PERPOS FCD-2 SYSTEM (SINGLE PATIENT USE), ANCHOR, STABILIZER

K090767 · Interventional Spine, Inc. · MRW · Jun 11, 2009 · OR

Device Facts

Record IDK090767
Device NamePERPOS FCD-2 SYSTEM (SINGLE PATIENT USE), ANCHOR, STABILIZER
ApplicantInterventional Spine, Inc.
Product CodeMRW · OR
Decision DateJun 11, 2009
DecisionSESU
Submission TypeSpecial
Device ClassClass U
AttributesTherapeutic

Indications for Use

The PERPOS™ FCD-2 is indicated for spondylolisthesis, spondylolysis, degenerative disc disease (DDD) as defined by back pain of discogenic origin as confirmed by radiographic studies, degeneration of the facets with instability and fracture. pseudoarthrosis, or failed previous fusion. The intended use of the PERPOS™ FCD-2 is to stabilize the spine as an aid to fusion through bilateral immobilization of the facet joints. The PERPOS FCD-2 may be used to supplement legally marketed anterior fusion products in order to create an anterior fixation fixation construction as an aid to fusion. The screws are inserted posteriorly through the superior side of the facet, across the facet joint and into the pedicle. The PERPOS™ FCD-2 is intended for lumbar bilateral facet fixation, with bone graft, at single or multiple levels from L1 to S1. The intended use of the associated manual surgical instruments is to aid in the implantation of the PERPOS™ FCD-2 Anchor and Stabilizer.

Device Story

PERPOS™ FCD-2 System is a single-patient-use spinal stabilization device; consists of double-helix facet screws with compression-locking collars, retaining rings, and self-retaining polymer washers; includes associated manual surgical instruments. Used by surgeons to stabilize the spine as an aid to fusion via bilateral immobilization of facet joints; screws inserted posteriorly through superior side of facet, across joint, into pedicle. May supplement anterior fusion products for anterior/posterior fixation. Benefits include spinal stabilization to facilitate fusion in patients with degenerative conditions or failed prior surgery.

Clinical Evidence

No clinical data. Substantial equivalence supported by design controls, risk analysis, and design verification/validation testing.

Technological Characteristics

System comprises double-helix facet screws, compression-locking collars, retaining rings, and polymer washers. Manual surgical instruments provided for implantation. Device is for single-patient use. No software or electronic components.

Indications for Use

Indicated for patients with spondylolisthesis, spondylolysis, degenerative disc disease (discogenic back pain), facet degeneration with instability/fracture, pseudoarthrosis, or failed previous fusion requiring lumbar bilateral facet fixation (L1-S1) as an aid to fusion.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 1690767 ### SECTION 5 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Special 510(k) ## SUBMITTER INFORMATION Company Name: Company Address: Company Phone: Company Facsimile: Interventional Spine, Inc 13700 Alton Parkway Suite 160 Irvine, CA 92618 (949) 472-0006 (949) 472-0016 Carol Emerson JUN 1 1 2009 DEVICE IDENTIFICATION Contact Person: Trade Name: Common Name: Classification Name: Product Code: Device Panel: Device Class: PERPOS™ FCD-2 System, Single Patient Use PERPOS™ FCD-2 Implant, ANCHOR AND STABILIZER Facet Screw and associated manual surgical instruments Unclassified, various manual surgical instruments MRW Orthopedic, General, and Plastic Surgery # PREDICATE DEVICES PERPOS™ PLS System previously cleared under 510(k) K082795 #### DEVICE DESCRIPTION The PERPOS™ FCD-2 Single Use System consists of two each of a double-helix facet screw with a compression-locking collar, retaining ring and a self-retaining polymer washer packaged in a single use tray with the associated manual instruments. Modifications from the predicate device include changes to the washer configuration, instrumentation and packaging. Unclassified Interventional Spine, Inc. Proprietary and Confidential {1}------------------------------------------------ ### INTENDED USE/INDICATIONS for USE The PERPOS™ FCD-2 System isindicated for spondylolisthesis, spondylolysis, degenerative disc disease (DDD) as defined by back pain of discogenic origin as confirmed by rediographic studies, degeneration of the facets with instability and fracture, pseudoarthrosis, or failed previous fusion. The intended use of the PERPOS™ FCD-2 Anchor and Stabilizer is to stabilize the spine as an aid to fusion through bilateral immobilization of the facet joints. The PERPOS™ FCD-2 may be used to supplement legally marketed anterior fusion products in order to create an anterior/posterior fixation construction as an aid to fusion. The PERPOS FCD-2 is indicated for bilateral facet fixation, with or without bone graft, at single or multiple levels from L1-S1. The intended use of the associated manual surgical instruments is to aid in the implantation of the PERPOS™ FCD-2 Anchor and Stabilizer. ### TECHNOLOGICAL CHARACTERISTICS and SUBSTANTIAL EQUIVALENCE Documentation was provided to demonstrate that the PERPOS™ FCD-2 System is identical or similar to the predicate device in technological characteristics. The PERPOS™ FCD-2 System is identical to the predicate device in intended use and identical or substantially equivalent to the predicate device in materials, design and technological characteristics. #### PERFORMANCE TESTING Risk analysis was conducted on the impact of the changes and appropriate design verification and validation was conducted under the company's design controls. #### CONCLUSION The results from design controls and the information provided in this submission support the conclusion that the PERPOS™ FCD-2 System is substantially equivalent-to-the-predicate-PERPOS™ PLS-System- {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring three stylized, overlapping shapes that resemble birds in flight. The emblem is positioned in the upper right portion of the circle. #### Public Health Service JUN 11 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Interventional Spine, Incorporated % Ms. Carol Emerson VP, Quality and Logistics 13700 Alton Parkway, Suite 160 Irvine, California 92618 K090767 Re: 11:21:3 Trade/Device Name: PerPos FCD-2 System Regulatory Name/Number: Unclassified Regulatory Class: Unclassified Product Code: MRW Dated: May 4, 2009 Received: May 12, 2009 Dear Ms. Emerson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling: The safety and effectiveness of this device for the indication of spinal stabilization without fusion have not been established. Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. {3}------------------------------------------------ Page 2-Ms. Carol Emerson The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to vour labeling. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/. {4}------------------------------------------------ Page 3-Ms. Carol Emerson You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Christy Foreman for Donna-Bea Tillman, Ph.D., M.P.A. Director Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### INDICATIONS FOR USE STATEMENT 510(k) Number (if known): K090767 Device Name: Single Use PERPOS™ FCD-2 System #### Indications For Use: The PERPOS™ FCD-2 is indicated for spondylolisthesis, spondylolysis, degenerative disc disease (DDD) as defined by back pain of discogenic origin as confirmed by radiographic studies, degeneration of the facets with instability and fracture. pseudoarthrosis, or failed previous fusion. The intended use of the PERPOS™ FCD-2 is to stabilize the spine as an aid to fusion through bilateral immobilization of the facet joints. The PERPOS FCD-2 may be used to supplement legally marketed anterior fusion products in order to create an anterior fixation fixation construction as an aid to fusion. The screws are inserted posteriorly through the superior side of the facet, across the facet joint and into the pedicle. The PERPOS™ FCD-2 is intended for lumbar bilateral facet fixation, with bone graft, at single or multiple levels from L1 to S1. The intended use of the associated manual surgical instruments is to aid in the implantation of the PERPOS™ FCD-2 Anchor and Stabilizer. Prescription Use ਮ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) E c.(Division, or Office) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices K090767 510(k) Number_ Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...