← Product Code [MQV](/productcode/MQV) · K090731

# NOVABONE POROUS - SYNTHETIC BONE GRAFT SCAFFOLD (K090731)

_Novabone Products, LLC · MQV · Apr 15, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K090731

## Device Facts

- **Applicant:** Novabone Products, LLC
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Apr 15, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

NovaBone Porous is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. NovaBone Porous is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

## Device Story

NovaBone Porous is a synthetic, resorbable, osteoconductive bone void filler composed of calcium phospho-silicate (45S5 Bioglass) particulate. The device is designed as a bulk porous scaffold with multidirectional interconnected porosity. Upon implantation, it undergoes surface modification to form a calcium phosphate layer, facilitating new bone infiltration and gradual resorption. The device is intended for use by clinicians experienced in bone grafting and internal/external fixation. It is not load-bearing and requires standard stabilization techniques for fracture fixation. The subject modification introduces a syringe delivery system to aid in material preparation and application, similar to existing predicate delivery methods. The device provides a scaffold for natural bone regeneration.

## Clinical Evidence

No clinical data provided; bench testing only. The submission relies on the established safety and efficacy of the predicate device and the similarity of the new syringe delivery format to existing cleared devices.

## Technological Characteristics

Single-phase bioactive glass (45S5 Bioglass) in particulate form. Osteoconductive, resorbable scaffold with multidirectional interconnected porosity. Delivered via syringe. Non-structural, non-load-bearing. No software or electronic components.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- NovaBone Porous - Bone Graft Scaffold ([K060432](/device/K060432.md))
- NovaBone Putty - Bioactive Synthetic Graft ([K060728](/device/K060728.md), [K080009](/device/K080009.md), [K082672](/device/K082672.md))
- NovaBone-AR - Resorbable Bone Graft Substitute ([K041613](/device/K041613.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for NOVABONE SYNTHETIC BONE GRAFT PRODUCTS. The word "NOVABONE" is in large, bold, black letters. Below the word is the phrase "SYNTHETIC BONE GRAFT PRODUCTS" in smaller, black letters.

Image /page/0/Picture/1 description: The image shows a sequence of handwritten characters, "K690731". The characters are written in a simple, clear style, with each digit and letter easily distinguishable. The sequence appears to be a code or identifier, possibly a serial number or a reference code.

APR 1 5 2009

# Special 510(k): Device Modification Summary NovaBone Porous - Bone Graft Scaffold

03/16/09

# Submitter Information: 1.

| Name:      | NovaBone Products, LLC                             |
|------------|----------------------------------------------------|
| Address:   | 13709 Progress Boulevard, #33<br>Alachua, FL 32615 |
| Telephone: | (386) 462-7660                                     |
| Facsimile: | (386) 418-1636                                     |
| Contact:   | David M. Gaisser                                   |

### 2. Name of Device:

| Trade Name:        | NovaBone Porous – Synthetic Bone Graft Scaffold                              |
|--------------------|------------------------------------------------------------------------------|
| Common Name:       | Osteoconductive Bone Void Filler<br>Synthetic Resorbable Bone Graft Material |
| Regulation Number: | 21 CFR 888.3045                                                              |
| Regulation Name:   | Bone Void Filler                                                             |

# 3. Legally Marketed Predicate Device:

| Predicate #1: | NovaBone Porous - Bone Graft Scaffold<br>[K060432]                        |
|---------------|---------------------------------------------------------------------------|
| Predicate #2: | NovaBone Putty - Bioactive Synthetic Graft<br>[K060728, K080009, K082672] |
| Predicate #3: | NovaBone-AR - Resorbable Bone Graft Substitute<br>[K041613]               |

# 4. Device Description

NovaBone Porous is an osteoconductive bioactive device. It is a one-component, resorbable bone void filler composed of a synthetic calcium phospho-silicate (Bioglass) particulate, fused into a bulk porous form having a multidirectional interconnected porosity. On implantation, NovaBone Porous undergoes a timedependent surface modification, resulting in the formation of a calcium phosphate layer on the device surfaces. The device acts as a scaffold, with new bone infiltrating the porous structure. NovaBone Porous is progressively resorbed and replaced by new bone tissue during the healing process.

# 5. Intended Use

NovaBone Porous is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. NovaBone Porous is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. the extremities and These defects may be surgically created osseous defects or osseous pelvis). defects created from traumatic injury to the bone. The product provides an 

{1}------------------------------------------------

osteoconductive scaffold that resorbs and is replaced with bone during the healing process.

#### Technological Characteristics Q.

The technological characteristics of the NovaBone Porous device are identical to the NovaBone Porous device as cleared per K060432, and are similar to those of the NovaBone Putty and NovaBone-AR predicates. The devices are designed as osteoconductive space-filling devices to be gently packed into defect sites and used as non-structural scaffolds for the body's natural healing and bone regeneration process. The device indications are the same as for the predicate. The device per this submission is modified to include a syringe package format to aid in material preparation and delivery.

The NovaBone Porous device is a single-phase bioactive glass (45S5 Bioglass®) device. The device implant material is unmodified from that described in K060432, with the device being in particulate form. The design modification made per this submission is to include a syringe package format to contain the device material, the syringe being similar to that for the NovaBone-AR predicate per K041613 and the NovaBone Putty predicate per K082672.

# 7. Warnings and Precautions

NovaBone Porous does not possess sufficient mechanical strength to support loadbearing defects prior to hard tissue ingrowth and is not indicated for use in loadbearing applications. NovaBone Porous is intended for use by clinicians familiar with bone grafting and internal/external fixation techniques. In cases of fracture fixation or where load support is required, standard internal or external stabilization techniques must be followed to obtain rigid stabilization in all planes. NovaBone Porous must not be used to gain screw purchase or to stabilize screw placement.

#### 8. Complications

Possible complications are the same as to be expected of autogenous bone grafting procedures. These may include: superficial wound infection, deep wound infection, deep wound infection with osteomyelitis, delayed union, loss of reduction, failure of fusion, loss of bone graft, graft protrusion and / or dislodgement, and general complications that may arise from anesthesia and / or surgery.

#### 9. Conclusion

The NovaBone Porous design modification subject to this submission is to include a filled-syringe package format to augment the current package formats. This device modification does not result in a change in technological characteristics of the device. NovaBone Porous continues to be safe and effective as a non-structural osteoconductive bone void filler for osseous defects.

\$\pi\$

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the caduceus. The logo is simple and monochromatic.

ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NovaBone Products, LLC % Mr. David M. Gaisser 13709 Progress Boulevard, #33 Alachua, Florida 32615

APR ] 5 2009

Re: K090731

Trade/Device Name: NovaBone Porous - Synthetic Bone Graft Scaffold Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MQV Dated: March 16, 2009 Received: March 19, 2009

Dear Mr. Gaisser:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. David M. Gaisser

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please . contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Barbare Brehm

Mark N. Mel kerson Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Special 510(k): Design Modification NovaBone Products, LLC NovaBone Porous - Synthetic Bone Graft Scaffold

# STATEMENT OF INDICATIONS FOR USE

K 01073 510(k) Number (if known):

NovaBone Porous - Synthetic Bone Graft Scaffold Device Name:

Indications For Use:

NovaBone Porous - Synthetic Bone Graft Scaffold is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. NovaBone Porous is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

Prescription Use XX

OR (Per 21 CFR 801.109) Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Clifford Buelow for MXM
Division Sign-Off

rative. and Neurological De

**510(k) Number** K090731

1-1

---

**Source:** [https://fda-staging.innolitics.com/device/K090731](https://fda-staging.innolitics.com/device/K090731)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
