HEMOGLOBIN CAPILLARY COLLECTION SYSTEM (HCCS), 1 TEST, HEMOGLOBIN CAPILLARY COLLECTION SYSTEM (HCCS), 5 TESTS, HEMOGLOBI

K090737 · Bio-Rad Laboratories, Inc. · JKA · Jun 3, 2009 · Clinical Chemistry

Device Facts

Record IDK090737
Device NameHEMOGLOBIN CAPILLARY COLLECTION SYSTEM (HCCS), 1 TEST, HEMOGLOBIN CAPILLARY COLLECTION SYSTEM (HCCS), 5 TESTS, HEMOGLOBI
ApplicantBio-Rad Laboratories, Inc.
Product CodeJKA · Clinical Chemistry
Decision DateJun 3, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Indications for Use

The Hemoglobin Capillary Collection System (HCCS) is intended for the collection of human whole blood for the percentage determination of hemoglobin A1c using Bio-Rad HPLC methods. For In Vitro diagnostic use The measurement of percent hemoglobin A1c is effective in monitoring long-term glucose control in individuals with diabetes mellitus.

Device Story

HCCS facilitates collection/preparation of capillary or venous whole blood for HbA1c analysis; system includes microvials with reagent (EDTA/potassium cyanide), plastic capillaries, capillary holder, and labels. Professional users collect blood via capillary; sample is diluted in reagent vial at time of collection. Prepared samples are stable for transport/storage (4 days at 42°C, 2 weeks at 15-30°C, or 4 weeks at 2-8°C). Samples are analyzed using Bio-Rad VARIANT II or D-10 Hemoglobin A1c Programs (HPLC). Output is percent HbA1c; used by clinicians to monitor long-term glucose control in diabetic patients.

Clinical Evidence

Method comparison study conducted at an external site. Compared HbA1c values from capillary samples (HCCS) vs. venous whole blood samples (predicates). Results showed no significant difference in HbA1c values between HCCS-prepared samples and predicate-diluted samples.

Technological Characteristics

Components: 1.5 mL microvials (11mm x 40mm) with pierceable caps, 5 µL plastic capillaries, capillary holder. Reagent: Aqueous solution of EDTA and potassium cyanide (0.25 mmol/L). Principle: Hemolysis and sample preservation for HPLC analysis. Shelf life: 24 months at room temperature.

