RESTORIS MCK is indicated for single or multi-compartmental knee replacement used in conjunction with RIO, the Robotic Arm Interactive Orthopedic System, in individuals with osteoarthritis or posttraumatic arthritis of the tibiofemoral and/or patellofemoral articular surfaces. The specific knee replacement configurations include: - Medial unicondylar . - . Lateral unicondylar - Patellofemoral - Medial bi-compartmental (medial unicondylar and patellofemoral) RESTORIS MCK is for single use only and is intended for implantation with bone cement.
Device Story
Restoris MCK is a modular knee prosthesis system comprising unicompartmental (Uni) and patellofemoral (PF) components. Used exclusively with MAKO Robotic Arm Interactive Orthopedic System (RIO) for surgical planning and execution. Components include femoral condyles, tibial inlays/baseplates, patellofemoral components, and patella components. Materials include CoCrMo alloy, UHMWPE, and Ti6Al4V ELI. System allows for medial/lateral unicondylar, patellofemoral, or medial bi-compartmental replacements. Instrumentation (sizers, templates, trials, drills) provided for use with RIO. Intended for cemented implantation. Benefits include precise component placement via robotic assistance, potentially improving alignment and patient outcomes in knee arthroplasty.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility testing per recognized ASTM standards and mechanical performance verification of the implant components and instrumentation.
Indicated for individuals with osteoarthritis or post-traumatic arthritis of tibiofemoral and/or patellofemoral articular surfaces requiring single or multi-compartment knee replacement. Configurations include medial/lateral unicondylar, patellofemoral, or medial bi-compartmental replacement. Must be used with RIO Robotic Arm Interactive Orthopedic System and bone cement.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold letters, with the words "SURGICAL CORP" in smaller letters underneath. Above the word "MAKO" is a stylized graphic that resembles a mountain range. The text "KD9" is written in the upper right corner of the image.
JUN 1 7 2009
2555 Davie Road . Et. Lauderdale, El, 33317. . . Phone 954,927,2044 . Fax 954,927,0446 . . www.makosurgical.com ...
### 510(K) SUMMARY
| Submitter: | MAKO Surgical Corp. |
|----------------------------|---------------------------------------------------------------------------------------------------------------------|
| Address: | 2555 Davie Road, Fort Lauderdale, FL 33317 |
| Phone number / Fax Number: | (Ph) 954-927-2044 x 605; (F) 954-927-0446 |
| Contact Person: | William F. Tapia |
| Date Prepared: | March 18, 2009 |
| Proprietary Name: | Restoris MultiCompartmental Knee System (Restoris MCK) |
| Common Name: | Compartmental Knee Prosthesis System |
| Classification: | Class II |
| Product Code / #: | KRR - Knee joint patellofemoral polymer/metal semi-constrained cemented<br>prosthesis; 21 CFR 888.3540 |
| | HSX - Knee joint femorotibial metal/polymer non-constrained cemented<br>prosthesis; 21 CFR 888.3520 |
| | NPJ - Knee joint patellofemorotibial polymer/metal/polymer semi-constrained<br>cemented prosthesis; 21 CFR 888.3560 |
| | HRY - Knee joint femorotibial metal/polymer semi-constrained cemented<br>prosthesis; 21 CFR 888.3530 |
Description: The Restoris MCK is composed of a unicompartmental implant system (Restoris MCK Uni) and a patellofemoral implant system (Restoris MCK PF). Restoris MCK Uni and Restoris MCK PF may be used in various combinations to create: a single unicompartmental femorotibial replacement for the medial or lateral side of the knee; a patellofemoral replacement; or a bicompartmental patellofemorotibial replacement for the medial side of the knee. Restoris MCK Uni is designed for use when load bearing ROM is expected to be less than or equal to 155 degrees.
