← Product Code [LSL](/productcode/LSL) · K090827

# BD PROBETEC NEISSERIA GONORRHOEAE (GC) QX AMPLIFIED DNA ASSAY (K090827)

_Becton, Dickinson & CO · LSL · May 27, 2009 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K090827

## Device Facts

- **Applicant:** Becton, Dickinson & CO
- **Product Code:** [LSL](/productcode/LSL.md)
- **Decision Date:** May 27, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3390
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K090827 · May 27, 2009 | BD PROBETEC NEISSERIA GONORRHOEAE (GC) QX AMPLIFIED DNA ASSAY | Becton, Dickinson & CO | Clinical specimens (endocervical and PreservCyt) collected from patients in routine clinical settings (family planning, OB/GYN, and STD clinics) | Clinical performance evaluation of the BD ProbeTec GC Qx Assay compared to a patient infected status (PIS) algorithm derived from reference NAAT methods. | Clinical performance; Routine clinical setting; Diagnostic accuracy |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Multi-site clinical evaluation of diagnostic performance | 2,074 female subjects attending family planning, OB/GYN, and sexually transmitted disease clinics; Sample Size: 2074; Number of Sites: 11 | Patient Infected Status (PIS) algorithm based on three reference NAAT methods | Sensitivity and specificity for detection of N. gonorrhoeae in PreservCyt specimens |

## Indications for Use

The BD ProbeTec™ Neisseria gonorrhoeae (GC) Q® Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Neisseria gonorrhoeae DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in PreservCyt® Solution using an aliquot that is removed prior to processing for additional gynecological testing. The assay is indicated for use with asymptomatic and symptomatic females and symptomatic males to aid in the diagnosis of gonococcal urogenital disease.

## Device Story

Assay uses Strand Displacement Amplification (SDA) to detect Neisseria gonorrhoeae DNA. Input: clinical specimens (swabs, urine, PreservCyt). Process: BD Viper System performs automated chemical lysis, DNA binding to para-magnetic particles, washing, and elution. Purified DNA is transferred to Priming Microwells (rehydration) then Amplification Microwells (enzymatic SDA). System monitors fluorescence via thermally-controlled readers. Output: qualitative result (positive/negative/EC failure) based on MaxRFU compared to threshold. Used in clinical labs; operated by technicians. Provides automated diagnostic aid for gonococcal urogenital disease; enables rapid, high-throughput testing.

## Clinical Evidence

Clinical study evaluated 2074 female subjects across 11 sites. Sensitivity and specificity calculated against a Patient Infected Status (PIS) algorithm (based on three reference NAATs). Overall sensitivity 95.3% (41/43), specificity 99.95% (2030/2031). Symptomatic sensitivity 100%, asymptomatic sensitivity 92.3%. Reproducibility study confirmed performance across three sites and systems with 100% correct results for target levels at/above LOD.

## Technological Characteristics

Strand Displacement Amplification (SDA) technology. Reagents: amplification primers, fluorescently-labeled detector probe, DNA polymerase, restriction endonuclease. Extraction: ferric oxide particles, magnetic separation. Instrument: BD Viper™ System (automated extraction and fluorescent reader). Connectivity: standalone system. Software: automated algorithm for signal processing and result reporting. Threshold: 125 MaxRFU.

## Regulatory Identification

Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identify Neisseria spp. from cultured isolates. Additionally, some of these reagents consist of Neisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence of Neisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genus Neisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.

## Predicate Devices

- BD ProbeTec™ Neisseria gonorrhoeae (GC) Qx Amplified DNA Assay ([K081825](/device/K081825.md))
- Gen-Probe APTIMA Assay for Neisseria gonorrhoeae (AGC) ([K062440](/device/K062440.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number: K090827

B. Purpose for Submission: Addition of the PreservCyt® Solution as a new specimen type taken from gynecological swab specimens stored in PreservCyt® Solution. The BD ProbeTec™ Neisseria gonorrhoeae (GC) Q® Amplified DNA Assay" and BD Viper™ System were previously cleared (K081825).