Indications for Use

Indicated for monitoring long-term glucose control in individuals with diabetes mellitus via measurement of percent hemoglobin A1c.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k090737 B. Purpose for Submission: New device – a whole blood, capillary and venous sample collection device for glycosylated hemoglobin C. Measurand: Glycosylated Hemoglobin (HbA1c) D. Type of Test: Quantitative, High Performance Liquid Chromatography (HPLC) E. Applicant: Bio-Rad Laboratories, Inc F. Proprietary and Established Names: Hemoglobin Capillary Collection System (HCCS) G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | LCP – Glycosylated Hemoglobin assay | II | 864.7470 | 81, Hematology | | JKA – Tubes, vials, systems, serum separators, blood collection | II | 862.1675 | 75, Chemistry | H. Intended Use: 1. Intended use(s): See indications for use below {1} 2. Indication(s) for use: The Hemoglobin Capillary Collection System (HCCS) is intended for the collection of human whole blood for the percentage determination of hemoglobin A1c using Bio-Rad HPLC methods. For In Vitro diagnostic use The measurement of percent hemoglobin A1c is effective in monitoring long-term glucose control in individuals with diabetes mellitus. 3. Special conditions for use statement(s): For prescription use 4. Special instrument requirements: Bio-Rad Variant II Hemoglobin A1c program and D-10 Hemoglobin A1c program I. Device Description: Each hemoglobin Capillary Collection System (HCCS) contains a combination of the following components: Sample Preparation Vials – clear microvials with blue pierceable caps, each contains 1.5 mL of HCCS reagent (aqueous solution of EDTA and potassium cyanide (.25mmol/L). The microvials are 11 mm x 40 mm and have a maximum volume of 2.0 mL. Capillaries – plastic capillaries (5 µL) in a dispenser Capillary holder – holder for manipulating the capillaries Labels – to label prepared samples J. Substantial Equivalence Information: 1. Predicate device name(s): VARIANT II Hemoglobin A1c Program and D-10 Hemoglobin A1c Program 2. Predicate 510(k) number(s): k070452 and k031043 respectively {2} 3. Comparison with predicate: | Similarities/Differences | | | | | --- | --- | --- | --- | | Item | Device | VARIANT II (k070452) | D-10 (k031043) | | Intended Use | The Hemoglobin Capillary Collection System (HCCS) is intended for the collection of human whole blood for the percent determination of hemoglobin A1c using Bio-Rad HPLC methods. For in vitro diagnostic use | The Bio-Rad VARIANT II Hemoglobin A1c Program is intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high performance liquid chromatography (HPLC). The Bio-Rad VARIANT II Hemoglobin A1c Program is intended for Prescription Use Only. | The Bio-Rad D-10 Hemoglobin A1c Program is intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high performance liquid chromatography (HPLC). The Bio-Rad D-10 Hemoglobin A1c Program is intended for Prescription Use Only. | | Indication for Use | Measurement of percent hemoglobin A1c is effective in monitoring long-term glucose control in individuals with diabetes mellitus. | Measurement of percent hemoglobin A1c is effective in monitoring long-term glucose control in individuals with diabetes mellitus. | Measurement of percent hemoglobin A1c is effective in monitoring long-term glucose control in individuals with diabetes mellitus. | | Formulation | Aqueous solution of EDTA and potassium cyanide | Deionized water with <0.05% sodium azide as a preservative. | Deionized water with <0.05% sodium azide as a preservative. | | Sample | Capillary or venous blood from plastic capillary | Venous whole blood sample collected in an EDTA vacuum collection tube. | Venous whole blood sample collected in an EDTA vacuum collection tube. | | Purpose of Reagent | Sample dilution for Hemoglobin A1c analysis using Bio-Rad HPLC methods. | Sample dilution for hemoglobin A1c analysis using Bio-Rad HPLC method. | Sample dilution for hemoglobin A1c analysis using Bio-Rad HPLC method | | Timing of Dilution Step | Sample dilution at time of sample collection. No further preparation is required; VARIANT II and D-10 Hemoglobin Testing Systems do not perform dilution on samples withdrawn | Dilution with the Wash/Diluent Solution is performed automatically by the VARIANT II Hemoglobin Testing System at time of analysis; when vacuum collection tubes are used. | Dilution with the Wash/Diluent Solution is performed automatically by the D-10 Hemoglobin Testing System at time of analysis; when vacuum collection tubes are used. | {3} | Similarities/Differences | | | | | --- | --- | --- | --- | | Item | Device | VARIANT II (k070452) | D-10 (k031043) | | | from sample vials | | | | Stability of Sample | Prepared samples can be passed on for analysis or shipped to another location. Samples are stable for: 4 days at 42°C, 2 weeks at 15 to 30°C, or 4 weeks at 2-8°C. | Whole blood specimens may be stored up to 7 days at 2-8°C. | Whole blood specimens may be stored up to 7 days at 2-8°C or 3 days at room temperature (15 to 30°C). | ## K. Standard/Guidance Document Referenced (if applicable): CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition CLSI EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach: Approved Guideline Guidance for Industry and FDA Staff; Replacement Reagent and Instrument Family Policy Draft Guidance Document for 510(k) Submission of Glycohemoglobin (Glycated or Glycosylated) Hemoglobin for IVD's Format for Traditional and Abbreviated 510(k)s – Guidance for Industry and FDA Staff ## L. Test Principle: The Hemoglobin Capillary Collection System (HCCS) reagent hemolyzes the Red Blood Cells in the blood sample, removes the Schiff base and preserves the sample. Once the sample is collected and prepared it is ready for immediate analysis, storage for future analysis or shipment for analysis at another site. