanThera - T35 Intended to provide ultraviolet light to tan the skin.
X : Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Niloenpurence of CDRH, Office of Device Evaluation (ODE) |
|---------------------|----------------------------------------------------------|
| (Division Sign-Off) | for mxm |
| | Division of Surgical, Orthopedic, |
| | and Restorative Devices |
| 510(k) Number | K090831 |
| | Page 1 of |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.