C. Measurand: Neisseria gonorrhoeae DNA

D. Type of Test: Qualitative determination of Neisseria gonorrhoeae DNA using the Strand Displacement Amplification technology

E. Applicant: BD Diagnostic System

F. Proprietary and Established Names: BD ProbeTec™ Neisseria gonorrhoeae (GC) Q® Amplified DNA Assay

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  LSL | Class II | 21CFR 866.3390 Neisseria spp. direct serological test reagents | Microbiology (83)  |

H. Intended Use:

The BD ProbeTec™ Neisseria gonorrhoeae (GC) Q® Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Neisseria gonorrhoeae DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in PreservCyt® Solution using an aliquot that is removed prior to processing for additional gynecological testing. The assay is indicated for use with asymptomatic and symptomatic females and symptomatic males to aid in the diagnosis of gonococcal urogenital disease.

3) Special conditions for use statement(s): For Prescription use only

4) Special instrument requirements: BD Viper™ System with automated nucleic acid extraction mode

{1}

I. Device Description: The BD ProbeTec GC Q® Amplified DNA Assay is based on the simultaneous amplification and detection of target DNA using amplification primers and a fluorescently-labeled detector probe. The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper™ System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, pre-warmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermally-controlled fluorescent readers. The presence or absence of N. gonorrhoeae DNA is determined by calculating the peak fluorescence (Maximum Relative Fluorescent Units (MaxRFU)) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.

In addition to the fluorescent probe used to detect amplified N. gonorrhoeae target DNA, a second labeled oligonucleotide is incorporated in each reaction. The Extraction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the N. gonorrhoeae -specific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is rehydrated upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper System and an automated algorithm is applied to both the EC and N. gonorrhoeae -specific signals to report results as positive, negative, or EC failure.

J. Substantial Equivalence Information:

a) Predicate device name (s):

BD ProbeTec™ Neisseria gonorrhoeae (GC) Q® Amplified DNA Assay Gen-Probe APTIMA Assay for Neisseria gonorrhoeae (AGC)

b) Predicate Numbers (s):

K081825

K062440

Comparison with predicate:

Device Comparison: GCQ Assay Specimen Collection and Processing on the BD Viper System in Extracted Mode

{2}

3

|   | BD ProbeTec GCQ Assay, PreservCyt Solution Specimens (Device) | BD ProbeTec GCQ Assay, Swab and Urine Specimens (K081825) | Gen-Probe AGC (K062440)  |
| --- | --- | --- | --- |
|  Specimen Types | • Same as K081825
• Gynecological specimen in PreservCyt® Solution | • Endocervical swab (females)
• Vaginal self-collected swab (in a clinical setting) (females)
• Urethral swab (males)
• Neat urine (female and male)
• UPT urine (female and male) | • Endocervical swab (females)
• Vaginal swab (females)
• Urethral swab (males)
• Neat urine (female and male)
• UTT urine (female and male)
• Gynecological specimen in PreservCyt® Solution  |
|  Specimen Collection and Transport Accessories | • Same as K081825
• Liquid Based Cytology Specimen (LBC) Dilution Tube | • Endocervical kit
• Urethral kit
• Vaginal kit
• UPT
• Neat urine (Qx Sample Tube) | • Unisex swab kit
• Vaginal swab kit
• Urine collection kit
• Specimen transfer kit (for gynecological specimen in PreservCyt® Solution)  |

## Device Comparison: Specimen Collection

|   | BD ProbeTec GCQ Assay, PreservCyt Solution Specimens (Device) | Gen-Probe AGC (K062440)  |
| --- | --- | --- |
|  Specimen Collection | • Gynecological specimen collected and placed in PreservCyt® Solution (per Cytyc Corporation’s instructions for use).
• Sample for CTQ/GCQ testing is drawn from original cytology specimen vial before the specimen is processed for cytology testing. | • Gynecological specimen collected and placed in PreservCyt® Solution (per Cytyc Corporation’s instructions for use).
• LBC specimen is first processed for cytology and then an aliquot is drawn from the remaining specimen in the vial for CT/GC testing.  |

## Device Comparison: Specimen Processing

|   | BD ProbeTec GCQ Assay, PreservCyt Solution Specimens (Device) | BD ProbeTec GCQ Assay, Swab and Urine Specimens (K081825)  |
| --- | --- | --- |
|  Specimen Processing | • Same as K081825 without the 15 minute pre-warm step for LBC Dilution Tube specimens
• Use of LBC Specimen Rack to prevent prewarming of LBC specimens | Pre-warm specimens (swabs and urines) for 15 minutes before running BD Viper System  |