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Precision studies were conducted by testing three patient whole blood samples at approximately the following levels; low 4-6%, mid 6-8% and high 8-12%. The samples were assayed twice a day for twenty days using one VARIANT II and one D-10 Hemoglobin testing systems. The results are presented in the table below: {4} | | VARIANT II | | | D-10 | | | | --- | --- | --- | --- | --- | --- | --- | | | Low | Mid | High | Low | Mid | High | | Mean - % HbA1c | 4.85 | 6.12 | 10.49 | 4.90 | 6.06 | 10.57 | | Within run % CV | 1.48 | 0.71 | 0.53 | 0.88 | 0.61 | 0.63 | | Between day % CV | 1.02 | 0.23 | 0.76 | 0.00 | 0.54 | 0.53 | | Between run % CV | 1.66 | 0.63 | 0.80 | 2.00 | 0.37 | 0.89 | | Total precision % CV | 2.45 | 0.98 | 1.22 | 1.92 | 0.90 | 1.21 | # b. Linearity/assay reportable range: Linearity across the assay range was confirmed by inter mixing the level 1 sample of the Bio-Rad Lyphochek hemoglobin A1c linearity set ( $\sim$ $3.5\%$ HbA1c) with a high patient sample ( $\sim$ $18.5\%$ HbA1c) creating three pooled sample sets each with 5 levels of HbA1c. One set of samples was diluted with the Hemoglobin Capillary Collection System (HCCS) reagent and two other sets of samples were diluted with the Wash/Diluent solutions one with the VARIANT II and the other with the D-10 Hemoglobin Testing Systems. Each sample was tested in duplicate on each system. Results are presented below: | VARIANT II | | | | --- | --- | --- | | Sample | Theoretical | Measured | | 1 | 3.25 | 3.10 | | 2 | 6.75 | 6.91 | | 3 | 10.70 | 10.72 | | 4 | 14.45 | 14.53 | | 5 | 18.45 | 18.34 | | Regression | y = 1.0026x - 0.1586 | | | D-10 | | | | --- | --- | --- | | Sample | Theoretical | Measured | | 1 | 3.45 | 3.26 | | 2 | 7.3 | 7.10 | | 3 | 11.15 | 10.94 | | 4 | 15 | 14.78 | | 5 | 18.85 | 18.62 | | Regression | y = 0.9974x - 0.181 | | The VARIANT II test system has a reportable range of $3.1 - 18.5\%$ A1c. The D-10 test system has a reportable range of $3.8 - 18.5\%$ A1c. {5} c. Traceability, Stability, Expected values (controls, calibrators, or methods): The sponsor has documented traceability to the NGSP’s recommended accuracy base for HbA1c by performing a direct comparison with a Secondary Reference Laboratory (SRL) using 40 fresh human specimens. At this time, the assay is NGSP certified. However, NGSP certification expires at the end of one year. For current certifications see the NGSP website at: http://www.ngsp.org/prog/index2.html Accelerated stability studies have been conducted. Protocols and acceptance criteria were described and found to be acceptable. When stored at room temperature the shelf life of the HCCS device is 24 months. Stability of whole blood samples prepared with the Hemoglobin Capillary Collection System (HCCS) has been conducted. Protocols and acceptance criteria were described and found to be acceptable. Samples when stored at 2-8°C are stable for 4 weeks, 15-30°C stable for 2 weeks and 42°C stable for 4 days. d. Detection limit: The reportable range for the VARIANT II is 3.1 to 18.5% HbA1c and the D-10 is 3.8 to 18.5% HbA1c. See the linearity study above for data on recovery of samples across the measuring range. e. Analytical specificity: A study was conducted to determine if the level of interference for Liable A1c and Carbamylated Hemoglobin in patient samples differs when diluting samples using the HCCS to diluting the sample with Wash/Diluent for the VARIANT II and D-10 Hemoglobin A1c Programs. Two venous whole blood samples representing normal and diabetic A1c levels were split into 2 aliquots. One aliquot of each pool was treated to elevate the Labile A1c. The other aliquot was treated to elevate the Carbamyated Hemoglobin. Next set was diluted using the HCCS reagent and the other with the Wash/Diluent solution of the assay being studied. Samples and controls were run in duplicate on each system. The sponsor’s acceptance criteria states that the % difference of HbA1c for samples diluted with HCCS method as compared to samples diluted with Wash/Diluent Solution must be less than ≤ 0.3 % HbA1c for normal samples and ≤ 0.5 % HbA1c for diabetic samples. The study showed that these criteria were met. {6} To evaluate the effect on hemoglobin variants such as Hb S, C, D and E when tested using the Hemoglobin Capillary Collection System (HCCS), samples with known hemoglobin variants were tested. One sample was prepared using the Wash/Diluent solution (control) and one using the HCCS device (candidate). The sponsor's acceptance criterion required that the % difference of HbA1c between the control and the candidate devices were within +/- 0.2 % HbA1c. The studies found that samples containing HbS, HbC, HbD and HbE showed no significant interference. f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: A method comparison study was conducted at an external site comparing % HbA1c in capillary samples collected in the HCCS to venous whole blood testing performed on the VARIANT II and D-10 Hemoglobin testing Systems. The following are the correlations: | | Slope | Intercept | R² | N | Range | | --- | --- | --- | --- | --- | --- | | VARIANT II | 1.0169 | -0.092 | 0.9984 | 52 | 3.2 – 18.5% | | D-10 | 1.0176 | -0.0342 | 0.9989 | 52 | 3.2 – 18.5% | b. Matrix comparison: Paired capillary and EDTA venous whole blood samples were diluted using the Hemoglobin Capillary Collection System (HCCS) and were assayed on the VARIANT II and D-10 Hemoglobin Testing Systems. The following are the correlations: | | Slope | Intercept | R² | N | Range | | --- | --- | --- | --- | --- | --- | | VARIANT II | 1.000 | 0.000 | 0.998 | 31 | 5 – 12.0% | | D-10 | 1.000 | -0.100 | 0.999 | 31 | 4.8 – 12.2% | 3. Clinical studies: a. Clinical Sensitivity: Not applicable {7} b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Hemoglobin A1c Expected Value Range was determined from literature (American Diabetes Association. Standards of Medical Care for Patients with Diabetes Mellitus. Diabetes Care 2001, 24 (Suppl. 1), 33-43). Hemoglobin A1c &gt; 8% Action is suggested. Hemoglobin A1c &lt; 7% is the goal. Hemoglobin A1c &lt; 6% is non-diabetic level. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...