| Feature | Restoris MCK |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Implant Components | Restoris MCK Uni Femoral Condyle component Tibial Inlay component Radiographic marker in Tibial Inlay component Tibial Onlay Insert component Tibial Baseplate |
| | Restoris MCK PF Patellofemoral component Patella component |
| Sizes | Restoris MCK Uni Femoral Condyle components are available in 8 sizes. Tibial components are available in 8 sizes. Restoris MCK PF Patellofemoral components available in 8 sizes. Patella components are available in 6 sizes. |
| Materials | Restoris MCK Uni Femoral Condyle component – CoCrMo alloy Tibia Inlay component – UHMWPE Radiographic marker in Tibial Inlay component – Unalloyed titanium wire Tibia Onlay Insert component – UHMWPE Tibial Baseplate – Ti6Al4V ELI alloy Restoris MCK PF Patellofemoral component – CoCrMo alloy Patella component – UHMWPE |
| Instrumentation | Provided separately in a re-usable/sterilizable tray. Tray includes various tools<br>(e.g., sizers, templates, trials, drills, gauges, impactors, inserters, extractors) used<br>during surgery. Restoris MCK can only be used with the MAKO Robotic Arm<br>Interactive Orthopedic System (RIO) |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold letters, with the words "SURGICAL CORP." in smaller letters underneath. Above the word "MAKO" is a stylized image of a mountain range. The mountain range is shaded with a gradient, with the top of the mountains being lighter than the bottom.
2555 Davie Road_ * . Et. Lauderdale, FL 33317 . • Phone 954.927.0446 . • Fax 954.927.0446 . • www.makosurgical.com...
| Sterilization and Packaging | Sterilization:<br>o All components – gamma radiation<br>o Instrumentation – steam sterilization<br>Packaging:<br>o All components are supplied in double sealed containers maintaining double<br>sterile barriers. |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | All components are made of materials for surgical implant applications per<br>recognized ASTM standards. |
Substantial Equivalence: Restoris MCK is substantially equivalent to the MAKO Modular Knee System (K082172).
Indications for Use:
RESTORIS MCK is indicated for single or multi-compartment used in conjunction with RIO, the Robotic Arm Interactive Orthopedic System, in individuals with osteoarthritis or post-traumatic arthritis of the tibiofemoral and/or patellofemoral articular surfaces. The specific knee replacement configurations include:
- Medial unicondylar · .
- Lateral unicondylar .
- Patellofemoral .
- Medial bi-compartmental (medial unicondylar and patellofemoral) .
RESTORIS MCK is for single use only and is intended for implantation with bone cement.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 1 7 2009
MAKO Surgical Corporation % Mr. William F. Tapia 2555 Davie Road Fort Lauderdale, Florida 33317
Re: K090763
Trade/Device Name: Restoris MultiCompartmental Knee System (Restoris MCK) Regulation Number: 21 CFR 888.3560
Regulation Name: Knee ioint patellofemorotibial polymer/metal/polymer semiconstrained cemented prosthesis
Regulatory Class: Class II Product Code: NPJ, KRR, HSX, HRY Dated: March 17, 2009 Received: March 23, 2009
Dear Mr. Tapia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish farther announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting
{3}------------------------------------------------
Page 2-Mr. William F. Tapia
(reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Barbara Buehr
Mark N. Mellman
Mark N. Melkerson Director . Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold letters, with the words "SURGICAL CORP" in smaller letters underneath. Above the word "MAKO" is a stylized graphic that resembles a shark fin or a mountain peak. The logo is simple and modern, and it conveys a sense of strength and precision.
# 2555 Davie Road
### INDICATIONS FOR USE
510(k) Number (if known): K090763
Device Name: Restoris MultiCompartmental Knee System (Restoris MCK)
Indications for Use:
RESTORIS MCK is indicated for single or multi-compartmental knee replacement used in conjunction with RIO, the Robotic Arm Interactive Orthopedic System, in individuals with osteoarthritis or posttraumatic arthritis of the tibiofemoral and/or patellofemoral articular surfaces. The specific knee replacement configurations include:
- Medial unicondylar .
- . Lateral unicondylar
- Patellofemoral
- Medial bi-compartmental (medial unicondylar and patellofemoral)
RESTORIS MCK is for single use only and is intended for implantation with bone cement.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart.C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Oputa
(Division Sign-Off)
(Division Sign-6 (Division Sign-Ord), Orthopedic, Divisionative Devices
510(k) Number K090763
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.