## K. Standard/Guidance Document Referenced (if applicable):

1. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005. http://www.fda.gov/cdrh/ode/guidance/337.pdf
2. CLSI EP5-A2 “Evaluation of Precision Performance of Quantitative Measurement Methods”,

{3}

3. CLSI EP12-A “User Protocol for Evaluation of Qualitative Test Performance”.

L. Test Principle:

The BD ProbeTec GC Q⁰ Amplified DNA Assay is designed for use with the BD ProbeTec Chlamydia trachomatis/Neisseria gonorrhoeae (CT/GC) Q⁰ specimen collection and transport devices, applicable reagents, the BD Viper System and BD Fox Extraction. Specimens are collected and transported in their respective transport devices which preserve the integrity of the N. gonorrhoeae DNA over the specified ranges of temperature and time.

Urine and swab specimens undergo a pre-warm step in the BD Viper Lysing Heater to dissolve mucus and homogenize the specimen. After cooling, the specimens are loaded onto the BD Viper System which then performs all the steps involved in extraction and amplification of target DNA, without further user intervention. The specimen is transferred to an Extraction Tube that contains ferric oxide particles in a dissolvable film and dried Extraction Control. A high pH is used to lyse the bacterial cells and liberate their DNA into solution. Acid is then added to lower the pH and induce a positive charge on the ferric oxide, which in turn binds the negatively charged DNA. The particles and bound DNA are then pulled to the sides of the Extraction Tube by magnets and the treated specimen is aspirated to waste. The particles are washed and a high pH Elution Buffer is added to recover the purified DNA. Finally, a Neutralization Buffer is used to bring the pH of the extracted solution to the optimum for amplification of the target.

The BD ProbeTec GC Q⁰ Amplified DNA Assay is based on the simultaneous amplification and detection of target DNA using amplification primers and a fluorescently-labeled detector probe. The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, pre-warmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermally-controlled fluorescent readers. The presence or absence of N. gonorrhoeae DNA is determined by calculating the peak fluorescence (Maximum Relative Fluorescent Units (MaxRFU)) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.

In addition to the fluorescent probe used to detect amplified N. gonorrhoeae target DNA, a second fluorescently-labeled oligonucleotide is incorporated in each reaction. The Extraction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the N. gonorrhoeae-specific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is re-hydrated upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper instrument and an automated

4

{4}

algorithm is applied to both the EC and  $N$ . gonorrhoeae-specific signals to report specimen results as positive, negative, or EC failure.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

A reproducibility study of the BD Viper System using the BD ProbeTec GC  $\mathbf{Q}^{\mathrm{x}}$  Assay was evaluated for Liquid Based Cytology (LBC) specimens at three clinical sites on one BD Viper System per site. A panel of simulated specimens comprising CT and GC organisms seeded into LBC Specimen Dilution Tubes containing PreservCyt Solution was tested with the BD ProbeTec GC  $\mathbf{Q}^{\mathrm{x}}$  Assay. Uninoculated LBC Specimen Dilution Tubes containing PreservCyt Solution were used for the GC negative samples. Nine replicates of each panel member were tested every day for five days on each BD Viper System. The data are summarized in the table below.

Summary of Reproducibility Data for LBC Specimens on the BD Viper System for the GC  $\mathbf{Q}^{\mathrm{x}}$  Assay

|   |   |   |   |   | Within Run |   | Between Runs Within Site |   | Between Site  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  CT EB's/mL | GC Cells/mL | % Correct | 95% CI | Mean MaxRFU | SD | %CV | SD | %CV | SD | %CV  |
|  0 | 0 | 100.0% (135/135) | (97.3% - 100.0%) | 1.21 | 4.00 | 330.38 | 0.00 | 0.00 | 0.00 | 0.00  |
|  30 | 0 | 100.0% (135/135) | (97.3% - 100.0%) | 0.98 | 7.47 | 761.30 | 0.00 | 0.00 | 0.17 | 17.04  |
|  0 | 100 | 100.0% (135/135) | (97.3% - 100.0%) | 1982.77 | 83.92 | 4.23 | 0.00 | 0.00 | 0.00 | 0.00  |
|  30 | 250 | 100.0% (135/135) | (97.3% - 100.0%) | 1983.66 | 87.76 | 4.42 | 0.00 | 0.00 | 24.8 | 1.25  |
|  75 | 100 | 100.0% (135/135) | (97.3% - 100.0%) | 1920.14 | 81.94 | 4.27 | 59.45 | 3.10 | 0.00 | 0.00  |

