K090793 · Bioalpha, Inc. · LYC · May 19, 2009 · Dental
Device Facts
Record ID
K090793
Device Name
BONGROS HA
Applicant
Bioalpha, Inc.
Product Code
LYC · Dental
Decision Date
May 19, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Bongros®-HA (HAGS, HAGM, HAGL) is intended for use in dental surgery. The products may be used in surgical procedures such as: · Augmentation or reconstructive treatment of the alveolar ridge · Filling of extraction sockets to enhance preservation of the alveolar ridge · Elevation of the maxillary sinus floor · Filling of peri-implant defects in conjunction with products intended for guided bone regeneration.
Device Story
Bongros®-HA is a synthetic bone grafting material composed of highly pure hydroxyapatite (Ca10(PO4)6(OH)2). It features a trabecular, 3D interconnected pore structure mimicking human cancellous bone. Used in dental surgery by clinicians, the material is placed in direct contact with viable bone; it acts as a scaffold where hard tissue infiltrates the reticulated spaces. Bone formation occurs in apposition to the surface and within the interstices of the implant skeleton. The device is provided sterile for single use and is available in three size models (HAGS, HAGM, HAGL). It aids in alveolar ridge preservation, sinus floor elevation, and defect filling, facilitating bone regeneration in dental procedures.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use comparisons to legally marketed predicate devices.
Technological Characteristics
Synthetic hydroxyapatite (Ca10(PO4)6(OH)2); trabecular 3D interconnected pore structure; sterile, single-use; available in three sizes (HAGS, HAGM, HAGL).
Indications for Use
Indicated for dental surgery patients requiring alveolar ridge augmentation/reconstruction, extraction socket filling for ridge preservation, maxillary sinus floor elevation, or filling of peri-implant defects in conjunction with guided bone regeneration products.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a sequence of handwritten characters, which appear to be a combination of letters and numbers. The sequence reads as 'K090793'. The characters are written in a cursive style, with some connections between the letters and numbers. The image is in black and white, and the characters are clearly visible against the background.
# 510(k) SUMMARY
MAY 19 2009
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
【
## Submitter:
| Name: | BioAlpha Inc. |
|----------|--------------------------------------------------------------------------|
| Address: | 223-23 Sangdaewon-Dong, Jungwon-Gu, Seongnam, Gyeonggi Do, 462-120 Korea |
### US Agent:
| Name: | Alex Chang |
|------------|-------------------------------------------------------------|
| Address: | Bionet America Inc. 2691 Dow Ave. Suite B, Tustin, CA 92780 |
| Phone No.: | 714-838-1274, 949-468-6312 |
| Fax No.: | 714-734-1761 |
| E-Mail: | achang@bionetus.com |
## Official Correspondent:
| Name: | Hyun-Seung Ryu, Ph.D |
|------------|-----------------------------------------------------------------------------|
| Address: | 223-23 Sangdaewon-Dong, Jungwon-Gu, Seongnam, Gyeonggi<br>Do, 462-120 Korea |
| Phone No.: | +82-31-746-5209 |
| Fax No.: | +82-31-749-5208 |
| E-Mail: | hsryu@daewoong.co.kr |
### Device Identification
| Proprietary Name: | Bongros®-HA |
|----------------------|----------------------------------------------|
| Common/Usual Name: | Bone grafting material |
| Classification Name: | Bone Grafting Material, Synthetic , 872.3930 |
| Product Code: | LYC |
| Review Panel: | Dental |
### Substantially Equivalent Predicate Legally Marketed Devices
The subject device is deemed to be substantially equivalent to those following devices manufactured and currently available in commercial distribution.
| Device Name | Osteograf | EndoBon | SynOss | Bio-Oss |
|---------------|-----------|---------|---------|---------|
| 510(k) Number | K981182 | K980679 | K072397 | K952619 |
| Decision | SE | SE | SE | SE |
| Product Code | LYC | LYC | LYC | LYC |
{1}------------------------------------------------
#### Device Description
Bongros®-HA is made of highly pure, synthetic hydroxyapatite (Ca10(PO4)6(OH)2). Bongros®-HA has trabecular structure that resembles the 3-dimensional interconnected pore structure of human cancellous bone. When Bongros -HA is placed in direct contact with viable bone, the reticulated spaces in the implant are infiltrated with hard tissue. Bone formation occurs in apposition to the Bongros® HA surface and within the interstices of the implant skeleton.
Bongros -HA have three specific models which are HAGS, HAGM and HAGL. The quality of the material of HAGS, HAGM and HAGL is equal but only the sizes of each model are different from each other.
Therewith, Bongros®-HA is provided sterile for single use.
#### Indications for Use
Bongros -HA (HAGS, HAGM, HAGL) is intended for use in dental surgery.
The products may be used in surgical procedures such as:
- · Augmentation or reconstructive treatment of the alveolar ridge
- · Filling of extraction sockets to enhance preservation of the alveolar ridge
- · Elevation of the maxillary sinus floor
- · Filling of peri-implant defects in conjunction with products intended for guided bone regeneration.
## Technological Characteristics and Substantial Equivalence
Bongros®-HA is substantially equivalent to predicate devices. Bongros®-HA and predicate devices are identical in intended use, indication and application. Therewith, Bongros -HA and predicate devices are biocompatible and have similar properties.
Based on the discussion above, BioAlpha Inc. believes that Bongros® HA is substantially equivalent.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling three overlapping human figures or ribbons, positioned to the right. To the left of the symbol is a circular seal containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.
Public Health Service
19 2009 MAY
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Hyun-Seung Ryu, Ph.D Director BioAlpha, Incorporated 223-23 Sangdaewon-Dong, Jungwon-Gu Seongnam, Gyeonggi-Do 462-120 KOREA
Re: K090793
Trade/Device Name: Bongros -HA Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: March 19, 2009 Received: March 24, 2009
Dear Dr. Ryu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2- Dr. Ryu
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Rummer
Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K090793 Device Name: Bongros -HA
Indications for Use:
Bongros®-HA (HAGS, HAGM, HAGL) is intended for use in dental surgery.
The products may be used in surgical procedures such as:
· Augmentation or reconstructive treatment of the alveolar ridge
· Filling of extraction sockets to enhance preservation of the alveolar ridge
· Elevation of the maxillary sinus floor
· Filling of peri-implant defects in conjunction with products intended for guided bone regeneration.
Rin Maly for MSR
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K090793
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.