Two additional levels were included in the panels to characterize the reproducibility of test results (i.e., proportion positive or negative) at target levels below the analytical Limit of Detection (LOD) of the BD ProbeTec GC  $\mathbf{Q}^{\mathrm{x}}$  Assay. These additional specimens comprised CT and GC organisms seeded into LBC Specimen Dilution Tubes containing PreservCyt Solution at dilutions of 1:10 and 1:100 of the respective analytical LODs of each analyte. These levels were selected to fall within the dynamic range of the analytical LOD curves for the BD ProbeTec CT  $\mathbf{Q}^{\mathrm{x}}$  and GC  $\mathbf{Q}^{\mathrm{x}}$  assays. Nine replicates of each panel member were tested every day for five days across the three BD Viper Systems. The data are summarized in the table below.

{5}

6

# Characterization of System Reproducibility at Target Levels below the Analytical Limit of Detection for the GC Q^x Assay for LBC Specimens

|  Dilution of Analytical LOD | % Positive | 95% CI (Positive) | MaxRFU Mean (Positive) | % Negative | 95% CI (Negative) | MaxRFU Mean (Negative)  |
| --- | --- | --- | --- | --- | --- | --- |
|  1:10 | 74.1 (100/135) | (65.8 - 81.2) | 1159.2 | 25.9 (35/135) | (18.8 - 34.2) | 21.2  |
|  1:100 | 8.9 (12/135) | (4.7 - 15.0) | 1136.5 | 91.1 (123/135) | (85.0 - 95.3) | 6.6  |

## b. Linearity/assay reportable range: NA

## c. Traceability, Stability, Expected values (controls, calibrators, or methods): A total of 142 positive and 142 negative assay controls were tested a sufficient number of times with acceptable results on all testing days at all testing sites. Additionally, a total of 71 processing controls were tested a sufficient number of times with acceptable results on all testing days at all testing sites.

## d. Detection limit: The Limits of Detection (LODs) for the GC Q^x Assay with Neisseria gonorrhoeae strain ATCC 19424 in PreservCyt specimens when extracted on the BD Viper System were determined to be ≤ 100 GC cells per mL. The GC Q^x Assay on the BD Viper System in extracted mode was able to detect 17 GC strains with ≥ 95% proportion positive at a concentration of 50 cells per mL in clean diluted PreservCyt Solution.

## e. Analytical specificity:

1. Cross Reactivity: Same as described in K081825 submission. Cross reactivity studies were not repeated for this submission.

2. Interference: The performance of the BD ProbeTec GC Q^x Assay on the BD Viper System in extracted mode was evaluated in the presence of potential interfering substances which may be encountered in swab, urine, and/or PreservCyt specimens. Potential interfering substances were spiked into UPT urine and vaginal swab specimen matrices as well as PreservCyt specimens in LBC Specimen Dilution Tubes, in both the presence and the absence of GC organisms (150 GC cells/mL in urine matrix and 300 GC cells/mL in swab/LBC Specimen Dilution Tube matrix). Results are summarized in the following table.

{6}

GC Q⁺ Interfering Substances.

|  Interpretation | Swab | Urine | PreservCyt  |
| --- | --- | --- | --- |
|  No Interference Observed | Blood (≤60%)
Seminal Fluid
Mucus
Over The Counter vaginal products and contraceptives
Hemorrhoidal cream
Prescription vaginal treatments
Leukocytes (1x10⁶ cells/mL)
1x10⁶ cells/mL Neisseria gonorrhoeae | Blood (≤1%)
Seminal fluid
Mucus
Antibiotics
Analgesics
Phenazopyridine
Over The Counter deodorant sprays and powders
Hormones
Leukocytes
Albumin <1 mg/mL
Glucose
Acidic urine (pH 4.0)
Alkaline urine (pH 9.0)
Bilirubin
1x10⁶ cells/mL Neisseria gonorrhoeae
Organisms associated with Urinary Tract Infections | Blood (≤1%)
Seminal Fluid
Mucus
Over The Counter vaginal products and contraceptives
Hemorrhoidal cream
Prescription vaginal treatments
Leukocytes (1x10⁶ cells/mL)
1x10⁶ cells/mL Neisseria gonorrhoeae  |
|  May cause extraction control (EC) failures | Blood (>60%) | Not applicable | Glacial Acetic Acid + Blood (≤5%/1% V/V)  |
|  May cause False Negative results | Not applicable | Not applicable | Glacial Acetic Acid + Blood (≤5%/1% V/V)  |

f. Assay cut-off: NA

g. Cross Contamination and Carryover: An internal study was conducted to evaluate the risk of producing a false positive result in either the same run on the BD Viper System in extracted mode (within run cross-contamination) or in a subsequent run (between run carryover). Testing was conducted using negative and positive samples on three BD Viper Systems. Negative samples consisted of CT/GC Q⁺ Swab Diluent/LBC Specimen Dilution Tube with PreservCyt Solution. Positive samples consisted of a representative analyte (10⁵ CT EB/mL) spiked into CT/GC Q⁺ Swab Diluent/LBC Specimen Dilution Tube with PreservCyt Solution. The overall rate of cross-contamination (i.e., with alternating columns of positive and negative samples and a prevalence of 50%) was 0.41% (9/2208) for the CT/GC Q⁺ Swab Diluent and 0.45% (5/1104) for the LBC Specimen Dilution Tube with PreservCyt Solution. The overall rate of carryover contamination (i.e., carryover between successive runs when the prevalence was 50% in the previous run) was 0.36% (8/2208) for the CT/GC Q⁺ Swab Diluent and 0.54% (6/1104) for the LBC Specimen Dilution Tube with PreservCyt Solution. Cross-contamination and carryover contamination across the three BD Viper Systems are summarized in the following table.

{7}

Cross Contamination and Carryover Contamination (PreservCyt)

|  Media Type | BD Viper System | Cross-Contamination |   |   | Carryover Contamination  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  n | Positive Results | Percent Positive | n | Positive Results | Percent Positive  |
|  PreservCyt Solution | 1 | 368 | 1 | 0.27 | 368 | 1 | 0.27  |
|   |  2 | 368 | 3 | 0.82 | 368 | 0 | 0.00  |
|   |  3 | 368 | 1 | 0.27 | 368 | 5 | 0.45  |
|   |  Overall | 1104 | 5 | 0.45 | 1104 | 6 | 0.54  |

2. Comparison studies:

a. Method comparison with predicate method: See below under Performance Characteristics
b. Matrix comparison: NA

3. Clinical studies:

a. Clinical Sensitivity: See under performance characteristics
b. Clinical specificity: See under performance characteristics

Performance Characteristics

PreservCyt Specimen Clinical Study

Endocervical swab specimens and PreservCyt specimens were collected from 2079 compliant female subjects attending family planning, OB/GYN, and sexually transmitted disease clinics at eleven geographically diverse clinical sites in North America. Subjects were classified as symptomatic if they reported symptoms such as dysuria, coital pain/difficulty/bleeding, abnormal vaginal discharge, or pelvic/uterine/adnexal pain. Subjects were classified as asymptomatic if they did not report symptoms. Two subjects were excluded due to an undetermined patient infected status. Three subjects did not have a PreservCyt specimen result. Therefore there were 2074 subjects evaluated.

Three randomized endocervical swab specimens and a PreservCyt specimen were collected from each female subject. The three reference endocervical swabs were tested with the BD ProbeTec ET CT/GC/AC assay, the BD ProbeTec GC Q⁰ assay, and another commercially available NAAT (Nucleic Acid Amplification Test). Sensitivity and specificity for PreservCyt specimens were calculated by comparing results to a patient infected status (PIS) algorithm. The designation of positive or negative PIS was based on the endocervical swab specimen results from the three reference methods. At least two positive reference results were required to establish a subject as PIS-positive. At least two negative reference results were required to establish a subject as PIS-negative. Data are presented in the following tables.

{8}

Summary of Cervical Sampling Devices Used in the PreservCyt Specimen Clinical Study

|  Cervical Sampling Device Used | Clinical Collection Site Number |   |   |   |   |   |   |   |   |   |   | Total  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12  |   |
|  Broom-Type Device | 89 | 0 | 0 | 45 | 16 | 464 | 272 | 83 | 0 | 99 | 0 | 1068  |
|  Spatula/Cytobrush | 74 | 154 | 95 | 0 | 0 | 52 | 0 | 209 | 282 | 0 | 145 | 1011  |

GC Q* Assay Performance for PreservCyt Specimens Compared to Patient Infected Status (by symptomatic status)

|   |   | Performance Compared to Patient Infected Status |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Symptomatic Status | n | Sensitivity | 95% C.I. | Specificity | 95% C.I. | PPV% | NPV% | Error Initial/Final  |
|  A | 1349 | 92.3% (24/26) | (74.9% - 99.1%) | 100.0% (1323/1323) | (99.7% - 100.0%) | 100% | 99.9% | 1/0  |
|  S | 725 | 100.0% (17/17) | (80.5% - 100.0%) | 99.9% (707/708) | (99.2% - 100.0%) | 95.9% | 100% | 0/0  |
|  Total | 2074 | 95.3% (41/43) | (84.2% - 99.4%) | 99.95% (2030/2031) | (99.7% - 100.0%) | 100% | 99.9% | 1/0  |

GC Q* Assay Performance for PreservCyt Specimens Compared to Patient Infected Status (by clinical site)

|   |   |   | Performance Compared to Patient Infected Status  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Collection Site | Prevalence | n | Sensitivity | 95% C.I. | Specificity | 95% C.I. | # CT (+) and GC (+) | PPV % | NPV %  |
|  1 | 5.5% | 163 | 88.9% (8/9) | (51.8% - 99.7%) | 100.0% (154/154) | (97.6% - 100.0%) | 5 | 100.0 % | 99.4%  |
|  2 | 5.2% | 154 | 100.0% (8/8) | (63.1% - 100.0%) | 99.3% (145/146) | (96.2% - 100.0%) | 1 | 88.7% | 100.0 %  |
|  3 | 3.2% | 95 | 100.0% (3/3) | (29.2% - 100.0%) | 100.0% (92/92) | (96.1% - 100.0%) | 2 | 100.0 % | 100.0 %  |
|  4 | 13.3% | 45 | 100.0% (6/6) | (54.1% - 100.0%) | 100.0% (39/39) | (91.0% - 100.0%) | 2 | 100.0 % | 100.0 %  |
|  5 | 0.0% | 16 | NA | NA | 100.0% (16/16) | (79.4% - 100.0%) | 0 | NA | NA  |
|  6 | 1.6% | 516 | 100.0% (8/8) | (63.1% - 100.0%) | 100.0% (508/508) | (99.3% - 100.0%) | 2 | 100.0 % | 100.0 %  |
|  7 | 2.9% | 272 | 87.5% (7/8) | (47.3% - 99.7%) | 100.0% (264/264) | (98.6% - 100.0%) | 3 | 100.0 % | 99.6%  |
|  8 | 0.0% | 292 | NA | NA | 100.0% (292/292) | (98.7% - 100.0%) | 0 | NA | NA  |

{9}

10

|   |   |   | Performance Compared to Patient Infected Status  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Collection Site | Prevalence | n | Sensitivity | 95% C.I. | Specificity | 95% C.I. | # CT (+) and GC (+) | PPV % | NPV %  |
|  9 | 0.0% | 282 | NA | NA | 100.0% (282/282) | (98.7% - 100.0%) | 0 | NA | NA  |
|  10 | 0.0% | 97 | NA | NA | 100.0% (97/97) | (96.3% - 100.0%) | 0 | NA | NA  |
|  11 | 0.7% | 142 | 100.0% (1/1) | (2.5% - 100.0%) | 100.0% (141/141) | (97.4% - 100.0%) | 0 | 100.0 % | 100.0 %  |

GC Q⁵ Assay Performance for PreservCyt Specimens Compared to Patient Infected Status (by clinic type)

|   |   |   | Performance Compared to Patient Infected Status  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Clinic Type | Prevalence | n | Sensitivity | 95% C.I. | Specificity | 95% C.I. | PPV % | NPV %  |
|  Family Planning | 0.7% | 1187 | 100.0% (8/8) | (63.1% - 100.0%) | 100.0% (1179/1179) | (99.7% - 100.0%) | 100.0 % | 100.0 %  |
|  OB/GYN | 3.0% | 367 | 90.9% (10/11) | (58.7% - 99.8%) | 100.0% (356/356) | (99.0% - 100.0%) | 100.0 % | 99.7%  |
|  STD | 4.6% | 520 | 95.8% (23/24) | (78.9% - 99.9%) | 99.8% (495/496) | (98.9% - 100.0%) | 95.9% | 99.8%  |

Analysis of GC Positive/Negative PreservCyt Specimens Based on Patient Infected Status

|   |   |   |   |   | Symptomatic Status |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  PIS GC | AC2 Swab | ProbeTec Swab | Viper Swab | PreservCyt | A | S | Table 13C: Total  |
|  + | NA | + | + | + | 1 | 3 | 4  |
|   |  + | - | + | - | 1 | 0 | 1  |
|   |  + | - | + | + | 1 | 0 | 1  |
|   |  + | + | NA | + | 1 | 0 | 1  |
|   |  + | + | + | - | 1 | 0 | 1  |
|   |  + | + | + | + | 21 | 14 | 35  |
|  Total PIS Positive |   |   |   |   | 26 | 17 | 43  |
|  - | NA | - | - | - | 181 | 79 | 260  |
|   |  - | I | - | - | 1 | 0 | 1  |
|   |  - | - | NA | - | 3 | 0 | 3  |
|   |  - | - | LE | - | 2 | 0 | 2  |
|   |  - | - | - | - | 1129 | 624 | 1753  |
|   |  - | - | - | + | 0 | 1 | 1  |
|   |  - | - | + | - | 2 | 0 | 2  |
|   |  - | + | - | - | 4 | 3 | 7  |
|   |  + | - | - | - | 1 | 1 | 2  |
|  Total PIS Negative |   |   |   |   | 1323 | 708 | 2031  |

{10}

4. Clinical cut-off: Clinical cut-off was validated during the swab and urine clinical trial (K081825). The cutoff was 125 MaxRFU (&gt; 125 MaxRFU = positive; &lt; 125 MaxRFU = negative). The same cutoff was utilized for the clinical trial to validate the performance of the PreservCyt Solution specimens when tested with the GCQ assay on the BD Viper System in extracted mode.

# 5. Expected values/Reference range:

A. Prevalence: The prevalence of positive  $N$  gonorrhoeae specimens in patient populations depends upon: clinic type, age, risk factors, gender, and test method. The prevalence observed with the GC  $Q^{\mathrm{x}}$  Assay during a multi-center clinical trial for PreservCyt specimens ranged from 0.0% to 13.3% based on the data presented in the table "GC  $Q^{\mathrm{x}}$  Assay Performance for PreservCyt Specimens Compared to Patient Infected Status (by clinical site)" under performance.

B. Positive and Negative Predictive Value: Hypothetical positive and negative predictive values (PPV &amp; NPV) for the GC  $Q^{\mathrm{x}}$  Assay from the multi-center clinical trial for PreservCyt specimens are shown in the following table. These calculations are based on hypothetical prevalence and overall sensitivity and specificity (compared to the patient infected status) of  $95.3\%$  and  $99.95\%$  for PreservCyt specimens.

GC Hypothetical Positive and Negative Predictive Values (PreservCyt) Compared to Patient Infected Status

|  Prevalence (%) | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%)  |
| --- | --- | --- | --- | --- |
|  2 | 95.3 | 99.95 | 97.5 | 99.9  |
|  5 | 95.3 | 99.95 | 99.0 | 99.8  |
|  10 | 95.3 | 99.95 | 99.5 | 99.5  |
|  20 | 95.3 | 99.95 | 99.8 | 98.8  |
|  30 | 95.3 | 99.95 | 99.9 | 98.0  |
|  40 | 95.3 | 99.95 | 99.9 | 97.0  |
|  50 | 95.3 | 99.95 | 99.9 | 95.5  |

N. Instrument: Same as described in K081825.

O. System Descriptions: Same as described in K081825

P. Other Supportive Device and Instrument Information: NA

Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K090827](https://fda-staging.innolitics.com/device/K090827)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com